(102 days)
The STARFIRE™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:
- degenerative disc disease (DDD; defined as back pain of discogenic origin . with degeneration of the disc confirmed by history and radiographic studies);
- spondylolisthesis; ●
- trauma (i.e., fracture or dislocation); ●
- . spinal stenosis;
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); .
- . tumor;
- pseudoarthrosis; and .
- . failed previous fusion
The Choice Spine STARFIRE™ Pedicle Screw System is a top-loading, polyaxial screw spinal fixation system including polyaxial screws, rods, and connectors. The components are provided clean and non-sterile. Various sizes of the implants are provided.
The following information is extracted from the provided text regarding the Choice Spine STARFIRE™ Pedicle Screw System:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Type | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Static Compression Bending | Equivalence in bending loads and stiffness to predicate systems. | Cannulated STARFIRE pedicle screw constructs were equivalent, with respect to bending loads and stiffness, to non-cannulated predicate pedicle fixation systems. |
| Compression Bending Fatigue | Equivalent or greater endurance strength compared to predicate systems. | Cannulated STARFIRE pedicle screw constructs had equivalent or greater endurance strength compared to non-cannulated predicate pedicle fixation systems. |
| Static Torsional | Ability to resist loads greater than the range required for the human lumbar spine. | Cannulated STARFIRE pedicle screw constructs, without cross-connectors, were able to resist loads greater than the range of resistance required for the human lumbar spine. |
2. Sample Size for Test Set and Data Provenance
The document does not explicitly state the sample size used for each specific test or the "data provenance" in terms of country of origin or retrospective/prospective nature. It refers to "Mechanical testing" and "Test results," which implies the testing was conducted on physical samples of the device and predicate devices.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable to this type of device. The "ground truth" for the performance tests of a pedicle screw system is based on objective mechanical testing data and established ASTM standards, not expert interpretations of clinical data or images.
4. Adjudication Method for the Test Set
This information is not applicable as the tests are objective mechanical performance tests, not subjective assessments requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This type of study is typically conducted for diagnostic imaging devices where human readers interpret medical images.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, a standalone algorithm performance study was not done. This device is a physical pedicle screw system, not an AI or software algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the performance of the STARFIRE™ Pedicle Screw System was established through objective mechanical testing according to the guidelines outlined in ASTM F1717-09 "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model."
8. The Sample Size for the Training Set
This information is not applicable. This device is a physical medical implant, not an AI model or software that requires a "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable as there is no "training set" for this device.
{0}------------------------------------------------
8. 510(k) Summary
Preparation Date
November 11, 2010
Sponsor
Choice Spine, LP 306 Erin Drive Knoxville, TN 37919 p: 865.246.3333 f: 865.588.4045
Contact
G. Todd Hawkins Director of Regulatory Affairs/Quality Assurance
Trade Name
Choice Spine STARFIRE™ Pedicle Screw System
Common Name
Pedicle Screw Spinal System; Pedicle Screws
Regulatory Classification & Device Product Codes
888.3070
MNI - Orthosis, Spinal Pedicle Fixation MNH - Orthosis, Spondylolisthesis Spinal Fixation NKB - Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease
Predicate Devices
OPTIMA™ Spinal System (K024096, K031585) Firebird™ Spinal Fixation System (K081684) Coral™ Spinal Fixation System (K041592) Expedium™ 6.35 Spinal Fixation System (K062174)
Device Description
The Choice Spine STARFIRE™ Pedicle Screw System is a top-loading, polyaxial screw spinal fixation system including polyaxial screws, rods, and connectors. The components are provided clean and non-sterile. Various sizes of the implants are provided.
Intended Use
The Choice Spine STARFIRE™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities
NOV 1 5 2010
{1}------------------------------------------------
or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
Materials
The Choice Spine STARFIRE™ Pedicle Screw System implants are manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136, commercially pure titanium (G2) as described by ASTM F67 and stainless steel (316 LVM) as described by ASTM F138.
Technological Characteristics
The Choice Spine STARFIRE™ Pedicle Screw System consists of a range of polyaxial screws, rods, and connectors identical to the identified predicate systems.
The Intended Use of the STARFIRE™ System is identical to the predicate systems. The STARFIRE™ System materials are similar to the predicate systems.
The STARFIRE™ Pedicle Screw System utilizes a set screw mechanism to secure the rod in the same manner as the predicate systems.
Substantial Equivalence
Documentation was provided that demonstrates the Choice Spine STARFIRE™ Pedicle Screw System to be substantially equivalent to previously cleared device systems. The substantial equivalence is based upon equivalence in intended use, indications, anatomic location, materials, and performance.
Mechanical testing was performed according to the guidelines outlined in ASTM F1717-09 "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model."
Static compression bending tests showed that cannulated STARFIRE pedicle screw constructs were equivalent, with respect to bending loads and stiffness, to noncannulated predicate pedicle fixation systems.
Compression bending fatigue tests demonstrated that the cannulated STARFIRE pedicle screw constructs had equivalent or greater endurance strength compared to non-cannulated predicate pedicle fixation systems.
In static torsional tests the cannulated STARFIRE pedicle screw constructs, without cross-connectors, were able to resist loads greater than the range of resistance required for the human lumbar spine.
Test results indicate that the STARFIRE™ System is equivalent to the predicates in each of the test modes. Documentation of the testing is available in Section 16.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three lines extending from its head, representing growth and development. To the left of the figure, the text "DEPARTMENT OF HEALTH &" is arranged vertically, followed by "HUMAN SERVICES" in a curved formation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Choice Spine, LP % Mr. G. Todd Hawkins Director of Regulatory Affairs/Quality Assurance 306 Erin Drive Knoxville, Tennessee 37919
NOV 1 5 2010
Re: K102204
Trade/Device Name: Choice Spine STARFIRE™ Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI Dated: October 21, 2010 Received: October 22, 2010
Dear Mr. Hawkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Mr. G. Todd Hawkins
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Barbara Muchup
Mark N. Melker Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
7. Indication for Use Statement
510(k) Number (if known): KLU2204
NOV 1 5 2010
Device Name: Choice Spine STARFIRE" Pedicle Screw System
Indications for Use:
The STARFIRE™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:
- degenerative disc disease (DDD; defined as back pain of discogenic origin . with degeneration of the disc confirmed by history and radiographic studies);
- spondylolisthesis; ●
- trauma (i.e., fracture or dislocation); ●
- . spinal stenosis;
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); .
- . tumor;
- pseudoarthrosis; and .
- . failed previous fusion
Prescription Use _ _ _ _ X _ Over-The-Counter Use (21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of 1
510(k) Number_________________________________________________________________________________________________________________________________________________________________
N/A