K Number
K102198

Validate with FDA (Live)

Date Cleared
2010-11-05

(93 days)

Product Code
Regulation Number
892.1560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Quantel Medical COMPACT TOUCH STS is intended to be used for diagnostic imaging and measurement of the eye including:
◆ Visualization of the interior of the eye by B scan
◆ Make measurements inside the eye (Sulcus to Sulcus, Irido Cornea Angle Left, Irido Cornea Angle Right, Lens curvature and Anterior Chamber Depth).

Device Description

The COMPACT TOUCH STS is an ultrasonic system designed for ophthalmic use. It performs B type scans for diagnostic imaging of the eye. It is not intended to be used for determining the power of implanted lenses, but rather is capable of making intra-ocular measurements. The system is composed of a main console controlled by a touch screen. It uses a B-scan probe (LIN50) which is a motor driven 50 MHz transducer.

AI/ML Overview

The provided text describes a 510(k) summary for the Quantel Medical COMPACT TOUCH STS Ophthalmic Ultrasound System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the device meets specific acceptance criteria through clinical studies. Therefore, much of the requested information regarding acceptance criteria, extensive performance data, and detailed study methodologies is not available in this document.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria for performance metrics. Instead, it relies on the concept of "substantial equivalence" to a predicate device.

Acceptance Criteria (Not explicitly stated/Quantitative)Reported Device Performance (Summary)
Implicit Acceptance: Device uses similar technology and physical output characteristics to predicate device.Device uses similar technology and physical output characteristics as the Sonomed Inc. VuMax Ophthalmic Ultrasound System (K060626).
Implicit Acceptance: Meets design specifications.Laboratory testing verifies the device meets design specifications.
Implicit Acceptance: Similar indications for use.Indications for use are similar to previously cleared devices.
Implicit Acceptance: Comparable risks and benefits.Risks and benefits are argued to be comparable to predicate devices.
Implicit Acceptance: No new questions of safety or efficacy.No new questions of safety or efficacy are raised.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not applicable. The submission states, "No Clinical testing is required." The evaluation was based on laboratory testing to verify design specifications and demonstrate substantial equivalence, not on a clinical test set with patient data.
  • Data Provenance: Not applicable, as no clinical test set data is described. The "laboratory testing" mentioned implies internal testing data, likely from the manufacturer's facilities.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable. Since no clinical test set was used and no "ground truth" for patient data was established in the context of this submission, the number and qualifications of experts are not relevant or provided.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. There was no clinical test set requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No. The document explicitly states, "No Clinical testing is required." Therefore, no MRMC study, either with or without AI assistance, was conducted or reported.
  • Effect size of human readers improving with AI vs without AI assistance: Not applicable, as no MRMC study was done and the device is a standalone ultrasound system, not an AI-assisted diagnostic tool.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Was a standalone study done? Yes, in a sense, but not in the context of an AI algorithm. The device itself, as an ophthalmic ultrasound system, is a standalone diagnostic tool. The "Performance Data" section describes "Laboratory testing following the guidance 'Information for manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound systems and Transducers' (Sept 9, 2000)" to verify it met design specifications. This implies testing the device's physical performance, imaging capabilities, and measurement accuracy in a controlled laboratory setting, independent of human interpretation for substantial equivalence purposes.

7. Type of Ground Truth Used

  • Type of Ground Truth: For the "laboratory testing," the ground truth would have been established through validated measurement standards, phantoms, and calibrated equipment, rather than clinical ground truth like pathology or expert consensus on patient cases. The goal was to verify the device's technical specifications and ensure it produced accurate ultrasound images and measurements in a controlled environment, consistent with its design.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. The COMPACT TOUCH STS is an ophthalmic ultrasound system, not an AI/ML-based device that requires a training set of data.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth Was Established: Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Section 5

OV - 5 2010

510(k) Summary

This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and 807.92. Summary preparation date 08-06-10 [21 CFR 807.92(a)(1)].

Applicant Name and Address [21 CFR 807.92(a)(1)] A.

Quantel Medical S.A.

21 rue Newton

ZI Le Brezet

63039 Clermont Ferrand Cedex 2

FRANCE

Tel: +33 (0) 473 745 745

Fax: +33 (0) 473 745 700

B. Contact Information

Quantel USA

601 Haggerty Lane

Bozeman, MT 59715

Tel: 406-586-0131

Fax: 406-586-2924

Contact person: Michael Johnson M.D.

{1}------------------------------------------------

C. Device Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade Name: COMPACT TOUCH STS Ophthalmic Ultrasound System

Device Common Name: Ophthalmic Ultrasound System

Classification Name: Ultrasonic Pulsed Echo Imaging System (21 CFR 892.1560) and Diagnostic Ultrasound Transducer (21 CFR 892.1570).

Product Code: IYO and ITX, respectively

Panel: 86 Ophthalmic

Device Classification: Class II

D. Predicate Devices [21 CFR 807.92(a)(3)]

The COMPACT TOUCH STS uses similar technology and physical output characteristics as the following predicate devices:

K060626 Sonomed Inc. VuMax Ophthalmic Ultrasound System

E. Device Description [21 CFR 807.92(a)(4)]

The COMPACT TOUCH STS is an ultrasonic system designed for ophthalmic use. It performs B type scans for diagnostic imaging of the eye. It is not intended to be used for determining the power of implanted lenses, but rather is capable of making intra-ocular measurements. The system is composed of a main console controlled by a touch screen. It uses a B-scan probe (LIN50) which is a motor driven 50 MHz transducer.

F. Device Specifications [21 CFR 807.92(a)(6)]

The COMPACT TOUCH STS system includes a B-scan ultrasonic probe. The Bscan probe operates at 50 MHz and has an active diameter of 4.5 mm.

5-2

{2}------------------------------------------------

Indications for Use [21 CFR 807.92(a)(5)] G.

The Quantel Medical COMPACT TOUCH STS is intended to be used for diagnostic imaging and measurement of the eye including:

◆ Visualization of the interior of the eye by B scan

◆ Make measurements inside the eye (Sulcus to Sulcus, Irido Cornea Angle Left, Irido Cornea Angle Right, Lens curvature and Anterior Chamber Depth).

H.

Performance Data [21 CFR 807.92(b)(2)]

Laboratory testing following the guidance "Information for manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound systems and Transducers" (Sept 9, 200) was conducted to verify that the COMPACT TOUCH STS met design specifications and was substantially equivalent to the predicate devices. No Clinical testing is required.

I. Conclusion [21 CFR 807.92(b)(3)]

Technologically, the COMPACT TOUCH STS was found to be substantially equivalent to the currently cleared K060626 Sonomed Inc. VuMax Ophthalmic Ultrasound System. The indications for use are similar to these previously cleared devices. The risks and benefits for the COMPACT TOUCH STS are argued to be comparable to the predicate devices. We believe that there are no new questions of safety or efficacy raised by the introduction of the COMPACT TOUCH STS.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in capital letters.

Food and Drug Administration 10903 New Hampshire Avenue Building 66 - Room 5645 Silver Spring, MD 20993-0002

Quantel Medical S.A. c/o Dr. Mike Johnson QUANTEL USA 601 Haggerty Lane BOZEMAN MT 59715

2010

Re: K102198/S001

Trade/Device Name: COMPACT ACT TOUCH STS Ophthalmic Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Il Product Code: IYO, and ITX Dated: October 1, 2010 Received: October 1 2010

Dear Dr Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the COMPACT TOUCH STS LIN50 ultrasound probe, as described in your premarket notification:

Transducer Model Number .

LIN50 ..

{4}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device ' results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.

Sincerely yours,

David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

Section 4

K102198

NOV - 5 2010

Indications for Use

510(k) Number (if known): not known i wi

Device Name: "COMPACT TOUCH STS" Ophthalmic Ultrasound System

Indications for Use Statement:

The Quantel Medical COMPACT TOUCH STS is intended to be used for diagnostic imaging and measurement of the eye including:

◆ Visualization of the interior of the eye by B scan

Concurrence of CDRH, Office of Device Evaluation

510K

◆ Make measurements inside the eye ( Sulcus to Sulcus, Irido Cornea Angle Left, Irido Cornea Angle Right, Lens curvature and Anterior Chamber Depth).

4-1

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Prescription Use X
(Per 21 CFR 801.109)

OR

Over-The Counter Use

(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

{6}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

Page __ 1 __ of ______________________________________________________________________________________________________________________________________________________________

510(k) Number (if known): NA

Device Name: COMPACT TOUCH STS

Intended Use: The Quantel Medical COMPACT TOUCH STS is intended to be used for diagnostic imaging and measurement of the eye including:

◆ Visualization of the interior of the eye by B scan

� Make measurements inside the eye ( Sulcus, Irido Cornea Angle Left, Irido Cornea Angle Right, Lens curvature and Anterior Chamber Depth).

CLINICALAPPLICATIONABMPWDCWDCOLORDOPPLERPOWER(AMPLITUDE)DOPPLERCOLORVELOCITYIMAGINGCOMBINED(SPECIFY)OTHER(SPECIFY)
OphthalmicP
Fetal
Abdominal
Intra-operative(specify)
Intra-operative
Neurological
Pediatric
Small Organ(Specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Trans-esophageal
Trans-rectal
Trans-vaginal
Trans-urethral
Intra-luminal
PeripheralVascular
Laparoscopic
Musculo-Skeletal
Other (Specify)*

Mode of Operation

Additional Comments:

4-2

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription UseX
(Per 21 CFR 810.109)

{7}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

Page _________________________________________________________________________________________________________________________________________________________________________

510(k) Number (if known): NA

Device Name: COMPACT TOUCH STS LIN50 ultrasonic probe.

Intended Use: The Quantel Medical COMPACT TOUCH STS LIN 50 ultrasonic probe is intended to be used for diagnostic imaging and measurement of the eye including:

◆ Visualization of the interior of the eye by B scan

� Make measurements inside the eye ( Sulcus to Sulcus, Irido Cornea Angle Left, Irido Cornea Angle Right, Lens curvature and Anterior Chamber Depth).

Mode of Operation
CLINICALAPPLICATIONABMPWDCWDCOLORDOPPLERPOWER(AMPLITUDE)DOPPLERCOLORVELOCITYIMAGINGCOMBINED(SPECIFY)OTHER(SPECIFY)
OphthalmicP
Fetal
Abdominal
Intra-operative(specify)
Intra-operativeNeurological
Pediatric
Small Organ(Specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Trans-esophageal
Trans-rectal
Trans-vaginal
Trans-urethral
Intra-luminal
Peripheral
Vascular
Laparoscopic
Musculo-Skeletal
Other (Specify)*

N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:

4-3

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription UseX (Per 21 CFR 810.109)

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.