K Number
K102083

Validate with FDA (Live)

Device Name
VELSCOPE VX
Manufacturer
Date Cleared
2010-11-18

(115 days)

Product Code
Regulation Number
872.6350
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VELscope Vx is intended to be used by a dentist or health-care provider as an adjunct to traditional oral examination by incandescent light to enhance the visualization of oral mucosal abnormalities that may not be apparent or visible to the naked eye, such as oral cancer and premalignant dysplasia.

VELscope Vx is further intended to be used by a surgeon to help identify diseased tissue around a clinically apparent lesion and thus aid in determining the appropriate margin for surgical excision.

Device Description

The VELscope Vx system is a natural tissue fluorescence direct visualization system to be used as an adjunctive tool for oral mucosal examination.

The main components of VELscope are the Handpiece, incorporating light source, viewing optics and rechargeable battery. Charging Cradle and external power supply. The VELscope Handpiece emits a safe, visible, blue light into the oral cavity, which excites the oral tissue and causes it to fluoresce. The oral cavity can then be examined in real time and suspicious tissue that may require further investigation can be quickly identified. When viewed through the VELscope Handpiece, abnormal tissue typically appears as an irregular, dark area that stands out against the otherwise normal green fluorescence pattern of surrounding health tissue.

AI/ML Overview

The document provided does not contain specific acceptance criteria outlined in a table or a detailed study proving the device meets particular numerical performance targets. Instead, it focuses on demonstrating substantial equivalence to a predicate device (VELscope, K070523).

Here's an analysis of the provided information related to your request:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria:
The core "acceptance criteria" presented in this 510(k) summary are based on demonstrating substantial equivalence to the predicate VELscope device (K070523). The key "criteria" for equivalence are:

  • Identical Indications for Use.
  • Equivalent essential performance specifications, specifically:
    • Excitation band (400-460nm) is the same as the predicate.
    • Optical output power in that excitation band is comparable.
    • Emission (viewing) optics are identical to the predicate.
  • Clinical equivalence demonstrated through comparative imaging of oral mucosal lesions between the new device and the predicate.

Reported Device Performance:

Acceptance Criteria (for Substantial Equivalence to Predicate)Reported Device Performance (VELscope Vx vs. VELscope - K070523)
Identical Indications for UseIdentical Indications for Use.
Equivalent Essential Performance Specifications:
- Excitation band (400-460nm)Same (400-460nm).
- Optical output power in excitation bandComparable.
- Emission (viewing) opticsIdentical.
Clinical Equivalence (through comparative imaging)Clinical photographs of a variety of oral mucosal lesions show substantial equivalence between VELscope Vx and predicate VELscope images.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document states that "Clinical photographs were taken of a variety of oral mucosal lesions." It does not specify the exact number of lesions or patients included in this "variety."
  • Data Provenance: The patients were "referred to oral medicine and oral dysplasia clinics," suggesting a clinical setting. The country of origin is not explicitly stated, but the submitter's address is in Burnaby, BC, Canada. The data appears to be prospective in the sense that images were specifically "acquired" for the comparison, but it's not clear if these were newly diagnosed cases or existing cases.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The study relies on "conventional (white light) as well as fluorescence photographs" and a comparison of these images. There is no mention of independent experts evaluating the images or establishing ground truth.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method. It states "A comparison of the predicate VELscope and VELscope Vx images supports the substantial equivalence," implying a qualitative assessment, but the process of this comparison (e.g., blinded, adjudicated, single reviewer) is not described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a MRMC comparative effectiveness study was not done. The study's purpose was to demonstrate substantial equivalence between the new VELscope Vx and its predicate (VELscope) through comparative imaging, not to evaluate the improvement of human readers with AI assistance. The VELscope Vx is described as an "adjunct to traditional oral examination" and a "natural tissue fluorescence direct visualization system," not an AI-assisted diagnostic tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

This question is not applicable as the VELscope Vx is a direct visualization device, not an algorithm. Its operation inherently involves a human observer (dentist or healthcare provider) viewing the fluorescence to identify abnormalities.

7. The Type of Ground Truth Used

The "ground truth" implicitly used is the visual identification of "oral mucosal lesions" and "diseased tissue" by conventional examination and the use of the predicate VELscope. The clinical context is "oral medicine and oral dysplasia clinics," suggesting that these lesions would have a known clinical or pathological status. However, the document does not explicitly state if pathology or outcomes data was used to confirm the nature of the "variety" of lesions imaged. The substantial equivalence is based on the visual comparability of images, not on a definitive ground truth of diagnosis for each lesion.

8. The Sample Size for the Training Set

This information is not applicable/not provided. The VELscope Vx is a medical device, not an AI/machine learning algorithm that requires a "training set." The engineering and design of the device would have been based on established principles of fluorescence and previous knowledge from the predicate device.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no "training set" in the context of this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains a sequence of digits. The digits are '1102083'. The digits are written in a simple, sans-serif font and are evenly spaced. The image appears to be a close-up of the digits.

510(k) Summary

NOV 1 8 2010

Submitter's Name: David Morgan, PhD

235 - 5589 Byrne Road Address: Burnaby BC V5J 3J1 Canada

(604) 434-4614, ext. 262 Phone:

Date of Submission: July 23 2010

Contact Person: David Morgan, PhD

E-mail: david.morgan@leddental.com

(604) 434-4612 Fax:

Device Name: VELscope Vx

Common Name: Oral Examination Light

Classification Name: Ultraviolet detector

Marketed Device of Equivalence: VELscope (K070523)

Description of Device:

The VELscope Vx system is a natural tissue fluorescence direct visualization system to be used as an adjunctive tool for oral mucosal examination.

The main components of VELscope are the Handpiece, incorporating light source, viewing optics and rechargeable battery. Charging Cradle and external power supply. The VELscope Handpiece emits a safe, visible, blue light into the oral cavity, which excites the oral tissue and causes it to fluoresce. The oral cavity can then be examined in real time and suspicious tissue that may require further investigation can be quickly identified. When viewed through the VELscope Handpiece, abnormal tissue typically appears as an irregular, dark area that stands out against the otherwise normal green fluorescence pattern of surrounding health tissue.

Intended Use:

VELscope Vx is intended to be used by a dentist or health-care provider as an adjunct to traditional oral examination by incandescent light to enhance the visualization of oral mucosal abnormalities that may not be apparent or visible to the naked eye, such as oral cancer and premalignant dysplasia.

VELscope Vx is further intended to be used by a surgeon to help identify diseased tissue around a clinically apparent lesion and thus aid in determining the appropriate margin for surgical excision.

{1}------------------------------------------------

Characteristics of VELscope compared to Predicate Device:

As compared to the predicate device, the VELscope Vx system has identical Indications for Use.

The main technological differences between the VELscope Vx and the predicate VELscope system are:

  • The light source has now been integrated into the Handpiece by replacing the . metal halide lamp originally situated in a separate Light Source Unit with a ring of blue light emitting diodes (LED's) now situated around the distal window of the Handpiece itself.
  • The VELscope Vx Handpiece can now operate in a completely cordless fashion . powered by a rechargeable lithium ion battery.

The essential performance specifications of the device are equivalent to the predicate VELscope device:

  • The excitation band (400-460nm) is the same as the predicate device and the . optical output power in that excitation band is comparable.
  • . The emission (viewing) optics are identical to the predicate device.

Non-Clinical Data

Spectral data comparing the optical intensity distribution of the VELscope Vx excitation light with that of the predicate VELscope are provided to support substantial equivalence.

Clinical Data

Clinical photographs were taken of a variety of oral mucosal lesions from patients referred to oral medicine and oral dysplasia clinics. Conventional (white light) as well as fluorescence photographs using both the predicate VELscope and the VELscope Vx were acquired. No adverse events or complications were reported. A comparison of the predicate VELscope and VELscope Vx images supports the substantial equivalence of the VELscope Vx with the predicate VELscope.

Conclusion

The VELscope Vx system has identical indications for use as the predicate VELscope system and comparative excitation spectral data and clinical fluorescence photographs of oral mucosal lesions support the fact that, despite the technological differences, the VELscope Vx system is substantially equivalent to the VELscope system already cleared under 510(k) - K070523.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized, curved lines that appear to be flowing or waving.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

David Morgan, PhD Chief Science Officer LED Dental Incorporated 235-5589 Byrne Road Burnaby, BC Canada V5J 3J1

NOV 1 8 2010

Re: K102083

Trade/Device Name: VELscope Vx Regulation Number: 21 CFR 872.6350 Regulation Name: Ultraviolet Detector Regulatory Class: II Product Code: NXV Dated: October 18, 2010 Received: October-29, 2010

Dear Dr. Morgan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Dr. Morgan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): 10 2083

VELscope Vx Device Name:

NOV 1 8 2010

Indications For Use:

VELscope Vx is intended to be used by a dentist or health-care provider as an adjunct to traditional oral examination by incandescent light to enhance the visualization of oral mucosal abnormalities that may not be apparent or visible to the naked eye, such as oral cancer and premalignant dysplasia.

VELscope Vx is further intended to be used by a surgeon to help identify diseased tissue around a clinically apparent lesion and thus aid in determining the appropriate margin for surgical excision.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Kure

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of 510(k) Number: __

§ 872.6350 Ultraviolet detector.

(a)
Identification. An ultraviolet detector is a device intended to provide a source of ultraviolet light which is used to identify otherwise invisible material, such as dental plaque, present in or on teeth.(b)
Classification. Class II.