K Number
K102919

Validate with FDA (Live)

Date Cleared
2011-03-11

(161 days)

Product Code
Regulation Number
870.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The venous hardshell cardiotomy reservoir is used to collect, store and filter blood in extracorporeal circulation in cardiopulmonary bypass operations on pediatric patients for up to 6 hours: The reservoir can also be employed postoperatively as drainage and autotransfusion reservoir (e.g., for thorax drainage) to return the autologous blood to the patient which was removed from the thorax for the volume exchange.

Device Description

The Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs are developed for the use in the field of cardiopulmonary bypass operation. They are used as blood buffer in the extracorporeal circuit and as collecting and defoaming device for sucked blood. The Neonatal and Pediatric Reservoirs may be applied during the surgery in the extracorporeal circulation for collecting venous blood by gravitation or vacuum assist venous drainage (VAVD), air removal in venous blood, as a volume depot of blood, for collecting, defoaming and filtering cardiotomy blood. The Reservoirs may also be applied after the surgery on the intensive care unit for vacuum operated thoracic drainage and for the autotransfusion of autologous blood. If the application occurs in the intensive care unit the reservoir which was employed in the operation is usually used. The blood contacting surfaces are coated optionally with SOFTLINE Coating. The Neonatal and Pediatric Reservoirs are sterile and non-pyrogenic devices, for single use only and are not to be re-sterilized by the user.

AI/ML Overview

Here's an analysis of the provided text, focusing on acceptance criteria and the study proving adherence:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative "acceptance criteria" in the sense of specific thresholds for a device's performance metrics (e.g., "filtration efficiency must be >X%"). Instead, the acceptance criteria are framed as compliance with recognized standards and demonstrating substantial equivalence to predicate devices. The "reported device performance" is essentially a statement of compliance.

Acceptance Criteria CategoryReported Device PerformanceComments
Biologic EvaluationComplies with ISO 10993-1This standard covers biocompatibility testing (e.g., cytotoxicity, sensitization, irritation). The reported performance indicates the device passed these tests.
Performance (System)Complies with ISO 15674This standard specifically addresses "Cardiovascular implants and artificial organs - Hard-shell cardiotomy / venous reservoir systems (with/without filter) and softbag venous reservoir bags." Compliance indicates the device meets the functional and safety requirements outlined in this standard for its intended use.
Substantial EquivalenceDemonstrated to be substantially equivalent to Capiox Reservoir RX 05 Baby (K022115) and Resevoir D101 Dideco Kids Infant (K072091).This is the overarching regulatory acceptance criterion for 510(k) clearance. The device's performance, integrity, biocompatibility, and intended use were compared to these predicate devices.
IntegrityEvaluated and met requirementsThis likely refers to physical integrity, resistance to leaks, and structural soundness under operational conditions.
SterilityEvaluated and met requirementsThis often involves demonstrating the ability to be sterilized to a specified Sterility Assurance Level (SAL) and maintaining sterility throughout its shelf-life.
Indications for UseEvaluated and met requirementsThis confirms the device functions as intended for its specified uses (collecting, storing, filtering blood in CPB for pediatric patients up to 6 hours, and postoperative drainage/autotransfusion).

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify a sample size for any "test set" in the traditional sense of a clinical trial or large-scale device performance study. The testing described is primarily laboratory and bench testing against international standards (ISO 10993-1, ISO 15674) and direct comparison to a predicate device.

  • Sample Size: Not specified.
  • Data Provenance: The studies were likely conducted internally by the manufacturer (Maquet Cardiopulmonary AG) or by contracted laboratories, as part of the regulatory submission process.
  • Retrospective or Prospective: Not explicitly stated, but typically, this type of non-clinical testing is prospective, meaning it's conducted specifically to support the 510(k) submission. There is no indication of patient data being used in this summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable and not provided in the document. The "ground truth" for this medical device (a cardiotomy reservoir) is established by adherence to recognized engineering standards (e.g., ISO) and functional performance metrics, not by expert interpretation of data like in an AI/diagnostic imaging context. The "experts" involved would be engineers, biocompatibility specialists, and quality control personnel performing the tests according to standard protocols.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

This information is not applicable and not provided. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in studies involving human interpretation (e.g., reading medical images) where there can be disagreement among experts. For device performance testing against objective standards (like ISO norms for a physical device), the results are typically quantitative measurements that either pass or fail the specified criteria, not subjective interpretations requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable and not provided. This document describes a physical medical device (a blood reservoir) and its regulatory clearance based on substantial equivalence and compliance with standards. It is not an AI algorithm or a diagnostic imaging device that would involve human readers or AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable and not provided. The device is a physical hard-shell cardiotomy reservoir, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device is based on:

  • Compliance with International Standards: Specifically ISO 10993-1 (Biologic Evaluation) and ISO 15674 (Hard-shell cardiotomy/venous reservoir systems). These standards define performance characteristics, safety requirements, and test methods.
  • Comparison to Predicate Devices: Demonstrating that the new device performs as safely and effectively as legally marketed predicate devices (Capiox Reservoir RX 05 Baby and Resevoir D101 Dideco Kids Infant) in terms of indications for use, integrity, performance, biocompatibility, and sterility.

No pathology, expert consensus, or outcomes data from patients are mentioned for establishing the "ground truth" of the device itself for this 510(k) submission.

8. The Sample Size for the Training Set

This information is not applicable and not provided. There is no "training set" as this is not an AI/machine learning device.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable and not provided. Since there is no training set for an AI model, there's no ground truth established in this context.

{0}------------------------------------------------

MAR 1 1 2011

510 (K) Summary [as required by 21 CFR 807.92(c) ]

K102919

Submitter: Maquet Cardiopulmonary AG Hechinger Strasse 38 72145 Hirrlingen Germany

Katrin Schwenkglenks Contact Person: Phone: +49 7478 921-151 +49 7478 921-400 Fax: E-mail: katrin.schwenkglenks@maquet-cp.com

September 28, 2010 Date Prepared:

Neonatal Venous Hardshell Cardiotomy Reservoir Device Trade Name: and Pediatric Venous Hardshell Cardiotomy Reservoir, with and without SOFTLINE Coating

Venous Hardshell Cardiotomy Reservoir Common/Usual name:

Cardiopulmonary bypass blood reservoir, Classification names: Cardiopulmonary bypass defoamer, Cardiopulmonary bypass cardiotomy suction line

Predicate Devices:

Capiox Resevoir RX 05R Baby, Terumo Cardiovascular Systems Corp (K022115),

Resevoir D101 Dideco Kids Infant with Ph.J.S.I.O. Coating, Sorin (K072091), Venous Hardshell Cardiotomy Reservoir with SOFTLINE Coating (K090534), MAQUET Cardiopulmonary AG.

{1}------------------------------------------------

Device Description:

The Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs are developed for the use in the field of cardiopulmonary bypass operation. They are used as blood buffer in the extracorporeal circuit and as collecting and defoaming device for sucked blood.

The Neonatal and Pediatric Reservoirs may be applied during the surgery in the extracorporeal circulation for collecting venous blood by gravitation or vacuum assist venous drainage (VAVD), air removal in venous blood, as a volume depot of blood, for collecting, defoaming and filtering cardiotomy blood.

The Reservoirs may also be applied after the surgery on the intensive care unit for vacuum operated thoracic drainage and for the autotransfusion of autologous blood. If the application occurs in the intensive care unit the reservoir which was employed in the operation is usually used.

The blood contacting surfaces are coated optionally with SOFTLINE Coating.

The Neonatal and Pediatric Reservoirs are sterile and non-pyrogenic devices, for single use only and are not to be re-sterilized by the user.

Indications for Use:

The venous hardshell cardiotomy reservoir is used to collect, store and filter blood in extracorporeal circulation in cardiopulmonary bypass operations on pediatric patients for up to 6 hours.

The reservoir can also be employed postoperatively as drainage and autotransfusion reservoir (e.g., for thorax drainage) to return the autologous blood to the patient which was removed from the thorax for the volume exchange.

Statement of Technical Comparison:

The Neonatal Venous Hardshell Cardiotomy Reservoir and Pediatric Venous Hardshell Cardiotomy Reservoir, with and without SOFTLINE Coating is comparable to the above named predicate devices which are also designed and intended to be used with neonate and pediatric patients. The Softline Coating is the same coating contained in the Venous Hardshell Cardiotomy Reservoir with SOFTLINE Coating.

{2}------------------------------------------------

GETINGE GROUP

Non-clinical Testing:

The Neonatal Venous Hardshell Cardiotomy Reservoir and Pediatric Venous Hardshell Cardiotomy Reservoir, with and without SOFTLINE Coating have been tested or evaluated for compliance to ISO 10993-1 Biologic Evaluation of Medical Devices and ISO 15674 "Cardiovascular implants and artificial orqans - Hard-shell cardiotomy / venous reservoir systems (with/without filter) and softbag venous reservoir bags" The products met these requirements.

Determination of Substantial Equivalence

Testing and evaluation on safety and effectiveness was conducted to demonstrate that the Neonatal Venous Hardshell Cardiotomy Reservoir and Pediatric Venous Hardshell Cardiotomy Reservoir, with and without SOFTLINE Coating is substantially equivalent to the Capiox Reservoir RX 05 Baby from Terumo (K022115) as well as to the Resevoir D101 Dideco Kids Infant with Ph.1.S.I.O. Coating from Sorin (K072091).

The following areas have been tested or evaluated:

  • Indications for Use
  • Integrity
  • Performance
  • Biocompatibility
  • Sterility

Conclusion

The data given demonstrate that the Neonatal Venous Hardshell Cardiotomy Reservoir and Pediatric Venous Hardshell Cardiotomy Reservoir, with and without SOFTLINE Coating is substantially equivalent to the named predicate devices which currently hold market clearance.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Maquet Cardiopulmonary AG c/o Ms. Katrin Schwenkglenks Hechinger Strasse 38 D-72145 Hirrlingen, Germany

MAR 1 1 2511

Re: K102919

Trade/Device Name: Neonatal Venous Hardshell Cardiotomy Reservoir (VHK11000), Neonatal Venous Hardshell Cardiotomy Reservoir with SOFTLINE Coating (BO-VHK 11000), Pediatric Venous Hardshell Cardiotomy Reservoir (VHK31000), and Pediatric Venous Hardshell Cardiotomy Reservoir with SOFTLINE (THRS1000), and Fedaring Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary bypass blood reservoir Regulatory Class: II Product Code: DTN, DTP, JOD Dated: March 1, 2011 Received: March 4, 2011

Dear Ms. Schwenkglenks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (fav the ucications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr w and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbanding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements af the Act

{4}------------------------------------------------

Page 2 – Ms. Katrin Schwenkglenks

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html fart the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

R. Vahner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K102912

Device Name: Neonatal Venous Hardshell Cardiotomy Reservoir and Pediatric Venous Hardshell Cardiotomy Reservoir, with and without SOFTLINE Coating

Indications for Use:

The venous hardshell cardiotomy reservoir is used to collect, store and filter blood in extracorporeal circulation in cardiopulmonary bypass operations on pediatric patients for up to 6 hours:

The reservoir can also be employed postoperatively as drainage and autotransfusion reservoir (e.g., for thorax drainage) to return the autologous blood to the patient which was removed from the thorax for the volume exchange.

Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Summa R. Varner

(Division Sian-Off) Division of Cardiovascular Devices

510(k) Number K102919

§ 870.4400 Cardiopulmonary bypass blood reservoir.

(a)
Identification. A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.(b)
Classification. Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.