K Number
K103120

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-02-28

(129 days)

Product Code
Regulation Number
888.3353
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Corin BIOLOX delta Modular Femoral Heads are used in combination with femoral hip stems and ultra high molecular weight polyethylene (UHMWPE) acetabular cup liners to reinstate function following the degenerative effects of:

  • Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
  • Rheumatoid arthritis
  • Correction of functional deformity and
  • Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH)
    The Corin BIOLOX delta Modular Femoral Heads are intended for single use only.
Device Description

BIOLOX delta material is an aluminum oxide / zirconia ceramic composite composed of approximately 75% aluminum oxide and approximately 25% zirconia.
The Corin BIOLOX delta Modular Femoral Heads are available in 28mm and 32mm diameters. The 28mm heads are available with short (-3.5mm), medium (0mm) and long (+3.5mm) offsets. The 32mm heads are available with short (-4.0mm), medium (0mm), long (+4mm) and extra long (+7mm) offsets.
The Corin Biolox Delta heads are compatible with Corin titanium stems (i.e. Tri-Fit, Metafix and MiniHip femoral stems having a 12/14 taper trunnion) and the Trinity acetabular system.

AI/ML Overview

The provided text describes the regulatory clearance for the Corin BIOLOX delta Modular Femoral Heads. This is a medical device application, and the "acceptance criteria" here refers to the requirements for demonstrating substantial equivalence to predicate devices, rather than performance metrics for an AI algorithm. Therefore, many of the requested fields (e.g., sample size for test/training sets, number of experts for ground truth, MRMC study, effect size, standalone performance) are not applicable in this context.

Here's an analysis of the available information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Demonstrating Substantial Equivalence)Reported Device Performance
Intended Use/Indications: Same as predicate devicesThe Corin BIOLOX delta Modular Femoral Heads have the same types of indications and intended uses as the predicates (Non-inflammatory degenerative joint disease, Rheumatoid arthritis, Correction of functional deformity, Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH)).
Technological Characteristics: Same as predicate devicesThe BIOLOX delta material is an aluminum oxide/zirconia ceramic composite (75% aluminum oxide, 25% zirconia). The heads are available in 28mm and 32mm diameters with various offsets. They are compatible with Corin titanium stems and the Trinity acetabular system. The document states "The technological characteristics are the same as the predicates."
Performance (Non-Clinical Testing): Meet suggested values in "Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems" and be within the range of legally marketed predicates.Burst Strength Testing: Performed on 28mm-12/14L and 32mm-12/14XL Biolox delta heads on Corin titanium taper trunnion (worst case construct). Fatigue Testing: Performed on the worst case construct. Post-Fatigue Burst Testing: Performed on the worst case construct. Pull-off Testing: Performed on the worst case construct. Results: "The results of the ceramic head testing meet the suggested values in the 'Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems.' The results of the preclinical data provided indicate that the subject system is within the range of legally marketed predicates."
Clinical Testing: Not necessary if substantial equivalence can be demonstrated through other means."Clinical testing was not necessary to determine substantial equivalence between the Corin BIOLOX delta Modular Femoral Heads and the predicate devices."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable as the described study is non-clinical testing of a hip implant device, not a diagnostic AI device requiring a clinical test set from human subjects. The testing involved physical ceramic heads and trunnions.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable. The "ground truth" for this device's performance is based on established engineering standards and material properties, not expert interpretation of medical images or patient outcomes in a clinical trial setting.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods are typically used in clinical studies involving interpretation of data (e.g., imaging) by multiple human readers, which is not the case here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This is not an AI-assisted diagnostic device, so an MRMC study is irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the non-clinical testing was based on established engineering standards, material specifications, and the performance characteristics of legally marketed predicate devices. The "Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems" provides the benchmark values.

8. The sample size for the training set

This information is not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable. This is not an AI/machine learning device.


Summary of the Study:

The study proving the device meets acceptance criteria was a series of non-clinical mechanical tests designed to assess the durability and structural integrity of the Corin BIOLOX delta Modular Femoral Heads.

  • Tests Performed: Burst strength testing, fatigue testing, post-fatigue burst testing, and pull-off testing.
  • Test Setup: The tests were performed on the ceramic heads in combination with Corin titanium taper trunnions, specifically focusing on "worst case constructs" (28mm-12/14L and 32mm-12/14XL Biolox delta heads).
  • Acceptance Criteria: The results of these tests were compared against "suggested values in the 'Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems'" and demonstrated that the subject system performed within the range of legally marketed predicate devices.
  • Ground Truth: The "ground truth" for these tests was derived from these established guidance documents and the performance of existing, legally marketed hip implant devices.
  • Clinical Testing: No clinical testing was deemed necessary as substantial equivalence was established through the non-clinical mechanical testing and comparison of material, geometric, and intended use characteristics with predicate devices.

{0}------------------------------------------------

K103/20
p.1/2

3. 510(K) SUMMARY
1. Applicant/Sponsor:Corin USA10500 University Center DriveSuite 190Tampa, Florida 33612Establishment Registration No.:FEB 2 8 2011
2. Contact Person:Lucinda Gerber BA (Hons)Regulatory Affairs AssociateCorin USA813-977-4469Lucinda.gerber@coringroup.com
3. Proprietary Name:Corin BIOLOX delta Modular Femoral Heads
4. Common Name:Femoral Head
5. Classification Name:Hip joint metal/ceramic/polymer semi-constrained cemented ornonporous uncemented prosthesis (21CFR 888.3353)
    1. Legally Marketed Devices to which Substantial Equivalence is claimed:
    • Zyranox Zirconia Ceramic Modular Heads (K992235)
    • Smith & Nephew BIOLOX delta Ceramic Femoral Heads (K100412) 트

7. Device Description:

BIOLOX delta material is an aluminum oxide / zirconia ceramic composite composed of approximately 75% aluminum oxide and approximately 25% zirconia.

The Corin BIOLOX delta Modular Femoral Heads are available in 28mm and 32mm diameters. The 28mm heads are available with short (-3.5mm), medium (0mm) and long (+3.5mm) offsets. The 32mm heads are available with short (-4.0mm), medium (0mm), long (+4mm) and extra long (+7mm) offsets.

The Corin Biolox Delta heads are compatible with Corin titanium stems (i.e. Tri-Fit, Metafix and MiniHip femoral stems having a 12/14 taper trunnion) and the Trinity acetabular system.

{1}------------------------------------------------

K103120
p. 2/2

8. Intended Use / Indications:

The Corin BIOLOX delta Modular Femoral Heads are used in combination with femoral hip stems and ultra high molecular weight polyethylene (UHMWPE) acetabular cup liners to reinstate function following the degenerative effects of:

  • Non-inflammatory degenerative joint disease including osteoarthritis and avascular 0 necrosis
  • Rheumatoid arthritis 0
  • Correction of functional deformity and 0
  • Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH) 0

The Corin BIOLOX delta Modular Femoral Heads are intended for single use only.

  • Summary of Technologies/Substantial Equivalence: 9.
    The Corin Biolox Delta modular femoral heads have the same types of indications and intended uses as the predicates. The technological characteristics are the same as the predicates. Based on the materials, geometry, mechanical testing and indications for use, the Biolox Delta heads are considered to be substantially equivalent to the predicate devices.

10. Non-Clinical Testing:

Nonclinical testing included burst strength testing on both the 28mm-12/14L and 32mm-12/14XL Biolox delta heads on the Corin titanium taper trunnion to determine the worst case construct. Subsequently, fatigue, post-fatigue burst and pull-off testing were then performed on the worst case construct. The results of the ceramic head testing meet the suggested values in the "Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems." The results of the preclinical data provided indicate that the subject system is within the range of legally marketed predicates.

11. Clinical Testing:

Clinical testing was not necessary to determine substantial equivalence between the Corin BIOLOX delta Modular Femoral Heads and the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002

Corin USA % Lucinda Gerber, BA (Hons) Regulatory Affairs Associate 10500 University Center Drive Suite 190 Tampa, Florida 33612

FEB 2 8 2011

Re: K103120

Trade/Device Name: Biolox Delta Modular Femoral Heads Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO Dated: February 09, 2011

Received: February 10, 2011

Dear Ms. Gerber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Lucinda Gerber, BA (Hons)

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

for Dothum ReopDic

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

2. INDICATIONS FOR USE

K103120 510{k} Number (if known): _

Device Name: Corin BIOLOX delta Modular Femoral Heads

Indications for Use:

The Corin BIOLOX delta Modular Femoral Heads are used in combination with femoral hip stems and ultra high molecular weight polyethylene (UHMWPE) acetabular cup liners to reinstate function following the degenerative effects of:

  • Non-inflammatory degenerative joint disease including osteoarthritis and avascular ্ necrosis
  • Rheumatoid arthritis 0
  • Correction of functional deformity and o
  • Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH) O

The Corin BIOLOX delta Modular Femoral Heads are intended for single use only.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for M. Melkerson

(Division Sign thopedic, Division of Surgical, and Restorative Device

510(k) Number K103120

Page 5 of 129

§ 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.

(a)
Identification. A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.(b)
Classification. Class II.