EMIT II PLUS6-ACETYLMORPHINE ASSAY; EMIT II PLUS 6-AM/ ECSTASY CALIBRATOR/ CONTROL LEVEL 1; EMIT II PLUS 6-AM / ECTASY C

K102779 · Siemens Healthcare Diagnostics, Inc. · DJG · Mar 18, 2011 · Clinical Toxicology

Device Facts

Record IDK102779
Device NameEMIT II PLUS6-ACETYLMORPHINE ASSAY; EMIT II PLUS 6-AM/ ECSTASY CALIBRATOR/ CONTROL LEVEL 1; EMIT II PLUS 6-AM / ECTASY C
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeDJG · Clinical Toxicology
Decision DateMar 18, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3650
Device ClassClass 2

Intended Use

Emit® II Plus 6-Acetylmorphine Assay: The Emit® II Plus 6-Acetylmorphine Assay is a homogeneous enzyme immunoassay with 10 ng/mL cutoff. The assay is intended for use in laboratories for the qualitative and/or semiquantitative analyses of 6-acetylmorphine (6-AM), a heroin metabolite, in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. Semiquantitative test results may be used to assess assay performance as part of a quality control program and to estimate a dilution of the specimen for confirmation by GC/MS. The Emit® II Plus 6-Acetylmorphine Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectroscopy (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used. Emit® II Plus 6-AM/Ecstasy Calibrators/Controls: When used as Calibrators, the materials are for the calibration of the Emit® II Plus 6-Acetylmorphine and Emit® II Plus Ecstasy Assays. When used as Controls, the materials may be used as quality control materials based on the specific Emit® II Plus 6-Acetylmorphine and Emit® II Plus Ecstasy Assay cutoffs.

Device Story

Homogeneous enzyme immunoassay for detection of 6-acetylmorphine in human urine; utilizes two-reagent system (mouse monoclonal antibodies, G6P, NAD, and 6-AM labeled rG6PDH). Operates on clinical chemistry analyzers; measures enzymatic rates at 340 nm. Provides qualitative or semiquantitative results; requires confirmation by GC/MS. Used in clinical laboratories by trained personnel. Results assist in identifying heroin metabolite presence; clinical judgment required for interpretation.

Clinical Evidence

Method comparison study using 105 unaltered human urine samples analyzed by Emit® II Plus 6-Acetylmorphine Assay vs. GC/MS. Qualitative agreement was 98% for positives and 100% for negatives. Semiquantitative agreement was 98% for positives and 100% for negatives. One discordant result reported.

Technological Characteristics

Homogeneous enzyme immunoassay; R1 liquid (monoclonal antibodies, G6P, NAD); R2 lyophilized (6-AM labeled rG6PDH). Requires chemistry analyzer with 340 nm measurement capability. Calibrators/Controls are human urine-based. Storage 2–8°C.

Indications for Use

Indicated for laboratory use for qualitative and/or semiquantitative analysis of 6-acetylmorphine (heroin metabolite) in human urine. Intended for use with chemistry analyzers to provide preliminary analytical results requiring confirmation by GC/MS.

Regulatory Classification

Identification

An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAR 1 8 2011 ## 510(k) Summary of Safety and Effectiveness for the ## Emit® II Plus 6-Acetylmorphine Assay ## Emit® II Plus 6-AM / Ecstasy Calibrators / Controls Levels 1 - 4 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR807.92. ### A. 510(k) Number: k102779 #### B. Date of Preparation: March 1, 2011 ## C. Proprietary and Established Names Emit® II Plus 6-Acetylmorphine Assay Emit® II Plus 6-AM / Ecstasy Calibrator / Control Level 1 Emit® II Plus 6-AM / Ecstasy Calibrator / Control Level 2 Emit® II Plus 6-AM / Ecstasy Calibrator / Control Level 3 Emit® II Plus 6-AM / Ecstasy Calibrator / Control Level 4 ## D. Applicant: Siemens Healthcare Diagnostics Inc. P.O. Box 6101, Newark, DE 19714-6101 Janet Fose, Regulatory Technical Specialist, Regulatory Affairs Office: (302) 631-8826 Fax: (302) 631-6299 ### E. Regulatory Information: Emit® II Plus 6-Acetylmorphine Assay: - 1. Regulation section: 21 CFR § 862.3650 Opiate test system - 2. Classification: Class II - 3. Product Code: DJG Enzyme Immunoassay, Opiates - 4. Panel: Clinical Toxicology (91) Emit® II Plus 6-AM / Ecstasy Calibrators / Controls Levels 1 - 4: #### Calibrator - 1. Regulation section: 21 CFR § 862.3200 Clinical toxicology calibrator - 2. Classification: Class II - 3. Product Code: DKB Calibrators, Drug Mixture - 4. Panel: Clinical Toxicology (91) {1}------------------------------------------------ #### Siemens Healthcare Diagnostics Inc. 510(k) Premarket Notification e-submission #### Control - 1. Regulation section: 21 CFR § 862.3280 Clinical toxicology control material - 2. Classification: Class I. reserved - 3. Product Code: DIF Drug Mixture Control Materials - 4. Panel: Clinical Toxicology (91) ## F. Predicate Device(s): Emit® II Plus 6-Acetylmorphine Assay: The Emit® II Plus 6-Acetylmorphine Assay is substantially equivalent to the Microgenics Corporation CEDIA® DAU 6-Acetylmorphine Assay cleared under k001178. Both are designed for use on a number of chemistry analyzers. Emit® II Plus 6-AM / Ecstasy Calibrators / Controls Levels 1 - 4: ### Calibrator The Emit® II Plus 6-AM / Ecstasy Calibrators / Controls are substantially equivalent to the Emit® II Plus Ecstasy Calibrators / Controls cleared under k043028. #### Control The Emit® II Plus 6-AM / Ecstasy Calibrators / Controls are substantially equivalent to the Emit® II Plus Ecstasy Calibrators / Controls cleared under k043028. The Emit® Calibrator / Control Level 0, which was cleared under k993755 will also be used with the Emit® II Plus 6- Acetylmorphine Assay. There was no change to the Calibrator Level 0. #### G. Device Description(s): Emit® II Plus 6-Acetylmorphine Assay: The Emit® II Plus 6-Acetylmorphine Assay is a homogenous enzyme immunoassay with a 10 ng/mL cutoff. The assay, used for the detection of 6-acetylmorphine (a heroin metabolite) in human urine, utilizes a two-reagent system. The Antibody/Substrate Reagent 1 is a liguid ready-to-use product comprised of mouse monoclonal antibodies to 6-acetylmorphine (6-AM), glucose-6-phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in a diluent containing bovine serum albumin (BSA), preservatives and stabilizers. The Enzyme Reagent 2 is a lyophilized product containing 6-AM labeled bacterial recombinant glucose-6-phosphate dehydrogenase (rG6PDH) in a diluent containing bovine serum albumin (BSA), Hepes buffer, preservatives and stabilizers. Reagent 2 is reconstituted with either deionized or distilled water. The assay kit consists of Reagent 1 and Reagent 2 in plastic containers (Reagent 2 is provided in a plastic bag with desiccant) and is available in two sizes. Emit® II Plus Assays are designed for use with a number of chemistry analyzers. {2}------------------------------------------------ Emit® II Plus 6-AM / Ecstasy Calibrators / Controls Levels 1 - 4: The Emit® II Plus 6-AM / Ecstasy Calibrators / Controls are in-vitro diagnostic products used in the calibration of the Emit® II Plus 6-Acetylmorphine Assay and the Emit® II Plus Ecstasy Assay. These materials may also be used as quality controls based on the specific 6-Acetylmorphine Assay or Ecstasy Assay cutoffs. The calibrator / control products have the same formulation as the existing Emit® II Plus Ecstasy Calibrators / Controls: cleared under k043028. The matrix is pooled, drug-free. human urine based product containing 6- acetylmorphine (6-AM), methylenedioxymethamphetamine (MDMA) and preservatives. The four levels of product are packaged separately in 15 mL plastic vials with a 10 mL fill per vial. The multi-analyte Calibrators / Controls Levels 1 through 4 contain 6-AM and MDMA at the following concentrations: | Calibrator /<br>Control | Targeted 6-AM<br>Concentration<br>(ng/mL) | Targeted MDMA<br>Concentration<br>(ng/mL) | |-------------------------|-------------------------------------------|-------------------------------------------| | Level 1 | 5 | 150 | | Level 2 | 10 | 300 | | Level 3 | 15 | 500 | | Level 4 | 20 | 1000 | The Emit® Calibrator / Control Level 0, which contains no drug and was cleared under k993755 will also be used with the Emit® II Plus 6-AcetyImorphine Assay. There was no change to the Calibrator Level 0 product. # H. Intended Use: Emit® II Plus 6-Acetylmorphine Assay: The Emit® II Plus 6-Acetylmorphine Assay is a homogeneous enzyme immunoassay with 10 ng/mL cutoff. The assay is intended for use in laboratories for the qualitative and/or semiquantitative analyses of 6-acetylmorphine (6-A.M), a heroin metabolite, in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. Semiquantitative test results may be used to assess assay performance as part of a quality control program and to estimate a dilution of the specimen for confirmation by GC/MS. The Emit® II Plus 6-Acetylmorphine Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectroscopy (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used. {3}------------------------------------------------ Emit® II Plus 6-AM/Ecstasy Calibrators/Controls: When used as Calibrators, the materials are for the calibration of the Emit® II Plus 6-Acetylmorphine and Emit® II Plus Ecstasy Assays. When used as Controls, the materials may be used as quality control materials based on the specific Emit® II Plus 6-Acetylmorphine and Emit® II Plus Ecstasy Assay cutoffs. ## I. Substantial Equivalence Information: The Emit® II Plus 6-AcetyImorphine Assay was compared to the Microgenics predicate device CEDIA® DAU 6-Acetylmorphine Assay (k001178). The Emit® II Plus 6-AM / Ecstasy Calibrators / Controls were compared to Siemens Healthcare Diagnostics Inc. Emit® II Plus Ecstasy Calibrators / Controls Levels 1 – 4 (k043028). A comparison of the important similarities and differences between the devices and the predicates is provided in the following tables: | Feature | Proposed Device<br>Emit® II Plus 6-Acetylmorphine Assay | Predicate<br>CEDIA® DAU 6-<br>Acetylmorphine Assay<br>(k001178) | |--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The Emit® II Plus 6-Acetylmorphine Assay<br>is a homogeneous enzyme immunoassay<br>with 10 ng/mL cutoff. The assay is<br>intended for use in laboratories for the<br>qualitative and/or semiquantitative analyses<br>of 6-acetylmorphine (6-AM), a heroin<br>metabolite, in human urine. Emit® II Plus<br>assays are designed for use with a number<br>of chemistry analyzers.<br>Semiquantitative test results may be used to<br>assess assay performance as part of a<br>quality control program and to estimate a<br>dilution of the specimen for confirmation by<br>GC/MS.<br>The Emit® II Plus 6-Acetylmorphine Assay<br>provides only a preliminary analytical test<br>result. A more specific alternative chemical<br>method must be used to obtain a confirmed<br>analytical result. Gas chromatography/mass<br>spectroscopy (GC/MS) is the preferred<br>confirmatory method. Other chemical<br>confirmation methods are available.<br>Clinical consideration and professional<br>judgment should be applied to any drug-of-<br>abuse test result, particularly when<br>preliminary positive results are used. | The CEDIA® Heroin Metabolite<br>(6-Acetylmorphine, or 6-AM)<br>Assay is an in vitro diagnostic<br>medical device intended for the<br>qualitative and semiquantitative<br>analysis of heroin metabolite (6-<br>AM) in human urine. *<br>The assay provides only a<br>preliminary analytical test result.<br>A more specific alternate<br>chemical method must be used to<br>obtain a confirmed analytical<br>result. Gas chromatography /<br>mass spectrometry (GC/MS) is<br>the preferred confirmatory<br>method. Clinical consideration<br>and professional judgment should<br>be applied to any drug of abuse<br>test result, particularly when<br>preliminary positive results are<br>observed. *<br>*From the predicate device IFU | ### Comparison of Assay Features {4}------------------------------------------------ . -- - - - --- | | | Predicate<br>CEDIA® DAU 6- | |------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature | Proposed Device<br>Emit® II Plus 6-Acetylmorphine Assay | Acetylmorphine Assay<br>(k001178) | | Assay<br>Methodology | Homogeneous enzyme immunoassay using<br>EMIT® technology | Homogeneous enzyme<br>immunoassay using CEDIA®<br>technology | | Antibody | Mouse monoclonal antibodies to 6-AM | Monoclonal antibodies to 6-AM | | Reference<br>Methodology | GC / MS | GC / MS | | Cutoff | 10 ng/mL | 10 ng/mL | | Sample Type | Human urine | Human urine | | Reagents:<br>Form | R1: Liquid - Ready to Use<br>R2: Lyophilized (Reconstitution required) | R1 & R2: Lyophilized<br>(Reconstitution<br>required) | | Stability<br>(Reconstituted) | R1: Until Exp. on vial<br>R2: 30 days | R1 & R2: 60 days | | Instrument | Chemistry analyzers capable of maintaining<br>constant reaction temperature, pipetting<br>specimens/reagents and measuring enzyme<br>rates at 340 nm, timing reaction accurately<br>and mixing reagent thoroughly | Clinical chemistry analyzers<br>capable of maintaining constant<br>temperature, pipetting samples,<br>mixing reagents, measuring<br>enzymatic rates at 570 nm and<br>timing the reaction accurately | . ## Comparison of Calibrator Features | | Proposed Device<br>Emit® II Plus 6-AM / Ecstasy<br>Calibrators / Controls | Predicate<br>Emit® II Plus Ecstasy Calibrators /<br>Controls (k043028) | |--------------------------------------|-----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature | | | | Indications for<br>Use | Calibrators are used in the calibration of<br>the Emit® II Plus 6-Acetylmorphine<br>and Emit® II Plus Ecstasy Assays. | The Emit® II Plus Ecstasy<br>Calibrators/ Controls are used in the<br>calibration of the Emit® II Plus<br>Ecstasy Assay. These standards may<br>also be used as quality control<br>materials based on the specific<br>Ecstasy Assay cutoff. | | Matrix | Human urine based | Human urine based | | Analyte | Contains 6-AM and MDMA | Contains MDMA | | Target<br>Concentrations<br>for 6-AM | Level 1: 5 ng/mL<br>Level 2: 10 ng/mL<br>Level 3: 15 ng/mL<br>Level 4: 20 ng/mL | None | | Preparation | Liquid -- Ready to Use | Liquid - Ready to Use | | Storage | 2 – 8°C | 2 – 8°C | {5}------------------------------------------------ | Feature | Proposed Device<br>Emit® II Plus 6-AM / Ecstasy<br>Calibrators / Controls | Predicate<br>Emit® II Plus Ecstasy Calibrators /<br>Controls (k043028) | |---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications For<br>Use | Controls may be used as quality control<br>materials based on the specific Emit® II<br>Plus 6-Acetylmorphine and Emit® II<br>Plus Ecstasy Assay cutoffs. | The Emit® II Plus Ecstasy<br>Calibrators / Controls are used in the<br>calibration of the Emit® II Plus<br>Ecstasy Assay. These standards may<br>also be used as quality control<br>materials based upon the specific<br>Ecstasy Assay cutoff. | | Analyte | Contains 6-AM and MDMA | Contains MDMA | | Positive<br>Quality Control<br>Level for<br>Qualitative<br>Analysis | Level 4 | Level 4 | ## Comparison of Control Features ## J. Method Comparison: ## Qualitative and Semiquantitative Results Split sample method comparison was conducted and one-hundred five (105) unaltered human urine samples were analyzed by the Emit® II Plus 6-Acetylmorphine Assay vs. GC/MS. The results are presented below: Emit® II Plus 6-Acetylmorphine Assay vs. GC/MS | | | GCMS | | | | | |---------------------|-----|---------------------------------------------------------------|----------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------------------|------------------| | | | LOW NEG<br>Less than<br>50% below<br>the cutoff<br>(<5 ng/mL) | NEG<br>Within 50% below<br>the cutoff<br>(5.0~9.9 ng/mL) | POS<br>Within 50% above<br>the cutoff<br>(10.0~ 15 ng/mL | HIGH POS<br>Greater than<br>50% above<br>the cutoff<br>(>15 ng/mL) | 0/0<br>Agreement | | Qualitative Summary | | | | | | | | Emit® | POS | 0 | | ો ર | ਤੇ ਕੇ | 98% | | | NEG | 49 | ર્ભ | 0 | 0 | 100% | | | | Semiquantitative Summary | | | | | | Emit® | POS | 0 | | ો.રે | 34 | 98% | | | NEG | ਧੇਰੇ | б | 0 | 0 | 100% | {6}------------------------------------------------ ## Discordant Result Summary | Cutoff Value<br>(10 ng/mL) | Qualitative Result<br>(POS/NEG) | | Semiquantitative Result<br>(ng/mL) | | |----------------------------|---------------------------------|-------|------------------------------------|-------| | | Emit® Assav | GC/MS | Emit® Assay | GC/MS | | Sample # 55 | | | 16.3 | 7.8 | ## K. Conclusion: The information provided in this pre-market notification, demonstrates the Emit® II Plus 6- Acetylmorphine Assay and the Emit® 6-AM / Ecstasy Calibrators / Controls are substantially equivalent to the legally marketed predicate devices for their general intended use. Substantial equivalence was demonstrated through comparison of intended use and technological features to the commercially available predicate devices and confirmed by gas chromatography/mass spectrometry (GC/MS), an independent analytical method. The information given in this pre-market notification provides reasonable assurance that the Emit® II Plus 6- Acetylmorphine Assay and Calibrators /Controls are safe and effective for their stated intended use. {7}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 Siemens Healthcare Diagnostics Inc. c/o Ms. Janet Fose Regulatory Technical Specialist. Regulatory Affairs P.O. Box 6101. Mailbox 514 Newark. DE 19714-6101 MAR : 8 2011 Re: k102779 Trade/Device Name: Emit® II Plus 6-Acetylmorphine Assay and Emit® II Plus 6-AM / Ecstasy Calibrator / Control Levels 1 - 4 Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG, DKB and DIF Dated: February 10, 2011 Received: February 11, 2011 Dear Ms. Fose: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {8}------------------------------------------------ Page 2 - If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (3011796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free mumber (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, C.C. Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {9}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Emit® II Plus 6-AcetyImorphine Assay Emit® II Plus 6-AM / Ecstasy Calibrator /Control Level 1 Emit® II Plus 6-AM / Ecstasy Calibrator /Control Level 2 Emit® II Plus 6-AM / Ecstasy Calibrator /Control Level 3 Emit® II Plus 6-AM / Ecstasy Calibrator /Control Level 4 Indications For Use: Emit® II Plus 6-Acetylmorphine Assay The Emit® II Plus 6-Acetylmorphine Assay is a homogeneous enzyme immunoassay with 10 ng/mL cutoff. The assay is intended for use in laboratories for the qualitative and/or semiquantitative analyses of 6-acetylmorphine (6-AM), a heroin metabolite, in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. Semiquantitative test results may be used to assess assay performance as part of a quality control program and to estimate a dilution of the specimen for confirmation by GC/MS. The Emit® II Plus 6-Acetylmorphine Assay provides only a preliminary analytical test result. A more specific alternative chemical must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clincal consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used. Emit® II Plus 6-AM/Ecstasy Calibrators/Controls Calibrators are used in the calibration of the Emit® II Plus 6-AcetyImorphine and Emit® II Plus Ecstasy Assays. Controls may be used as quality control materials based on the specific Emit® II Plus 6-Acetylmorphine and Emit® II Plus Ecstasy Assay cutoffs. Prescription Use ______________________________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) C Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) k102779
Innolitics FDA 510(k) Search
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%