K Number
K102751

Validate with FDA (Live)

Date Cleared
2013-02-11

(872 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LabonaCheck Gluppy Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The LabonaCheck Gluppy Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The LabonaCheck Gluppy Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The LabonaCheck Gluppy Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use.

The LabonaCheck Gluppy Blood Glucose Test Strips are for use with the LabonaCheck Gluppy Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.

The LabonaCheck Gluppy Control Solution is for use with the LabonaCheck Gluppy Blood Glucose Monitoring System as a quality control check to verify that the meter and test strips are working together properly.

Device Description

The system consists of Test Meter, Test Strips, Code key, Lancing Device, lancets, 3V battery and Carrying Case.

The system measures the amount of glucose (sugar) in whole blood. Blood is applied to the absorbent hole of the test strip and automatically drawn into the reaction zone where reaction between reagent and glucose occurs.

AI/ML Overview

The provided text is a 510(k) summary for the LabonaCheck Gluppy Blood Glucose Monitoring System. It describes the device, its intended use, and compares it to a predicate device to establish substantial equivalence. However, it does not contain any information about acceptance criteria, a specific study proving the device meets those criteria, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, or training set details.

The document is a regulatory submission for premarket notification, which focuses on demonstrating substantial equivalence to a legally marketed device rather than providing detailed performance study results with specific acceptance criteria and statistical analysis.

Therefore, I cannot fulfill your request for the specific points about acceptance criteria and detailed study information based on the provided text. The document primarily focuses on:

  • Device Name: LabonaCheck Gluppy Blood Glucose Monitoring System
  • Manufacturer: CERAGEM Medisys Inc.
  • Submission Date: February 11, 2013
  • Predicate Device: RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM, Model GM 100 (K081451)
  • Indication for Use: Quantitative measurement of glucose in fresh capillary whole blood samples from the fingertip for self-testing by people with diabetes at home (not for diagnosis, screening, or neonatal use).
  • Similarities to Predicate: Amperometry detection method, Glucose Oxidase enzyme, Potassium ferricyanide mediator, test range (20-600 mg/dL), temperature range (10-40°C), noble metal electrode, 3-year warranty, CR2032 battery.
  • Differences from Predicate: Sample volume (1.0 uL vs 1.4 uL), humidity range (Below 80% vs 10-90%), open use time for test strips (4 months vs 3 months), coding (Code Key vs No coding), test time (5 seconds vs 8 seconds), hematocrit range (20-60% vs 30-55%), memory (500 results vs 10 results).

To answer your questions, a different type of document, such as a validation study report or a more detailed performance specification document, would be required.

{0}------------------------------------------------

K102751

FEB 1 1 2013

CERAGEM Medisys Inc. www.ceragemmedisys.com

510(k) Summary

In accordance with the requirements of 21 CFR.807.92, the following information about 510(k) safety and effectiveness is being submitted.

CERAGEM

MEDISYS

Submitter CERAGEM Medisys Inc. #103-703, SK Ventium, 522 Dangjeong-dong, Gunpo-si, Gyeonggi-do, 435-776, Republic of Korea Phone : (82) 41-529-8420 Fax : (82) 41-551-0667 Contact Person: Mi - Suk Park Date Prepared 2. February 11, 2013 3. Device Name Common name : LabonaCheck Gluppy Blood Glucose Monitoring System LabonaCheck Gluppy Control Solution Classification : Class II (Regulation: 21 CFR § 862.1345) Class I (Regulation: 21 CFR & 862.1660) Product Codes : NBW(Blood Glucose Test System, Over the Counter), CGA 11X Predicate Device 4. LabonaCheck® Gluppy Blood Glucose Monitoring System is substantially equivalent to The RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM described as below. (1) Device Name: RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM, (2) Model: GM 100 (3) Manufacturer: BIONIME CORPORATION (4) 510(K) Number: K081451 Device Description 5. The system consists of Test Meter, Test Strips, Code key, Lancing Device, lancets, 3V battery and Carrying Case.

The system measures the amount of glucose (sugar) in whole blood. Blood is applied to the

Total 3 pages

{1}------------------------------------------------

absorbent hole of the test strip and automatically drawn into the reaction zone where reaction between reagent and glucose occurs.

  1. Indication for use

MEDISYS

CERAGEN

The LabonaCheck® Gluppy Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The LabonaCheck® Gluppy Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The LabonaCheck® Gluppy Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The LabonaCheck® Gluppy Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The LabonaCheck® Gluppy Blood Glucose Test Strips are for use with the LabonaCheck® Gluppy Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.

The LabonaCheck® Gluppy Control Solution is for use with the LabonaCheck® Gluppy Blood Glucose Monitoring System as a quality control check to verify that the meter and test strips are working together properly.

Comparison
ItemDevicePredicate
Device NameLabonaCheck® GluppyRIGHTEST GM100
Similarities
Detection MethodAmperometryAmperometry
EnzymeGlucose Oxidase(Aspergillus niger)Glucose Oxidase(Aspergillus niger)
MediatorPotassium ferricyanidePotassium ferricyanide
Test range20~600 mg/dL20~600 mg/dL
Temperature range50104°F1040°C50104°F1040°C

Comparison to Predicate Device ` 7.

Total 3 pages

{2}------------------------------------------------

ర్తార్థిt for1are in1216प
y C

CERAGEM Medisys Inc. www.ceragemmedisys.com

Sample Volume1.0 uL1.4 uL
Humidity rangeBelow 80%10~90%
ElectrodeNoble metal electrodeNoble metal electrode
Warranty(Test Meter)3 years3 years
Power(Battery)CR2032CR2032
Differences
Open use time (Test Strips)4 months3 months
CodingCode KeyNo coding
Test Time5 second8 second
Hematocrit range20~60%30-55%
Memory500 blood glucose test resultswith date and time10 blood glucose test resultswith date and time

Conclusion

As shown in the comparison table, the LabonaCheck® Gluppy Blood Glucose Monitoring System and the RIGHTEST GM100 have the same detection method, test range, temperature range, electrode and batteries. Furthermore, the LabonaCheck® Gluppy Blood Glucose Monitoring System also uses the same enzyme and mediator as the RIGHTEST GM100. To sum up the similarities, the LabonaCheck® Gluppy Blood Glucose Monitoring System is similar with the predicate device because most of the specifications deciding the characteristics of the device are the same. In conclusion, despite the differences such as open use time, coding, test time, etc. the LabonaCheck® Gluppy Blood Glucose Monitoring System is substantially equivalent to the predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms or branches extending upwards, representing health and well-being. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the figure.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 11, 2013

Ceragem Medisys Inc. c/o Grace Kim Ceragem International Inc. 3699 Wilshire Blvd., Suite 930 Los Angeles, CA 90010

Re: K102751

Trade/Device Name: LabonaCheck Gluppy Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA, JJX Dated: January 28, 2013 Received: February 07, 2013

Dear Ms. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2— Kim Grace

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol C. Benson for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K102751

Device Name: LabonaCheck Gluppy Blood Glucose Monitoring System

Indications for Use:

The LabonaCheck Gluppy Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The LabonaCheck Gluppy Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The LabonaCheck Gluppy Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The LabonaCheck Gluppy Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use.

The LabonaCheck Gluppy Blood Glucose Test Strips are for use with the LabonaCheck Gluppy Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.

The LabonaCheck Gluppy Control Solution is for use with the LabonaCheck Gluppy Blood Glucose Monitoring System as a quality control check to verify that the meter and test strips are working together properly.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use _ × (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR)

Katherine Serrano

Division Sign-Off Office of In Vitro Devices and Radiologic Health

510(k) K102751

Page 1 of 1

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.