K Number
K103145

Validate with FDA (Live)

Date Cleared
2011-02-09

(107 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To provide fixation of soft-tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL) and/or posterior cruciate ligament (PCL) reconstruction.

Device Description

The Biomet Sports Medicine TunneLoc™ Tibial Fixation Device is a non-resorbable intratunnel implant intended to aid in arthroscopic ACL and/or PCL reconstructions. The inserter instrument provides a means to apply and hold tension to the soft tissue, align and drive the PEEK implant. A nitinol quide wire instrument is provided to aid in implant placement. The TunneLoc™ Tibial Fixation Devices provide the surgeon with a means to aid in the management of soft tissue to bone reattachment procedures.

AI/ML Overview

This document describes the TunneLoc™ Tibial Fixation Device, a non-resorbable intratunnel implant for ACL and/or PCL reconstructions. The device is intended for soft tissue fixation within the tibial tunnel.

Here is an analysis of the provided information regarding acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Equivalent pullout strength to predicate devices"The test results indicate that the TunneLoc™ Tibial Fixation Devices provide equivalent pullout strength to the predicate devices and would be functional within their intended use."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample size (number of devices, test repetitions) for the non-clinical pullout tests.

  • Data Provenance: The tests were conducted using porcine tibia and bovine tendon test medium. This indicates an in vitro or ex vivo setting, not human data. The country of origin of the data is not specified, but the applicant is based in Indiana, USA. The testing is retrospective in the sense that it evaluates the device after its design, but not on human patients.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • This information is not provided. Non-clinical mechanical tests typically rely on established engineering standards and methodologies for "ground truth" (e.g., force transducers, load cells, material properties testing machines), rather than human expert consensus.

4. Adjudication Method for the Test Set

  • This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are typically used for clinical studies involving human interpretation or subjective assessments. Mechanical testing follows predefined protocols and measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical Testing: None provided as a basis for substantial equivalence."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This question is not applicable. The TunneLoc™ Tibial Fixation Device is a medical implant, not an algorithm or AI system. Its performance is evaluated through mechanical testing, not algorithmic output.

7. The Type of Ground Truth Used

  • The ground truth for the non-clinical testing was established through objective measurement of pullout strength using force testing equipment. This relies on the physical properties of the materials and the mechanical forces applied, rather than expert consensus, pathology, or outcomes data, which are relevant for clinical studies or diagnostic devices.

8. The Sample Size for the Training Set

  • This question is not applicable. This is a medical device, not an AI or machine learning system that requires a "training set."

9. How the Ground Truth for the Training Set was Established

  • This question is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Biomet Sports Medicine. The logo consists of the word "BIOMET" in a bold, blocky font, with a small registered trademark symbol to the right of the "T". Below the word "BIOMET" is the phrase "SPORTS MEDICINE" in a smaller, sans-serif font.

FEB - 9 2011

510(k) Summary

Preparation Date:February 9, 2010
Applicant/Sponsor:Biomet Sports Medicine
Contact Person:Robert R. FriddleRegulatory Affairs Specialist
Proprietary Name:TunneLoc™ Tibial Fixation Device
Common Name:Soft tissue fixation device
Classification Name:Fastener, Fixation, Nondegradable, Soft Tissue (21CFR§888.3040) MBIScrew, Fixation, Bone (21CFR §888.3040) HWC

Legally Marketed Devices To Which Substantial Equivalence Is Claimed:

K041274Resorbable Interference Screw
K982497Arthrotek Interference Screw
K083607AperFix™ Tibial Implant with Inserter
K983560Intratunnel Tibial Fixation Fastener

Device Description:

The Biomet Sports Medicine TunneLoc™ Tibial Fixation Device is a non-resorbable intratunnel implant intended to aid in arthroscopic ACL and/or PCL reconstructions. The inserter instrument provides a means to apply and hold tension to the soft tissue, align and drive the PEEK implant. A nitinol quide wire instrument is provided to aid in implant placement. The TunneLoc™ Tibial Fixation Devices provide the surgeon with a means to aid in the management of soft tissue to bone reattachment procedures.

Intended Use:

The TunneLoc™ Tibial Fixation Device is intended for soft tissue fixation for the following indications:

To provide fixation of soft-tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL) and/or posterior cruciate ligament (PCL) reconstruction.

Mailing Address:
P.O. Box 587
Warsaw, IN 46581-0587
Toll Free: 800.348.9500
Office: 574.267.6639
Main Fax: 574.267.8137
www.biomet.com

Shipping Address:
56 East Bell Drive
Warsaw, IN 46582

{1}------------------------------------------------

K103145(²/₂)

Summary of Technologies:

The TunneLoc™ Implant design includes technological characteristics and indications similar or identical to the predicate devices. The conical shell with blunt wedge tip is similar to the tapered sheath of the predicate Intratunnel Tibial Fixation Fastener (K983560), while the cannulation and insertion over a quide wire are similar to the predicate Gentle Threads™ (K041274). The TunneLoc™ implant is composed of PEEK material like the predicate AperFix™ (K083607). Dimensional characteristics of a 7-10mm diameter range is available in the predicate Gentle Threads™ (K041274) and 30mm length available in all predicates. The TunneLoc™ inserter instrument provides a means to apply and hold tension to the soft tissue graft prior to and during implant insertion similar to the predicate Intratunnel Tibial Fixation Fastener (K983560) and AperFix™ (K083607) Instruments.

Non-Clinical Testing:

Non-clinical laboratory testing was performed to verify the fixation strength of the TunneLoc™ Tibial Fixation Devices in pullout tests using porcine tibia and bovine tendon test medium as compared to the predicate devices for specific indications for use. The test results indicate that the TunneLoc™ Tibial Fixation Devices provide equivalent pullout strength to the predicate devices and would be functional within their intended use.

Clinical Testing:

None provided as a basis for substantial equivalence.

All trademarks are the property of Biomet, Inc. except AperFix™ which is a trademark of Cayenne Medical, Inc.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake coiled around it, and three lines extending from the top of the staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

Biomet Sports Medicine % Mr. Robert Friddle Regulatory Specialist 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0587

FEB 0 9 2011

Re: K103145

Trade/Device Name: TunneLoc™ Tibial Fixation Devices Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, HWC Dated: February 4, 2011 Received: February 7, 2011

Dear Mr. Friddle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Ating B. Rata
for

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): كل 103146)5

Device Name: TunneLoc™ Tibial Fixation Device

Indications For Use:

To provide fixation of soft-tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL) and/or posterior cruciate ligament (PCL) reconstruction.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

(Division Sign-Off)
for M.Mellors

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K103145

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.