K Number
K102977

Validate with FDA (Live)

Date Cleared
2010-12-20

(75 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.

AI/ML Overview

The provided text describes the acceptance criteria and performance of a medical device, specifically Powder Free Vinyl Patient Examination Gloves. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicsStandard / Acceptance CriteriaDevice Performance
DimensionASTM standard D 5250-06 e1Meets
Physical PropertiesASTM standard D 5250-06 e1Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06 e1 and D6124-06Meets (<2mg/glove)
Biocompatibility (Irritation)Primary Skin Irritation in rabbits (ISO 10993-10)Passes (Not a Primary Skin Irritation)
Biocompatibility (Sensitization)Dermal sensitization in the guinea pig (ISO 10993-10)Passes (Not a Dermal Sensitization)

2. Sample size used for the test set and the data provenance:

The provided text does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It simply states that the device "meets" the specified standards for various characteristics.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The provided text does not mention using experts to establish ground truth for a test set. This type of device (patient examination gloves) typically relies on standardized laboratory testing rather than expert-based evaluation for its performance characteristics.

4. Adjudication method for the test set:

The provided text does not describe any adjudication method for a test set. Adjudication is not applicable in this context as performance is assessed against objective technical standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

A multi-reader multi-case (MRMC) comparative effectiveness study is not applicable to this device. This is a medical glove, not an AI-powered diagnostic or imaging tool that would involve human readers. Therefore, there is no mention of such a study or an effect size for AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

A standalone algorithm performance study is not applicable to this device. This product is a physical medical device (gloves) and does not involve any algorithms.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth used for evaluating this device's performance is based on objective technical standards and laboratory testing specified by ASTM and FDA regulations. For example:

  • Dimensional and Physical Properties: Defined by ASTM standard D 5250-06 e1.
  • Freedom from pinholes: Defined by 21 CFR 800.20.
  • Powder Residual: Defined by ASTM standard D 5250-06 e1 and D6124-06.
  • Biocompatibility: Defined by ISO 10993-10 for primary skin irritation and dermal sensitization.

8. The sample size for the training set:

The provided text does not mention a "training set" as this is not a machine learning or AI-driven device. Performance is evaluated against established physical and chemical standards.

9. How the ground truth for the training set was established:

The concept of a "training set" and its ground truth establishment does not apply to this device as it is not an AI/ML product. The "ground truth" for the device's characteristics is defined by the accepted industry and regulatory standards.

{0}------------------------------------------------

Section C 510(k) Summary (21 CFR 807.92)

DEC 2 0 2010

510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: 1029177 " (applicant leave blank)

Premarket Notification [510(k)] Summary

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :Heze Juxinyuan Food Co.,Ltd.
Submitter's address :Jinzhuang Zone, Suji, Caoxian , Heze City,Shandong Province, 274415,China
Phone number :0086-530-5022111
Fax number :0086-530-5311988
Name of contact person:Mr. Wenzhong Zhang
Date the summary was prepared:Sep.19, 2010

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves,Clear (non-colored)
Proprietary/Trade name:Powder Free Vinyl Patient Examination Gloves,Clear (non-colored)Other clients private labeling
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital(80)
Product Code:LYZ

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.

Section C

{1}------------------------------------------------

Predicate device: FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .

[(a)(4)] A description of the device

Device Description: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-06 e1.Meets
Physical PropertiesASTM standard D 5250-06 e1.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06 e1and D6124-06Meets<2mg/glove
BiocompatabilityPrimary Skin Irritation in rabbitsISO 10993-10PassesNot a Primary Skin Irritation
Dermal sensitization in the guinea pigISO 10993-10PassesNot a Dermal sensitization

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), meet requirements per ASTM D5250-06 el, per ASTM D6124-06, per 21 CFR 800.20 and FDA recognition number 2-87: ISO 10993-10:2002.

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims.

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is as safe, as effective, and performs as well as the predicate device, FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908

Section C

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Heze Juxinyuan Food Company, Limited C/O Mr. Chu Xiaoan Beijing Easy-Link Company, Limited Room 1606 Building 1 Jian Xiang Yuan N209 Bei Si Huan Beijing China 100083

DEC 2 0 2010

Re: K102977

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: September 27, 2010 Received: October 6, 2010

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In

{3}------------------------------------------------

Page 2- Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Jacus J. Robertson
for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices

Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{4}------------------------------------------------

DEC 2 0 2010 INDICATIONS FOR USE

Heze Juxinyuan Food Co.,Ltd. Applicant:

510(k) Number (if known): * 1029 77

Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)

Indications For Use:

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Edith S. (Clemis-Will

(Division Sign-Off)
(Division Sign-Off)
Division of Anesthesinlogy, General Hospital
Provision of Anesthesinlogy, General Devices Division of Antoonomics Infection Control Devices

510(k) Number: K102977

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.