K Number
K103247

Validate with FDA (Live)

Device Name
VASCUCOMP
Manufacturer
Date Cleared
2010-12-10

(37 days)

Product Code
Regulation Number
870.5800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VascuComp™ AI Mode: The VascuComp™ compression system in AI Mode is used as an adjunct therapy for patients with ischemic disease of the lower extremities, due to one or more of the following causes: Minor Amputations, Angioplasty / Stent Failure, Arteriopathic Wounds, Graft Failure, Intermittent Claudication, Ischemia, Night Pain, Rest Pain, Small Vessel Discase, Ulcers. VascuComp™ DVT Mode: Decrease the risk of deep venous thrombosis (DVT). Aids the blood flow back to the heart. Treat and assist healing of cutaneous ulceration (wounds), reduce wound healing time, enhance arterial circulation (blood flow), reduce compartmental pressures, reduce edema (swelling), reduce the need for anticoagulant (blood thinning) medications. VascuComp™ L Mode: Treatment of Venous disorders and dysfunction of the "muscle pump".

Device Description

The VascuComp™ device is comprised of a reusable air-pump with user adjustable pressures, and various single-patient sleeves or garments containing discrete, segmental inflatable chambers to be externally applied over the affected extremity of the pationt. The Vascu Comp™ provides calibrated, intermittent, sequential compression.to the treatment site for optimum therapy for patients with ischemic disease of the lower extremities, by increasing arterial circulation, and to reduce the risk of the formation of DVT and to reduce or control peripheral edema. The Vascu Comp™ systems utilize microprocessor control with multiple sensors to ensure patient safety and system functionality, and to provide consistent and repeatable therapy modalities. Alarms are both visual on the unit display and audible. Alarms are in place to detect a potentially unsafe situation and to terminate therapy to protect the patient and the system.

AI/ML Overview

The ThermoTek VascuComp™ is an intermittent, sequential pneumatic compression device. The provided text describes non-clinical performance testing to demonstrate substantial equivalence to predicate devices, rather than a study proving the device meets specific acceptance criteria in a clinical context.

The "acceptance criteria" in this context are the published and observed therapy modalities (pressure levels and cycle times) and safety standards of the predicate devices. The "reported device performance" is the VascuComp™'s ability to match these parameters.

1. Table of Acceptance Criteria and Reported Device Performance

Parameter (Acceptance Criteria Based on Predicate Device)VascuComp™ Reported Device Performance
VascuComp™ AI Mode (Predicate: BioArterial Plus K072666)
Inflation Pressure
CalfMatches 120 mmHg
FootMatches 120 mmHg
Cycle Time
CompressionMatches 3 seconds (± 0.5 sec)
Non-CompressionMatches 17 seconds (± 0.5 sec)
Delay between Foot & CalfMatches 1 second (± 0.5 sec)
Total CycleMatches 20 seconds / 3 Cycles per minute
Electrical SafetyMatches IEC 60601-1, UL60601-1, CAN/CSA 22.2, IEC 60601-1-2
VascuComp™ DVT Mode (Predicate: VascuTherm K061866)
Inflation Pressure
CalfMatches 45 mmHg
FootMatches 100 mmHg
Cycle Time
InflationMatches 30 seconds
DeflationMatches 30 seconds
Electrical SafetyMatches IEC 60601-1, UL60601-1, CAN/CSA 22.2, IEC 60601-1-2
VascuComp™ L Mode (Predicate: PBS Model 701A K060220)
Inflation PressureMatches 20 to 80 mmHg
Electrical SafetyMatches IEC 60601-1, UL60601-1, CAN/CSA 22.2, IEC 60601-1-2
General Safety (IEC 60601-1 & IEC60601-1-2)Met construction, dielectric testing, electrical leakage, and pre-scan data for conducted and radiated emissions.

2. Sample Size Used for the Test Set and Data Provenance

The document describes non-clinical performance testing.

  • Sample Size: The testing for the pneumatic compression therapy functionality (pressure and cycle times) involved "a minimum of 2 compression cycles" for the L Mode, "a minimum of two compression cycles" for the DVT Mode, and "a minimum of three compression cycles" for the AI Mode.
  • Data Provenance: The data is generated from laboratory testing of the VascuComp™ device and its accessories. It is considered prospective in the context of this specific device's evaluation, as the tests were conducted specifically to demonstrate its performance against established predicate device parameters. The country of origin of the data is not specified but is implicitly the location of ThermoTek, Inc. (Flower Mound, Texas).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Experts were not used to establish ground truth for this non-clinical performance test. The "ground truth" (acceptance criteria) was derived from the published and observed therapy modalities and technical specifications/standards of the predicate devices (BioArterial Plus K072666, VascuTherm K061866, and PBS Model 701A K060220).

4. Adjudication Method for the Test Set

No adjudication method was mentioned as this was non-clinical performance testing, not a clinical study involving human assessment or interpretation. The performance was measured against quantitative specifications.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This device is an intermittent pneumatic compression machine, not an AI-assisted diagnostic or interpretive tool that would involve human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is a standalone device in the sense that its performance was measured directly based on its outputs (pressure and cycle times). There is no "algorithm only" performance separate from the device's physical operation, as it is a mechanical device, not a software-only product.

7. The Type of Ground Truth Used

The ground truth used was pre-established physical specifications and performance characteristics of legally marketed predicate devices, along with international electrical safety and electromagnetic compatibility standards (IEC 60601-1, IEC 60601-1-2, UL and CAN/CSA equivalents).

8. The Sample Size for the Training Set

There is no "training set" in the context of this device evaluation. The device's operation is based on electromechanical principles, not machine learning or AI that requires a training dataset.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

K103247

DEC 1 0 2010

Section 5. 510(k) Summary

APPLICANT

Telephone Number:

Company Contact:

Address:

[As Required by 21 CFR 807.92 a(1)] Applicant:

ThermoTek, Inc.

1200 Lakeside Parkway, #200 Flower Mound, Texas 75028

972-874-4949 972-874-4945 (fax)

Niran Balachandran Director of Engineering

Date:

April 28, 2010

DEVICE NAME [As Required by 21 CFR 807.92 a(2)]

Proprietary Name:

Common Name:

Device Classification: Regulation Number: Product Code: Device Panel:

VascuComp™

Intermittent, External Pncumatic Compression Device

Class II 870.5800 JOW Cardiovascular

K072666

IDENTIFICATION OF PREDICATE DEVICES [As Required by 21 CFR 807.92 a(3)]

BioCompression Systems, Inc., BioArterial Plus

Thermo Tek Inc., VascuThermK061860
PBS (Petit Basci System) Model 701AK060220

{1}------------------------------------------------

DEVICE DESCRIPTION [As Required by 21 CFR 807.92 a(4)]

Intended Use

The VascuComp™-01 and VascuComp™-02 is a new device that is intended to function as an intermittent, sequential pneumatic compression device. It is intended to provide the following therapeutic functions:

    1. VascuComp™ AI Mode: Increase arterial inflow to lower limbs of those with ischemic disease of the lower extremities.
  • VascuComp™ DVT Mode: Decrease the risk of deep venous thrombosis (DVT) 2 by aiding blood flow back to the heart via lower extremity limb compression.
    1. VascuCompl'M L Mode: Intended to aid in the treatment of Venous disorders and dysfunction of the "muscle pump".

Physical Description

The VascuComp™ device is comprised of a reusable air-pump with user adjustable pressures, and various single-patient sleeves or garments containing discrete, segmental inflatable chambers to be externally applied over the affected extremity of the pationt.

Therapy Modality Used

The Vascu Comp™ provides calibrated, intermittent, sequential compression.to the treatment site for optimum therapy for patients with ischemic disease of the lower extremities, by increasing arterial circulation, and to reduce the risk of the formation of DVT and to reduce or control peripheral edema.

Safety Features

The Vascu Comp™ systems utilize microprocessor control with multiple sensors to ensure patient safety and system functionality, and to provide consistent and repeatable therapy modalities. Alarms are both visual on the unit display and audible. Alarms are in place to detect a potentially unsafe situation and to terminate therapy to protect the patient and the system,

STATEMENT OF INTENDED USE [As Required by 21 CFR 807.92 a(5)]

Vascu Comp™ Indications for Use: VascuComp™ AI Mode:

The VascuComp™ compression system in AI Mode is used as an adjunct therapy for patients with ischemic disease of the lower extremities, due to one or more of the following causes:

{2}------------------------------------------------

  • Minor Amputations .
  • Angioplasty / Stent Failure .
  • Arteriopathic Wounds .
  • Graft Failure .
  • Intermittent Claudication .
  • Ischemia .
  • Night Pain .
  • Rest Pain .
  • Small Vessel Discasc
  • Ulcers .

VascuComp™ DVT Mode:

  • . Decrease the risk of deep venous thrombosis (DVT).
  • Aids the blood flow back to the heart. .
  • Treat and assist healing of cutaneous ulceration (wounds), reduce wound healing . time, enhance arterial circulation (blood flow), reduce compartmental pressures, reduce edema (swelling), reduce the need for anticoagulant (blood thinning) medications.

VascuComp™ L Mode:

  • . Treatment of Venous disorders and dysfunction of the "muscle pump".

TECHNOLOGICAL CHARACTERISTICS

[As Required by 21 CFR 807.92 a(6)]

The Vascu Comp™ compression device has the same technical and performance characteristics as the predicate devices.

The pneumatic control circuitry is a microprocessor-controlled system. Multiple safety redundancies are built into the system including: alarms for open and blocked airflow detection and for unit malfunction situations.

The Vascu Comp™ compression device is powered from a 110 or a 220 ± 10% VAC source. It also incorporates an ergonomic, intuitive user interface for case of use and to customize treatment parameters.

{3}------------------------------------------------

Comparison of features and principles of operation between Vascu Comp™ and predicate devices per Section 510(k) of the Act.


ParameterVascuComp™ Compression DeviceBio Compression Systems, Inc.,BioArterial Plus (K072666)
Intended useIntermittent Sequential Pneumatic CompressionIntermittent Sequential Pneumatic Compression
Indicationsfor UseThe VascuComp™ compression system in Al Mode is used as an therapy device for patients with ischemic disease of the lower extremities, due to one or more of the following causes:Minor AmputationsAngioplasty / Stent FailureArteriopathic WoundsGraft FailureIntermittent ClaudicationIschemiaNight PainRest PainSmall Vessel DiseaseUlcersThe BioArterial Plus Arterial Blood Flow Enhancement System is intended as adjunct therapy for patients with ischemic disease of the lower extremities, due to one or more of the following causes:Minor AmputationsAngioplasty / Stent FailureArteriopathic WoundsGraft FailureIntermittent ClaudicationIschemiaNight PainRest PainSmall Vessel DiseaseUlcers.
InflationPressureCalf: 120 mmHgFoot: 120 mmHgCalf: 120 mmHgFoot: 120 mmHg
Cycle Time(approximate)Sequential compression of the limbs (foot followed by calf)Compression: 3 seconds (± 0.5 sec)Non-Compression: 17 seconds (± 0.5 sec)Delay between Foot & Calf: 1 second (± 0.5 sec)Cycle: 20 seconds / 3 Cycles per minuteSequential compression of the limbs (foot followed by calf)Compression: 3 seconds (± 0.5 sec)Non-Compression: 17 seconds (± 0.5 sec)Delay between Foot & Calf: 1 second (± 0.5 sec)Cycle: 20 seconds / 3 Cycles per minute
ElectricalSafetyIEC 60601-1 (safety)UL60601-1 (safety) Class II,Type BCAN/CSA 22.2IEC 60601-1-2 (Emissions)IEC 60601-1-2 (Immunity)UL Std: 60601-1, Class II, Type BEN60601-1-2:2001
Power RatingVascuComp - 01Voltage: 100-120 VACFreq: 60Hz Ph: 1Current: 3.0ABioArterial PlusVoltage: 120 VACFreq: 60HzCurrent: 0.5A
VascuComp - 02Voltage: 220-240 VACFreq: 50/60Hz Ph: 1Current: 1.5A

VascuComp™ AI Mode: .

.

{4}------------------------------------------------

:

:


VascuComp™ DVT Mode:

.

. . . .

ParameterVascuComp™ CompressionDeviceThermoTek Inc., VascuTherm(K061866)
Intended useIntermittent PneumaticCompressionIntermittent Pneumatic Compression
Indicationsfor UseDecrease the risk of deep venousthrombosis (DVT).Decreases the risk of deep venousthrombosis, DVT
Aids the blood flow back to theheart.Aids the blood flow back to the heart.
Treat and assist healing ofcutaneous ulceration (wounds),reduce would healing time, enhancearterial circulation (blood flow),reduce compartmental pressures,reduce edema (swelling), reduce theneed for anticoagulant (bloodthinning) medications.Treat and assist healing of cutaneousulceration (wounds), reduce woundhealing time, enhance arterialcirculation (blood flow), reducecompartmental pressures, reduceedema (swelling), reduce the need foranticoagulant (blood thinning)medications.
Inflation45mmHg (calf)45 mmHg (calf)
PressureCycle Time100mmHg (foot)Inflation: 30 seconds100 mmHg (foot)
(approximate)Deflation: 30 secondsInflation: 30 secondsDeflation: 30 seconds
ElectricalIEC 60601-1 (safety)UL Std: 60601 Class II, Type B
SafetyUL60601-1 (safety) Class II,Type BCAN/CSA 22.2IEC 60601-1-2 (Emissions)IEC 60601-1-2 (Immunity)IEC 60601-1-2 (Emissions)IEC 60601-1-2 (Immunity)
Power RatingVascuComp - 01Voltage: 100-120 VACFreq: 60Hz Ph: 1Current: 3.0AVascuComp - 02Voltage: 220-240 VACFreq: 50/60Hz Ph: 1VascuThermVoltage: 100-120 VACFreq: 60Hz Ph: 1Current: 4.5A

.

.

{5}------------------------------------------------

ParameterVascuComp™ CompressionDevicePBS (Petit Basci System)Model 701A (K060220)
Intended useIntermittent, Sequential, PneumaticCompressionIntermittent, Sequential, PneumaticCompression
Indicationsfor UseTreatment of Venous disorders anddysfunction of the "muscle pump"Treatment of Venous disorders anddysfunction of the "muscle pump"
InflationPressure20 to 80 mmHg20 to 80 mmHg
ElectricalSafetyIEC 60601-1 (safety)UL60601-1 (safety) Class II,Type BCAN/CSA 22.2IEC 60601-1-2 (Emissions)IEC 60601-1-2 (Immunity)UL Std: 60601-1, Class II,CAN/CSA C22.2
Power RatingVascuComp - 01Voltage: 100-120 VACFreq: 60Hz Ph: 1Current: 3.0APBS Model 701AVoltage: 115 VACFreq: 50/60 HzCurrent: 2.0A
VascuComp - 02Voltage: 220-240 VACFreq: 50/60Hz Ph: 1Current: 1.5A

n'I'M T Mode

CONTRAINDICATIONS FOR USE

The patient should not use the Vascu Comp™ compression therapy system if the patient is suspected of or observed to have any of the following:

VascuComp™ AI Mode:

  • 트 Undesirable venous and lymphatic return (as with congestive heart failure)
  • 트 Deep vein thrombosis (suspected or present)
  • " Inflammatory Phlebitis
  • . Episodes of pulmonary embolism
  • 트 Sepsis in the limb
  • Cellulitis without appropriate antibiotic coverage ..
  • I Immediately following skin grafts in/around treatment sides
  • Pulmonary edema associated with congestive heart failure 트
  • Acute thrombophlebitis 트

{6}------------------------------------------------

VascuComp™ DVT Mode:

  • 내 Presumptive evidence of congestive heart failure
  • 디 Suspected / observed pre-existing deep vein thrombosis or pulmonary emblosim
  • Suspected/observed deep acute venal thrombosis (phlebothrombosis) 비
  • I Suspected/observed inflammatory phlebitis process
  • 피 Suspected/observed pulmonary edema
  • Suspected/observed acute inflammations of the veins (thrombophlebitis)
  • 트 Suspected/observed decompensated cardiac insufficiency
  • Suspected/observed arterial dysregulation
  • I Suspected/observed erysipelas
  • 買 Suspected/observed carcinoma and carcinoma metastasis in the affected extremity
  • 트 Suspected/observed decompensated hypertonia
  • 피 Suspected/observed acute inflammatory skin diseases or infection
  • . Suspected/observed venous or arterial occlusive disease
  • 로 Determine venous and lymphatic return is undesirable
  • 트 Poor Peripheral Circulation
  • 에 Severe Arteriosclerosis, or active infection

VascuComp™ L Mode:

  • ロ Known or suspected deep venous thrombosis
  • 피 Inflammatory Phlebitis
  • 노 Episodes of pulmonary embolism
  • 지 Pulmonary edema or congestive heart failure
  • 미 Severe arteriosclerosis or other ischemic vascular disease.
  • 제 Any circumstance where increased venous and lymphatic return is undesirable.

PERFORMANCE DATA

Non clinical performance testing was performed as part of a sequence of verification testing to validate the subject device, the VascuComp, meets the required specifications when compared to the predicates the PBS (Petit Basic System) Model 701A (K060220), ThermoTek VascuTherm (K061866) and Bio Compression Systems, Inc., BioArterial Plus (K072666) predicate devices.

The objective of the first tests were to verify that the therapy pressure level and cycle times in regards to pneumatic compression therapy functionality for the treatment of edema and lymphedema on the VascuComp and four-chamber full leg garment is satisfactory as compared to the published and observed therapy modality of the PBS (Petit Basci System) Model 701A pedicate device.

{7}------------------------------------------------

Conclusions

Over a minimum of 2 compression cycles, VascuComp with a the four-chamber full leg garment showed that the cvele pressure and time passed all preset criteria from the PBS (Petit Basci System) Model 701A predicate device literature.

The objectives of the second tests were to verify that the therapy pressure level and cycle times in regards to pneumatic compression therapy functionality to reduce the risk of DVT formation on the VascuComp and ThermoTek predicate foot wrap and call wrap is satisfactory as compared to the published and observed therapy modality of the ThermoTek, VascuTherm predicate device.

Conclusions

Use of the ThermoTek predicate foot wrap and calf wrap accessory on the VascuComp device vielded compression data that meets the preset criteria for proper DVT therapy modality from the ThermoTek predicate device. The VascuComp DVT wraps are identical in treatment area and therapy modality to the predicate device, the VascuTherm. The data from this test showed that the VascuComp device passed all preset test criteria as compared to ThermoTek, VascuTherm (K061866) predicate device literature.

The objectives of the third tests were to verify that the therapy pressure level and cycle times in regards to pneumatic compression therapy functionality for the enhancement of arterial blood flow on the VascuComp and a sequential compression foot - cal! garment is satisfactory as compared to the published and observed therapy modality of the Bio Compression Systems, Inc., BioArterial Plus (K072666) predicate device.

Conclusions

Over a minimum of three compression cycles, VascuComp with a foot-calf garment showed that the treatment pressure and time passed all preset criteria form the Bio Compression Systems, Inc., BioArterial Plus (K072666) predicate device literature,

The objectives of the fourth test were to verify conformance of the VascuComp I and II devices to IEC 60601-1 safety and IEC60601-1-2 radiated and conducted emissions standard.

Conclusions

The device met the construction, dialetric testing and electrical leakage as specificd in the IEC 60601-1 standard. Pre scan data of conducted and radiated emissions of the device also confirmed that it met the IEC60601-1-2 EMC/EMI standards.

{8}------------------------------------------------

STATEMENT OF SUBSTANTIAL EQUIVALENCE:

·

Based upon the safety and performance testing, compliance with voluntary standards, and comparison to the predicate devices in terms of features, functionality and non clinical comparison testing, the manufacture believes that the VascuComp™ therapy system is substantially equivalent to the predicate devices, and does not raise any new risks of safety of effectiveness.

{9}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Thermo Tek, Inc. c/o Mr. Bhavesh V. Sheth Regulatory Reviewer Intertek Testing Services, Inc. 2307 E. Aurora Rd. Unit B7 Twinsburg, OH 44087

DEC 1 0 2010

Re: K103247

VascuComp (Models 01 and 02) Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: November 2, 2010 Received: November 3, 2010

Dear Mr. Sheth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{10}------------------------------------------------

Page 2 - Mr. Bhavesh V. Sheth

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

R. Vohrer

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Section 4. Indications for Use Statement

DEC 1 0 2010

4103247 510(k) Number (if known):

VascuComp™-01 and VascuComp™-02 Device Name:

Indications for Use:

VascuComp™ AI Mode:

The VascuComp™ compression system in AI Mode is used as an adjunct therapy for patients with ischemic disease of the lower extremities, due to one or more of the following causes:

  • . Minor Amputations
  • Angioplasty / Stent Failure .
  • Arteriopathic Wounds .
  • Graft Failure .
  • Intermittent Claudication .
  • Ischemia .
  • Night Pain .
  • Rest Pain .
  • Small Vessel Discase
  • · · Ulcers

VascuComp™ DVT Mode:

  • . Decrease the risk of deep venous thrombosis (DVT).
  • Aids the blood flow back to the heart. .
  • . Treat and assist healing of cutaneous ulceration (wounds), reduce wound healing time, enhance arterial circulation (blood flow), reduce compartmental pressures, reduce edema (swelling), reduce the need for anticoagulant (blood thinning) medications.

VascuComp™ L Mode:

  • Treatment of Venous disorders and dysfunction of the "muscle pump". .

{12}------------------------------------------------

Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE of NEEDED) .

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 2 of 2

Duma R. Vimes

:

(Division Sign-Off) (Division of Cardiovascular Devices

510(k) Number_KIO 3247

5

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).