The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
The Varian Eclipse™ Treatment Planning System (Eclipse TPS) provides software tools for planning the treatment of malignant or benign diseases with radiation. Eclipse TPS is a computerbased software device used by trained medical professionals to design and simulate radiation therapy treatments. Eclipse TPS is capable of planning treatments for external beam irradiation with photon, electron, and proton beams, as well as for internal irradiation, (brachytherapy) treatments.
The provided text describes the Eclipse Treatment Planning System (Eclipse TPS) and compares its features to a predicate device. However, it does not contain a study or data proving the device meets specific acceptance criteria in the context of a clinical performance evaluation.
Instead, the document is a Premarket Notification 510(k) Summary for the Eclipse TPS, which focuses on demonstrating substantial equivalence to a legally marketed predicate device. This is a regulatory submission process, not a clinical study.
Here's a breakdown of why the requested information cannot be fully provided from the given text:
- No "Acceptance Criteria" as clinical performance metrics: The table presented in the document is a "Substantial Equivalence Comparison Chart" which outlines features and functionalities of the Eclipse TPS and compares them to its predicate device. These are design and functional specifications, not clinical acceptance criteria with performance targets (e.g., sensitivity, specificity, accuracy, or reader agreement).
- No "Study" in the sense of a clinical trial: The document does not describe any clinical study, experimental design, or results that evaluate the device's performance against clinical endpoints. The "study" mentioned here refers to the 510(k) submission process itself, which relies on demonstrating equivalence rather than proving de novo clinical performance.
Therefore, for aspects related to "acceptance criteria," "study," "sample size," "experts," "ground truth," "MRMC," and "standalone performance," the information is not available in the provided text.
Here's what can be extracted based on the nature of the document:
1. Table of Acceptance Criteria and Reported Device Performance
As explained above, the provided document does not present clinical acceptance criteria with performance metrics. Instead, it provides a "Substantial Equivalence Comparison Chart" detailing functional specifications and comparing the proposed device (Eclipse TPS) to a predicate device (Eclipse Treatment Planning System, K091492).
| Feature/Characteristic | Acceptance Criteria (Predicate) | Reported Device Performance (Eclipse TPS) |
|---|---|---|
| Indication for Use | See above | See above |
| General Usage | ||
| External beam PHOTON planning | yes | yes |
| External beam PHOTON inverse planning | yes | yes |
| External beam ELECTRON planning | yes | yes |
| External beam PROTON planning | yes | yes |
| External beam OCULAR PROTON planning | yes | yes |
| Internal BRACHYTHERAPY planning | yes | yes |
| Stereotactic Frame Localization | yes | yes |
| Supported External Beams & Accessories | ||
| Photon beams | yes | yes |
| Electron beams | yes | yes |
| Proton beams | yes | yes |
| Coplanar fields | yes | yes |
| Non-coplanar fields | yes | yes |
| Multi-leaf Collimators | yes | yes |
| Asymmetric collimators | yes | yes |
| Stereotactic Cone collimators | yes | yes |
| Arc fields | yes | yes |
| Poured Blocks | yes | yes |
| Compensators | yes | yes |
| Physical wedges | yes | yes |
| Dynamic wedges | yes | yes |
| Rotating treatment couch | yes | yes |
| Supported Brachytherapy Sources & Accessories | ||
| Plan for high dose rate afterloader | yes | yes |
| Manual low dose rate brachytherapy:seeds, line sources, wire | yes | yes |
| Applicator library | yes | yes |
| Needle templates | yes | yes |
| Graphical User Interface | ||
| Multiple-instance application | yes | yes |
| Multiple-workspace layout | yes | yes |
| Graphical display/editing of field parameters | yes | yes |
| Beam's-Eye-View display | yes | yes |
| 3D patient image display | yes | yes |
| Model for human Eye | yes | yes |
| SRS Localization application | yes | yes |
| SRS Planning application | yes | yes |
| Biological Optimization application | yes | yes |
| Biological Evaluation application | yes | yes |
| 3D Conformal Optimization application | yes | yes |
| Image Processing | ||
| Orthogonal image displays (3) | yes | yes |
| Oblique image display | yes | yes |
| Edge enhancement filters | yes | yes |
| Image smoothing filters | yes | yes |
| CT/MR/PET Image Registration | yes | yes |
| Image blending utility | yes | yes |
| 4D image display (registration of time series of 3D images) | yes | yes |
| Digitally reconstructed radiographs | yes | yes |
| Enclosed Volume measurement | yes | yes |
| Stereotactic Frame Coordinate transformation | yes | yes |
| Image Segmentation | ||
| Geometrical shapes | yes | yes |
| Manual editing and manipulation tools | yes | yes |
| Automatic /semi-automatic tools | yes | yes |
| Automatic/semi-automatic on-demand and post-processing tools for individual organs/structures | yes | yes |
| Automatic on-demand and pre-processing tools for multiple organs/structures | yes | yes |
| 3D Automargin | yes | yes |
| Logical operators | yes | yes |
| Dose Calculation | ||
| Distributed Calculation Framework | yes | yes |
| Photon calculation | yes | yes |
| Energy Range (Photon) | 1 MV - 50 MV | 1 MV - 50 MV |
| CT-based volumetric calculation | yes | yes |
| Non-CT based IRREG calculation | yes | yes |
| Convolution method | yes | yes |
| Combined electron/photon scatter | yes | yes |
| Directional heterogeneity correction | yes | yes |
| Treatment head modeling | yes | yes |
| Photon Monitor Unit calculation | yes | yes |
| Beam Angle Optimization (GEOS) | yes | yes |
| Leaf Motion Sequencing | yes | yes |
| Dose Dynamic Arc planning | yes | yes |
| Cone Dose Calculation | yes | yes |
| Biological optimization | yes | yes |
| 3D Conformal Optimization | yes | yes |
| AcurosXB dose calculation algorithm | no | yes |
| Electron calculation | yes | yes |
| Energy Range (Electron) | 1 MeV - 50 MeV | 1 MeV - 50 MeV |
| Gaussian Pencil Beam Model | yes | yes |
| Electron Monte Carlo algorithm | yes | yes |
| Electron Monitor Unit calculation | yes | yes |
| Proton calculation | yes | yes |
| Energy Range (Proton) | 50 MeV - 300 MeV | 50 MeV - 300 MeV |
| Brachytherapy calculation | yes | yes |
| AAPM TG 43 compliant | yes | yes |
| Point Dose calculation | yes | yes |
| Optimization to point dose constraints | yes | yes |
| Geometric optimization | yes | yes |
| Acuros dose calculation algorithm | yes | yes |
| Eclipse Algorithm Application Programming Interface (EAAPI) | no | yes |
| Dose Evaluation | ||
| Dose color wash | 2D, 3D | 2D, 3D |
| Isodose levels | 2D, 3D | 2D, 3D |
| Isodose Surface | 3D | 3D |
| Reference point dose summary | yes | yes |
| Dose Volume Histogram plot | yes | yes |
| Plan summing tool | yes | yes |
| Plan comparison tools | yes | yes |
| Evaluation using biological models | yes | yes |
| Plan Output - Hardcopy | ||
| Graphics window screen dump | yes | yes |
| Patient administration data | yes | yes |
| Plan parameters | yes | yes |
| Geometrical displays of plan data | yes | yes |
| Dose distribution | yes | yes |
| DVH plot | yes | yes |
| BEV display | yes | yes |
| Patient orientation | yes | yes |
| User Configurable hardcopy layouts | yes | yes |
| Import/Export Interfaces | ||
| VARIS/Vision database integration | yes | yes |
| DICOM RT / 3.0 | yes | yes |
| Other image formats | yes | yes |
| Electromagnetic Digitizer | import | import |
| Film Scanner | import | import |
| Export field coordinates to Laser System | export | export |
2. Sample size used for the test set and the data provenance:
- Sample size: Not applicable/not provided. This document doesn't describe a clinical test set or study.
- Data provenance: Not applicable/not provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/not provided.
4. Adjudication method for the test set:
- Not applicable/not provided.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was done/reported in this document. The device is a treatment planning system, not an AI-assisted diagnostic device for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable/not provided as a standalone performance study. The document focuses on the functional equivalence of the entire system for treatment planning, which inherently involves human professional users.
7. The type of ground truth used:
- Not applicable/not provided in the context of a performance study. For regulatory purposes, the "ground truth" for substantial equivalence is the predicate device's established performance and specifications.
8. The sample size for the training set:
- Not applicable/not provided. This document is not describing a machine learning or AI algorithm development study that would involve a training set.
9. How the ground truth for the training set was established:
- Not applicable/not provided.
{0}------------------------------------------------
Premarket Notification 510(k) Summary
Submitter's Name:
Varian Medical Systems, Inc. 3100 Hansen Way E-110 Palo Alto, CA 94304 Contact Name: Vy Tran Phone: (650) 424-5731 (650) 424-5040 Fax: Date: July 2010
Proprietary Name:
Eclipse Treatment Planning System
Classification Name:
.
Medical charged-particle radiation therapy system, 21 CFR §892.5050, MUJ, Class II
Eclipse Treatment Planning System, K091492
irradiation, (brachytherapy) treatments.
Common/Usual Name:
Eclipse TPS
Predicate Devices:
Device Description:
Statement of Indications for Use:
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
The Varian Eclipse™ Treatment Planning System (Eclipse TPS) provides software tools for planning the treatment of malignant or benign diseases with radiation. Eclipse TPS is a computerbased software device used by trained medical professionals to design and simulate radiation therapy treatments. Eclipse TPS is capable of planning treatments for external beam irradiation with photon, electron, and proton beams, as well as for internal
Technological Characteristics:
Refer to the attached Substantial Equivalence Comparison Chart.
SEP 3 .
{1}------------------------------------------------
| 1. Indication for Use | See above | See above | |
|---|---|---|---|
| 2. General Usage | |||
| External beam PHOTON planning | yes | yes | |
| External beam PHOTON inverse planning | yes | yes | |
| External beam ELECTRON planning | yes | yes | |
| External beam PROTON planning | yes | yes | |
| External beam OCULAR PROTONplanning | yes | yes | |
| Internal BRACHYTHERAPY planning | yes | yes | |
| Stereotactic Frame Localization | yes | yes | |
| 3. Supported External Beams &Accessories | |||
| Photon beams | yes | yes | |
| Electron beams | yes | yes | |
| Proton beams | yes | yes | |
| Coplanar fields | yes | yes | |
| Non-coplanar fields | yes | yes | |
| Multi-leaf Collimators | yes | yes | |
| Asymmetric collimators | yes | yes | |
| Stereotactic Cone collimators | yes | yes | |
| Arc fields | yes | yes | |
| Poured Blocks | yes | yes | |
| Compensators | yes | yes | |
| Physical wedges | yes | yes | |
| Dynamic wedges | yes | yes | |
| Rotating treatment couch | yes | yes | |
| 4. Supported Brachytherapy Sources &Accessories | |||
| Plan for high dose rate afterloader | yes | yes | |
| Manual low dose rate brachytherapy:seeds, line sources, wire | yes | yes | |
| Applicator library | yes | yes | |
| Needle templates | yes | yes | |
| 5. Graphical User Interface | |||
| Multiple-instance application | yes | yes | |
| Multiple-workspace layout | yes | yes | |
| Graphical display/editing of fieldparameters | yes | yes | |
| Beam's-Eye-View display | yes | yes | |
| 3D patient image display | yes | yes | |
| Model for human Eye | yes | yes | |
| SRS Localization application | yes | yes | |
| SRS Planning application | yes | yes | |
| Biological Optimization application | yes | yes | |
| Biological Evaluation application | yes | yes | |
| 3D Conformal Optimization application | yes | yes | |
| 6. Image Processing | |||
| Orthogonal image displays (3) | yes | yes | |
| Oblique image display | yes | yes | |
| Edge enhancement filters | yes | yes | |
| Image smoothing filters | yes | yes | |
| CT/MR/PET Image Registration | yes | yes | |
| Image blending utility | yes | yes | |
| 4D image display (registration of timeseries of 3D images) | yes | yes | |
| Digitally reconstructed radiographs | yes | yes | |
| Enclosed Volume measurement | yes | yes | |
| Stereotactic Frame Coordinatetransformation | yes | yes | |
| 7. Image Segmentation | |||
| Geometrical shapes | yes | yes | |
| Manual editing and manipulation tools | yes | yes | |
| Automatic /semi-automatic tools | yes | yes | |
| Automatic/semi-automatic on-demand andpost-processing tools for individualorgans/structures | yes | yes | |
| Automatic on-demand and pre-processingtools for multiple organs/structures | yes | yes | |
| 3D Automargin | yes | yes | |
| Logical operators | yes | yes | |
| 8. Dose Calculation | |||
| Distributed Calculation Framework | yes | yes | |
| Photon calculation | yes | yes | |
| Energy Range: | 1 MV - 50 MV | 1 MV - 50 MV | |
| CT-based volumetric calculation | yes | yes | |
| Non-CT based IRREG calculation | yes | yes | |
| Convolution method | yes | yes | |
| Combined electron/photon scatter | yes | yes | |
| Directional heterogeneity correction | yes | yes | |
| Treatment head modeling | yes | yes | |
| Photon Monitor Unit calculation | yes | yes | |
| Beam Angle Optimization (GEOS) | yes | yes | |
| Leaf Motion Sequencing | yes | yes | |
| Dose Dynamic Arc planning | yes | yes | |
| Cone Dose Calculation | yes | yes | |
| Biological optimization | yes | yes | |
| 3D Conformal Optimization | yes | yes | |
| AcurosXB dose calculation algorithm | no | yes | |
| Electron calculation | yes | yes | |
| Energy Range: | 1 MeV - 50 MeV | 1 MeV - 50 MeV | |
| Gaussian Pencil Beam Model | yes | yes | |
| Electron Monte Carlo algorithm | yes | yes | |
| Electron Monitor Unit calculation | yes | yes | |
| Proton calculation | yes | yes | |
| Energy Range: | 50 MeV - 300 MeV | 50 MeV - 300 MeV | |
| Brachytherapy calculation | yes | yes | |
| AAPM TG 43 compliant | yes | yes | |
| Point Dose calculation | yes | yes | |
| Optimization to point dose constraints | yes | yes | |
| Geometric optimization | yes | yes | |
| Acuros dose calculation algorithm | yes | yes | |
| Eclipse Algorithm ApplicationProgramming Interface (EAAPI) | no | yes | |
| 9. Dose evaluation | |||
| Dose color wash | 2D, 3D | 2D, 3D | |
| Isodose levels | 2D, 3D | 2D, 3D | |
| Isodose Surface | 3D | 3D | |
| Reference point dose summary | yes | yes | |
| Dose Volume Histogram plot | yes | yes | |
| Plan summing tool | yes | yes | |
| Plan comparison tools | yes | yes | |
| Evaluation using biological models | yes | yes | |
| 10. Plan Output - Hardcopy | |||
| Graphics window screen dump | yes | yes | |
| Patient administration data | yes | yes | |
| Plan parameters | yes | yes | |
| Geometrical displays of plan data | yes | yes | |
| Dose distribution | yes | yes | |
| DVH plot | yes | yes | |
| BEV display | yes | yes | |
| Patient orientation | yes | yes | |
| User Configurable hardcopy layouts | yes | yes | |
| 11. Import/Export Interfaces | |||
| VARIS/Vision database integration | yes | yes | |
| DICOM RT / 3.0 | yes | yes | |
| Other image formats | yes | yes | |
| Electromagnetic Digitizer | import | import | |
| Film Scanner | import | import | |
| Export field coordinates to Laser System | export | export |
{2}------------------------------------------------
{3}------------------------------------------------
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the text "DEPARTMENT OF HEALTH". To the left of the text is a symbol that appears to be three stylized human figures stacked on top of each other. The text "DEPARTMENT OF HEALTH" is in a bold, sans-serif font. The overall design is simple and professional.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Vy Tran Official Correspondent Varian Medical Systems, Inc. 3100 Hansen Way PALO ALTO CA 94304-1038
2010 SEP 3
Re: K102011
Trade/Device Name: Eclipse Treatment Planning System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ, LHN Dated: July 15, 2010 Received: July 16, 2010
Dear Ms. Tran:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act . The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. .
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 200). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, vermits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796 5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollon the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Donald J. Trump
Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
2010
SEP 3
Indications for Use
510(k) Number (if known):
Eclipse Treatment Planning System Device Name:
Indications for Use:
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
D.W.B
(Division Sign-On)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of 1
ETPS 510(k) Indications
510K
Page 1 of 1
Date: June 30, 2010
N/A