K Number
K102586

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-03-18

(190 days)

Product Code
Regulation Number
868.5905
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The S9 VPAP Adapt is indicated for the treatment of patients weighing more than 66 lb (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. It is intended for home and hospital use.

Device Description

S9 VPAP Adapt System (S9 VPAP Adapt with H5i) is similar to the predicate device(s), using a blower based positive pressure system with an integrated heater controller. The device platform is similar to the S8 Aspen with H4i Plus (K091947) and contains a Micro-processor controlled blower system that generates controlled positive airway pressure between 3-25 cmHzO as required to maintain an "air splint" for effective treatment of OSA. The system comprises the flow generator, patient interface) and humidifier.

The S9 VPAP Adapt is a flow generator device designed to provide adaptive servo-ventifation therapy (ASV) to stabilize a patient's ventilation during sleep. The device continually measures the patient's instantaneous ventilation, and calculates a target ventilation based on to the patient's recent average. It then adjusts the degree of pressure support to servo-control the patient's ventilation to at least equal the target ventilation. Therapy modes contained in the S9 VPAP Adapt are CPAP, CPAP with EPR, and Auto Servo Ventilation (ASV). Therapy modes come from the VPAP Tx system (K092186) and VPAP Adapt (K051364).

The functional characteristics of the S9 VPAP Adapt system includes all the dinician and user friendly features of the predicate devices.

AI/ML Overview

The S9 VPAP Adapt is a non-continuous ventilator indicated for the treatment of patients weighing over 30 kg with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. It is intended for home and hospital use.

The provided document describes a 510(k) submission for the S9 VPAP Adapt, focusing on demonstrating substantial equivalence to predicate devices rather than a standalone clinical study with detailed performance metrics against specific acceptance criteria.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly provide a table of acceptance criteria with corresponding performance results for the S9 VPAP Adapt. Instead, it states that "All tests (predicate bench testing) confirmed the product met the predetermined acceptance criteria." However, specific quantitative criteria or performance values are not listed.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "predicate bench testing and clinical studies" were used to show substantial equivalence. However, it does not provide any details regarding the sample size used for the test set or the data provenance (e.g., country of origin of the data, retrospective or prospective) for these studies.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

There is no information provided in the document about experts used to establish ground truth for any test set or their qualifications. The evaluation primarily relies on "predicate bench testing and clinical studies" for substantial equivalence.

4. Adjudication Method for the Test Set

The document does not describe any adjudication method (e.g., 2+1, 3+1, none) for a test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study or any effect size of how much human readers improve with AI vs. without AI assistance. The S9 VPAP Adapt is described as a device that provides adaptive servo-ventilation therapy, not an AI-based diagnostic or assistive tool for human readers.

6. Standalone Performance Study (Algorithm Only)

While the device includes a "Micro-processor controlled blower system" and "adjusts the degree of pressure support to servo-control the patient's ventilation," the document does not describe a standalone performance study for an algorithm in isolation without human interaction. The device's operation is inherently integrated into a patient's breathing cycle.

7. Type of Ground Truth Used

The document does not explicitly state the type of ground truth used in the "clinical studies" mentioned. Given the nature of the device (ventilator for sleep apnea), it is most likely that ground truth would be established through polysomnography (PSG) or other objective measures of respiratory events and ventilation effectiveness, but this is not specified.

8. Sample Size for the Training Set

The document does not provide information on the sample size used for any training set.

9. How Ground Truth for the Training Set Was Established

The document does not provide information on how ground truth was established for any training set.

Summary of Evidence Regarding Acceptance Criteria and Studies:

The submission for the S9 VPAP Adapt focuses on demonstrating substantial equivalence to pre-existing predicate devices (VPAP Tx, S8 Aspen with H4i Plus, VPAP Adapt). The core argument for meeting acceptance criteria is that "All tests (predicate bench testing) confirmed the product met the predetermined acceptance criteria" and that "predicate bench testing and clinical studies were used to show substantial equivalence." However, the abstract does not provide the detailed quantitative acceptance criteria, the specific results against those criteria, or the methodology and sample sizes of the mentioned "clinical studies." The document primarily affirms that the new device has "not altered the safety and effectiveness of CPAP treatment" and complies with relevant FDA guidance documents.

{0}------------------------------------------------

510(k) Summary - S9 VPAP Adapt

MAR 1 8 2011

102

86

Date Prepared

21* Feb, 2011

Official Contact David D'Cruz VP, US Medical and Regulatory Affairs ResMed Corp 9001 Spectrum Center Blvd. San Diego CA 92123 USA

Tel:(858) 836-5984
Fax:(858) 836-5522
Classification Reference21 CFR 868.5905
Product Code73 BZD
Common/Usual NameNon continuous ventilator (IPPB).
Proprietary NameS9 VPAP Adapt
Predicate Device(s)VPAP Tx (K092186)S8 Aspen with H4i Plus (K091947)VPAP Adapt (K051364)
Reason for submissionNew Device

......

{1}------------------------------------------------

Indication for Use

The S9 VPAP Adapt is indicated for the treatment of patients weighing more than 66 lb (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. It is intended for home and hospital use.

Substantial Equivalence

The new device has the following similarities to the previously cleared predicate devices.

  • Same intended use

  • Same operating principle >
  • Similar technologies A
  • Same manufacturing process >

Design and Verification activities were performed on the S9 VPAP Adapt System as a result of the risk and design requirements. All tests (predicate bench testing) confirmed the product met the predetermined acceptance criteria. ResMed has determined that the new device has not altered the safety and effectiveness of CPAP treatment for patients with Obstructive Sleep Apnoea (OSA), central and/or mixed apneas, or periodic breathing who weigh more than 66 lb (>30 kg). The new device complies with the applicable requirements referenced in the FDA guidance documents:

  • FDA Reviewer Guidance for Premarket Notification Submissions (November 1993) A
  • FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)

Predicate bench testing and clinical studies were used to show substantial equivalence between the SS VPAP Adapt and VPAP Tx system (K092186).

Device Description

S9 VPAP Adapt System (S9 VPAP Adapt with H5i) is similar to the predicate device(s), using a blower based positive pressure system with an integrated heater controller. The device platform is similar to the S8 Aspen with H4i Plus (K091947) and contains a Micro-processor controlled blower system that generates controlled positive airway pressure between 3-25 cmHzO as required to maintain an "air splint" for effective treatment of OSA. The system comprises the flow generator, patient interface) and humidifier.

The S9 VPAP Adapt is a flow generator device designed to provide adaptive servo-ventifation therapy (ASV) to stabilize a patient's ventilation during sleep. The device continually measures the patient's instantaneous ventilation, and calculates a target ventilation based on to the patient's recent average. It then adjusts the degree of pressure support to servo-control the patient's ventilation to at least equal the target ventilation. Therapy modes contained in the S9 VPAP Adapt are CPAP, CPAP with EPR, and Auto Servo Ventilation (ASV). Therapy modes come from the VPAP Tx system (K092186) and VPAP Adapt (K051364).

The functional characteristics of the S9 VPAP Adapt system includes all the dinician and user friendly features of the predicate devices.

Conclusion

The S9 VPAP Adapt is substantially equivalent to the Predicate devices, VPAP Tx system (K092186), VPAP Adapt (K051364) and S8 Aspen with H4i Plus (K091947).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

ResMed Limited C/O Mr. David D'Cruz V.P., Clinical & Regulatory Affairs ResMed Corporation 9001 Spectrum Center Boulevard San Diego, California 92123

MAR 1 8 2011

Re: K102586

Trade/Device Name: S9 VPAP Adapt Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: February 21, 2011 Received: March 11, 2011

Dear Mr. D'Cruz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class 111 (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. D'Cruz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Nh for

Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indication for Use

510(k) Number (if known):

Device Name: S9 VPAP Adapt

Indication for Use.

The S9 VPAP Adapt is indicated for the treatment of patients weighing more than 66 Ib (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. It is intended for horne and hospital use.

Prescription Use _ X ___ AND/OR Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of Device Evaluation (ODE)

Page 1 of__1_

L. Schultheis

(Division Sign-Off) Division of Ancsthesiology, General P Infection Control and Demail D 510(k) Number:

Page 12

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).