K Number
K103344

Validate with FDA (Live)

Device Name
INFUSION SETS
Manufacturer
Date Cleared
2010-12-10

(25 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Infusion sets are intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood and blood products via parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural or irrigation routes of administration. These sets are intended for use primarily in the hospital setting and can be used in other acute and non-acute care areas, such as, but not limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, Hospice, Subacute facilities, Outpatient / Surgical Centers, Long Term Care, Urgent Care, Transport and in Physician offices.

Device Description

Hospira infusion sets are intended for use as gravity sets or with dedicated Hospira Infusion Pumps. Hospira infusion sets are disposable devices for single patient use, which incorporate various set configurations and components. These components including the back check valve which is the subject of this submission may be shared across Hospira set families and used with sets designed for gravity or other pump platforms.

AI/ML Overview

The acceptance criteria and the study proving the device meets these criteria are detailed below, based on the provided document.

Acceptance Criteria and Device Performance

The Hospira Infusion Sets with an alternate fixed diaphragm Back Check Valve were evaluated against various performance characteristics, primarily focusing on biocompatibility and sterility. The acceptance criteria are implicitly defined by the successful "Pass" result for each test.

CharacteristicAcceptance Criteria (Standard/Test Method)Reported Device Performance
Biocompatibility: CytotoxicityISO 10993-5: 2009Pass
Biocompatibility: SensitizationISO 10993-10: 2002Pass
Biocompatibility: Irritation / Intracutaneous ReactivityISO 10993-10: 2002Pass
Biocompatibility: Systemic Toxicity (Acute)ISO 10993-11:2006Pass
Biocompatibility: HemocompatibilityISO 10993-4:2002Pass
SterilityISO 11137-2:2006 (SAL 10-6)Pass
Set FunctionalityMeets all performance V&V testingSame (as predicate device)
Back Check Valve PerformanceReduction of backflowDemonstrated excellent

Study Details

  1. Sample Size used for the test set and the data provenance:
    The document does not specify the exact sample sizes used for each individual test (e.g., number of devices tested for cytotoxicity, sensitization, etc.). It only states that the "Hospira Infusion sets with an alternate Back Check Valve passed all specified test requirements."
    The data provenance is not explicitly stated as country of origin, nor is it explicitly mentioned whether the data is retrospective or prospective. However, given this is a 510(k) submission for a modified device, the testing would typically be prospective for the modified component.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not provided in the document. The tests performed are laboratory-based performance and safety tests, not clinical evaluations requiring expert consensus on a 'ground truth' in the context of diagnostic algorithms.

  3. Adjudication method for the test set:
    Not applicable, as the tests are laboratory-based performance and safety tests with objective pass/fail criteria according to international standards.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This submission is for a physical medical device (infusion set) and not an AI/software as a medical device (SaMD) that involves human readers or AI assistance.

  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
    Not applicable, as this is a physical medical device.

  6. The type of ground truth used:
    The ground truth for the device's performance is established by the specified international standards (ISO series) for biocompatibility and sterility. For the back check valve performance, it's assessed against "reduction of backflow" and "user needs and design inputs for an Infusion set." These are objective, measurable criteria defined by regulatory and engineering standards.

  7. The sample size for the training set:
    Not applicable, as this is a physical medical device and not an AI/machine learning algorithm that requires a training set.

  8. How the ground truth for the training set was established:
    Not applicable, for the same reason as above.

{0}------------------------------------------------

K103344

November 12, 2010

·

Section 6:

510(k) SUMMARY

DEC 1 0 2010

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92. Hospira Infusion Sets with alternate fixed diaphragm Back Check Valve

Submitter Information
NameRebecca Andersen
Address275 North Field Drive
Phone number224-212-5270
Fax number224-212-5401
EstablishmentRegistration NumberOwner/Operator #9063339
Name of contact personRebecca Andersen
Date preparedNovember 12, 2010
Name of device
Trade or proprietarynameHospira Infusion Sets
Common or usual nameIV Set
Classification nameInfusion Sets Class II
Classification panel80-FPA - General Hospital
Regulation21 CFR Part 880.5440
Product Code(s)80-FPA
Legally marketed device(s) towhich equivalence is claimedSymbiq sets as cleared in K041550
Reason for 510(k) submissionBack Check Valve component
Device descriptionHospira infusion sets are intended for use as gravity sets or with dedicatedHospira Infusion Pumps. Hospira infusion sets are disposable devices forsingle patient use, which incorporate various set configurations andcomponents. These components including the back check valve which is thesubject of this submission may be shared across Hospira set families andused with sets designed for gravity or other pump platforms.
Intended use of the deviceInfusion sets are intended for the delivery of fluids, solutions, drugs, agents,nutritionals, electrolytes, blood and blood products via parenteral, enteral,intravenous, intra-arterial, subcutaneous, epidural or irrigation routes ofadministration. These sets are intended for use primarily in the hospital setting andcan be used in other acute and non-acute care areas, such as, but not limited toHome Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers,Hospice, Subacute facilities, Outpatient / Surgical Centers, Long Term Care, UrgentCare, Transport and in Physician offices.
Indications for useThese are general use sets which are intended for use primarily in the hospitalsetting and can be used in other acute and non-acute care areas, such as, but notlimited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory InfusionCenters, Hospice, Subacute facilities, Outpatient / Surgical Centers, Long TermCare, Urgent Care, Transport and in Physician offices.

{1}------------------------------------------------

November 12, 2010

Infusion Set Modification

Special 510(k)

510(k) SUMMARY

Section 6:510(k) SUMMARY
Summary of the technological characteristics of the device compared to the predicate device
CharacteristicProposed DeviceHospira Infusion SetPredicate Symbiq Infusion SetK041550
Intended UseSameThese sets are intended for use primarily inthe hospital setting and can be used in otheracute and non-acute care areas, such as, butnot limited to Home Care, Nursing Homes,Mobile Intensive Care, Ambulatory InfusionCenters, Hospice, Subacute facilities,Outpatient / Surgical Centers, Long TermCare, Urgent Care, Transport and inPhysician offices.
Set FunctionalitySameMeets all performance V&V testing
Visual characteristicsSimilar Back Check valve is ofsimilar size, but is blue and clearBack check valve is clear / white
Back Check ValveNormally closed - One way checkvalve in two piece housing withfixed silicone diaphragmNormally closed - One way check valve in twopiece housing with floating siliconediaphragm
Bonding processesSameSolvent bond / over post or socket bond
BiocompatibilitySameExternal Communicating - Blood path indirect-Prolonged contact
Principle of OperationSameOne way pressure differential
SUMMARY OF NON CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
Performance Test Summary-New Device
CharacteristicStandard/Test MethodStandard / Test Title Device Performance
BiocompatibilityISO 10993-5: 2009Cytotoxicity Pass
BiocompatibilityISO 10993-10: 2002Sensitization Pass
BiocompatibilityISO 10993-10: 2002Irritation / IntracutaneousReactivity Pass
BiocompatibilityISO 10993-11:2006Systemic Toxicity (Acute) Pass
BiocompatibilityISO 10993-4:2002Hemocompatibility Pass
SAL 10-6ISO 11137-2:2006Sterility Pass

Summary discussion of Bench Performance Data

The Hospira Infusion sets with an alternate Back Check Valve passed all specified test requirements. Testing The Prospira Infosion sels with respect to reduction of backflow. The validation and verification testing acmonotiate oxeculier. Street user needs and design inputs for an Infusion set.

mod these doness most as attributes and device performance meet requirements of fire standards istead inthe

{2}------------------------------------------------

Section 6:

510(k) SUMMARY

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Statement of Safety and Efficacy:

The Hospira Infusion Sets with the alternate back check valve meet the functional claims, and intended use as described in the product labeling. The safety and effectiveness, are substantially equivalent to the predicate Hospira Infusion Sets as cleared in K041550

The claim for substantial equivalence is supported by the information provided in this Special 510(k) submission.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Rebecca Anderson Associate Director, Global Regulatory Affairs, Devices HOSPIRA, Incorporated 275 North Field Drive Lake Forest, Illinois 60045

DEC 1 0 2000

Re: K103344

Trade/Device Name: Hospira Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: November 12, 2010 Received: November 15, 2010

Dear Ms. Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of Amendinente, or to arrivand Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, lecturents for allinar regainst misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 w/1), it thay of surfet is alla of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not Fleast oc advisod that 1 Dr. 5 lookines on that your device complies with other requirements Incall that FDA has made a decommanent in administered by other Federal agencies. of the Act of ally I oderal bates as requirements, including, but not limited to: registration 1 ou must comply with an the Fiel singles and 801); medical device reporting and fishing (21 CFR Pat 607), lacenad (21 CFR 803); good manufacturing (reporting of medical device the quality systems (QS) regulation (21 CFR Part 820); practice requirements as sectionic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm113809.htm for inttp://www.ida.gov//About Health's (CDRH's) Office of Compliance. Also, the Somer the regulation entitled, "Misbranding by reference to premarket notification" prease note the regarding the reporting the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the I ou him of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S.,M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Infusion Set Modification Special 510(k)

. !

November 12, 2010

Section 5: Indications for Use

DEC 1 0 2010

K103344.

510(k) Number (unknown at this time)

Device Name: Hospira Infusion Set

Infusion sets are intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood and blood products via parenteral, intravenous, intra-arterial, subcutaneous, epidural or irrigation routes of administration. These sets are intended for use primarily in the hospital setting and can be used in other acute and non-acute care areas, such as, but not limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, Hospice, Subacute facilities, Outpatient / Surgical Centers, Long Term Care, Urgent Care, Transport and in Physician offices.

Prescription UseXAND/OROver-The-Counter Use______
(Part 21 CFR 801 Subpart D)(Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rill C. Chapman 12/10/10

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K103344

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.