RAPIDVUE HCG TEST

K102175 · Quidel Corporation · JHI · Aug 25, 2010 · Clinical Chemistry

Device Facts

Record IDK102175
Device NameRAPIDVUE HCG TEST
ApplicantQuidel Corporation
Product CodeJHI · Clinical Chemistry
Decision DateAug 25, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1155
Device ClassClass 2

Intended Use

The RapidVue hCG test is an immunoassay intended for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine for the early detection of pregnancy. The test is intended for use by health care professionals.

Device Story

RapidVue hCG test is a lateral-flow immunoassay for qualitative detection of hCG in urine; used by healthcare professionals in clinical settings. Device consists of a dipstick; user dips device into urine specimen. If hCG concentration is ≥25 mIU/mL, monoclonal antibody-coupled red-colored beads bind to hCG, forming a pink-to-red Test (T) Line on a nitrocellulose membrane; a blue procedural Control (C) Line confirms proper flow. Absence of T-line indicates hCG levels below 25 mIU/mL or absence of hCG. Results are visually interpreted by the clinician within 3 minutes to confirm pregnancy status.

Clinical Evidence

Bench testing only. Studies included assay sensitivity verification, prozone effect evaluation, urinary pH interference testing, procedural flex testing, clinical sample testing, and a Physician Office Laboratory (POL) reproducibility study. Results demonstrated performance consistency with the predicate device.

Technological Characteristics

Lateral-flow immunoassay (dipstick). Components: nitrocellulose membrane with immobilized polyclonal goat anti-alpha hCG antibody (Test Line) and protein for blue latex binding (Control Line). Indicator: monoclonal anti-hCG antibody coupled to red-colored beads. Sensitivity: 25 mIU/mL. Visual readout.

Indications for Use

Indicated for the qualitative detection of hCG in urine for early pregnancy detection in patients. Intended for use by healthcare professionals.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) SUMMARY # 102175 ANG 25 2010 Submitted By: Quidel Corporation 10165 McKellar Court San Diego, California 92121 Telephone: 858-552-1100 x1015 858-646-8045 Fax: Submission Contact: Michelle Bodien August 23, 2010 RapidVue® hCG Test hCG pregnancy test JHI Device Trade Name: Common Name: Date Prepared: Predicate Device: Intended Use: Product Code Device Classification/Name: 21 CFR 862.1155 / Human chorionic gonadotropin (hCG) Test System The RapidVue hCG test is similar to other FDAcleared devices used for the qualitative detection of human chorionic gonadotropin (HCG) for the early detection of pregnancy. QuickVue One-Step hCG Urine Test (K020799) The Food and Drug Administration has classified test systems for the detection of pregnancy as Class II. The RapidVue hCG test is an immunoassay intended for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine for the early detection of pregnancy. The test is intended for use by health care professionals Human chorionic gonadotropin is a hormone Physiologic Basis of the Test: produced by the placenta shortly after implantation. Since hCG is present in the urine of pregnant women, it is an excellent marker for confirming pregnancy. {1}------------------------------------------------ ### Device Description: ﻟﺮﺳ The RapidVue hCG test uses a monoclonal antibody specific to the beta subunit of hCG in a single-step technology to accurately detect hCG. The Dipstick is dipped in urine. If hCG is present in the specimen at a level of 25 mlU/mL or greater, a pink-to-red Test (T) Line will appear along with a blue procedural Control (C) Line in the test result area. If hCG is present at lower levels, or not present in the specimen, only a blue procedural Control Line will appear in the test result area. #### Device Comparison: | Features | RapidVue hCG<br>(Proposed) | QuickVue One-Step hCG Urine<br>(K020799) | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The RapidVue hCG test is an<br>immunoassay intended for the<br>qualitative detection of human<br>Chorionic Gonadotropin (hCG) in<br>urine for the early detection of<br>pregnancy. The test is intended for<br>use by health care professionals | The QuickVue One-Step hCG-Urine test<br>is a one-step immunoassay for the<br>qualitative detection of human<br>Chorionic Gonadotropin (hCG) in<br>urine for the early detection of<br>pregnancy. The test is intended for<br>use by health care professionals. | | Analyte | Human Chorionic Gonadotropin | Human Chorionic Gonadotropin | | Specimen Type | Urine | Urine | | Format | Lateral-flow immunoassay (Dipstick) | Lateral-flow immunoassay (Cassette) | | Total steps | 1 | 1 | | Read Time | 3 minutes | 3 minutes | | Sensitivity | 25 mIU/mL | 25 mIU/mL | | Test Interpretation | Blue procedural control line<br>Pink-to-red test line | Blue procedural control line<br>Pink-to-red test line | | Test strip<br>components | Test Line*<br>Polyclonal goat anti-alpha hCG<br>antibody is immobilized in the test<br>zone on the nitrocellulose membrane<br>Indicator*<br>Monoclonal anti-hCG antibody<br>coupled to red-colored beads is<br>incorporated into the Label Pad. | Test Line<br>Polyclonal goat anti-alpha hCG<br>antibody is immobilized in the test<br>zone on the nitrocellulose membrane<br>Indicator<br>Monoclonal antibody specific to the<br>beta subunit of hCG is incorporated<br>into the Label Pad. | | | Control Line*<br>An unrelated protein capable of<br>binding the blue latex is spotted in the<br>control zone on the nitrocellulose<br>membrane. | Control Line<br>An unrelated protein capable of<br>binding the blue latex is spotted in the<br>control zone on the nitrocellulose<br>membrane. | {2}------------------------------------------------ *Note: The monoclonal antibodies used for the Test Line and the Indicator in the RapidVue hCG test are identical to those used in the predicate QuickVue One-Step hCG Urine test. The components that generate the Control Line are also identical. #### Summary of Performance Data: Numerous analytical studies were undertaken to document the performance characteristics and the substantial equivalence of the RapidVue hCG test to the predicate device. These studies included the following: - 1. Assay Sensitivity: to verify that the sensitivity of the RapidVue hCG test is consistent with QuickVue One-Step hCG Urine test. - 2. Prozone Effect: to verify that no prozone effect is observed with the RapidVue hCG test as is consistent with QuickVue One-Step hCG Urine test. - 3. Urinary pH: to verify that the RapidVue hCG test is not affected by variations in urinary pH. - 4. Procedural Flex Testing: to demonstrate the robustness of the RapidVue hCG test and its test procedure. - 5. Clinical Sample Testing: to verify that the accuracy of the RapidVue hCG test is consistent with QuickVue One-Step hCG Urine test using procured clinical urine samples. - ్. Physician Office Laboratory (POL) Reproducibility Study: to evaluate the performance of the RapidVue hCG test by physician office personnel. #### Conclusion: The results of these studies demonstrate that the RapidVue hCG test is substantially equivalent with the predicate QuickVue One-Step hCG Urine test. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES DEPARTMENT OF HEALTH & HUMAN SERVICES - USA Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993 AUG 2 5 - 2010 Quidel Corporation c/o Ms. Michelle Bodien Senior Regulatory Affairs Specialist 10165 McKellar Court San Diego. CA 92121 Re: k102175 > Trade Name: RapidVue hCG test Regulation Number: 21 CFR §862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system. Regulatory Class: Class II Product Codes: JHI Dated: July 30, 2010 Received: August 2, 2010 Dear Ms. Bodien: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Signature Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 1102175 AN 2017 2 510(k) Number (if known): K102175 Device Name: RapidVue hCG test Indications For Use: The RapidVue hCG test is an immunoassay intended for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine for the early detection of pregnancy. The test is intended for use by health care professionals. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Signature Division Slan- Office of In Vitro Digane Device Evaluation and Safety 510(K) Page 1 of
Innolitics FDA 510(k) Search
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%