K Number
K102400

Validate with FDA (Live)

Date Cleared
2011-08-01

(342 days)

Product Code
Regulation Number
866.3830
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Syphilis Health Check is a qualitative rapid membrane immunochromatographic assay for the detection of Treponema pallidum (syphilis) antibodies in human whole blood, serum or plasma. This product can be used as an initial screening test or in conjunction with a non-treponemal laboratory test and clinical findings, may aid in the diagnosis of syphilis infection. This test is not intended for use in screening blood or plasma donors.

Device Description

SYPHILIS HEALTH CHECK is a rapid qualitative screening test for detection of human antibodies to TP in serum, plasma or whole blood. The method employs an unique combination of anti-human immunoglobulins gold conjugate and highly purified TP recombinant proteins to specifically detect anti-TP antibodies. The test mainly detects IgG and IgM will also react in case of high concentrations. As the samples flow through the absorbent device, the anti-human immunoglobulins/protein A dye conjugate binds to the human immunoglobulins forming an antigen-antibody complex. This complex binds to the recombinant protein in the positive reaction zone and produces a pink-rose colored band. In the absence of anti TP antibodies, there is no line in the positive reaction zone. The reaction mixture continues flowing through the absorbent device past the reaction and control zones. Unbound conjugate binds to the reagents in the control zone producing a pink-rose color band, demonstrating that the reagents are functioning correctly.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Syphilis Health Check device, based on the provided 510(k) summary:

Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Predicate Device K001552)Reported Device Performance (Syphilis Health Check)
Sensitivity (Treponemal tests)> 97%> 98% (overall: 98.5%)
Specificity (Treponemal tests)> 97%> 98% (overall: 97.3%)
Positive Agreement (RPR total)Not explicitly stated (predicate is EIA)95.6% (95% C.I. = 93.4 - 97.2%)
Negative Agreement (RPR total)Not explicitly stated (predicate is EIA)90.5% (95% C.I. = 88.2 - 92.4%)
Overall Agreement (RPR total)Not explicitly stated (predicate is EIA)92.4% (95% C.I. = 90.8 - 93.8%)
Positive Agreement (Treponemal total)Not explicitly stated (predicate is EIA)98.5% (95% C.I. = 97.1 - 99.4%)
Negative Agreement (Treponemal total)Not explicitly stated (predicate is EIA)97.3% (95% C.I. = 95.9 - 98.4%)
Overall Agreement (Treponemal total)Not explicitly stated (predicate is EIA)97.8% (95% C.I. = 96.9 - 98.6%)

Note: The predicate device uses Enzyme ImmunoAssay (EIA) with a stated sensitivity and specificity of >97%. The Syphilis Health Check is a rapid immunochromatographic membrane assay, so direct acceptance criteria for RPR comparisons are not explicitly stated for the predicate, but the overall performance values of the new device are provided for both RPR and treponemal comparisons.

Study Information

2. Sample size used for the test set and the data provenance:

  • Total Test Sample Size: 1292 samples
    • Prospective Samples: 880 patient samples
      • Provenance: Collected from five clinical study sites (four STD clinics and one hospital clinic) in the U.S. Patients were suspected positive for syphilis and exhibiting symptoms.
    • Retrospective Samples: 412 frozen samples
      • Provenance: Purchased from outside commercial vendors and blood centers. These included 149 banked RPR and treponemal reactive serum samples, 28 serum samples from primary/secondary patients (treated but symptomatic), and 138 frozen serum/plasma samples.
    • Clinically Diagnosed Samples: 164 well-characterized serum samples (from treated and untreated patients with primary, secondary, and latent syphilis infections).
      • Provenance: Obtained from a clinic serving a population with various infectious diseases.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The document does not specify the number of individual experts or their qualifications (e.g., radiologist with 10 years of experience) used to establish the ground truth for the test set.
  • Ground truth was established by "reference methods" such as RPR (Non-Treponemal), TPPA, TPHA, FTA, ELISA, and MHA-TP. These are established laboratory tests, implying that the ground truth was determined by qualified laboratory personnel performing these assays, not necessarily by clinical experts adjudicating cases beyond the initial patient suspicion and clinical diagnosis for the clinically diagnosed samples.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • The document does not explicitly describe an adjudication method for conflicting results between reference methods or between the device and reference methods (e.g., 2+1, 3+1).
  • For the "Suspected and Known Positive Syphilis Samples" retrospective study, it mentions that some samples with discrepancies between initial RPR/Treponemal results and Syphilis Health Check were "further tested by an outside laboratory for TPPA titer and Syphilis Health Check results." This implies a form of re-testing or confirmatory testing for discordant results but not a structured expert adjudication panel.
  • For the "Clinically Diagnosed" samples, a panel of 164 samples was "well-characterized" and tested with RPR, TPPA, and FTA-ABS as reference assays, suggesting these reference methods collaboratively defined the "known clinical status."

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC comparative effectiveness study was done. This device is a rapid diagnostic test (immunochromatographic assay) designed for standalone use, not a diagnostic imaging AI tool requiring human-in-the-loop performance measurement or evaluation of human reader improvement. The "readers" of this test are individuals visually interpreting a pink-rose colored band.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, a standalone performance study was done. The entire performance evaluation presents the Syphilis Health Check test's results in isolation, comparing its output directly to that of established reference methods (RPR, TPPA, TPHA, FTA, ELISA, MHA-TP). The visual interpretation of the test result (presence or absence of a colored band) is the direct output of the device, without human interpretive assistance beyond reading the test itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth was established by reference laboratory assays (established Treponemal and non-Treponemal tests). These include RPR, TPPA, TPHA, FTA, ELISA, and MHA-TP.
  • For the "clinically diagnosed" samples, the ground truth was based on "well-characterized clinically diagnosed serum samples" using these reference assays to confirm clinical status (primary, secondary, or latent syphilis).

8. The sample size for the training set:

  • The document does not specify a training set sample size. As an immunochromatographic rapid diagnostic test, it is a chemical/biological assay and not an AI algorithm that typically undergoes a distinct "training phase" with a labeled dataset in the same way machine learning models do. Its "development" would involve optimization of reagents, concentrations, and assay design rather than data-driven machine learning training.

9. How the ground truth for the training set was established:

  • Since there's no explicit "training set" for an AI algorithm outlined, this question is not directly applicable. The performance characteristics of the device were evaluated against established reference methods, as described in point 7.

{0}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K102400

Submitted By:Diagnostics Direct, LLC376 96th StreetStone Harbor, NJ 08247
Telephone(609) 368-6912

Official Correspondent: Gary Lehnus, Ph.D. Lehnus & Associates Consulting

(570) 620-0198 Telephone

FAX (570) 620-0199

August 24, 2010 Date Prepared:

Trade or Proprietary Name: Syphilis Health Check

Common or Usual Name: Rapid immunochromatographic membrane assay, Treponema pallidum antibody.

Regulation21 CFR 866.3830
-----------------------------

Class II (Special Controls) Classification

Product Codes LIP

Panel: Immunology and Microbiology (83)

Predicate Device: K001552 - Phoenix Bio-Tech - TREPCHEK TREPONEMAL ANTIBODY EIA

Device Description SYPHILIS HEALTH CHECK is a rapid qualitative screening test for detection of human antibodies to TP in serum, plasma or whole blood.

The method employs an unique combination of anti-human immunoglobulins gold conjugate and highly purified TP recombinant proteins to specifically detect anti-TP antibodies. The test mainly detects IgG and IgM will also react in case of high concentrations.

As the samples flow through the absorbent device, the anti-human immunoglobulins/protein A dye conjugate binds to the human immunoglobulins forming an antigen-antibody complex. This complex binds to the recombinant protein in the positive reaction zone and produces a pink-rose colored band. In the absence of anti TP antibodies, there is no line in the positive reaction zone. The reaction mixture continues flowing through the absorbent device past the reaction and control zones. Unbound conjugate binds to the reagents in the control zone producing a pink-rose color band, demonstrating that the reagents are functioning correctly.

Intended Use Syphilis Health Check is a qualitative rapid membrane immunochromatographic assay for the detection of Treponema pallidum (syphilis) antibodies in human

1

{1}------------------------------------------------

whole blood, serum or plasma. This product can be used as an initial screening test or in conjunction with a non-treponemal laboratory test and clinical findings, may aid in the diagnosis of syphilis infection. This test is not intended for use in screening blood or plasma donors.

Indications for Use Syphilis Health Check is a qualitative rapid membrane

immunochromatographic assay for the detection of Treponema pallidum (syphilis) antibodies in human whole blood, serum or plasma. This product can be used as an initial screening test or in conjunction with a non-treponemal laboratory test and clinical findings, may aid in the diagnosis of syphilis infection. This test is not intended for use in screening blood or plasma donors.

TECHNOLOGICAL CHARACTERISTICS

The following tables summarize similarities and differences between Syphilis Health Check and the current TREPCHEK Treponemal Antibody EIA (K001552):

ITEMPREDICATE DEVICENEW DEVICE
Device nameTREPCHEKSyphilis Health Check
AnalyteT. pallidum antibodiesT. pallidum Ig antibodies(Detection of IgG and IgMisotypes)
SpecimensSerum and plasmaWhole Blood, Serum, andPlasma
MethodEnzyme ImmunoAssayRapid Immunochromato-graphic membrane
Test principleTREPCHEK is an enzyme-immunoassay utilizing specificrecombinant treponemalantigens immobilized tomicrowells. Anti-treponemalspecific antibodies in thepatient's serum or plasma bindto the antigens. Non-boundproteins are removed duringthe washing step. HorseradishPeroxidase-labelled goat anti-human IgG is added to bindwith the human antibodies inthe well. The unboundconjugate is removed by asecond washing step; then thechromogenic substrate TMB isadded. After an appropriateincubation period, a stopsolution is added and theabsorbance of the resultingcolour is measuredphotometrically with an EIAreader (450 nm). Colourintensity is proportional to theamount of specific antibodypresent in the patient's sample.Specific, recombinant treponemalantigens are immobilized on themembrane. Patient samples orcontrols are added to the samplewell, followed by a buffer, and asthe sample flows through theabsorbent device, the humanimmunoglobulins bind to theantigens forming an antigen-antibody complex. The antigen-antibody complex is furtherreacted with a unique combinationof colloidal gold conjugatedProtein A and anti-humanimmunoglobulins producing a pink-rose colored band that can be readvisually indicating antibody ispresent in the patient's sample. Aseparate Control Line indicatesreagents are functioning correctly

{2}------------------------------------------------

ITEMPREDICATE DEVICENEW DEVICE
Antigen UsedRecombinant proteinssame
Detectioncolorimetric detection - opticaldensityColloidal Gold - pink-red linevisually read
CalculationQualitative determination withratiometric values CalibratorQualitative determination fromthe Test Line
Quality control2 Controls at different levelsPositive and Negative Controls
Indication foruseAid in the diagnosis of syphilisdisease.same
SampleDilutionDiluted 100 µL per wellUndiluted 25 µL or 50µL per testwell
IncubationTime30 min / 30 min / 15 min10 - 15 min.
IncubationTemperatureRoom Temperature (18 - 25°C)Room Temp (18 - 25°C)
ConjugateAnti-human IgG - HRPOProtein A / Anti-human Ig -Colloidal gold
Sensitivity> 97%>98%
Specificity> 97%>98%

Performance: The following are performance characteristics of Syphilis Health Check

In order to evaluate the performance of the Syphilis Health Check test as series of 880 patient samples were obtained that were from prospective study sites and 412 frozen retrospective samples purchased from outside commercial vendors and blood centers. In addition, a series of 164 clinically diagnosed samples were obtained that were known primary, secondary, and latent both treated and untreated. The total of 1292 known retrospective samples (576) and prospectively collected (880) patient samples were used to demonstrate the performance of the Syphilis Health Check test to Non-treponemal RPR and other treponemal tests.

Prospective Studies were conducted at five clinical study sites which compared the Syphilis Health Check to RPR, a non-treponemal test, and treponemal tests such as TPPA, TPHA, or ELISA, using specimens from patients coming into four STD clinics and one hospital clinic. The patients enrolled in the study were identified by medical associates as suspected positive for syphilis and exhibiting symptoms.

Initial evaluations were performed at a university clinic and a hospital clinic to assess the performance of the Syphilis Health-Check test versus RPR and their reference treponemal tests -FTA and TPHA. Only gender and age were collected from these patients. A more comprehensive study was performed at three study sites to collect further patient history information in order to identify a broader range of STD related patients. The information and data collected from these sites is presented below.

The Syphilis Health Check assay demonstrated 95.6% and 98.5% Percent Positive Agreement versus the non-treponemal test and treponemal tests respectively, and 90.5% and 97.3% Percent Negative Agreement respectively.

CUMULATIVE COMPARISON RESULTS

The results obtained for the prospective samples vields the following results compared to Non-treponemal RPR and treponemal tests.

{3}------------------------------------------------

TOTAL Non-Treponemal Comparison - total combined sample results are presented from the 5 prospective sites and the frozen known and suspected positive samples.

RPR - Non-Treponemal
PositiveNegativeTotal
Syphilis HealthCheckPositive47376549
Negative22721743
Total4957971292

RPR - Non-Treponemal

Percent Positive Agreement: 473/495 = 95.6% (95% C.I. = 93.4 - 97.2%) Percent Negative Agreement: 721/797 = 90.5% (95% C.I. = 88.2 - 92.4%) Percent Overall Agreement: 1194/1292 = 92.4% (95% C.I. = 90.8 - 93.8%)

TOTAL Treponemal Comparison - total combined sample results are presented from the 5 sites and the frozen known and suspected positive samples

Reference Treponemal tests
PositiveNegativeTotal
Syphilis HealthCheckPositive53120551
Syphilis HealthCheckNegative8733741
Total5397531292

Percent Positive Agreement: 531/539 = 98.5% (95% C.I. = 997.1 - 99.4%) Percent Negative Agreement: 733/753 = 97.3% (95% C.I. = 95.9 - 98.4%) Percent Overall Agreement: 1264/1292 = 97.8% (95% C.I. = 96.9 - 98.6%)

PROSPECTIVE STUDIES

University Clinic site

RPR
posnegTotal
Syphilispos32133
HealthCheckneg066
Total32739

Percent Positive Agreement = 100.0% (95% C.I. = 89.1 - 100%) Percent Negative Agreement = 85.7% (95% C.I. = 42.1 - 99.6%) Percent Overall Agreement = 97.4% (95% C.I. = 86.5 - 99.9%)

FTA
posnegTotal
27633
066
271239

100% (95% C.I. = 87.2 - 100%) 50.0% (95% C.I. = 21.1 - 78.9%) 84.6% (95% C.I. = 69.5 - 94.1%)

{4}------------------------------------------------

Hospital Clinic site

RPR
posnegTotal
Syphilispos336
HealthCheckneg04444
Total34750
TPHA
posnegTotal
606
04444
64450

Percent Positive Agreement = 100.0% (95% C.I. = 29.2 - 100.0% (95% C.I. = 54.1 - 100%) Percent Negative Agreement = 93.6% (95% C.I. = 82.5 - 98.7%) 100.0% (95% C.I. = 92.0 - 100%) Percent Overall Agreement = 94.0% (95% C.I. = 83.5 - 98.7%) 100.0% (95% C.I. = 92.9 - 100%)

Study Site 1

RPR
posnegTotal
SyphilisHealthCheckpos131528
neg2370372
Total15385400
TPPA
posnegTotal
21829
6365371
27373400

Percent Positive Agreement = 86.7% (95% C.I.= 59.5 - 98.3) Percent Negative Agreement = 96.1% (95% C.I.= 93.7 - 97.8) Percent Overall Agreement = 95.8% (95% C.I.= 93.3 - 97.5)

77.8% (95% C.I.= 57.7 - 91.4) 97.9%(95% C.I.= 95.8 - 99.1) 96.5% (95% C.I.= 94.2 - 98.1)

It should be noted that two (2) of the false negative samples vs TPPA were only positive for TPPA but negative with the other two reference methods (RPR and FTA). Three samples were positive for TPPA and FTA, but negative for Syphilis Health Check and RPR.

Study Site 2

RPR
SyphilisHealthCheckposnegTotal
pos224
neg08585
Total28789

Percent Positive Agreement = 100.0% (95% C.I. = 15.8 - 100) Percent Negative Agreement = 97.7% (95%C.I. = 91.9 - 99.7) Percent Overall Agreement = 97.8% (95% C.I. = 92.1 - 100)

TPPA
posnegTotal
404
08585
48589

100.0% (95% C.I. = 39.8 - 100%) 100.0% (95% C.I. = 95.8 - 100%) 100.0% (95% C.I. = 95.9 - 100%)

{5}------------------------------------------------

Study Site 3

RPR
posnegTotal
SyphilisHealthCheck6410
0195195
Total6199205

P P A = 100.0% (95% C.I.= 54.1 - 100) P N A = 98.0% (95% C.I.= 94.9 - 99.4) P O A = 98.0% (95% C.I.= 95.1 - 99.5)

EIA
posnegTotal
9211
1193194
10195205

90.0% (95% C.I.= 55.5 - 99.7%) 99.0% (95% C.I.= 96.3 - 99.9%) 98.5% (95% C.I.= 95.8 - 99.7%)

RETROSPECTIVE STUDIES

Suspected and Known Positive Syphilis Samples

A series of 412 total samples were purchased from serum and blood center suppliers consisting of 149 banked RPR and treponemal reactive serum samples and 28 serum samples that were requested to be Primary or Secondary Patients, treated, but exhibiting a Syphilitictype lesion or rash were purchased from a serum supplier, and 138 frozen serum and plasma samples were obtained from a blood center. The samples were found to be RPR and treponemal reactive and having mixed titers. Another series of 97 samples being highly suspected of having a syphilis infection were obtained from a serum supplier that were obtained from various laboratories around the U.S., and were submitted to the laboratories for testing. The samples were tested by RPR. TPPA. and MHA-TP as reference methods for comparison to Syphilis Health Check test.

The samples were further tested by an outside laboratory for TPPA titer and Syphills Health Check results. 19 samples were found to be Negative for TPPA and Syphilis Health Check and Reactive RPR, four samples were negative by TPPA, but Positive by Syphilis Health Check and RPR, one sample was Syphilis Health Check negative and positive RPR and TPPA, and one sample was RPR neq but TPPA and Syphilis Health Check positive. The remaining two hundred eighty-nine samples remained Positive by all four methods. In the Suspected Positive patients two samples were negative with Syphilis Health Check, with one sample low reactive with RPR, and both samples Non-Reactive with TPHA. The results of the testing are shown below.

Known Positives

RPR
posnegTotal
SyphilisHealthCheckpos2931294
neg20121
Total3132315

Percent Positive Agreement = 93.4% (95%C.I. = 89.8 - 96.0%) Percent Negative Agreement = 100.0% (95%C.I. =. 100%) Percent Overall Agreement = 93.4% (95%C.1. = 89.8 - 96.0%)

TPPA
posnegTotal
2904294
12021
29124315

99.6% (95%C.I.= 97.9 - 100.0%) 85.7% (95%C.I.= 63.7 - 97.0%) 98.6% (95%C.1.= 96.5 - 99.6%)

6

{6}------------------------------------------------

Suspected Positive

RPR
posnegTotalpos
SyphilisHealthCheckpos62258787
neg010100
Total62359787
TPHA
posnegTotal
87087
01010
871097

PPA = 100.0% (95% C.I. = 94.2 - 100%) PNA = 28.6% (95% C.I. = 14.6 - 46.3%) POA = 74.2% (95% C.I. = 64.3 - 82.6%)

100.0% (95% C.I. = 95.8 - 100%) 100.0% (95% C.I. = 69.2 - 100%) 100.0% (95% C.I. = 96.3 - 100%)

neg Total pos 87 0 87 10 0 10 87 10 97

MHTP

100.0% (95% C.I. = 95.8 - 100%) 100.0% (95% C.l. = 69.2 - 100%) 100.0% (95% C.I. = 96.3 - 100%)

Clinically Diagnosed

A panel of one hundred sixty-four (164) well-characterized clinically diagnosed serum samples from treated and untreated patients with primary, secondary, and latent syphilis infections were obtained from a clinic serving a population of individuals with a variety of infectious diseases. The samples were tested with reference assay tests for RPR, TPPA, and FTA-ABS. The samples were then tested with Syphilis Health Check and the results are summarized below.

Known Clinical StatusRPRTPPAFTA-ABSSyphilisHealthCheckNo.%Agree-ment95% C.I.
UntreatedPrimaryReactReactReactReact2310085.2 - 100%
SecondaryReactReactReactReact2510086.3 - 100%
LatentReactReactReactReact2210084.6 - 100%
NRReactReactReact310029.2 - 100%
TreatedPrimaryReactReactReactReact2810087.8 - 100%
SecondaryReactReactReactReact2610086.8 - 100%
LatentReactReactReactReact1810081.5 - 100%
NRReactReactReact1910082.4 - 100%
Total16410097.8 - 100%

{7}------------------------------------------------

Interference Study

.

Interference testing was conducted using serum. Concentrates of the compounds were prepared and diluted to multiple concentrations into eight sera with different levels of syphilis reactivity. The following results were obtained:

Hemoglobin: No effect was observed up to 1000 mg/dL of hemoglobin;

Bilirubin (total): No effect was observed up to 40 mg/dL of total bilirubin;

Triglycerides: No effect was observed up to 3000 mg/dL of triglycerides.

Cholesterol (total): No effect was observed up to 400mg/dL of cholesterol

Albumin: No effect was observed up to 1000 mq/dL of albumin

Gamma-globulin: No effect as observed up to 5000 mg/dL of gamma-globulin

Potential Cross Reactors:

Panels of samples were obtained to evaluate potential interference from different disease conditions confirmed positive and containing different concentrations of potentially cross-reactive antibodies and were analyzed with the Syphilis Health Check test. From 151 (138 Males) prospectively collected drug users eleven (7%) were found syphilis positive by reference methods and Syphilis Health Check, with two additional samples found positive for Syphilis Health Check. Two samples from Lyme disease and HSV showed a positive result with Syphilis Health Check but could not be confirmed. One sample each from CMV positive and heterophile positive patients were found Syphilis Health Check positive, but non-reactive by reference methods . Two HPV patients were found Syphilis Health Check positive with one confirmed by reference methods. Two co-infected Chlamydia/GC patients were found Syphilis Health Check and one confirmed positive by TPPA. The number of disease condition cateqories and reactive results obtained is listed in the following table.

Cross-reactorNumber# positive bySyphilisHealth Check# positive byreferencemethod(s)
Self-reported DrugUsers1511311
ANA Positive240
RF Positive400
U.S. Lyme DiseaseIgG & IgM251Not confirmed
HSV241Not confirmed
CMV1021
EBV210
HAV250

{8}------------------------------------------------

HIV 1 & 21411
HTLV1411
Heterophile3210
HCV2411
Anti-HBs2522
Other STD - GC,Chlamydia, HPV,Trichomonas,7842

Syphilis Positive HIV / HCV / HBV Patients

A series of 13 patient EDTA Plasma samples from a blood center were identified that were screened RPR positive in patients that were also confirmed positive for HIV. HCV and/or Hepatitis B virus. Three of these patients were further identified as previously testing positive for syphilis and treated at that time. All of the samples were tested with TPPA and Syphilis Health Check to evaluate syphilis reactivity. All 13 samples were positive for RPR. TPPA, and Syphilis Health Check.

To further evaluate the influence of HIV on Syphilis Health Check results, a series of 24 banked serum containing high levels of viral load for HIV were tested with Syphilis Health Check. The HIV positive patient samples were purchased from a commercial serum supplier that were known to be RPR and/or Treponema positive. The samples were tested by an outside lab with the Syphilis Health Check test to assess reactivity. Six of these samples were Nonreactive by RPR but reactive with TPHA. One sample was RPR positive, but TPHA nonreactive. All 24 samples were Syphilis Health Check Positive.

Pregnant Women

A series of sixty-nine (69) pregnant female samples were purchased from a vendor that had known trimester, age, and ethnicity. An additional set of 93 pregnant women serum samples were obtained from a commercial source that were identified as syphilis positive by RPR screening and further tested with TPPA and Syphilis Health Check. Age and trimester were known, but ethnicity was not identified. Three samples out of the 162 total samples were RPR low positive but nonreactive by the treponemal methods.

Pregnant Women

Summary

RPR
posnegTotal
SyphilisHealthCheckpos91091
neg36871
Total9468162
TPPA
posnegTotal
94094
06868
9468162

9

{9}------------------------------------------------

Percent Positive Agreement = 96.8% (95% C.I.= 91.0 - 99.3%)100.0% (95% C.I.= 96.2 - 100%)
Percent Negative Agreement = 100.0% (95% C.I.= 94.7 - 100%)100.0% (95% C.I.= 94.7 - 100%)
Percent Overall Agreement = 98.1% (95% C.I.= 94.7 - 99.6%)100.0% (95% C.I.= 97.7 - 100%)

PRECISION and REPRODUCIBILITY

Studies were performed to demonstrate the Intra-Assay, and Inter-Lot reproducibility of the Syphilis Health Check Test kit.

To demonstrate the reproducibility of the Syphilis Health Check test three laboratory sites using two trained technicians at each site to perform the testing for within-Run and between day reproducibility using a panel of 6 pooled samples. Each testing site conducted reproducibility studies using a supplied panel ranging from non-reactive to highly reactive, i.e. one nonreactive, one borderline, one borderline reactive, one moderate reactive, and two mid-high to high reactive in addition to the kit controls. Each site ran these panel member solutions for at least 5 days, twice per day. Each site performed one Within-Run assay by each of the operators each run for 10 times on one day. The one negative and three low to high positive samples had 100% agreement. The two borderline samples near the cut-off vielded the following results.

The following summary demonstrates the Reproducibility of the two critical borderline samples obtained by the Technicians.

Intra-Run

Panel D (Borderline Reactive)

PositiveNegativeExpected% AgreeDiscrepant
Site 1 (2 operators)57360 positive95.03
Site 2 (2 operators)58260 positive96.72
In-house (2 operators)59160 positive98.31

Panel E (Borderline Nonreactive)

PositiveNegativeExpected% AgreeDiscrepant
Site 115960 negative98.31
Site 225860 negative96.72
In-House06060 negative100.00

{10}------------------------------------------------

Inter-Day (5 days)

Panel D (Borderline Reactive)

PositiveNegativeExpected% AgreeDiscrepant
Site 1 (2 operators)57360 positive95.03
Site 2 (2 operators)58260 positive96.72
In-house (2 operators)57360 positive95.03

Panel E (Borderline Nonreactive)

PositiveNegativeExpected% AgreeDiscrepant
Site 125860 negative96.72
Site 225860 negative96.72
In-House15960 negative98.31

Inter-Lot

Studies were performed to demonstrate the Inter-Lot assay reproducibility of the Syphilis Health Check Test kits. The Inter-Lot study used four dilutions of a positive pool and negative sera samples. These five samples were run in duplicate on 5 different lots to demonstrate lot-lot reproducibility run on the same day by the same technician.

The following table demonstrates the reproducibility of the Syphilis Health Check Test over 5 lots of kits.

DilutionResults after 10 min.
ExpectedresultsLot ILot IILot IIILot IVLot V
Negative serum------
1/104------
1/103+/-+/-+/-+/-+/-+/-
1/102++++++
1/10++++++

{11}------------------------------------------------

EXPECTED VALUES

To assess the normal range (Expected Values) of the Syphilis Health Check test a series of ninety-eight (98) samples were obtained from various hospital laboratories in different U.S. geographical locations. The samples came from a general "presumed" healthy normal population (ages 20 - 66) whose serum or plasma were collected by hospital laboratories for routine serology testing, not related to STD.

All of the samples were tested with RPR, TPPA and Syphilis Health Check tests. Two samples were found positive by the treponemal tests with both samples non-reactive with RPR. These 2 samples were confirmed with MHA-TP.

The results of the testing of a presumed normal population obtained the following results for the Syphilis Health Check test: This resulted in a Overall Agreement of 100%. Therefore from this study of 98 presumed normal samples 2 were confirmed positive resulting in a 2% positivity rate.

In a prospectively collected population of drug users visiting STD clinics, of 138 males and 13 females (ages 18 - 61 yrs) thirteen (8.6%) were found to be Syphilis Health Check positive with eleven of those also reference method positive.

In a series of 69 pregnant women, ages 20 - 40 yrs, that were tested for routine screening, one sample (1.4%) was found to be Syphilis Health Check positive and confirmed by reference methods.

Eight hundred eighty (880) patients were prospectively collected in a population of individuals visiting STD and a hospital clinics and POC sites complaining and/or exhibiting signs and symptoms of STD infections, ages 16 - 81 vrs, and having 53% to 47% male to female ratio. It was found that a range of 3% - 6% of the patients identified as suspected syphilis were found positive by Syphilis Health Check and yielded a Percent positive agreement to RPR of 98.3% and 95.7% to treponemal tests. Percent Negative Agreement to RPR and Treponemal test was 93.4% and 97.8% respectively. Overall Agreement to RPR and Treponemal tests was 94.1% and 97.4% respectively.

These results are consistent with published rates for prevalence of antibody in the adult population. Prevalence may vary depending on a variety of factors such as geography, age, socio-economic status, ethnic background, type of test employed, specimen collection and handling procedures, clinical and epidemiological history.

{12}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Diagnostics Direct, LLC c/o Gary Lehnus President Lehnus & Associates Consulting 150 Cherry Lane Road East Stroudsburg, PA 18301

AUG - 1 2011

Re: K102400

Trade/Device Name: Syphilis Health Check Treponemal Antibody Test; Rapid immunochromatographic membrane assay, Treponema pallidum Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: Class II Product Code: LIP Dated: July 21, 2011 Received: July 27, 2011

Dear Mr. Lehnus:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000

{13}------------------------------------------------

  1. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Sef atten

Sally A. Hoivat, M.S Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

. Indications for Use

510(k) Number (if known):

Device Name: Syphilis Health Check Test

Indications For Use:

Syphilis Health Check is a qualitative rapid membrane immunochromatographic assay for the detection of Treponema pallidum (syphilis) antibodies in human whole blood, serum or plasma. This product can be used as an initial screening test or in conjunction with a non-treponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection. This test is not intended for use in screening blood or plasma donors.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (OIVD)

Division Sign-Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(K) K102400.

Page 1 of 1

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).