(51 days)
The Decompression' of Choice (DOC) and AAAR 2K10 is a non invasive method of applying distractive forces to the spine through controlled tensions. It is designed to apply decompression (decompression is unloading due to distraction and positioning and/or as non-surgical in nature) forces to intervertebral discs. The Decompression' of Choice (DOC) and AAAR 2K10 may be used for back pain, neck pain, hemiated discs, protruding discs, degenerative disc disease, sciatica and posterior facet syndromes.
The Devices are a non surgical traction or also called decompression' systems. It accurately controls tensions designed to relax Para spinal musculature and to allow distractive- traction forces to decompress inter vertebral spinal disk space. The Devices do this by creating a negative pressure in the disc due to the distractive forces and treatment protocols. Bulges or herniation are reduced back into the center or annular part of the disc. Fibroblasts then form a matrix over the protryded area to form a "seal" over the previously injured area. The load cell constantly updates the user interface provided which controls a gear-motor/linear actuator to immediately and safely apply forces as determined by qualified healthcare personnel. The load cell feedback is also used to further verify and adjust traction forces allowing for variations in patient postures and outside influencing forces. This method of operation allows safe, continuous and smooth operation of the unit and proper comfortable and safe appliance of said forces to the patient. An extensive manual of suggested treatment protocols and safety warnings is furnished to the end user.
The provided text describes a medical device called "Decompression of Choice (DOC) and AAAR 2K10" (referred to as "Devices") which is a power traction equipment for spinal decompression. The study described focuses on proving equivalence to predicate devices rather than establishing specific performance metrics against pre-defined acceptance criteria for a novel functionality.
Here's a breakdown of the information requested, based only on the provided text:
Acceptance Criteria and Device Performance
The document does not explicitly list quantitative "acceptance criteria" for the device itself in terms of clinical efficacy or diagnostic accuracy (as would be common for AI/imaging devices). Instead, the performance testing described is focused on demonstrating the device's ability to deliver and measure specific forces accurately, which is a functional requirement for this type of equipment. The primary "acceptance" is based on substantial equivalence to predicate devices and compliance with safety standards.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Accuracy of Applied Force | Demonstrated accurate force delivery and measurement on cervical and lumbar sections using a calibrated scale. |
| Safety Standards Compliance | In compliance with IEC 60601-1, IEC 60601-2-46, IEC 60601-1-4, UL 60601-1, and IEC 60601-1-2. |
| Equivalence of Traction Pulls | Performs similar static, intermittent, and cyclic ramp up, treatment, and ramp down traction pulls as predicate K053223. |
| Maximum Traction Force | Lumbar section maximum traction: 200 lbs (matches predicate K053223). Cervical section maximum traction: 50 lbs (for patient safety). |
| Biocompatibility | Materials in contact with patient (table top, cervical capture) are commonly used and have no known adverse effects (biocompatibility testing not warranted). |
Study Details
The provided text describes a functional and safety performance test, not a clinical study to assess diagnostic or therapeutic efficacy in patients.
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable in the traditional sense of patient data. The "test set" involved the device itself being tested with a calibrated scale. No specific number of "tests" or "measurements" is provided, but it implies a series of measurements to confirm force accuracy across different settings.
- Data Provenance: The testing was likely conducted in a controlled laboratory environment by the manufacturer or a testing facility during device development/validation. Country of origin not specified, but the applicant is from the USA (Watertown, SD). This is prospective testing of the device's physical function.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The "ground truth" for the force measurements would be the readings from the calibrated scale, which is an objective instrument. No human experts were involved in establishing this ground truth.
-
Adjudication method for the test set: Not applicable. The test involved physical measurements against a calibrated instrument.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is therapeutic power traction equipment, not an AI-assisted diagnostic or imaging device. No human reader studies were conducted or described.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This device is mechanical/electronic equipment, not an algorithm. Its operation inherently involves a human operator (qualified healthcare personnel) who sets the parameters.
-
The type of ground truth used: For the force measurement performance, the ground truth was objective measurements from a calibrated scale. For safety, compliance with listed safety standards served as the "ground truth." For substantial equivalence, the characteristics and performance of the predicate devices served as the comparative ground truth.
-
The sample size for the training set: Not applicable. This device does not use machine learning or AI, so there is no "training set."
-
How the ground truth for the training set was established: Not applicable. There is no training set for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Pivotal Health Solutions. The logo consists of a stylized human figure on the left, followed by the word "PIVOTAL" in large, bold letters. Below "PIVOTAL" are the words "Health Solutions" in a smaller font. The logo is black and white.
Section 5: 510(k) Summary
-
- Submitted By:
Phillip Klein Pivotal Health Solutions 724 Oakwood Road Watertown SD, 57201 Phone: 605-753-0110 Fax: 605-882-8398
- Submitted By:
AUG 27 2010
Date Prepared: November 23, 2009
-
- Trade Name: Decompression' of Choice (DOC) and AAAR 2K10 (referred to collectively as the "Devices" in this document) Common Name: Power Traction Equipment Classification Name: TTH, Power Traction Equipment, 21 CFR, Section 890-5900
3. Predicate Device Information:
The Devices are substantially equivalent to the Saunders Group 3D Active Trac. K001712 and Chattanooga Group Triton/Tru-Trac/Tx/Triton Dts Traction K053223.
- Device Description: বঁ
The Devices are a non surgical traction or also called decompression' systems. It accurately controls tensions designed to relax Para spinal musculature and to allow distractive- traction forces to decompress inter vertebral spinal disk space. The Devices do this by creating a negative pressure in the disc due to the distractive forces and treatment protocols. Bulges or herniation are reduced back into the center or annular part of the disc. Fibroblasts then form a matrix over the protryded area to form a "seal" over the previously injured area.
The load cell constantly updates the user interface provided which controls a gear-motor/linear actuator to immediately and safely apply forces as determined by qualified healthcare personnel. The load cell feedback is also used to further verify and adjust traction forces allowing for variations in patient postures and outside influencing forces. This method of operation allows safe, continuous and smooth operation of the unit and proper comfortable and safe appliance of said forces to the patient.
An extensive manual of suggested treatment protocols and safety warnings is furnished to the end user.
- ഗ് Intended Use Statement:
The Decompression' of Choice (DOC) and AAAR 2K10 is a non invasive method of applying distractive forces to the spine through controlled tensions.
It is designed to apply decompression' forces to intervertebral discs.
The Decompression' of Choice (DOC) and AAAR 2K10 may be used for back pain, neck pain, hemiated discs, protruding discs, degenerative disc disease, sciatica and posterior facet syndromes.
1 Decompression is unloading due to distraction and positioning and/or as non-surgical in nature.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Pivotal Health Solutions. The logo consists of a stylized human figure on the left, followed by the word "PIVOTAL" in large, bold letters. Below "PIVOTAL" are the words "Health Solutions" in a smaller font.
6. Technological Characteristics Comparison Summary
The Devices and predicate device K001712 are similar in shape and can be divided into similar smaller sections. They have a base, elevation, cervical, thoracic, pelvic and a controls section. The base supports the lifting mechanism and also provides a stable surface during table operation. The elevation section lowers the table surface so it is easier to get on and off of the table. The elevation also elevates so the operator is working in an ergonomically correct position. The cervical section applies distractive forces and is capable of P-A flexion for targeting. The thoracic section is stationary and provides a mounting place for the cervical and pelvic section applies distractive forces and is capable of P-A flexion and axial rotation for targeting. The controls sections move and monitor the table during treatment.
The Devices and predicate device K001712 use similar material and construction methods. They are constructed from structural steel that has been welded and bolted together. All of the steel has been coated with a protective coating to prevent rust. The cushions are constructed from plywood, foam and vinyl. The Plywood is the base of the cushion and mounts directly to the steel. The foam is secured to the plywood and provides a comfortable surface for the patient to lie on. Vinyl covers the foam and is secured to the plywood.
The Devices and predicate devices K001712 use different sources of power for the distractive forces. The Devices use a power supply and linear actuator to generate the force for distraction. The predicate device uses an air compressor and pneumatic cylinder. The force delivered to the patient is measured using a load cell, similar to a scale, and a pic interface for the devices. The predicate device uses an air pressure regulator and calculation to create the required force.
The Devices and predicate device K053223 perform similar traction pulls. Both the Devices and predicate K053223 have static, intermittent and cyclic ramp up, treatment and ramp down. They have similar hold period and rest period settings as well as ramp up and ramp down steps. The maximum traction of the predicate K053223 is 200 lbs. and the lumbar section of the devices is 200 lbs. The cervical section of the devices is limited to 50 lbs for safety of the patient.
- Performance and Compliance Data
To prove the Devices are effective and as safe as the predicate performance testing was done on the cervical and lumbar sections. A calibrated scale was placed between the cervical and lumbar sections to simulate a patient. The scale was used to measure the force being applied to the patient. Using manual distraction the table was ramped up to target poundage and the readings where noted.
The Devices are in compliance with the following safety standards.
-
- IEC 60601-1. Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1. 1991-11. Amendment 2, 1995
-
- IEC 60601-2-46, Medical Electrical Equipment; Part 2-46: Particular requirements for safety of Operating Tables, First Edition: 1998
-
- IEC 60601-1-4:2000 Consol. Ed. 1.1, Medical electrical equipment Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems, edition 1.1
-
- UL 60601-1, Medical Electrical Equipment; Part 1: General Requirements for Safety, 1ª Edition: 2003 with revisions dated April 26, 2006.
-
- IEC 60601-1-2, (Second Edition, 2001), Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Pivotal Health Solutions. The logo consists of a stylized human figure on the left, followed by the word "PIVOTAL" in large, bold letters. Below "PIVOTAL" are the words "Health Solutions" in a smaller font. The logo is black and white.
8. Biocompatibility Information
Materials that come in direct contact with the patient are the table top and cervical capture. The materials of those devices are used on marketed device with no known adverse effects. Because the materials are commonly used on other products biocompatibility testing is not warranted.
..
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002
Pivotal Health Solutions % Intertek Testing Services NA, Inc. Mr. William J. Sammons 2307 E. Aurora Road Unit B7 Twinsburg, OH 44087
AUG 27 2010
Re: K101889
Trade/Device Name: Decompression of Choice and AAAR 2K10 Regulation Number: 21 CFR 890.5900 Regulation Name: Power traction equipment Regulatory Class: II Product Code: ITH Dated: August 11, 2010 Received: August 13, 2010
Dear Mr. Sammons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 - Mr. William J. Sammons
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Mink M. Mllanson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Pivotal Health Solutions. The logo consists of a stylized human figure on the left, with the word "PIVOTAL" in large, bold letters to the right. Below "PIVOTAL" are the words "Health Solutions" in a smaller font. The logo is black and white.
K101 889
Indications for Use
K101889 510(k) Number:
The Decompression' of Choice (DOC) and AAAR 2K10 Device Name:
Indications For Use:
The Decompression' of Choice (DOC) and AAAR 2K10 is a non invasive method of applying distractive forces to the spine through controlled tensions.
It is designed to apply decompression is unloading due to distraction and positioning and/or as non-surgical in nature) forces to intervertebral discs.
The Decompression' of Choice (DOC) and AAAR 2K10 may be used for back pain, neck pain, hemiated discs, protruding discs, degenerative disc disease, sciatica and posterior facet syndromes.
1 Decompression is unloading due to distraction and positioning and/or as non-surgical in nature.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101899
N/A