K Number
K103207

Validate with FDA (Live)

Date Cleared
2011-05-20

(200 days)

Product Code
Regulation Number
866.3500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Q Fever Detection Kit is a qualitative real-time polymerase chain reaction (PCR) test kit intended to identify and detect target DNA sequence from Coxiella burnetii in serum collected from individuals suspected of having acute Q fever, typically 7-10 days after onset of symptoms or before antibody formation. This in vitro diagnostic (IVD) test is intended to aid in the diagnosis of Q fever in individuals presenting with signs and symptoms of acute Q fever when used in conjunction with other clinical and laboratory findings. This kit is only intended to aid in the diagnosis of Q fever of patients presenting in the acute stage of the disease. Negative results do not preclude C. burnetii infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.

The JBAIDS Q Fever assay is run on the JBAIDS instrument using the Diagnostic Wizard. Results are for the presumptive identification of C. burnetii in conjunction with serology and/or other laboratory tests. The following considerations also apply:

  • The diagnosis of acute Q fever must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of C. burnetii from serum specimens.
  • Sensitivity is decreased by ~25-40%, with no change in specificity, if the sample is collected after the patient has formed specific antibodies to C. burnetii, typically 7-10 days after onset of symptoms.
  • The definitive identification of C. burnetii from serum specimens requires additional testing and confirmation procedures in consultation with public health or other authorities for whom reports are required.
Device Description

The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Q Fever Detection Kit is a fully integrated in-vitro diagnostic (IVD) system composed of the JBAIDS instrument with laptop computer, software, a freeze-dried reagent assay for the qualitative detection of pathogenic Coxiella burnetii, and 2 different sample preparation protocols for isolating target DNA from serum. Use of the JBAIDS DNA Extraction Control kit (not included) is also recommended.

The JBAIDS instrument, using Polymerase Chain Reaction (PCR) technology, is a portable thermocycler and real-time fluorimeter. The JBAIDS Q Fever Detection Kit is specially designed for PCR in glass capillaries using the JBAIDS instrument and hydrolysis probes for sequence-specific detection of the C. burnetii IS1111 DNA target.

The reagent kit contains 3 different types of freeze-dried reagent vials: Positive Controls, Negative Controls, and Unknowns (used for testing the patient sample). Each JBAIDS assay requires a Positive and Negative Control, and each sample is tested using an Unknown reagent vial which contains a multiplexed target and inhibition control (IC) assay.

Prior to testing, serum samples are purified using the Idaho Technology IT 1-2-3™ QFLOW or IT 1-2-3™ Platinum Path Sample Purification Kit. The resulting purified sample is added to an Unknown reagent vial, along with reconstitution buffer. A Positive Control and a Negative Control vial are prepared using reconstitution buffer and reagent grade water. Aliquots from each reagent vial are transferred to 2 reaction capillaries that are tested together in the JBAIDS instrument. The instrument is programmed to perform heating and cooling cycles that drive the PCR process. The heating and cooling cycles are generated using a heating coil and varying fan speeds. Fluorescence emission is monitored over 1 of 3 wavelengths, and the instrument software interprets the change in fluorescence to determine whether the target DNA is present.

When the organism is present, a fragment of C. burnetii DNA is amplified using specific primers. The amplicon is detected by fluorescence using a specific hydrolysis probe. The hydrolysis probe contains a short oligonucleotide that hybridizes to an internal sequence of the amplified fragment during the annealing phase of the PCR cycle. This probe has the 5' and 3' ends labeled with a reporter dye and a quenching dye, respectively. When the probe hybridizes to the specific DNA target, the Taq polymerase enzyme replicating the targetspecific DNA hydrolyzes the probe, which separates the two fluorophores, thus allowing the reporter dye to fluoresce. Fluorescence is monitored at a wavelength of 530 nm. Inhibition is monitored using an internal inhibition control. The inhibition control consists of a linearized plasmid containing an artificial intervening sequence flanked by target assay primer sequences. Similar to the target, the IC probe also has the 5' and 3' ends labeled with a reported and quenching dye, respectively. Hydrolysis of the IC probe during amplification is monitored at a wavelength of 705 nm.

The level of fluorescence from each unknown sample and control is measured by the JBAIDS instrument. JBAIDS Software analyzes fluorescence amplification curves and reports results as Positive, Negative, Inhibited, or Uncertain. A failure of the Positive or Negative Control will result in the entire run being called Invalid. Failure of the Inhibition Control when no target amplification is observed yields an Inhibited result for the associated sample and requires retesting of that sample. A positive result for the target will override an inhibited result.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the JBAIDS Q Fever Detection Kit, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal "acceptance criteria" in a quantitative, pre-defined manner for the clinical study. Instead, it presents performance characteristics (sensitivity and specificity) from a clinical trial. The analytical performance (LoD and exclusivity) directly reports observed results without listing specific target criteria.

Performance MetricAcceptance Criteria (Not explicitly stated as formal "acceptance criteria" in the doc, but implied by data presentation)Reported Device Performance (JBAIDS Q Fever Detection Kit)
Analytical Studies
• Limit of Detection (LoD)Able to yield positive results for target concentration (Implied: High detection rate at LoD)- 20/20 (100%) positive with JBAIDS Q Fever assay for serum samples spiked with 10 TCID50/mL live C. burnetii (using both IT 1-2-3™ Platinum Path and IT 1-2-3™ QFLOWd Purification Kits).- 10/10 (100%) of isolates in C. burnetii inclusivity panel correctly identified at determined system LoD.
• Exclusivity (Specificity)No cross-reactivity with phylogenetically related or commonly found organisms.- 22 of 22 non-C. burnetii strains tested in exclusivity panel were negative at high concentration.- In silico evaluation of Ehrlichia and Neorickettsia indicated no cross-reactivity.
• ReproducibilityHighly reproducible test results at or above LoD (Implied: High agreement with expected positive results; low %CV for Cp values).- Agreement with Expected Positive Results (Detection ≥ LoD): 100% (204/204) across all sites and both purification kits (95% CI: 98.2-100.0%).- Agreement at LoD/15 (low concentration): 52% (53/102) across all sites and both purification kits (95% CI: 41.8-62.0%).- Average Cp %CV: Low %CV values for 5X LoD, LoD, and LoD/15, indicating consistent Cycle threshold values. Percent CV for 5X LoD and LoD was generally below 3%.
Clinical Studies
• Clinical SensitivityDetect C. burnetii in samples from acute Q fever patients (Implied: Reasonable detection rate, especially prior to antibody formation).For samples prior to seroconversion (Acute Q Fever, Seronegative):- Site 1: Platinum Path: 47% (9/19; 95% CI=24-71%); QFLOWdna: 29% (5/17; 95% CI=10-56%).- Site 2: Platinum Path: 77% (20/26; 95% CI=56-91%); QFLOWdna: 81% (22/27; 95% CI=62-94%). (Note: Sensitivity decreased significantly in seropositive samples).
• Clinical SpecificityNo false positives in samples without C. burnetii (Implied: High negative predictive value).For serology negative samples (no antibody to C. burnetii):- Site 1: Platinum Path: 100% (40/40; 95% CI=91-100%); QFLOWdna: 100% (39/39; 95% CI=91-100%).- Site 2: Platinum Path: 100% (71/71; 95% CI=95-100%); QFLOWdna: 100% (72/72; 95% CI=95-100%).- Overall, 306/307 (99.7%) of serology negative samples gave expected negative results. One false positive was observed from a sample that failed to show a four-fold antibody titer rise.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Description: Banked, frozen serum specimens.
  • Data Provenance: The specimens were sequentially received during a specified time period. The clinical study involved three independent laboratories located in distinct geographical regions: Australia, France, and the Netherlands. This indicates the data is retrospective.
  • Total Subjects Evaluated: 749 subjects.
  • Subjects Included in Analysis: 465 subjects.
  • Exclusions:
    • 86 for inconclusive or incomplete serology.
    • 21 due to environmental contamination.
    • 7 for inconclusive JBAIDS results.
    • All 170 samples tested at Site 3 due to deviations from the study protocol.
  • Total Final Test Results (after purification):
    • Platinum Path: 324 results (170 from serology positive, 154 from serology negative).
    • QFLOWdna: 324 results (171 from serology positive, 153 from serology negative).
      (Note: 183 specimens had adequate volume for purification by both methods, contributing to the total results.)
  • Breakdown for Sensitivity/Specificity Calculation:
    • Seronegative, No Antibody to C. burnetii (Specificity):
      • Site 1: 40 (Platinum Path) + 39 (QFLOWdna) = 79 samples
      • Site 2: 71 (Platinum Path) + 72 (QFLOWdna) = 143 samples
      • Total: 222 samples
    • Seronegative, Failed 4-fold Antibody Rise (where 1 false positive occurred): 85 samples
    • Seropositive (Sensitivity, where seroconversion occurred):
      • Site 1: 19 (Platinum Path) + 17 (QFLOWdna) = 36 samples
      • Site 2: 26 (Platinum Path) + 27 (QFLOWdna) = 53 samples
      • Total: 89 samples (for samples exhibiting seroconversion)

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth for the clinical study was established by standard paired serological testing for Q fever, which had been previously performed on the banked specimens.
The document does not specify the number or qualifications of experts who performed or interpreted this serological testing. It refers to "standard paired serological testing," implying established laboratory procedures.

4. Adjudication Method for the Test Set

The document does not describe an explicit "adjudication method" involving independent review of results for the clinical study. The ground truth was based on the outcome of "standard paired serological testing." The JBAIDS results were compared against this serological ground truth. There is no mention of multiple experts reviewing individual cases or a conciliation process if disagreements arose.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done.
The JBAIDS Q Fever Detection Kit is a real-time PCR assay, an in vitro diagnostic (IVD) device, not an AI-powered diagnostic imaging or interpretation system that would typically involve human "readers." The device provides automated test interpretation and report generation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance evaluation was done. The JBAIDS Q Fever Detection Kit is described as a "fully integrated in-vitro diagnostic (IVD) system" with "automated test interpretation and report generation." The clinical and analytical studies measure the performance of this system operating in a standalone capacity, producing "Positive, Negative, Inhibited, or Uncertain" results based on fluorescence amplification curves. While human operators are involved in sample preparation and loading, the interpretation of the PCR data is handled by the instrument's software algorithms, without human interpretative input into individual test results.

7. The Type of Ground Truth Used

The ground truth used for the clinical study was based on standard paired serological testing for Q fever.
Specifically, for establishing sensitivity, it relied on samples exhibiting a four-fold rise in antibodies (indicating seroconversion, a definitive sign of infection). For establishing specificity, it relied on samples with no antibody to C. burnetii confirmed by serology.

8. The Sample Size for the Training Set

The document does not specify the sample size for a training set. As this is a molecular diagnostic kit (PCR), the "training" would typically involve optimization and validation of primers, probes, and reaction conditions rather than machine learning on a dataset with ground truth. The performance data presented (LoD, exclusivity, reproducibility, clinical studies) are evaluations of the finalized kit.

9. How the Ground Truth for the Training Set was Established

Since no explicit "training set" for an algorithm is mentioned or implied, the question of how its ground truth was established is not applicable in the typical sense of machine learning. The kit's design and analytical performance (e.g., selection of primers and probes for C. burnetii IS1111 DNA sequences) would have been developed against known strains and samples, but these are part of assay development and analytical validation, not a separate "training set" with ground truth in the context of an AI/ML device.

{0}------------------------------------------------

K103207

いつきのですが、そのためになるところです。 ときになると、いつきはあり、ありましていたい

MAY 2 0 2011

510(k) Summary 1 19949

510(k) Summary Idaho Technology Inc. JBAIDS Q Fever Detection Kit

Introduction: According to the requirements of 21 CFR 807.92, the following informationprovides sufficient detail to understand the basis for a determination ofsubstantial equivalence.
Submittedby:Idaho Technology Inc.390 Wakara WaySalt Lake City, UT 84108Telephone: 801-736-6354Facsimile: 801-588-0507Contact Person: Beth Lingenfelter, ext. 407Date Prepared: Oct. 29, 2010
DeviceName:Trade Name:JBAIDS Q Fever Detection KitCommon Name:Real-time PCR amplification and detection system for targeted Coxiellaburnetii DNA sequences

Classification Name:

Reagent Kit: Rickettsia serological reagents (CFR 866.3500)

:
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

{1}------------------------------------------------

The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Q Device Description: Fever Detection Kit is a fully integrated in-vitro diagnostic (IVD) system · composed of the JBAIDS instrument with laptop computer, software, a freezedried reagent assay for the qualitative detection of pathogenic Coxiella burnetii, and 2 different sample preparation protocols for isolating target DNA from serum. Use of the JBAIDS DNA Extraction Control kit (not included) is also recommended.

The JBAIDS instrument, using Polymerase Chain Reaction (PCR) technology, is a portable thermocycler and real-time fluorimeter. The JBAIDS Q Fever Detection Kit is specially designed for PCR in glass capillaries using the JBAIDS instrument and hydrolysis probes for sequence-specific detection of the C. burnetii IS1111 DNA target.

The reagent kit contains 3 different types of freeze-dried reagent vials: Positive Controls, Negative Controls, and Unknowns (used for testing the patient sample). Each JBAIDS assay requires a Positive and Negative Control, and each sample is tested using an Unknown reagent vial which contains a multiplexed target and inhibition control (IC) assay.

Prior to testing, serum samples are purified using the Idaho Technology IT 1-2-3™ QFLOW or IT 1-2-3™ Platinum Path Sample Purification Kit. The resulting purified sample is added to an Unknown reagent vial, along with reconstitution buffer. A Positive Control and a Negative Control vial are prepared using reconstitution buffer and reagent grade water. Aliquots from each reagent vial are transferred to 2 reaction capillaries that are tested together in the JBAIDS instrument. The instrument is programmed to perform heating and cooling cycles that drive the PCR process. The heating and cooling cycles are generated using a heating coil and varying fan speeds. Fluorescence emission is monitored over 1 of 3 wavelengths, and the instrument software interprets the change in fluorescence to determine whether the target DNA is present.

When the organism is present, a fragment of C. burnetii DNA is amplified using specific primers. The amplicon is detected by fluorescence using a specific hydrolysis probe. The hydrolysis probe contains a short oligonucleotide that hybridizes to an internal sequence of the amplified fragment during the annealing phase of the PCR cycle. This probe has the 5' and 3' ends labeled with a reporter dye and a quenching dye, respectively. When the probe hybridizes to the specific DNA target, the Taq polymerase enzyme replicating the targetspecific DNA hydrolyzes the probe, which separates the two fluorophores, thus allowing the reporter dye to fluoresce. Fluorescence is monitored at a wavelength of 530 nm. Inhibition is monitored using an internal inhibition control. The inhibition control consists of a linearized plasmid containing an artificial intervening sequence flanked by target assay primer sequences. Similar to the target, the IC probe also has the 5' and 3' ends labeled with a reported and quenching dye, respectively. Hydrolysis of the IC probe during amplification is monitored at a wavelength of 705 nm.

{2}------------------------------------------------

The level of fluorescence from each unknown sample and control is measured by the JBAIDS instrument. JBAIDS Software analyzes fluorescence amplification curves and reports results as Positive, Negative, Inhibited, or Uncertain. A failure of the Positive or Negative Control will result in the entire run being called Inyalid. Failure of the Inhibition Control when no target amplification is observed yields an Inhibited result for the associated sample and requires retesting of that sample. A positive result for the target will override an inhibited result.

The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Q Intended Use: Fever Detection Kit is a qualitative real-time polymerase chain reaction (PCR) test kit intended to identify and detect target DNA sequence from Coxiella burnetii in serum collected from individuals suspected of having acute Q fever, typically 7-10 days after onset of symptoms or before antibody formation. This in vitro diagnostic (IVD) test is intended to aid in the diagnosis of Q fever in individuals presenting with signs and symptoms of acute Q fever when used in conjunction with other clinical and laboratory findings. This kit is only intended to aid in the diagnosis of Q fever of patients presenting in the acute stage of the disease. Negative results do not preclude C. burnetii infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.

The JBAIDS O Fever assay is run on the JBAIDS instrument using the Diagnostic Wizard. Results are for the presumptive identification of C. burnetii in conjunction with serology and/or other laboratory tests. The following considerations also apply:

  • The diagnosis of acute O fever must be made based on history, signs, . symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of C. burnetii from serum specimens.
  • Sensitivity is decreased by ~25-40%, with no change in specificity, if . the sample is collected after the patient has formed specific antibodies to C. burnetii, typically 7-10 days after onset of symptoms.
  • The definitive identification of C. burnetii from serum specimens . requires additional testing and confirmation procedures in consultation with public health or other authorities for whom reports are required.

The test performance characteristics for this system were established with banked, frozen serum specimens that were sequentially received during a specified time period. The safety and effectiveness of other types of tests or sample types have not been established.

All users, analysts, and any person reporting diagnostic results from use of this device should be trained to perform and interpret the results from this procedure by JBAIDS instructors or designees prior to use. Use of this device is limited to designated Department of Defense (DoD) laboratories equipped with the JBAIDS instruments.

{3}------------------------------------------------

SubstantialEquivalence:The JBAIDS Q Fever Detection System is substantially equivalent to otherproducts in commercial distribution intended for similar use. The JBAIDSinstrument has been previously cleared under K051713.
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

The predicate for the JBAIDS Q Fever Detection Kit is the Focus Diagnostics IgM and IgG IFA tests cleared under K922374 and K913906. The following table compares the JBAIDS Q Fever Detection System with the predicate test.

Table 1. Similarities and Differences between the JBAIDS Q Fever Detection Kit and the Focus Diagnostics
IFA tests
Company of Children Children Children Children Children Children
ELEMENTJBAIDS Q Fever Detection KitFocus Diagnostics Q Fever IgG and IgM tests
Intended UseQualitative detection of C. burnetiiDetection and semi-quantitation of the humanIgG and IgM antibody response to phase Iand phase IICoxiella burnetii antigens and as an aid in thediagnosis of Q fever.
Indications for UseIdentification of C. burnetii in individualssuspected of having Acute Q Fever.Serodiagnosis of both acute and chronic C.burnetii infections.
TechnologicalPrinciplesReal-time PCR using hydrolysis probes.Microscopic visualization of humanantibodies for C. burnetii bacteria via a twostage "sandwich" procedure using afluorescently labeled antibody that binds tothe human antibodies.
Assay TargetIS1111 DNA sequences unique to C. burnetii.Human antibody response to C. burnetii.
Specimen TypesSerumSame
InstrumentationJBAIDS instrument (K051713)Fluorescent microscope
Time Required forAnalysis ofSpecimenLess than 3 hoursSame
Sample PreparationMethodUp front sample processing is required toextract nucleic acid.Diluted samples are tested directly.
Sample ControlsInternal control is multi-plexed with the targettest.No sample specific control.
Testing strategyOne time point testing during the symptomaticphase of the disease.Two serology time points are suggested. Oneduring the symptomatic phase of the diseasewith follow-up testing 2-3 weeks later.
Optimal Window ofDetectionEarly in the disease course prior to antibodyformation.Later in the disease course due to the timerequired for antibody formation.
Test InterpretationAutomated test interpretation and reportgeneration.Subjective interpretation by user.
Physical PropertiesFreeze dried reagents with reconstitutionbuffer and water provided in kit.Liquid reagents. IgM and IgG are separatetest kits.
StorageRoom temperature (18-28°C)Refrigerator temperature (2-8°C)

The JBAIDS Q Fever Detection System is intended for the qualitative IVD detection of target DNA sequence from the C. burnetii pathogen. The system can be used to test human serum. The results from the PCR tests are used in conjunction with serology and

{4}------------------------------------------------

other laboratory tests and clinical information as an aid in the diagnosis of systemic Q Fever infection in individuals suspected of having the disease.

The predicate device and the JBAIDS system have a similar intended use; they both provide test results that aid in the diagnosis of O Fever when considered with other clinical and laboratory evidence.

Performance Data

LoD

The JBAIDS Q Fever Detection System vielded positive results 20/20 (100%) with the JBAIDS Q Fever assay for serum samples spiked with the limit of detection level (10 TCID50 mL) of live C. burnetii for both the IT 1-2-37M Platinum Path and IT 1-2-3TM QFLOWand purification kits. In addition 10/10 (100%) of the isolates (see Table 2) included in the C. burnetii inclusivity panel were correctly identified by the JBAIDS assay the determined system LoD.

Strain IDGroup
Nine Mile Phase II
QiYiI
RSA334I
Henzerling Phase I SalkII
M44 – GritaII
Idaho QIII
Q173-PIV
Q154-kAVIV
WAVV
Q229V

Exclusivity

The analytic specificity evaluation of the JBAIDS Q Fever Detection System was conducted with organisms that are phylogenetically related to C. burnetii, as well as with unrelated organisms that are likely to be found in clinical samples.

The JBAIDS Q Fever Detection System assays also proved to be very specific.

  • . 22 of 22 non-C. burnetii strains (see Table 3) tested in the exclusivity panel were negative when tested at high concentration.
  • . 2 of 2 phylogenetically related genera. Ehrlichia and Neorickettsia, were evaluated in silico. This evaluation indicated that the primers and probes in the JBAIDS Q Fever assay would not cross-react with organisms from these genera.

{5}------------------------------------------------

OrganismRelevanceConc. Tested CFU/mL
Legionella pneumophilaNearest Neighbor$10^6$
Legionella longbeacheaeNearest Neighbor$10^6$
Legionella micdadeiNearest Neighbor/Cross reactive with serology$10^6$
Bartonella henselaeCross reactive with serology$10^6$
Rickettsia prowazekiiOther small obligate intracellular bacteria, agents of zoonoses$10^6$
Mycobacterium tuberculosisSimilar symptoms$10^6$
Brucella meliteusisSimilar symptoms4.2 x $10^4$
Orientia tsutsugamushiSimilar symptoms$10^6$
Francisella tularensisSimilar symptoms4.5 x $10^5$
Salmonella entericSimilar symptoms$10^6$
Listeria monocytogenesSimilar symptoms$10^6$
Haemophilus influenzaeSimilar symptoms$10^6$
Streptococcus pyogenesCommonly encountered$10^6$
Aggregatibacter actinomycetemcomitansCommonly encountered$10^6$
Haemophilus parainfluenzaeCommonly encountered7.8 x $10^4$
Acinetobacter baumanniiCommonly encountered$10^6$
Vibrio cincinnatiensisCommonly encountered$10^6$
Escherichia coliCommonly encountered$10^6$
Staphylococcus aureusCommonly encountered$10^6$
Staphylococcus epidermidisCommonly encountered$10^6$
Pseudomonas aeruginosaCommonly encountered$10^6$
Serratia marcescensCommonly encountered$10^6$

.

Table 3, Exclusivity Panel

ldaho Technology Inc. 510(k) JBAIDS Q Fever Detection Kit

: : :

{6}------------------------------------------------

Reproducibility

A multicenter reproducibility study was conducted in which a panel of specimens was tested twice a day for five days at two test sites and for seven days at a third test site. Results are summarized in Tables 4 and 5, and demonstrate the JBAIDS Q Fever Detection kit provides highly reproducible test results when testing at or above the established system LoD.

Table 4. Reproducibility Study Summary (Agreement with Expected Positive Results) for the JBAIDS Q Fever
Detection Kit
Test LevelIT 1-2-3 Platinum Path SamplePurification KitIT 1-2-3 QFLOWdna SamplePurification KitBothPurificationKits, AllSites95% CI
Site 1Site 2Site 3All SitesSite 1Site 2Site 3All Sites
5X LoD21/2115/1515/1551/51100%21/2115/1515/1551/51100%102/102100%96.5-100.0
LoD21/2115/1515/1551/51100%21/2115/1515/1551/51100%102/102100%96.5-100.0
LoD/1511/218/1513/1532/5163%7/218/156/1521/5141%53/10252%41.8-62.0
Detection ≥LoD42/42100%30/30100%30/30100%102/102100%42/42100%30/30100%30/30100%102/102100%204/204100%98.2-100.0
Detection allLevels53/6384%38/4584%43/4596%134/15388%49/6378%38/4584%36/4580%123/15380%257/30684%79.4 -87.9

Table 5. Reproducibility Study Summary (Average Cp and %CV) for the JBAIDS Q Fever Detection Kit

TestLevelIT 1-2-3 Platinum Path Sample Purification KitIT 1-2-3 QFLOWdna Sample Purification Kit
Test LocationTest LocationTest LocationAll SitesTest LocationTest LocationTest LocationAll Sites
Ave.Cp%CVAve.Cp%CVAve.Cp%CVAve.Cp%CVAve.Cp%CVAve.Cp%CVAve.Cp%CVAve.Cp%CV
5XLoD30.942.530.812.430.301.330.712.331.531.231.251.331.531.331.451.3
LoD32.732.032.541.832.422.532.582.133.351.633.181.433.531.133.351.4
LoD/1535.271.235.143.234.942.135.112.335.362.335.333.135.331.635.342.4

Note: Only the results for positive capillaries are included.

Clinical Studies

In addition to analytic studies, a multisite clinical trial was conducted.

The clinical performance of the JBAIDS Q Fever Detection Kit was performed using banked, frozen serum specimens that were sequentially received during a specified time period for which standard paired serological testing for Q fever had been previously performed. Three independent laboratories (located in distinct geographical regions; Australia, France and the Netherlands) were included in this

:

{7}------------------------------------------------

study. Serum specimens from a total of 749 subjects were evaluated in the study, however 284 were excluded (86 for inconclusive or incomplete serology, 21 due to environmental contamination, 7 for inconclusive JBAIDS results and all 170 samples tested at site 3 due to deviations from the study protocol). Table 6 provides a summary of demographic information for the 465 subjects included in this study.

SiteSite 1Site 2
Number of Subjects303162
SexFemale (%)43%31%
Male (%)57%69%
AgeMean5045
Min75
Max8782
Specimen Collection Period4/14/2008-10/15/200912/13/2005-6/26/2010

Table 6. Demographic summary for JBAIDS Q Fever clinical study

Prior to testing with the JBAIDS Q Fever Detection Kit, serum samples were purified using either the IT 1-2-3 Platinum Path or the IT 1-2-3 QFLOWand sample purification kit. Out of the 465 specimens, 183 specimens had adequate volume to be purified using both purification methods. This resulted in a total of 324 final test results for Platinum Path (170 from serology positive specimens and 154 from serology negative specimens) and 324 final test results for QFLOW (171 from serology positive specimens and 153 from serology negative specimen).

Of the 307 tests of serology negative samples, 222 had no antibody to C. burnetti and 85 failed to show a 4-fold rise in antibody titer between the paired acute and convalescence samples. All but one serology negative sample, (306/307, 99.7%) gave the expected negative result when tested with the JBAIDS Q Fever Detection Kit. The one false positive was obtained from one of the samples that failed to show a fourfold rise in antibody titer.

Of the 252 tests on samples that were serology positive based upon a four-fold rise in antibodies, the JBAIDS Q Fever Detection system correctly identified the presence of C. burnetii in 17.8% (45/252) samples. However, the positive JBAIDS test result rate for those samples which exhibited seroconversion was significantly higher at 62.9% (56/89). These data are consistent with several published reports indicating that PCR detection of C. burnetii declines significantly once specific antibodies are detected by serology.

Clinical sensitivity ranged from 29-81% (variation between sites) on specimens collected prior to seroconversion. As shown in Table 7, the clinical sensitivity for the JBAIDS Q Fever Detection Kit was different at the two study sites. Site 1 had a

{8}------------------------------------------------

lower detection rate (29% for specimens purified using the QFLOWd00 extraction kit and 47% for specimens purified using the Platinum Path purification kit) when compared to the detection rate at site 2 (81% for specimens purified using the QFLOWdna extraction kit and 77% for specimens purified using the Platinum Path purification kit). The differences observed at the two study sites may be the result of differences in organism strains, patient population or patient management. The clinical specificity was 100% at both sites for serology negative samples that had no antibody to C. burnetii.

Table 7. Sensitivity and Specificity for the JBAIDS Q Fever Detection Kit in Samples Tested for Acute Q Fever that were Seronegative

SitePurification KitTP/(TP + FN)SensitivityPercentTN/(TN + FP)SpecificityPercent
1Platinum Path9/1947%(95% CI=24-71%)40/40100%(95% CI=91-100%)
1QFLOWdna5/1729%(95% CI=10-56%)39/39100%(95% CI=91-100%)
2Platinum Path20/2677%(95% CI=56-91%)71/71100%(95% CI=95-100%)
2QFLOWdna22/2781%(95% CI=62-94%)72/72100%(95% CI=95-100%)

Idaho Technology Inc. 510(k) JBAIDS Q Fever Detection Kit

...

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or other bird with outstretched wings.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Idaho Technology, Inc. c/o Beth Lingenfelter Director of Regulated Products 390 Wakara Way Salt Lake City, Utah 84108

MAY 20 2011

Re: K103207

Trade/Device Name: JBAIDS Q Fever Detection Kit Regulation Number: 21 CFR 866.3500 Regulation Name: Rickettsia serological reagents Regulatory Class: Class I Product Code: OVF Dated: May 17, 2011 Received: May 18, 2011

Dear Ms. Lingenfelter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

{10}------------------------------------------------

Page 2 - Beth Lingenfelter

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Hay aAgus

Sally A. Hojvat, M.Sc. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K103207 Device Name: JBAIDS O Fever Detection Kit

The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Q Fever Detection Kit is a qualitative real-time polymerase chain reaction (PCR) test kit intended to identify and detect target DNA sequence from Coxiella burnetii in serum collected from individuals suspected of having acute Q fever, typically 7-10 days after onset of symptoms or before antibody formation. This in vitro diagnostic (IVD) test is intended to aid in the diagnosis of O fever in individuals presenting with signs and symptoms of acute Q fever when used in conjunction with other clinical and laboratory findings. This kit is only intended to aid in the diagnosis of Q fever of patients presenting in the acute stage of the disease. Negative results do not preclude C. burnetii infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.

The JBAIDS Q Fever assay is run on the JBAIDS instrument using the Diagnostic Wizard. Results are for the presumptive identification of C. burnetii in conjunction with serology and/or other laboratory tests. The following considerations also apply:

  • The diagnosis of acute Q fever must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of C. burnetii from serum specimens.
  • i Sensitivity is decreased by ~25-40%, with no change in specificity, if the sample is collected after the patient has formed specific antibodies to C. burnetii, typically 7-10 days after onset of symptoms.
  • .. The definitive identification of C. burnetii from serum specimens requires additional testing and confirmation procedures in consultation with public health or other authorities for whom reports are required.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Ludolie K. Poole

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)_ R103267

{12}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K103207 Device Name: JBAIDS Q Fever Detection Kit

The test performance characteristics for this system were established with banked, frozen serum specimens that were sequentially received during a specified time period. The safety and effectiveness of other types of tests or sample types have not been established.

All users, analysts, and any person reporting diagnostic results from use of this device should be trained to perform and interpret the results from this procedure by JBAIDS instructors or designees prior to use. Use of this device is limited to designated Department of Defense (DoD) laboratories equipped with the JBAIDS instruments.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Freddie L. Poole

:

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 183207 510(k)

§ 866.3500 Rickettsia serological reagents.

(a)
Identification. Rickettsia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rickettsia in serum. Additionally, some of these reagents consist of rickettsial antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify rickettsia directly from clinical specimens. The identification aids in the diagnosis of diseases caused by virus-like bacteria belonging to the genusRickettsiae and provides epidemiological information on these diseases. Rickettsia are generally transmitted by arthropods (e.g., ticks and mosquitoes) and produce infections in humans characterized by rash and fever (e.g., typhus fever, spotted fever, Q fever, and trench fever).(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.