ELECSYS CA 125 II CALCHECK 5

K102086 · Roche Diagnostics · JJX · Nov 1, 2010 · Clinical Chemistry

Device Facts

Record IDK102086
Device NameELECSYS CA 125 II CALCHECK 5
ApplicantRoche Diagnostics
Product CodeJJX · Clinical Chemistry
Decision DateNov 1, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

The Elecsys CA 125 II CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys CA 125 II reagent on the indicated Elecsys and cobas e immunoassay analyzers, for in vitro diagnostic use only.

Device Story

Elecsys CA 125 II CalCheck 5 is a lyophilized, multi-level assayed control product; used for calibration verification and assay range verification of Elecsys CA 125 II reagents on Elecsys and cobas e immunoassay analyzers. Product consists of five levels; level 1 uses equine serum matrix; levels 2-5 use human serum matrix; analyte is spiked into matrix during manufacture. Device is intended for professional use in clinical laboratory settings. Healthcare providers reconstitute the lyophilized material with distilled or deionized water; mix by inversion; and process on the analyzer. Output provides verification that the analyzer's calibration and assay range are performing within established specifications, ensuring accuracy of patient CA 125 test results.

Clinical Evidence

No clinical data. Performance evaluation was limited to bench testing, specifically focusing on value assignment and stability testing.

Technological Characteristics

Lyophilized assayed control; five levels. Matrix: Equine serum (Level 1), Human serum (Levels 2-5). Reconstitution: 1.0 mL distilled/deionized water. Stability: 2-8°C unopened; 4 hours at 20-25°C after reconstitution. Analyte: CA 125. Intended for use on Elecsys and cobas e immunoassay analyzers.

Indications for Use

Indicated for use as an assayed control for calibration verification and assay range verification of the Elecsys CA 125 II reagent on Elecsys and cobas e immunoassay analyzers. For in vitro diagnostic use.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102086 ## Introduction According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. Submitter Roche Diagnostics name, address, 9115 Hague Road, P.O. Box 50416 contact Indianapolis, IN 46250-0416 317-521-3501 Contact Person: Kelly French Phone: 317-521-3208 Fax: 317-521-2324 Email: kelly.french@roche.com Secondary Contact: Stephanie Greeman Phone: 317-521-2458 Fax: 317-521-2324 Email: stephanie.greeman@roche.com Date Prepared: June 30, 2010 Device Name Proprietary name: Elecsys CA 125 II CalCheck 5 Common name: CA 125 II CalCheck 5 Classification name: Single (specified) analyte controls (assayed and unassayed) Predicate The Elecsys CA 125 II CalCheck 5 is substantially equivalent to other device products in commercial distribution intended for similar use. We claim equivalency to the currently marketed Elecsys C-Peptide CalCheck 5 (K100810) and the Elecsys CA 125 II CalCheck (K003967). Device The Elecsys CA 125 II CalCheck 5 is a lyophilized product consisting of Description · equine serum in level 1 and human serum matrix for levels 2-5. During manufacture, the analyte is spiked into the matrix at the desired concentration levels. Intended use The Elecsys CA 125 CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys CA 125 II reagent on the indicated Elecsys and cobas e immunoassay analyzers, for in vitro diagnostic use only. ## 510(k) Summary NOV - 1 2010 {1}------------------------------------------------ Comparison Table The table below compares Elecsys CA 125 CalCheck 5 with the predicate devices, Elecsys C-Peptide CalCheck 5 (K100810) and the CA 125 II CalCheck (K003967). > Please note that the use of two predicates has been utilized for this submission based on previous FDA feedback to several of the most recently FDA-cleared CalCheck 5 products. The first predicate shows that the CA 125 II CalCheck 5 is substantially equivalent to another CalCheck 5 product. The Elecsys CA 125 II CalCheck 5 is also substantially equivalent to the second predicate, CA 125 II CalCheck, with several key similarities, especially the analyte. The shaded fields indicate similar characteristics between the candidate device and a predicate device. | Characteristic | Elecsys C-Peptide<br>CalCheck 5 (K100810) | Elecsys CA 125 II CalCheck<br>5<br>(Candidate Device) | Elecsys CA 125 II CalCheck<br>(K003967) | |----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The Elecsys C-Peptide CalCheck<br>5 is an assayed control for use in<br>calibration verification and for<br>use in the verification of the<br>serum and plasma assay range<br>established by the Elecsys<br>C-Peptide reagent on the.<br>indicated Elecsys and cobas e<br>immunoassay analyzers. | The Elecsys CA 125 II Calcheck<br>5 is an assayed control for use in<br>calibration verification and for<br>use in the verification of the<br>assay range established by the<br>Elecsys CA 125 II reagent on the<br>indicated Elecsys and cobas e<br>immunoassay analyzers, for in<br>vitro diagnostic use only. | For use in verification of the<br>calibration established by the<br>Elecsys CA 125 II reagent on the<br>Elecsys 2010 and 1010<br>Immunoassay analyzers. | | Analyte | C-Peptide | CA 125 | CA 125 | | Levels | Five | Five | Three | | Format | Lyophilized | Same | Same | | Handling | Reconstitute Check 1, Check 2,<br>Check 3, Check 4, and Check 5<br>with exactly 1.0 mL distilled or<br>deionized water. Allow to stand<br>closed for 15 minutes, then mix<br>gently by inversion. | Reconstitute Check 1, Check 2,<br>Check 3, Check 4, and Check 5<br>with exactly 1.0 mL distilled or<br>deionized water. Allow to stand<br>closed for 15 minutes, then mix<br>gently by inversion. | Reconstitute Check 1, Check 2,<br>and Check 3 with exactly 1.0 mL<br>distilled or deionized water.<br>Allow to stand closed for 15<br>minutes, then mix gently by<br>inversion. | | Stability | Unopened:<br>• Store at 2-8°C until expiration<br>date<br>Reconstituted:<br>• 20-25°C: 4 hours | Unopened:<br>• Store at 2-8°C until expiration<br>date<br>Reconstituted:<br>• 20-25°C: 4 hours | Unopened:<br>• Store at 2-8°C until expiration<br>date<br>Reconstituted:<br>• 20-25°C: 4 hours | | Matrix | Equine serum matrix | Level 1: Equine serum<br>Levels 2-5: Human serum matrix | Level 1: Equine serum<br>Levels 2 & 3: Human serum<br>matrix | Performance Characteristics The Elecsys CA 125 II CalCheck 5 was evaluated for value assignment and stability. > : .............................................................................................................................................................................. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the eagle. The text is in all caps and is relatively small compared to the eagle symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 Roche Diagnostics c/o Ms. Kelly French Regulatory Affairs Consultant Roche Professional Diagnostics 9115 Hague Road P.O. Box 50416 Indianapolis, IN 46250-0416 NOV 0 1 2010 Re: k102086 Trade/Device Name: Elecsys CA 125II CalCheck 5 Regulation Number: 21CFR 8862.1660 Regulation Name: Quality control material, assayed and unassayed Regulatory Class: Class I (Reserved) Product Code: JJX Dated: October 8, 2010 Received: October 12, 2010 Dear Ms. French: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does net mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 – Ms Kelly French comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. http://www.fda.gov/medicaldevices/resourcesforindustry/default.htm Sincerely yours, Tharai Im Chen Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ k102086 510(k) Number (if known): NOV 0 1 2010 Device Name: Elecsys CA 125 II CalCheck 5 Indication For Use: The Elecsys CA 125 II CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys CA 125 II reagent on the indicated Elecsys and cobas e immunoassay analyzers, for in vitro diagnostic use only. Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use __ i (21 CFR Part 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Bema Philip Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) k102086 Page 1 of 1