K Number
K102724

Validate with FDA (Live)

Date Cleared
2010-12-20

(90 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

Device Description

Classified by FDA's General and Plastic Surgery Device panel as Class I. 21 CFR 880.6250. Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirement of ASTM Standard D5250-06.

AI/ML Overview

This document is a 510(k) summary for a medical device: "Zibo Yuqun New Material Science and Technology Co. Synthetic Vinyl Patient Examination Gloves - Powder Free".

Here's an analysis of the provided text in relation to the requested information:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Physical and Dimensions Testing: (Per ASTM D5250-06)Meets requirements for Physical and Dimensions Testing (Inspection Level S-2, AQL 2.5)
Pinhole Detection: (Water Fill Test, FDA 1000 ml.)Meets requirements (AQL 2.5, Inspection Level I)
Biocompatibility (Primary Skin Irritation):No primary skin irritant reactions
Biocompatibility (Skin Sensitization/Allergic Contact Dermatitis):No sensitization reactions
Powder Content: (No more than 2 mg powder per glove)Meets "powder-free" claims (contains no more than 2 mg powder per glove)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not explicitly stated for each test, but "samplings" were conducted for the Water Fill Test. For Physical and Dimensions Testing, an "Inspection Level S-2, AQL 2.5" is mentioned, and for the Water Fill Test, an "AQL 2.5, Inspection Level I" is mentioned. These are statistical sampling plans, but the exact number of gloves tested is not provided in this summary.
  • Data Provenance: The manufacturing company is Zibo Yuqun New Material Science and Technology Co. in Zibo, Shandong, China. It is highly probable the testing was conducted in China or by labs affiliated with the manufacturer. The summary does not specify if the testing was retrospective or prospective, but product testing for premarket notification is typically prospective to demonstrate manufacturing controls and product quality.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This device is a patient examination glove, and the "ground truth" is established through standardized physical, chemical, and biological testing as outlined in ASTM standards and FDA guidelines, not by expert medical image interpretation. Therefore, this question is not applicable in the context of this device and study. The "experts" would be the industrial hygienists, chemists, and technicians performing the standard tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This device's performance is determined by objective, standardized tests (e.g., water fill for pinholes, tensile strength, powder content), not by human interpretation or adjudication in the medical context of image analysis.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical product (gloves), not an AI-powered diagnostic or assistive technology. Therefore, MRMC studies and AI-related performance metrics are irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an algorithm or software device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device is based on:

  • Pre-defined Physical Standards: Established in ASTM D5250-06 (e.g., specific dimensions, tensile strength, elongation).
  • Performance Standards: Such as the maximum allowable defect rate for pinholes (AQL 2.5) and maximum powder content (2 mg per glove).
  • Biocompatibility Testing: Standards for primary skin irritation and skin sensitization reactions.

These "truths" are determined through objective laboratory testing against established specifications, not through expert consensus or medical outcomes data in the way it applies to diagnostic devices.

8. The sample size for the training set

Not applicable. This is a physical product, not a machine learning model. There is no "training set."

9. How the ground truth for the training set was established

Not applicable. As this is a physical product and not an AI/ML model, there is no "training set" or ground truth establishment relevant to an algorithm.

{0}------------------------------------------------

K102724

510 (K) SUMMARY

®EC 2 0 2010

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.

1. Submitter's Identification:

3

Zibo Yugun New Material Science and Technology Co. Zhuiiang East Road Zhangdian Economic Development Zone Zibo, Shandong, China

Contact Person:

Mr. Cheng, Fanghai, Assistant General Manager Phone Number: 011-86-15725803988

Date summary prepared: September 17, 2010

2. Name of the Device:

Zibo Yugun New Material Science and Technology Co. Synthetic Vinyl Patient Examination Gloves - Powder Free

3. Predicate Device Informaton:

Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free (K992821)

4. Device Description:

Classified by FDA's General and Plastic Surgery Device panel as Class I. 21 CFR 880.6250. Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirement of ASTM Standard D5250-06.

5. Intended Use:

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

{1}------------------------------------------------

K102724

Comparison to Predicate Devices: 6.

Zibo Yuqun New Material Science and Technology Co. Synthetic Vinyl Patient Examination Gloves, Powder-Free are substantially equivalent in safety and effectiveness to the Shijiazhuang Hongxiang Plastic Products Co., Ltd. and Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, Powder free.

Discussion of Non-Clinical tests performed for Determination of 7. Substantial Equivalence are as follows:

The standards used for Zibo Yuqun New Material Science and Technology Co. glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AOL 2.5.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

There are no special labeling claims and we do not claim our gloves as hypoallergenic is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).

Discussion of Clinical Tests Performed: 8.

Not Applicable - There is no hypoallergenic claim.

Conclusions: 9.

Zibo Yuqun New Material Science and Technology Co. Synthetic Vinyl Patient Examination Gloves, Powder-Free conform fully to ASTM-D-5250-06 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Zibo Yuqun New Material Science and Technology Company C/O Mr. Frank Liu Basic Medical Industries, Incorporated 12390 East End Avenue Chino, California 91710

DEC 2 0 2010

Re: K102724

Trade/Device Name: Patient Vinyl Examination Gloves, Powder Free, Non-Sterile, Non-Colored Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: November 17, 2010 Received: November 18, 2010

Dear Mr. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. .

{3}------------------------------------------------

Page 2- Mr. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other l'ederal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-522 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K102724

Zibo Yuqun New Material Science and Technology Co.

Zhujiang East Road Zhangdian Economic Development Zone Zibo, Shandong, China

DEC 2 0 2010

INDICATIONS FOR USE

Applicant: Zibo Yuqun New Material Science and Technology Co.

510(k) Number: K | 02724

Patient Vinyl Examination Gloves, Powder free, Non-Sterile, Device Name: Non Colored

Indications of Use:

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

Prescription Use

Over the Counter Use X

Factory Initials

Elizabeth S. Lawrence-Webb

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K102724

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.