K Number
K102595

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-12-27

(109 days)

Product Code
Regulation Number
870.2100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medi-Stim VeriQC System is intended for use as an intraoperative system utilizing ultrasonography to visualize blood flow and to guide surgeons to successfully plan and accomplish surgical interventions. The clinical indications for the device are:

  1. Accurate transit time blood volume and Doppler velocity flow measurements during cardiovascular-, vascular- and transplantation-surgery.
  2. Simultaneous measurements of blood pressure, vascular resistance, interfaced physiological signals and other derived parameters during these procedures.
  3. Detection of normal and abnormal blood volume and Doppler velocity flow patterns during these procedures.
  4. Provides guidance to prepare surgical plans at the initiation of surgery and to support the successful accomplishment of surgery including detection and location of vessels during surgical procedures.
  5. Detection and quantification of the degree of stenosis in arteries by using the Doppler velocity profile.
Device Description

The VeriQC system incorporates several ultrasound modalities that can be used during a variety of surgical interventions. The system utilizes the well established technology of transit-time flow measurements to accurately measure blood flow in veins and arteries intraoperatively. In addition, the system provides ultrasound echocardiography for intraoperative imaging of the cardiovascular circulatory system. The system supports both B-Mode (2D) grayscale imaging, color Doppler (CFM) and pulsed wave Doppler (PW-Doppler). The system also has the ability to connect other external physiological signals such as blood pressure, ECG and other auxiliary signals provided by other monitoring systems.
The VeriQC transducer selection includes both transit-time probes for blood flow measurements on various vessel sizes and types as well as an intraoperative imaging probe.

AI/ML Overview

This 510(k) summary for the MediStim VeriQC System primarily focuses on demonstrating substantial equivalence to predicate devices based on intended use, technology, and safety testing rather than a clinical performance study with specific acceptance criteria.

Therefore, many of the requested sections regarding clinical study design for performance acceptance criteria cannot be extracted directly from this document.

Here's an attempt to answer based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

This document does not specify quantitative clinical acceptance criteria or a direct clinical performance study against such criteria. The "Summary of Testing" states: "Testing included EMC, Safety testing and Acoustic Safety Measurements. The MediStim VeriQC has also been verified and validated according to functional and software requirements." This refers to engineering and regulatory compliance, not clinical performance metrics like sensitivity, specificity, or accuracy compared to a ground truth in a patient population.

Therefore, a table cannot be populated as requested. The submission focuses on substantial equivalence to predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

No clinical test set is described or referenced in this document for the purpose of evaluating clinical performance against specific acceptance criteria. The submission relies on the established safety and effectiveness of its predicate devices and its own technical validation.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. No clinical test set with ground truth established by experts is described for performance evaluation.

4. Adjudication Method for the Test Set

Not applicable. No clinical test set requiring adjudication is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Reader Improvement

No MRMC comparative effectiveness study is mentioned or implied in this 510(k) summary. The submission focuses on the device's technical capabilities and its equivalence to other devices, not on improving human reader performance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

This device (MediStim VeriQC System) is an intraoperative ultrasound and flow measurement system. Its primary function involves direct measurements and imaging, often in real-time by a surgeon or clinician. It is not an AI algorithm that would typically undergo standalone performance testing as understood in the context of diagnostic AI. The "algorithm" here refers to the underlying system's measurement and imaging processes, which are verified through functional and software requirements.

7. The Type of Ground Truth Used

For the functional and software verification/validation mentioned, the ground truth would typically be established by:

  • Engineering specifications and standards: For EMC, safety, and acoustic safety.
  • Reference measurements/phantoms: For validation of flow measurements and imaging capabilities against known physical properties.
  • Comparison to predicate device performance (implicitly): The claim of substantial equivalence relies on the predicate devices having established safety and effectiveness, meaning their performance served as an implicit "ground truth" for what is acceptable in the market.

No explicit clinical ground truth (e.g., pathology, clinical outcomes, expert consensus on patient data) is described as being used for performance testing in this summary.

8. The Sample Size for the Training Set

Not applicable. This document does not describe the development or training of an AI algorithm in the contemporary sense. The system's "training" would be through its engineering design, calibration, and software development processes to meet its stated functionalities.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there's no mention of a training set for an AI model, the method for establishing ground truth for such a set is not relevant here.

{0}------------------------------------------------

510(k) Summary

K102595 DEC 2 7 2010

Submitter: MediStim ASA Fernanda Nissensgt. 3 N-0484 Oslo Norway +47 2305 9660

Contact Information: Constance G. Bundy C. G. Bundy Associates, Inc. 435 Rice Creek Terrace NE Fridley, MN 55432 763-574-1976

Submission Date: September 3, 2010

Device Name and Classification: MediStim VeriQC System, Medical ultrasonic imaging and volume flow-meter with probes, Class II

Product Code:

CFR NumberProduct Code
Cardiovascular blood flowmeter870.2100DPW
Ultrasonic Pulsed Doppler Imaging System892.1550IYN
Ultrasonic Pulsed Echo Imaging System892.1560IYO
Diagnostic Ultrasound Transducer892.1570ITX

Equivalent Device Identification:

The MediStim VeriQ is substantially equivalent to the MediStim VeriQ model VQ4122 (K040228) manufactured by MediStim ASA, Fernanda Nissensgate 3, N-0484 OSLO, with added features equivalent to the GE Vivid 7 Diagnostic Ultrasound System with Diagnostic ultrasound transducers (K041552) manufactured by GE Medical systems, Ultrasound and Care Diagnostics, LLC PO Box 414, Milwaukee, WI 53201.

{1}------------------------------------------------

Device Description:

The VeriQC system incorporates several ultrasound modalities that can be used during a variety of surgical interventions. The system utilizes the well established technology of transit-time flow measurements to accurately measure blood flow in veins and arteries intraoperatively. In addition, the system provides ultrasound echocardiography for intraoperative imaging of the cardiovascular circulatory system. The system supports both B-Mode (2D) grayscale imaging, color Doppler (CFM) and pulsed wave Doppler (PW-Doppler). The system also has the ability to connect other external physiological signals such as blood pressure, ECG and other auxiliary signals provided by other monitoring systems.

The VeriQC transducer selection includes both transit-time probes for blood flow measurements on various vessel sizes and types as well as an intraoperative imaging probe.

Intended Use:

The Medi-Stim VeriQC System is intended for use as an intraoperative system utilizing ultrasonography to visualize blood flow and to guide surgeons to successfully plan and accomplish surgical interventions. The clinical indications for the device are:

  1. Accurate transit time blood volume and Doppler velocity flow measurements during cardiovascular-, vascular- and transplantation-surgery.

  2. Simultaneous measurements of blood pressure, vascular resistance, interfaced physiological signals and other derived parameters during these procedures.

  3. Detection of normal and abnormal blood volume and Doppler velocity flow patterns during these procedures.

  4. Provides guidance to prepare surgical plans at the initiation of surgery and to support the successful accomplishment of surgery including detection and location of vessels during surgical procedures.

  5. Detection and quantification of the degree of stenosis in arteries by using the Doppler velocity profile.

{2}------------------------------------------------

Comparison Table:

MediStim VeriQCMediStim VeriQGE Vivid 7
Indications for use"The Medi-Stim VeriQCSystem is intended foruse as an intraoperativesystem utilizingultrasonography tovisualize blood flow andto guide surgeons tosuccessfully plan andaccomplish surgicalinterventions."The Medi-Stim VeriQSystem is aninteraoperative diagnosticsystem that utilizesultrasonography to guidesurgeons to successfullyplan and accomplishsurgical interventions.The device is intendedfor use by a qualifiedphysician for ultrasoundevaluation of Fetal;Abdominal (includingrenal and GYN);Pediatric; Small Organ(breast, testes, thyroid);Neonatal Cephalic; AdultCephalic, Cardiac (adultand pediatric); PeripheralVascular (PV); Musculo-skeletalConventional;Urology(including prostate),Transesophageal;Transrectal (TR);Transvaginal (TV); andInteraoperative(abdominal, thoracic andvascular)
The clinical indicationsfor the device are:1) Accurate transit timeblood volume andDoppler velocity flowmeasurements duringcardiovascular-, vascular-and transplantation-surgery.2) Simultaneousmeasurements of bloodpressure, vascularresistance, interfacedphysiological signals andother derived parametersduring these procedures.3) Detection of normaland abnormal bloodvolume and Dopplervelocity flow patternsduring these procedures.4) Provides guidance toprepare surgical plans atthe initiation of surgeryand to support thesuccessfulThe clinical indicationsfor the device are:1) Accurate transit timeblood volume andDoppler velocity flowmeasurements duringcardiovascular-, vascular-, transplantation- andneuro-surgery.2) Simultaneousmeasurements of bloodpressure, vascularresistance, interfacedphysiological signals andother derived parametersduring these procedures.3) Detection of normaland abnormal bloodvolume and Dopplervelocity flow duringthese procedures.
accomplishment ofsurgery includingdetection and location ofvessels during surgicalprocedures.5) Detection andquantification of thedegree of stenosis inarteries by using theDoppler velocity profile."4) Provides guidance toprepare surgical plans atthe initiation of surgeryand to support thesuccessfulaccomplishment ofsurgery includingdetection and location ofvessels during surgicalprocedures5) Detection andquantification of thedegree of stenosis inarteries by using theDoppler velocity profile.
Patient groupAdult and pediatricAdult and pediatricNeonatal, adult andpediatric
Ultrasound modalitiesTransit-time, B-mode,Color Doppler, PowerDoppler and PW Dopplerultrasound, CombinedMode.Transit-time, PWDopplerB-Mode, M-Mode, PWDoppler, Color Doppler,Color M Doppler, PowerDoppler, Combinedmodes, Coded Pulse
Other inputsBlood pressure, ECG,Auxiliary inputsBlood pressure, ECG,Auxiliary inputsNot available in summary

{3}------------------------------------------------

Summary of Testing: Testing included EMC, Safety testing and Acoustic Safety Measurements. The MediStim VeriQC has also been verified and validated according to functional and software requirements.

Conclusion: The indication for use statements of the two predicate devices with regard to patient groups and surgical procedures are equivalent to the indications for use statement of the Medi-Stim VeriQC System.

The main clinical applications for all devices are the same namely cardiovascular-, and transplantation-surgery.

The technology platform for the VeriQ and the VeriQ predicate device are identical with respect to computing platform, operating system and TTFM acquisition hardware. The ultrasound frontend of the VeriQ is based on the same principles as the predicate GE device, providing a modern fully digitized ultrasound acquisition module. The devices share the same imaging modalities and base the acoustic safety system on the same principles.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC 2 7 2010

MediStim ASA C/O Constance G. Bundy C. G. Bundy Associates, Inc. 435 Rice Creek Terrace NE Fridley, MN 55432

Re: K102595

Trade/Device Name: MediStim VeriQC System Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular blood flowmeter Regulatory Class: Class II (Two) Product Code: DPW, IYO, IYN, and ITX Dated: December 12, 2010 Received: December 15, 2010

Dear Ms. Bundy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{5}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

K102595

510(k) Number (if known): K102595

DEC 2 7 2010

Device Name: MediStim VeriOC system

Indications For Use:

The Medi-Stim VeriQC System is intended for use as an intraoperative system utilizing ultrasonography to visualize blood flow and to guide surgeons to successfully plan and accomplish surgical interventions." The clinical indications for the device are:

    1. Accurate transit time blood volume and Doppler velocity flow measurements during cardiovascular-, vascular- and transplantation-surgery.
    1. Simultaneous measurements of blood pressure, vascular resistance, interfaced physiological signals and other derived parameters during these procedures.
    1. Detection of normal and abnormal blood volume and Doppler velocity flow patterns during these procedures.
    1. Provides guidance to prepare surgical plans at the initiation of surgery and to support the successful accomplishment of surgery including detection and location of vessels during surgical procedures.
    1. Detection and quantification of the degree of stenosis in arteries by using the Doppler velocity profile.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number K102595

§ 870.2100 Cardiovascular blood flowmeter.

(a)
Identification. A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.(b)
Classification. Class II (performance standards).