K Number
K103015

Validate with FDA (Live)

Date Cleared
2011-04-13

(183 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ODRC-DHS System is intended for use in fixation of fractures to the proximal femur. The system is indicated for use in trochanteric, pertrochanteric, intertrochanteric, and basilar neck fractures.

Device Description

The ODRC-Dynamic Hip Screw System consists of non-sterile ODRC-DHS Plate, ODRC-DHS Screw, ODRC-DHS Blades, compression screw, and 4.5 mm cortex screw. The ODRC-DHS Plate is a straight with an angled barrel portion that accepts ODRC-DHS Screw and ODRC-DHS Blades. The barrel angle is 135° and length of 36 mm. The plates are available in multiple lengths (lengths ranging from 46 mm to 206 mm) to accommodate differing patient anatomy. The plate accepts DHS Helix Screw (lengths ranging from 75 mm to 130 mm) and 4.5 mm cortex screws.

AI/ML Overview

Here's an analysis of the provided text regarding the ODRC-Dynamic Hip Screw System, focusing on acceptance criteria and performance study details:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Mechanical Performance: - Static Four-Point Bending Test (ASTM F382) - Dynamic Four-Point Bending Test (ASTM F382) - Static Axial Compression Bending Test (ASTM F384) - Dynamic Axial Compression Bending Test (ASTM F384)The ODRC-Dynamic Hip Screw System performs as well as or better than the predicate device (SYNTHES DHS System K981757) in all conducted mechanical tests. (Specific quantitative results are not provided in the summary).
Material: - Medical grade 316L stainless steelMeets ASTM F138 & ASTM F139/lSO 5832-1.
Technical Characteristics: - Plate barrel angle: 135° - Plate barrel length: 36 mm - Plate lengths: 46 mm to 206 mm - DHS Helix Screw lengths: 75 mm to 130 mmThe device meets these specified technical characteristics. (Implicitly, as it forms part of the device description and is compared to the predicate).
Intended Use: - Fixation of fractures of the proximal femur (femoral neck, trochanteric, pertrochanteric, or Intertrochanteric zones).The device's indications for use are consistent with those of the predicate device, indicating it meets the intended use criteria.
Substantial Equivalence:Demonstrated to be substantially equivalent to the predicate system with respect to technical characteristics, performance, and intended use.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified. The document states "Mechanical testing... was conducted," but does not provide details on the number of components or systems tested.
  • Data Provenance: The testing was conducted by the "ORTHOPAEDIC DEVICE RESEARCH CENTER" at "National Yang-Ming University, Taipei 11221, Taiwan." This suggests the tests were performed in Taiwan. The study appears to be retrospective in the sense that it's a comparison to an existing predicate device, and the bench testing itself is a controlled laboratory study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This study involves mechanical bench testing of a medical device, not a diagnostic or clinical study requiring expert consensus for ground truth. The "ground truth" here is objective physical measurements and material specifications.

4. Adjudication method for the test set

Not applicable. As this is mechanical testing, there is no human adjudication process involved in establishing results. The results are obtained directly from test equipment based on established ASTM standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a submission for a mechanical fixation device, not an AI-assisted diagnostic tool. Therefore, no MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device. The performance evaluated is the physical, mechanical behavior of the hip screw system itself.

7. The type of ground truth used

The ground truth for the acceptance criteria and performance assessment is based on objective mechanical properties and material specifications as defined by relevant ASTM and ISO standards (ASTM F382, ASTM F384, ASTM F138, ASTM F139/lSO 5832-1). The performance is measured against these established engineering standards and compared directly to a legally marketed predicate device.

8. The sample size for the training set

Not applicable. There is no concept of a "training set" in this type of submission for a mechanical device. The device's design and manufacturing are based on established engineering principles and materials science, not machine learning.

9. How the ground truth for the training set was established

Not applicable. As there is no training set, this query is irrelevant to the provided information.

{0}------------------------------------------------

K1030156pg 1/2)

Chapter 1 510(k)Summary

APR 1 3 2011

510(K) SUMMARY

This summary regarding 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21CFR 807.92.

Sponsor :ORTHOPAEDIC DEVICE RESEARCH CENTERNo. 155. Sec. 2 Li-Nung St.Orthopaedic Device Research CenterNational Yang-Ming UniversityTaipei 11221TaiwanPhone:886-2-28267020Fax:886-2-28228557
Contact Person:Cheng-Kung ChengDirect, Orthopaedic Device Research Center
Device Name:ODRC-Dynamic Hip Screw System
Classification:Class II : Plate, Fixation, BoneHRS-CFR 888.3030 Class II : Screw, Fixation, BoneHWC — CFR 888.3040 Class II : Appliance, fixation, nail/blade/platecombination, multiple componentKTT — CFR 888.3030
Predicate Devices :SYNTHES DHS System (K981757)

Material :

The ODRC-Dynamic Hip Screw System components are manufactured from medical grade 316L stainless steel that meets ASTM F138 & ASTM F139/lSO 5832-1.

{1}------------------------------------------------

Device Description :

The ODRC-Dynamic Hip Screw System consists of non-sterile ODRC-DHS Plate, ODRC-DHS Screw, ODRC-DHS Blades, compression screw, and 4.5 mm cortex screw. The ODRC-DHS Plate is a straight with an angled barrel portion that accepts ODRC-DHS Screw and ODRC-DHS Blades. The barrel angle is 135° and length of 36 mm. The plates are available in multiple lengths (lengths ranging from 46 mm to 206 mm) to accommodate differing patient anatomy. The plate accepts DHS Helix Screw (lengths ranging from 75 mm to 130 mm) and 4.5 mm cortex screws.

Indication for Use :

The ODRC-Dynamic Hip Screw System is provided non-sterile. The device may be used for fixation of fractures of the proximal femur, such as femoral neck, trochanteric, pertrochanteric, or Intertrochanteric zones.

Performance Data :

Mechanical testing (static and dynamic four point bending test (ASTM F382), and static and dynamic axial compression bending test (ASTM F384)) was conducted to demonstrate substantial equivalence to the predicate system. The results demonstrate that the ODRC-Dynamic Hip Screw System performs as well as or better than the predicate device.

Basis of Substantial Equivalence :

The ODRC-Dynamic Hip Screw System has been demonstrated to be substantially equivalent to predicate system with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject device to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the address for the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Mail Center - WO66-G609, Silver Spring, MD 20993-0002. The address is likely for a specific department or division within the FDA.

Orthopaedic Device Research Center % Cheng-Kung Cheng No. 155, Sec 2. Li-Nung Street Taipei China (Taiwan) 112

APR 1 3 2011

Re: K103015

Trade/Device Name: ODRC -- Dynamic Hip Screw System Regulation Number: 888.3040 Regulation Name: Regulatory Class: Class II Product Code: KTT, HRS, HWC Dated: March 31, 2011 Received: April 8, 2011

Dear Mr.Cheng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the iddictions for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Foond, Or ug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

Image /page/2/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

{3}------------------------------------------------

Page 2 - Mr. Cheng - Kung Cheng

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Alig B. rita
for

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATION FOR USE

Indications for Use

510(k) Number (K103015):

Device Name: ODRC-DHS System

Indications for Use:

The ODRC-DHS System is intended for use in fixation of fractures to the proximal femur. The system is indicated for use in trochanteric, pertrochanteric, intertrochanteric, and basilar neck fractures.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ ( Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for M.Meukerson
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number

4-4

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.