K Number
K102298

Validate with FDA (Live)

Date Cleared
2011-01-07

(147 days)

Product Code
Regulation Number
888.3040
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TransTend Anchor is indicated for use in soft tissue to bone fixation in association with post-operative immobilization as follows: Shoulder (Rotator Cuff, Partial Thickness Rotator Cuff, Biceps Tenodesis, Acromio-Clavicular Separation, Deltoid Repair), Foot/Ankle (Lateral Stabilization, Medial Stabilization, Mid-foot Reconstruction, Hallux Valgus Repair, Metatarsal Ligament/Tendon Repairs), Knee (Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis), Elbow (Lateral Epicondylitis Repair), Wrist (Scapholunate Ligament Reconstruction), Hip (Capsular Repair, Acetabular Labral Repair).

Device Description

The proposed TransTend Anchors are suture anchors offered in three different materials, namely Biocryl Rapide, Titanium or Peek. The anchor comes preloaded on a disposable inserter assembly and is intended for fixation of size #2 suture to bone. The suture option is provided without needles. The TransTend Anchors are provided as size 3.4 mm in Biocryl Rapide, 3.4mm in Peek and 2.9 mm in Titanium. Each TransTend Anchor is provided sterile and is for single patient use only. The TranTend anchors will also be offered in a convenience kit and be packaged with a cannula assembly along with one or two TransTend Anchor(s).

AI/ML Overview

The provided 510(k) summary for the TransTend Anchors does not include explicit acceptance criteria with numerical thresholds, nor a specific study demonstrating compliance to such criteria in the manner typically seen for diagnostic AI/ML devices.

Instead, this submission is for a medical device (surgical suture anchors) and focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing. The "acceptance criteria" here are implicitly tied to the performance characteristics of the predicates and general safety and performance standards for such devices.

Here's an breakdown based on the information provided, framed as closely as possible to your request, but acknowledging the difference in device type:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria Category (Implied by Submission)Reported Device Performance (Summary)
Material EquivalenceMaterials (Biocryl Rapide, Titanium, Peek) are the same as predicates.
Design EquivalenceDesign characteristics (e.g., preloaded on disposable inserter, intended for fixation of #2 suture) are similar to predicates.
Packaging EquivalencePackaging is the same as predicates.
Functional Performance (Mechanical)Pull-out testing and insertion torque testing were performed. Results deemed suitable for intended use.
BiocompatibilityBiocompatibility testing was performed. Results deemed suitable for intended use.
SterilizationSterilization testing was performed. Results deemed suitable for intended use.
Safety and Performance"Results of performance and safety testing have demonstrated that the proposed devices are suitable for their intended use."

2. Sample size used for the test set and the data provenance

The document does not specify the sample sizes for the mechanical, biocompatibility, or sterilization tests. It describes "Verification activities" and "Testing," but no detailed protocols or sample numbers are provided for these non-clinical tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable to the type of device and testing described. The "ground truth" for a bone anchor is its physical and biological performance, which is assessed through engineering and biological tests, not expert interpretation of diagnostic images or clinical scenarios.

4. Adjudication method for the test set

This question is not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in clinical or image-based studies where expert consensus is needed. For mechanical and biological testing of a physical device, the "truth" is determined by the objective measurements of the tests themselves, against established standards or predicate performance.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. MRMC studies are specific to diagnostic AI/ML devices where human readers interpret cases, and the AI's impact on their performance is evaluated. This submission is for a surgical implant (bone anchor), not a diagnostic AI tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not applicable. The TransTend Anchors are a physical surgical device, not an algorithm, so there is no "standalone" algorithm performance to assess.

7. The type of ground truth used

For this device, the "ground truth" is established through:

  • Engineering Test Results: Objective measurements from mechanical tests (pull-out strength, insertion torque) compared against functional requirements and predicate device performance.
  • Biological Test Results: Objective measurements from biocompatibility tests, following established standards (e.g., ISO 10993).
  • Sterilization Validation: Objective verification of sterility assurance levels.

8. The sample size for the training set

This question is not applicable. As a physical medical device, there is no "training set" in the context of machine learning algorithms. The design and manufacturing process are informed by general engineering principles, material science, and the performance characteristics of predicate devices, rather than a data-driven training set.

9. How the ground truth for the training set was established

This question is not applicable for the same reasons as #8.

{0}------------------------------------------------

K102298
p. 1 of 2

510(k) SUMMARY

TransTend Anchors

JAN - 7 2011

Submitter's Nameand Address:DePuy Miteka Johnson & Johnson company325 Paramount DriveRaynham, MA 02767
Contact PersonKristine ChristoRegulatory Affairs Project ManagerDePuy Mitek, Inc.a Johnson & Johnson company325 Paramount DriveRaynham, MA 02767, USATelephone: 508-828-3359Facsimile: 508-977-6955e-mail: KChristo@its.jnj.comPrepared 10/30/2010
Name of MedicalDeviceClassification Name: Single/multiple component metallic bone fixationappliances and accessories
Common/Usual Name: Bone Anchor
Proprietary Name: TransTend Anchors
SubstantialEquivalenceThe proposed TransTend Anchors are substantially equivalent to:K100012 Gryphon Br - Hip (March 10, 2010) K090124 Gryphon P BR Anchor (March 11, 2009) K073412 Gryphon BR / Healix BR Anchors (January 17, 2008) K082282 Healix Ti Anchor (November 7, 2008) K071481 Healix Peek Anchor (August 9, 2007) K082810 Arthrex BioComposite Corkscrew Anchor (January 23, 2009) K081598 KFX Suture lock Nail Bone Anchor (July 3, 2008) K020159 Twin Fix Ti Quick T 3.5 mm Fixation System ( March 26, 2010)
DeviceClassificationThese devices carry an FDA product code HWC, and subsequent codes MBI andMAI, regulated under 21 CFR 888.3040 and 21 CFR 888.3030.
The proposed TransTend Anchors are suture anchors offered in three different

{1}------------------------------------------------

K102298
p. 2012

Device Description

materials, namely Biocryl Rapide, Titanium or Peek. The anchor comes preloaded on a disposable inserter assembly and is intended for fixation of size #2 suture to bone. The suture option is provided without needles. The TransTend Anchors are provided as size 3.4 mm in Biocryl Rapide, 3.4mm in Peek and 2.9 mm in Titanium. Each TransTend Anchor is provided sterile and is for single patient use only. The TranTend anchors will also be offered in a convenience kit and be packaged with a cannula assembly along with one or two TransTend Anchor(s).

Technologies characteristics including material for design, and packaging are the same as the predicates cleared devices.

The TransTend Anchor is indicated for use in soft tissue to bone fixation in Indications for association with post-operative immobilization as follows: Use

IndicationBRPEEKTitanium
ShoulderRotator CuffXXX
Partial Thickness Rotator CuffXXX
Biceps TenodesisXX
Acromio-Clavicular SeparationXX
Deltoid RepairXX
Foot/AnkleLateral StabilizationXX
Medial StabilizationXX
Mid-foot ReconstructionXX
Hallux Valgus RepairXX
Metatarsal Ligament/Tendon RepairsXX
KneeMedial Collateral Ligament RepairXX
Lateral Collateral Ligament RepairXX
Posterior Oblique Ligament RepairXX
Iliotibial Band TenodesisXX
ElbowLateral Epicondylitis RepairXX
WristScapholunate Ligament ReconstructionXX
HipCapsular RepairXX
Acetabular Labral RepairXX

Non clinical Testing

Verification activities were performed on the implant or its predicates. Testing includes pull out testing, insertion torque testing, sterilization and biocompatibility.

Results of performance and safety testing have demonstrated that the proposed Safety and devices are suitable for their intended use. Performance

Based on the indications for use, technological characteristics, and comparison to the predicate devices, the proposed TransTend Anchors Anchor (s) have shown to be substantially equivalent to the predicate devices under the Federal Food, Drug and Cosmetic Act.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The logo is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DePuy Mitek, Inc. % Ms. Kristine Christo Regulatory Affairs Project Manager 325 Paramount Drive Raynham, Massachusetts 02767

JAN - 7 2011

Re: K102298

Trade/Device Name: TransTend (BR, Ti, Peek) Anchors Regulation Number: 21 CFR 888,3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC, MBI, MAI Dated: December 30, 2010 Received: January 03, 2011

Dear Ms. Christo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Ms. Kristine Christo

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

For Peter Dunn

Mark N. Melkerson Director Division of Surgical. Orthonedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K102298

Indications for Use

510(k) Number (if known): K102298

JAN - 7 2011

· Device Name: TransTend (BR, Ti, Peek) Anchors

The TransTend Anchor is indicated for use in soft tissue to bone fixation in association with postoperative immobilization as follows:

IndicationBRPEEKTitanium
ShoulderRotator CuffXXX
Partial Thickness Rotator CuffXXX
Biceps TenodesisXX
Acromio-Clavicular SeparationXX
Deltoid RepairXX
Foot/AnkleLateral StabilizationXX
Medial StabilizationXX
Mid-foot ReconstructionXX
Hallux Valgus RepairXX
Metatarsal Ligament/Tendon RepairsXX
KneeMedial Collateral Ligament RepairXX
Lateral Collateral Ligament RepairXX
Posterior Oblique Ligament RepairXX
Iliotibial Band TenodesisXX
ElbowLateral Epicondylitis RepairXX
WristScapholunate Ligament ReconstructionXX
HipCapsular RepairXX
Acetabular Labral RepairXX

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of_1

for M.Mdkerrm
(Division Sign Off)

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K102298

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.