K Number
K102283

Validate with FDA (Live)

Date Cleared
2011-02-10

(182 days)

Product Code
Regulation Number
866.3830
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ZEUS ELISA Treponema pallidum IgG Test System is intended for the qualitative detection of specific IgG class antibodies to T. pallidum in human serum. The test may be used in conjunction with non treponemal testing and clinical findings to provide serological evidence of infection with T. pallidum. This test is for in vitro diagnostic use only.

This test is not intended for screening blood or plasma donors.

Device Description

The ZEUS ELISA Treponema pallidum IgG Test System is designed to detect IgG class antibodies in human sera to Treponema pallidum. Wells of plastic microwell strips are sensitized by passive absorption with T. pallidum antigen. The test procedure involves three incubation steps:

  1. Test sera are diluted with SAVE Diluent. During sample incubation, any antigen specific lgG antibody in the sample will bind to the immobilized antigen. The plate is washed to remove unbound antibody and other serum components.
  2. Peroxidase conjugated goat anti-human IgG is added to the wells and the plate is incubated. The conjugate will react with IgG antibody immobilized on the solid phase in step 1. The wells are washed to remove unbound conjugate.
  3. The microwells containing immobilized peroxidase conjugate with peroxidase substrate solution. Hydrolysis of the substrate by peroxidase produces a color change. After ten minutes have elapsed, the reaction is stopped and the color intensity of the solution is measured photometrically. The color intensity of the solution depends on upon the antibody concentration in the original test sample.
AI/ML Overview

The Zeus Scientific Inc. ZEUS ELISA Treponema pallidum IgG Test System is intended for the qualitative detection of specific IgG class antibodies to T. pallidum in human serum. The test may be used in conjunction with non-treponemal testing and clinical findings to provide serological evidence of infection with T. pallidum. It is for in vitro diagnostic use only and not intended for screening blood or plasma donors.

Acceptance Criteria and Device Performance

The acceptance criteria for the ZEUS ELISA Treponema pallidum IgG Test System are implicitly demonstrated by the performance against the predicate device and the specified reproducibility metrics. While explicit quantitative acceptance criteria for clinical performance (e.g., specific sensitivity, specificity thresholds) are not individually listed for each study, the submission aims to demonstrate substantial equivalence to the predicate device.

The internal and external reproducibility studies provide quantitative acceptance criteria for precision (Total CV < 15% for reactive samples, < 25% for negative samples for repeatability; < 15% for reactive samples, < 50% for negative samples for reproducibility).

The reported device performance across various clinical studies is summarized below, with performance relative to the predicate device (TPPA) or clinical diagnosis.

Table 1: Device Performance Summary

Study Population (Test Set Size)Ground Truth MethodPredicate Device (if applicable)Reported Device Performance (Interpretation Criteria: <= 0.90 Negative, 0.91-1.1 Equivocal, >1.1 Positive)
Intended Use Populations
500 patients with syphilis test orderedTPPA (treponemal reference assay)Phoenix Bio-Tech Trep-ChekPPA: 80.0% (95% CI: 28.4-99.5%), NPA: 99.2% (95% CI: 97.9-99.8%)
500 pregnant women with syphilis test orderedTPPA (treponemal reference assay)Phoenix Bio-Tech Trep-ChekPPA: 75.0% (95% CI: 19.4-99.4%), NPA: 100.0% (95% CI: 99.4-100%)
Other Clinical Performance Studies
1000 unselected hospitalized patientsTPPA (treponemal reference assay)Phoenix Bio-Tech Trep-ChekPPA: Not fully reported (21 Positive, 1 Equivocal by predicate resulted in 13 Positive by device), NPA: 97.1% (95% CI: 95.9-98.1%)
223 Retrospective HIV-1 Positive SamplesTPPA (treponemal reference assay)Phoenix Bio-Tech Trep-ChekPPA: 85.4% (95% CI: 72.2-93.9%), NPA: 99.4% (95% CI: 96.9-100%)
280 Retrospective TPPA/RPR Positive SamplesTPPA/RPR (reference assays)Phoenix Bio-Tech Trep-ChekPPA: 98.5% (95% CI: 96.2-99.6%), NPA: 70.6% (95% CI: 46.9-98.7%)
277 Retrospective TPPA Negative/Positive Pregnant WomenTPPA/RPR (reference assays)Phoenix Bio-Tech Trep-ChekPPA: 92.9% (95% CI: 76.5-99.1%), NPA: 99.6% (95% CI: 97.8-100%)
157 CDC Syphilis PanelClinical DiagnosisN/A% Agreement with Clinical Diagnosis: 93.0% (95% CI: 87.8-96.5%)
Precision (Repeatability)
15 samples (Negative, High Negative, Near Cut-off, Low Positive, High Positive)Internal testing, 48 replicates eachN/ATotal CV for reactive samples: 3.5% to 7.0% (all < 15%); Total CV for negative samples: 7.7% to 13.1% (all < 25%)
Precision (Reproducibility)
15 samples (Negative, High Negative, Near Cut-off, Low Positive, High Positive)Internal and external sites, 180 replicates eachN/ATotal CV for reactive samples: 5.9% to 10.9% (all < 15%); Total CV for negative samples: 7.3% to 36.6% (all < 50%)

Study Details

  1. Sample Size Used for the Test Set and Data Provenance:

    • Prospective Studies:
      • 500 unselected samples from patients with a syphilis test ordered (US Mid-Atlantic, Northeast, and Zeus laboratories).
      • 500 unselected samples from pregnant women with a syphilis test ordered (US Mid-Atlantic, Northeast, and Zeus laboratories).
      • 1000 unselected samples from hospitalized patients (US: Monmouth, Mainline laboratories, and Zeus).
    • Retrospective Studies:
      • 223 banked known HIV-1 positive samples (New York vendor).
      • 280 RPR/TPPA positive samples (New York vendor).
      • 250 TPPA negative and 27 TPPA positive samples from pregnant women (New York vendor).
    • Reference Panels:
      • 157 samples from the CDC Syphilis Panel (origin unspecified, but typically US-based reference laboratories).
    • Precision/Reproducibility: 15 samples per study (prepared by Zeus Scientific, Inc.).
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications: Not applicable. The ground truth for comparative clinical studies was established using a reference assay (TPPA or RPR) or clinical diagnosis, not by individual expert review of each case. For the CDC Syphilis Panel, the ground truth was "Clinical Diagnosis," implying a consensus or established diagnosis from medical records.

  3. Adjudication Method for the Test Set: Not applicable. The studies relied on a reference assay (TPPA) for concordance analysis or established clinical diagnoses for the CDC panel. There is no mention of an adjudication process for discordant results between the device and the reference standard, nor for expert consensus in establishing the ground truth.

  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done: No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is an ELISA test system, which provides an objective quantitative (OD value) and qualitative (Positive/Negative/Equivocal) result, not requiring human interpretation in the way an imaging AI algorithm would. Therefore, the concept of "human readers improving with AI vs. without AI assistance" does not apply here.

  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done: Yes, the studies described are standalone performance evaluations of the ZEUS ELISA Treponema pallidum IgG Test System. The device output (OD value and interpretation) is generated directly by the assay without human interpretation as part of the primary diagnostic step. The results are compared against a reference standard or clinical diagnosis.

  6. The Type of Ground Truth Used:

    • Reference Assay: For most clinical performance studies, the ground truth was established by a treponemal reference assay, specifically TPPA (Treponema pallidum Particle Agglutination assay), or in some cases, RPR (Rapid Plasma Reagin) for the retrospective RPR/TPPA positive samples.
    • Clinical Diagnosis: For the CDC Syphilis Panel, the ground truth was based on "Clinical Diagnosis" of syphilis disease.
  7. The Sample Size for the Training Set: Not applicable in the context of an ELISA assay. ELISA assays are biochemical tests with fixed reagents and protocols, not machine learning algorithms that require a "training set" in the same sense. The linearity, analytical specificity (interfering substances, cross-reactivity), and detection limits studies are part of the analytical validation, not algorithm training.

  8. How the Ground Truth for the Training Set was Established: Not applicable, as there is no "training set" for an ELISA assay as understood for AI/ML algorithms. The assay's performance characteristics are inherent to its biochemical design.

{0}------------------------------------------------

Zeus Scientific Inc. 510(k) Summary ZEUS ELISA Treponema pallidum IgG FEB 1 0 2011 Test System K102283

Zeus Scientific, Inc. (Zeus) PO Box 38, Raritan, NJ 08869 (908)526-3744 Contact: Ewa Nadolczak, Manager of Clinical Affairs, Direct (609) 408-1331 enadolczak@zeusscientific.com

Measurand: Treponema pallidum IgG antibodies. Type of Test: ELISA. Proprietary Name: ZEUS ELISA Treponema pallidum IgG Test System.

Section 1: Regulatory Information

  • Device Classification: Enzyme Linked Immunoabsorbent Assay, Treponema pallidum 1.
    1. Regulation Description: Treponema pallidum treponemal test reagents
    1. Class: 2
    1. Product Code: LIP
    1. Panel: Microbiology
  • Regulation Number: 866.3830 ହ.

Section 2: Intended Use

The ZEUS EUSA Treponema pallidum IgG Test System is intended for the qualitative detection of specific IgG class antibodies to T. pollidum in human serum. The test may be used in conjunction with non treponemal testing and clinical findings to provide serological evidence of infection with T. pallidum. This test is for in vitro diagnostic use only.

This test is not intended for screening blood or plasma donors.

{1}------------------------------------------------

Section 3: Indications for Use

Indications for Use: This test system when used in conjunction with non-treponemal based assays provides serological evidence of infection with Treponema pallidum.

Special Conditions for Use Statements:

This test is for in vitro diagnostic use only. · This test is for prescription use only. This test is not intended for screening blood or plasma donors.

Section 4: Substantial Equivalence

Examination of enclosed data indicates that the ZEUS ELISA Treponema pallidum IgG Test System for the qualitative detection of specific human IgG class antibodies to treponema pallidum in human serum is substantially equivalent to a commercially marketed test system which has been previously cleared by the FDA for in vitro diagnostic use.

T. pallidum IgG Result Interpretation:
OD ValueResultInterpretation
≤ 0.90NegativeA result of ≤0.90 indicates no detectable IgG antibodies to T. pallidum and should be reportedas non-reactive for IgG antibody to T. pallidum.
0.91 - 1.1EquivocalSpecimens with results in the equivocal range (0.91 - 1.1) should be retested in duplicate. Anytwo of the three results which agree should be reported. Specimens that remain equivocalafter repeat testing may be tested by an alternate serologic procedure and/or re-evaluated bydrawing another sample one to three weeks later.
>1.1PositiveA result of > 1.1 indicates that the specimen is positive for IgG antibody to T. pallidum, thecausative agent for syphilis.
A positive test result presumes a current or past infection with T. pallidum, and should bereported as reactive for IgG antibody to T. pallidum. Other T. pallidum serology assays shouldbe performed to confirm or rule out a current case of active syphilis.

Section 5: Interpretation of Results

{2}------------------------------------------------

Section 5A: Comparison of Investigational Device to Predicate Device

The comparisons of the ZEUS ELISA T.pallidum IgG Test System to the predicate device follow, including intended use and various aspects of the procedure.

CharacteristicZEUS ELISA T.pallidum IgGPheonix Bio-Tech Trep-Chek
UseIntended UseFor in vitro diagnostic use onlyIntended Use: The ZEUS ELISA Treponemapallidum ELISA Test System is intended forthe qualitative detection of specific IgG classantibodies to T. pallidum in human serum.The test system is intended to be used as anaid in the laboratory diagnosis of Syphilisdisease caused by T. pallidum spirochete.For in vitro diagnostic use onlyThe Phoenix Bio-Tech Corp. Syphilis Trep-Chek Test Kit is a confirmatory immunoassayfor the qualitative detection of Treponemapallidum IgG antibodies in human serum.
AssayDetection MethodSolid PhaseAntigen UsedImmunoassayColormetricPolystyrene 96 well platerecombinant p 17 Treponema pallidumantigenImmunoassayColormetricPolystyrene 96 well plateRecombinant Treponema pallidum antigen
Specimen TestedHuman SerumHuman Serum
ControlsOne Positive Control, one Negative ControlOne Positive Control, one Negative Control
CalibrationIncludes Calibrator (human serum)Includes Calibrator (human serum)
Analyte MeasuredSample DilutionHuman IgG1:21 in SAVe DiluentHuman IgG1:20 in phosphate buffer based diluent
Sample Incubation25 +/- 5 minutes at room temperature30 +/- 2 minutes at room temperature
Post Sample Wash5x wash (dispense/aspirate)4x wash (dispense/aspirate)
ConjugateGoat anti-human IgG; Fc chain specificGoat anti-human IgG; y chain specific
Conjugate LabelHorse radish peroxidaseHorse radish peroxidase
Conjugate Incubation25 +/- 5 minutes at room temperature30 +/- 2 minutes at room temperature
Post Conjugate Wash5x wash (dispense/aspirate)4x wash (dispense/aspirate)
SubstrateTMBTMB
ReadingData PointsRead the color change (optical density) ofthe wellsRead one OD value for each control andsampleRead the color change (optical density) ofthe wellsRead one OD value for each control andsample
MathSingle point curveSingle point curve
ScaleCalculate the index value of unknownsamples by comparing their OD to the cutoff ODCalculate the index value of unknownsamples by comparing their OD to the cutoff OD
Interpretation CriteriaNegative is <= 0.90, Positive is >= 1.10 andEquivocal is 0.90 - 1.09Negative is <= 0.90, Positive is >= 1.10 andEquivocal is 0.90 - 1.09

{3}------------------------------------------------

Section 6: Test Principle

The ZEUS ELISA Treponema pallidum IgG Test System is designed to detect IgG class antibodies in human sera to Treponema pallidum. Wells of plastic microwell strips are sensitized by passive absorption with T. pallidum antigen. The test procedure involves three incubation steps:

    1. Test sera are diluted with SAVE Diluent. During sample incubation, any antigen specific lgG antibody in the sample will bind to the immobilized antigen. The plate is washed to remove unbound antibody and other serum components.
    1. Peroxidase conjugated goat anti-human IgG is added to the wells and the plate is incubated. The conjugate will react with IgG antibody immobilized on the solid phase in step 1. The wells are washed to remove unbound conjugate.
    1. The microwells containing immobilized peroxidase conjugate with peroxidase substrate solution. Hydrolysis of the substrate by peroxidase produces a color change. After ten minutes have elapsed, the reaction is stopped and the color intensity of the solution is measured photometrically. The color intensity of the solution depends on upon the antibody concentration in the original test sample.

Section 7: Linearity

The WHO International Standard was diluted serially. Each dilution was tested in duplicate, the mean calculated and the result plotted.

Image /page/3/Figure/7 description: The image is a graph titled "T.pallidum IgG Linearity". The x-axis is labeled "Numerical Interpretation of Dilution Factor" and ranges from 0 to 0.6. The y-axis is labeled "Mean Index Value" and ranges from 0.00 to 3.50. The graph shows a linear relationship with an R-squared value of 0.9971.

Section 8: Analytical Specificity

Interfering Substances

The effect of potential interfering substances on sample results generated using the test system was evaluated with the following possible interfering substances: albumin, bilirubin, cholesterol, hemoglobin, triglycerides and intralipids.

{4}------------------------------------------------

The quantity of analyte in each interfering substance is as follows:

Bilirubin: 1mg/dL (low), 15 mg/dL (high) Albumin: 3.5 g/dL (low), 5 g/dL (high) Cholesterol: 150 mg/dL (low), 250 mg/dL (high) Triglycerides: 150 mg/dL (low), 500 mg/dL (high) Hemoglobin: 20 g/dL (low), 20 g/dL (high) Intralipid: 300 mg/dL (low), 750 mg/dL (high)

Three samples for Treponema pallidum IgG were chosen based on their performance on the Treponema pallidum IgG ELISA test system: positive, borderline and negative. The samples were exposed to the possible interfering substances and tested. An increase or reduction of signal equal to or less than 20% is considered acceptable. The negative sample may have a signal change greater than 20% if there is no change in the qualitative result of the sample. All signal changes greater than 20 % must be specified in the package insert.

All positive samples showed less than a 20% recovery of signal.

The borderline samples showed a recovery of signal <20% of with the exception of the high spike of hemoglobin (25.2%).

The negative sample showed a change of signal (>20%) with the high and low spikes of albumin, hemoglobin, intralipid, bilirubin, cholesterol and triglycerides. The negative sample results in each instance stayed below the cut-off and the change in signal did not affect the qualitative result.

Cross Reactivity

Ten samples minimally for each cross-reactant were tested. The results presented were obtained by testing the analytes against high concentrations of possible cross reactants. The results of this study are summarized in the following table.

Treponema pallidum IgGCross reactivity Study
Analytepositive/tested
EBV0/10
ANA0/10
RF IgM0/10
Rubella0/10
HIV0/10
HSV 10/10
HSV 20/10
Sera from Pregnant Patients0/10
Hepatitis B0/10
VZV0/10
VZV IgM0/10
CMV0/10
Toxoplasma0/10
Lyme G/M0/10
Hepatitis C0/10

{5}------------------------------------------------

Detection Limits

This is a qualitative assay (Positive, Negative or Equivocal results) so there is no limit of detection study.

Section 9: Clinical Performance

Clinical Data Generated for Submission: Method Comparison with Predicate Device

    1. Prospectively Collected Intended Use Populations:
    • . 500 unselected samples from patients with a syphilis test ordered. The samples were submitted for syphilis antibody testing, sequentially numbered, de-identified and archived.
    • . 500 unselected samples from pregnant women with a syphilis test ordered. The samples were submitted for syphilis antibody testing, sequentially numbered, de-identified and archived.

After procurement, the samples were tested at a hospital laboratory located in the Mid-Atlantic, the Northeast and at Zeus. The hospital laboratories tested 200 samples from patients with a syphilis test ordered and 200 samples from pregnant women. Zeus tested 100 samples from patients with a syphilis tests ordered and 100 samples from pregnant women.

    1. Prospectively Collected Population of Unselected Hospitalized Patients: Additional clinical performance was assessed in a population of 1000 hospitalized patients. These samples were pulled from a hospital laboratory routine workload of patient testing.
    1. Prospective HIV-1 Positive Samples: 223 banked known positive HIV-1 samples were acquired from a New York vendor.
  • Retrospective TPPA /RPR Positive: 280 samples requested to be RPR/TPPA positive were 4. purchased from a New York vendor.
    1. Retrospective TPPA Negative Collected from Pregnant Women: 250 samples requested to be collected from pregnant women and requested to be syphilis antibody negative were purchased from a New York vendor.
  • Retrospective TPPA Positive Collected from Pregnant Women: 27 samples requested to be 6. collected from pregnant women and requested to be RPR/TPPA positive were purchased from a New York vendor.
    1. CDC Syphilis Panel: 157 samples of various reactivity to syphilis were evaluated.

{6}------------------------------------------------

    1. Precision and Reproducibility: Precision of the device was assessed using fifteen samples. These repeatability studies were performed internally at the manufacturing site's laboratory. Reproducibility was assessed using fifteen samples tested at two external sites and internally at Zeus.

Prevalence and Expected Values

To determine expected values in the populations tested, internal investigators assessed the device's performance with 500 masked samples prospectively collected from individuals and 500 samples from pregnant women. The samples were requested to be random, unselected sera submitted for syphilis antibody testing. Additional studies were conducted in a population of 1000 unselected hospitalized patients.

In the 500 prospectively collected samples from patients ranging in age from <1 to >70 years of age, 7 tested positive. The overall observed prevalence in this group was 1.4% (7/500 samples).

In the 500 samples collected from pregnant women ranging in age from 15 to 48, 3/498 samples tested positive. The observed prevalence in this group was 0.6%. Two samples were excluded due to questionable age.

In the group of 1000 samples from unselected hospitalized patients ranging in age from <1 to >70 years of age, 32 tested positive. The overall observed prevalence in this group was 3.2% (32/999 samples). One sample was QNS for testing.

Clinical Performance and Comparative Testing for Intended Use Populations in a Prospective Study

The comparative study for the Intended Use Populations consisted of 500 unselected serum samples from patients with a syphilis test ordered. 500 purchased serum samples from pregnant women were also tested.

PERFORMANCE CHARACTERISTICS

The clinical study consisted of 2,937 serum samples evaluated at three sites located in the United States. All serum samples evaluated for concordance were tested with the treponemal (TPPA) reference assay. Samples that were positive by TPPA were reference assay positive. Samples that were negative by TPPA were reference assay negative. The testing sites, sample volumes are presented in the following table:

Sites, Populations and Amounts of Samples Tested
Prospective SamplesRerospective Samples
patients withpregnantsera frompurchasedpurchasedpurchasedsera requestedCDC
syphilis testwomen withunselectedsera fromsera fromsera requestedto beSyphilis
testsyphilis testhospitalizedHIV positivepregnantto becharacterized

{7}------------------------------------------------

SitesorderedorderedpatientspatientswomenRPR/TPPA +samples
Zeus100100350223000
Monmouth20020035002772800
Mainline2002003000000
CDC000000157
Total5005001000223277280157

Prospectively Collected Intended Use Populations:

500 unselected samples from patients with a syphilis test ordered. The samples were submitted for syphilis antibody testing, sequentially numbered, de-identified and archived.

Banked sera from patients with syphilis test ordered
Predicate
PositiveEquivocalNegativeSite TotalPPA/NPA95% CI
T.pallidum IgG ELISAPositive403780.0%28.4-99.5%
Equivocal1012
Negative0049149199.2%97.9-99.8%
Site Total50495500

500 unselected samples from pregnant women with a syphilis test ordered. The samples were submitted for syphilis antibody testing, sequentially numbered, de-identified and archived.

Banked purchased sera from pregnant women with syphilis test ordered
Predicate
PositiveEquivocalNegativeSite TotalPPANPA95% CI
T.pallidum IgG ELISAPositive300375.0%19.4-99.4%
Equivocal0000
Negative10494495100.0%99.4-100%
Site Total40494498

*2 samples excluded due to unverifiable age

Prospectively Collected Population of Unselected Hospitalized Patients: Additional clinical performance was assessed in a population of 1000 hospitalized patients. These samples were pulled from a hospital laboratory routine workload of patient testing and were tested at the three clinical sites.

Image /page/7/Figure/8 description: The image shows a table with three rows. The first row contains the text "Unselected hospitalized patients". The second row contains the text "Predicate". The third row contains the text "PPA".

{8}------------------------------------------------

PositiveEquivocalNegativeSite TotalNPA95% CI
T.pallidum IgG ELISAPositive131183261.9%38.4-81.9%
Equivocal10910
Negative7095095797.1%95.9-98.1%
QNS0011
Site Total2119781000

Retrospective HIV-1 Positive Samples: 223 banked known positive HIV-1 samples were acquired from a New York vendor.

Banked purchased known HIV-1 positive serum samples
Predicate
PositiveEquivocalNegativeSite TotalPPANPA95% CI
T.pallidum IgGELISAPositive41014285.4%72.2-93.9%
Equivocal1001
Negative4217418099.4%96.9-100%
Site Total462175223

Retrospective TPPA /RPR Positive: 280 samples requested to be RPR/TPPA positive were purchased from a New York vendor.

Banked purchased sera requested to be RPR/TPPA reactive
Predicate
PositiveEquivocalNegativeSite TotalPPANPA95% CI
T.pallidum IgGELISAPositive2591426498.5%96.2-99.6%
Equivocal11
Negative3121570.6%46.9-98.7%
Site Total263116280

Retrospective TPPA Negative and TPPA Positive Samples Collected from Pregnant Women: 250 samples requested to be collected from pregnant women and requested to be syphilis antibody negative were purchased from a New York vendor. Only 27 samples requested to be collected from pregnant women and requested to be RPR/TPPA positive were available. These samples were purchased from a New York vendor.

Banked purchased sera from pregnant women requested to be TPPA positive (27)

{9}------------------------------------------------

RPR/TPPA non-reactive (250)
Predicate
PositiveEquivocalNegativeSite TotalPPA/NPA95% CI
T.pallidum IgG ELISAPositive26102792.9%76.5-99.1%
Equivocal0000
Negative2024825099.6%97.8-100%
Site Total281248277

CDC Syphilis Panel: 157 samples from patients with various clinical diagnoses with syphills disease were evaluated. The % agreement of the Treponema pallidum IgG ELISA kit with the clinical diagnosis is presented in the following table.

Clinical DiagnosisT. pallidum IgG ELISA Results% Agreement with Clinical DiagnosisPresented with 95% CI
PositiveEquivocalNegativeTotal
Primary Treated110011100% (11/11)76.2-100%
Secondary Untreated41024395.3% (41/43)84.2-99.4%
Secondary Treated390039100% (39/39)92.6-100%
Latent Untreated6051154.5% (6/11)23.4-83.3%
Latent Treated48025096.0% (48/50)86.3-99.5%
Congenital111333.3% (1/3)0.84-90.6%
Total14611015793.0% (146/157)87.8-96.5%

PRECISION and REPRODUCIBILITY

Precision

Precision was evaluated at the manufacturer site. The study was conducted as follows: fifteen samples were identified and/or prepared (by Zeus Scientific, Inc.) for use in the study based upon their activity on the ZEUS ELISA Treponema pallidum IgG assay. Three samples each were selected that were negative, high negative, near cut-off, low positive. On each day of testing, the samples were diluted twice and tested. This was repeated in a second run on the same day by a different technologist for a total of twelve days. Precision is considered acceptable for the reactive samples if the Total CV is <15%. The reproducibility for the negative samples is considered acceptable if the Total CV is <25%. This study is summarized below:

Summary Of In-House Repeatability
PanelMemberSampleNMeanAU/mLWithin-RunWithin-DayBetween-RunTotal
SD%CVSD%CVSD%CVSD%CV
Negative 1480.080.0033.8%0.0068.0%0.0066.8%0.01113.1%
Negative 2480.120.0054.0%0.0036.5%0.0097.0%0.01310.5%
Negative 3480.500.0173.6%0.0349.5%0.0203.9%0.0459.0%
High Negative 1480.750.0577.6%0.0308.6%0.0192.5%0.0587.7%
High Negative 2480.720.0465.4%0.0147.1%0.0152.0%0.0527.2%

{10}------------------------------------------------

High Negative 3480.740.0152.0%0.0184.7%0.0111.4%0.0344.5%
Near Cut-off 1480.920.0283.0%0.0255.0%0.0363.9%0.0566.1%
Near Cut-off 2481.040.0222.1%0.0143.9%0.0333.1%0.0454.3%
Near Cut-off 3480.950.0373.9%0.0256.4%0.0363.8%0.0616.4%
Low Positive 1481.480.0292.0%0.0293.9%0.0140.9%0.0583.9%
Low Positive 2481.430.0261.8%0.0202.5%0.0171.2%0.0503.5%
Low Positive 3481.650.0271.6%0.0374.2%0.0181.1%0.0784.7%
High Positive 1485.430.1312.4%0.1543.8%0.275.0%0.387.0%
High Positive 2484.850.1102.3%0.1763.6%0.173.6%0.296.0%
High Positive 3484.740.1362.8%0.1894.9%0.173.5%4.745.2%
Non-Reactive Control480.130.0043.3%0.0086.23%0.0075.39%0.0108.2%
Reactive Control 1485.630.0497.5%5.55.15%0.295.11%0.427.51%

Reproducibility

Reproducibility was evaluated internally and at two external clinical sites. The study was conducted as follows: fifteen samples were identified and/or prepared (by Zeus Scientific, Inc.) for use in the study based upon their activity on the assay. Three samples each were selected that were negative, high negative, near cut-off, low positive and high positive. To assess reproducibility, on each day of testing, each sample was diluted twice and each dilution was run in triplicate. This was repeated in a second run by a second technologist resulting in twelve results per day. This was repeated for five days at each site and the resulting data used. Reproducibility is considered acceptable for the reactive samples if the Total CV is <15% and if the Total CV for the borderline and positive samples do not congregate at the high end of acceptability but show a spread of results of at least 3%. The reproducibility for the negative sample is considered acceptable if the Total CV is <50% and shows no change in the qualitative outcome.

PanelMemberSampleNMeanIndex valueWithin-RunWithin-DayBetween-RunBetween-SiteTotal
SD%CVSD%CVSD%CVSD%CVSD%CV
Negative 11800.060.0114.7%0.0119.3%0.0110.7%0.0229.1%0.0236.6%
Negative 21800.080.019.7%0.0112.2%0.018.3%0.0114.6%0.0115.5%
Negative 31800.310.038.6%0.0310.7%0.026.6%0.0412.7%0.0413.3%
High Negative 11800.800.045.1%0.056.3%0.034.2%0.067.3%0.067.3%
High Negative 21800.740.045.1%0.045.8%0.023.0%0.057.0%0.057.4%
High Negative 31800.760.045.0%0.045.6%0.212.7%0.056.5%0.057.2%
Near Cut-off 11801.050.076.3%0.077.3%0.033.6%0.099.0%0.1110.3%
Near Cut-off 21801.130.054.7%0.065.4%0.043.1%0.076.0%0.086.7%
Near Cut-off 31800.950.055.6%0.076.7%0.044.0%0.088.5%0.099.9%
Low Positive 11801.450.096.2%0.117.6%0.064.4%0.138.9%0.149.6%
Low Positive 21801.770.115.9%0.147.8%0.105.8%0.158.3%0.169.2%
Low Positive 31801.930.147.1%0.178.9%0.125.9%0.199.7%0.2110.7%
Positive1803.60.205.7%0.226.2%0.102.8%0.308.4%0.3710.2%
Positive 21803.10.206.2%0.227.3%0.134.4%0.289.0%0.3410.9%
Positive 31803.10.185.7%0.226.9%0.164.9%0.268.4%0.3110.2%
Non-Reactive Control1800.090.0110.6%0.0112.8%0.016.2%0.0115.9%0.0221.0%
Reactive Control1803.90.164.1%0.125.1%0.143.5%0.225.7%0.025.9%

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three legs, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Zeus Scientific Inc. c/o Ms. Ewa Nadolczak Manager Clinical Affairs 200 Evans Way Branchburg, NJ 08876

FEB 10 201

Re: K102283 Trade/Device Name: Treponema pallidum IgG ELISA Test System Regulation Number: 21CFR §866.3830 Regulation Name: Treponema pallidum treponemal test reagents. Regulatory Class: Class II Product Code: LIP Dated: January 25, 2011 Received: January 26, 2011

Dear Ms. Nadolczak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section

{12}------------------------------------------------

Page 2 - Ewa Nadolczak

510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Sally A. Hojvat, M.S., Ph.D.

Sally A. Hoivat, M.Sc . Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

Indications for Use

510(k) Number: K102283

Device Name: ZEUS ELISA Treponema pallidum IgG Test System

Indications for Use:

The ZEUS ELISA Treponema pallidum IgG Test System is intended for the qualitative detection of specific IgG class antibodies to T. pallidum in human serum. The test may be used in conjunction with non treponemal testing and clinical findings to provide serological evidence of infection with T. pallidum. This test is for in vitro diagnostic use only.

This test is not intended for screening blood or plasma donors.

Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Leeddi m. Poole
Division Sign-Off

Page 1 of __ 1

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102283

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).