K Number
K102291

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-09-15

(34 days)

Product Code
Regulation Number
878.4750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Endo GIA™ Radial Reload with Tri-Staple™ Technology has application in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection.

It may be used for transection of liver substance, hepatic vasculature and biliary structures.

Device Description

The Endo GIA™ Radial Reload with Tri-Staple™ Technology is a curved stapler cartridge that places three staggered rows of staples on either side of a cut line and simultaneouly divides (cuts) the tissue between the rows.

Covidien Reloads with Tri-Staple™ Technology apply height progressive rows of titanium staples on either side of the cut line. The shortest staple length is located in the first staggered row nearest the knife, an intermediate staple length is located in the second row, and the longest staple length is located in the third (outside) staggered row.

The Endo GIA™ Radial Reload with Tri-Staple™ Technology is compatible with the Endo GIA™ ULTRA Universal Stapler and GIA™ and Endo GIA™ Universal Stapler handles, and incorporates the functionality of those handles including interlocks to prevent firing a previously fired cartridge.

AI/ML Overview

The provided text describes a medical device, the "Covidien Endo GIA™ Radial Reload with Tri-Staple™ Technology," which is a surgical stapler cartridge. The document is a 510(k) summary for premarket notification, indicating its purpose is to demonstrate substantial equivalence to legally marketed predicate devices, not necessarily to independently establish de novo acceptance criteria with a comprehensive study design as one might find for a novel diagnostic AI algorithm.

Therefore, many of the requested categories (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, training set details) are not applicable in this context, as they relate to the validation of AI/machine learning diagnostic devices or studies involving human interpretation. This document describes a surgical stapling device, which is validated through bench and animal performance evaluations.

Here's an attempt to answer the questions based only on the provided text, noting where information is not available or not relevant to this type of device submission:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative "acceptance criteria" in a table format that would typically be associated with AI/diagnostic device performance (e.g., specific sensitivity, specificity, or F1-score targets). Instead, the performance evaluations verify that the device is "safe and effective and performs as intended" through various bench and in vivo tests. The "reported device performance" is a general statement that these tests were "completed to verify" the safety and effectiveness.

Performance Evaluation CategoryStated Purpose/VerificationOutcome
Bench Tests:
In vitro staple formationVerify safe and effective performanceCompleted to verify
Firing forceVerify safe and effective performanceCompleted to verify
Staple line pull-apart forceVerify safe and effective performanceCompleted to verify
Animal Model Tests:
In vivo staple formationVerify safe and effective performanceCompleted to verify
Free bleedVerify safe and effective performanceCompleted to verify
Burst strengthVerify safe and effective performanceCompleted to verify
Air leakVerify safe and effective performanceCompleted to verify
Tissue graspingVerify safe and effective performanceCompleted to verify
TraumaVerify safe and effective performanceCompleted to verify

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified. The document states "Bench and animal model performance evaluations were completed," but does not provide the number of samples, animals, or tests performed for each category.
  • Data Provenance: The studies were conducted as "Bench and animal model performance evaluations." The country of origin is not specified, but the submission is to the U.S. FDA. The studies are prospective in the sense that they were designed tests rather than retrospective analysis of existing data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This is a surgical stapling device, not a diagnostic device requiring expert interpretation for "ground truth" establishment in a test set. The performance evaluations involved physical and biological tests.

4. Adjudication method for the test set

  • Not Applicable. See point 3. Adjudication methods like 2+1 or 3+1 are used for resolving disagreements in expert interpretations, which is not relevant to the physical and biological performance testing of a stapling device.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI-assisted diagnostic device. No human reader studies (MRMC) were conducted or are relevant to demonstrate the safety and effectiveness of a surgical stapler.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This device is a physical surgical stapler, not an algorithm. Therefore, "standalone" algorithm performance is not relevant.

7. The type of ground truth used

  • Inferred Ground Truth: For the performance evaluations, the "ground truth" would be the direct measurement of physical properties (e.g., staple formation, firing force, pull-apart force, burst strength, air leak) and direct observation of biological outcomes (e.g., free bleed, tissue grasping, trauma) in controlled bench and animal models. This is not "expert consensus," "pathology," or "outcomes data" in the typical sense of AI/diagnostic device validation but rather objective measurement against predefined engineering and biological success criteria, though these criteria themselves are not explicitly detailed in the provided text beyond the general statement of "safe and effective."

8. The sample size for the training set

  • Not Applicable. This is not a machine learning device. The concept of a "training set" does not apply. Design and development of the stapler would involve iterative testing and refinement, but not in the sense of a machine learning training set.

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

K102291

510(k) Summary of Safety and Effectiveness

SUBMITTER:Surgical Devices, a global business unitof Tyco Healthcare Group LP (d/b/a Covidien)60 Middletown AvenueNorth Haven, CT 06473Tel. No.: (203) 492-5352SEP 15 2010
CONTACT PERSON:Tim M. Lohnes, Manager, Regulatory Affairs
DATE PREPARED:August 10, 2010
TRADE/PROPRIETARY NAME:Covidien Endo GIA™ Radial Reload with Tri-Staple™Technology
COMMON/USUAL NAME:Surgical Stapler with Implantable Staple
CLASSIFICATION NAME:Staple, Implantable
PREDICATE DEVICE(S):Autosuture™ ENDO GIA™ Stapler (K083519),Ethicon Contour Curved Cutter Stapler & Reloads (K091322)

DEVICE DESCRIPTION:

The Endo GIA™ Radial Reload with Tri-Staple™ Technology is a curved stapler cartridge that places three staggered rows of staples on either side of a cut line and simultaneouly divides (cuts) the tissue between the rows.

Covidien Reloads with Tri-Staple™ Technology apply height progressive rows of titanium staples on either side of the cut line. The shortest staple length is located in the first staggered row nearest the knife, an intermediate staple length is located in the second row, and the longest staple length is located in the third (outside) staggered row.

The Endo GIA™ Radial Reload with Tri-Staple™ Technology is compatible with the Endo GIA™ ULTRA Universal Stapler and GIA™ and Endo GIA™ Universal Stapler handles, and incorporates the functionality of those handles including interlocks to prevent firing a previously fired cartridge.

INTENDED USE:

The Endo GIA™ Radial Reload with Tri-Staple' Technology has application in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection.

It may be used for transection of liver substance, hepatic vasculature and biliary structures.

. Page 28

{1}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS:

This 510(k) reports a modification of our currently marketed Autosuture™ Endo GIA™ Staplers (K085319), namely, the Endo GIA™ Radial Reload with Tri-Staple" Technology applies staple radial, rather than linear staple lines from a curved, rather than straight, cartridge. The subject Endo GIA™ Radial Reload with Tri-Staple™ Technology incorporates identical Tri-Staple™ Technology and uses the existing Endo GIA™ Stapler handles for operation. It will be offered in a 60 mm length as a Purple "Medium-Thick" cartridge.

The Endo GIA™ Radial Reload Purple with Tri-Staple™ Technology is identical to the predicate Autosuture™ device with regard to stapling technologies. The curved cartridge facilitates access in restricted anatomy, for example low in the pelvis during lower anterior resection (LAR) with total mesorectal excision (TME) procedures. The Endo GIA™ Radial Reloads Purple with Tri-Staple™ Technology are compatible with the existing Endo GIA ™ Stapler handles for operation and incorporate the functionality of those handles, including interlocks to prevent firing a previously fired cartridge.

The subject device is equivalent to the predicate Ethicon Contour® Curved Cutter Stapler, which applies radial staple rows on either side of a cut line and divides between them, creating a 40mm curved transection. The Endo GIA™ Radial Reload Purple with Tri-Staple™ Technology places three radial staple rows 60mm in length and simultaneouly divides (cuts) between the rows, creating a 60mm curved transection.

MATERIALS:

All components of the Endo GIA™ Radial Reloads with Tri-Staple™ Technology are comprised of materials that are in accordance with ISO Standard 10993-1.

PERFORMANCE DATA:

Bench and animal model performance evaluations were completed to verify that the EndoGIA™ Radial Reload Purple with Tri-Staple™ Technology is safe and effective and performs as intended. Testing consisted of evaluation for in vitro stable formation, firing force and staple line pull-apart force, as well as in vivo staple formation, free bleed, burst strength, air leak, tissue grasping, and trauma.

Page 29

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Covidien % Mr. Tim M. Lohnes Manager, Regulatory Affairs 60 Middleton Avenue North Haven, Connecticut 06473

SEP 1 5 2010

Re: K102291

Trade/Device Name: Covidien Endo GIA™ Radial Reload with Tri-Staple™ Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: August 10, 2010 Received: August 12, 2010

Dear Mr. Lohnes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Tim M. Lohnes

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely yours.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

K102291 SEP 1 5 2010

Device Name:

Indications For Use

510(k) Number (if known): Device Name: Covidien Endo GIA™ Radial Reload with Tri-Staple™ Technology

K102291

The Endo GIA™ Radial Reload with Tri-Staple™ Technology has application in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. Iow anterior resection.

It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures.

X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Kane for MKY

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K/02291

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.