K Number
K102214

Validate with FDA (Live)

Device Name
ENDO FLIP
Manufacturer
Date Cleared
2010-12-16

(132 days)

Product Code
Regulation Number
876.1725
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EndoFLIP® EF-620 catheter is indicated for use as an adjunct to a bougie for measuring the size of a gastric sleeve created during bariatric surgery. It is suitable for diameter measurements for 22 to 60 French sleeves.

The EF-620 catheter is to be used only with the EndoFlip System.

Device Description

The EndoFLIP® comprises a measuring system and a single use catheter to assist in measuring the gastric sleeve. In practice, the EndoFLIP® balloon catheter is attached to a syringe, pre-filled with a diluted saline solution, which is inserted into the syringe pump on the front of the EndoFLIP® system. The deflated balloon is inserted into the stomach by an anesthesiologist with placement being confirmed by the surgeon under direct laparoscopic visualization. One the balloon has been correctly located, it is then inflated with the diluted saline solution to user programmed volume. The display estimates the gastric sleeve diameter at sixteen points along the balloon. The system also allows snapshots to be taken and compared to real-time images.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Estimated balloon diameter (Dest) at 16 points in the balloon, displayed in numeric and graphical form.Yes, the device provides this functionality.
Range: 7 to 20 mmRange: 7 to 20 mm (Matches acceptance criteria)
Resolution: 0.1 mmResolution: 0.1 mm (Matches acceptance criteria)
Accuracy: ± 1 mm (at 95% confidence) rounded to nearest integerAccuracy: ± 1 mm (at 95% confidence) rounded to nearest integer (Matches acceptance criteria)

Study Details

The provided document describes a 510(k) submission, which focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive clinical or standalone performance study in the way one might evaluate a new AI algorithm.

1. Sample sized used for the test set and the data provenance:

  • The document states, "We performed the appropriate bench testing to demonstrate that the new catheter in combination with the EndoFLIP® controller performs within the specifications of the predicate catheter with the EndoFLIP® controller."
  • The exact sample size for the "bench testing" is not directly specified in the document.
  • Data provenance is not specified. The testing appears to be internal bench testing, likely conducted in a lab environment.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This device measures physical dimensions using a specific electromechanical system. The "ground truth" for its accuracy would likely be established using calibrated physical measurement tools (e.g., precise calipers, micrometers, or reference gauges) in a laboratory setting, not expert human readers. There is no mention of human experts being involved in establishing ground truth for the diameter measurements.

3. Adjudication method for the test set:

  • Not applicable. As the "ground truth" is established by physical measurement tools, there is no need for expert adjudication.

4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No such study was conducted or reported. The EndoFLIP® system is a measurement device that provides data; it does not involve AI assistance for human readers in interpreting images or making diagnoses.

5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • While the device itself is a "standalone" system in terms of performing its measurement function, the "performance" section refers to the system's ability to accurately measure diameter, which is its core function. There isn't an "algorithm only" component that would perform a task typically done by a human, making this question less applicable in the context of this specific device's function. The stated accuracy reflects its standalone measurement capability.

6. The type of ground truth used:

  • The ground truth for the diameter measurements would be based on calibrated physical measurements (e.g., using precision measurement tools) of known diameters against which the device's readings are compared.

7. The sample size for the training set:

  • Not applicable. The EndoFLIP® system, as described, is an electromechanical measurement device. It does not utilize machine learning or AI that would require a "training set" in the conventional sense. Its function is based on established physical principles and calibrated components.

8. How the ground truth for the training set was established:

  • Not applicable. As no training set for a machine learning model is involved, this question is not relevant.

{0}------------------------------------------------

K102214
Page 1 of 3

510(k) Summary Page 1 of 3 9-Dec-10

DEC 1 6 2010

CompanyCrospon Ltd.Galway Business ParkDanganGalway, Ireland
Tel - 011 [353] (91) 519882Fax - 011 [353] (91) 519873
Official Contact:John O'Dea PhD
Proprietary or Trade Name:EndoFLIP®
Common/Usual Name:Gastrointestinal motility monitoring system
Classification / CFR:FFX / CFR 876.1725
Device:EndoFLIP® System and Catheter
Predicate Devices:K092850 – Crospon – EndoFLIP®

Device Description:

The EndoFLIP® comprises a measuring system and a single use catheter to assist in measuring the gastric sleeve. In practice, the EndoFLIP® balloon catheter is attached to a syringe, pre-filled with a diluted saline solution, which is inserted into the syringe pump on the front of the EndoFLIP® system. The deflated balloon is inserted into the stomach by an anesthesiologist with placement being confirmed by the surgeon under direct laparoscopic visualization. One the balloon has been correctly located, it is then inflated with the diluted saline solution to user programmed volume. The display estimates the gastric sleeve diameter at sixteen points along the balloon. The system also allows snapshots to be taken and compared to real-time images.

Indications for Use:

The EndoFLIP® EF-620 catheter is indicated for use as an adjunct to a bougie for measuring the size of a gastric sleeve created during bariatric surgery. It is suitable for diameter measurements for 22 to 60 French sleeves.

The EF-620 catheter is to be used only with the EndoFlip System.

Patient Population:Patients undergoing bariatric surgery
Environment of Use:Hospitals and Surgery Centers
Contraindications:The EndoFLIP® System is contraindicated where endoscopy iscontraindicated.

{1}------------------------------------------------

510(k) Summary Page 2 of 3

9-Dec-10

Device Attributes:

.

. . . . .

:

11.000

.

Design
Principle of OperationSyringe pump device that targets an inflate volume within a balloon
inserted into the lumen of an organ. Provides an Estimated
Diameter (Dest) of the balloon at 16 points along its length when
inflated with custom conductive solution.
Dest electrodes are located at 16-points along the catheter inside a
balloon.
Energy Used And/OrNo electrical energy is delivered into the patient
Delivered
Human FactorsSpecified for 20 to 40°C operating environment
Touch screen user interface
On screen keypad and external keypad interface provided
Data RecordingData is saved internally and can be exported to a USB mass storage
device or printed to an attached USB printer
Electrical SafetyIEC60601-1 2nd Ed. + Am.1 + Am.2
Mechanical SafetyTravel limits are detected by mechanical switches
Syringe is automatically put in its home position at power on
Maximum volume is set by the volume of solution in the syringe
which is factory filled. This volume matches the balloon size on the
catheter. The Balloon Inflate Volume cannot be set above the
factory set syringe volume.
The Deflate button is available after inflation has commenced.
The syringe can be manually retracted at any time.
Rigid syringe used to avoid potential volume errors
Stepper motor driven lead screw is used to maintain accurate
control over the syringe piston position. Piston movement
resolution is 0.003175 mm (one step)
No calibration required
Chemical SafetyConductive solution inside balloon is diluted saline
Thermal SafetyInternal cooling fan with enclosure temperature monitoring. Alarm
if temperature exceeds limits.
BiocompatibilityAll materials have passed biocompatibility tests in accordance with
ISO 10993-1
Compatibility With TheEndoFLIP operates with custom catheters only
Environment And Other
Devices
SterilityAccessories are supplied non-sterile and are single patient use
PerformanceEndoFLIP ® passed the performance testing :
Estimated balloon diameter (Dest) at 16 points in the balloon,
displayed in numeric and graphical form
Range: 7 to 20 mm
Resolution: 0.1 mm
Accuracy: ± 1 mm (at 95% confidence) rounded to nearest integer

{2}------------------------------------------------

510(k) Summary Page 3 of 3 9-Dec-10

Substantial Equivalence:

The EndoFLIP® system with EF-620 catheter is viewed as substantially equivalent to the predicate device because:

Indications:

Similar to predicate - K092850 - Crospon EndoFLiP® system with catheter measure the diameter of the stoma during gastric band procedures while the proposed indication of the EndoFLIP® EF-620 catheter is for use in measuring the size of a gastric sleeve created during bariatric surgery.

Technology:

Identical technology to predicate - K092850 - Crospon EndoFLIP® system with catheter

Both catheters can only be used with the predicate EndoFLIP® controller and the construction and materials of the new catheter are identical to the predicate.

Materials:

The materials in contact with the patient are identical to predicate device K092850 – Crospon EndoFLIP® system with catheter.

Environment of Use:

Identical to predicate - K092850 -- Crospon EndoFLIP® system with catheter

Patient Population:

Identical to predicate - K092850 - Crospon EndoFLIP® system with catheter

Comparative Performance and Specifications:

We performed the appropriate bench testing to demonstrate that the new catheter in combination with the EndoFLIP® controller performs within the specifications of the predicate catheter with the EndoFLIP® controller.

Therefore one can find that the EndoFLIP® system with new catheter and the new indications for use are viewed as substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's emblem, which is a stylized depiction of an eagle or bird in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged around the upper half of the circle, following its curvature.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Crospon, Ltd. c/o Mr. Paul E. Dryden President and Regulatory Consultant ProMedic, Inc. 24301 Woodsage Drive BONITA SPRINGS FL 34134-2958

DEC 1 6 2010

ﺮ ﺟﺰﻳﺮ ﺍﻟ

Re: K102214

Trade/Device Name: EndoFLIP® System

Regulation Number: 21 CFR §876:1725

Regulation Name: Gastrointestinal motility and monitoring system

Regulatory Class: II

Product Code: FFX

Dated: December 9, 2010 Received: December 10, 2010

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability . warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{4}------------------------------------------------

Page 2

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Hukal Lemur MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number: K102214

Device Name:

EndoFLIP® System

Indications for Use:

The EndoFLIP® EF-620 catheter is indicated for use as an adjunct to a bougie for measuring the size of a gastric sleeve created during bariatric surgery. It is suitable for diameter measurements for 22 to 60 French sleeves.

The EF-620 catheter is to be used only with the EndoFlip System.

Prescription Use XX (Part 21 CFR 801 Subpart D)

or

Over-the-counter use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ductive, Gastro-Renal, and

§ 876.1725 Gastrointestinal motility monitoring system.

(a)
Identification. A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.(b)
Classification. Class II (performance standards).