K Number
K101860

Validate with FDA (Live)

Date Cleared
2010-09-01

(61 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NovaBone Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. NovaBone Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. the extremities, posterolateral spine, and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

NovaBone Putty is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.

Device Description

NovaBone Putty is an osteoconductive, bioactive, bone void filler device. It is composed of a calcium-phosphorus-sodium-silicate (Bioglass) particulate mixed with a synthetic binder that acts as a temporary binding agent for the particulate. The particulate and binder are provided premixed as a pliable cohesive material. The mixed device is supplied sterile, packaged in a disposable plastic syringe with one or more attachable cannula allowing for placement of the putty at a predetermined distance from the syringe. On implantation, the binder is absorbed to permit tissue infiltration between the Bioglass particles. The particles then are slowly absorbed and replaced by new bone tissue during the healing process.

AI/ML Overview

This document is a 510(k) summary for a medical device modification ("NovaBone Putty - Bioactive Synthetic Bone Graft"). It does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the typical sense of a clinical trial or performance study comparing a device's effectiveness against predefined metrics.

This submission is specifically for a modification to an already cleared device. The core of the document states that the modification is solely in terms of the device packaging (to permit attachment of an extension cannula) and that "the device itself is unchanged." Therefore, the submission's focus is on demonstrating that this packaging change does not alter the fundamental technological characteristics, safety, or effectiveness of the device.

Here's an analysis based on the provided text, highlighting why a traditional "acceptance criteria" and "study proving device meets acceptance criteria" are not present for this specific submission:

1. A table of acceptance criteria and the reported device performance

Not applicable. This submission is for a packaging modification. There are no performance metrics or clinical acceptance criteria being evaluated or reported for this specific change. The core device's performance was established in prior 510(k) clearances (K060728, K080009, K082672).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. No "test set" in the context of clinical performance or diagnostic accuracy is described. The submission focuses on the implications of a packaging change.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. There is no "ground truth" to be established by experts as this is not a diagnostic or efficacy study.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No test set for adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a bone void filler, not an AI diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

Not applicable. No ground truth in the context of clinical or diagnostic performance is applicable to this documentation.

8. The sample size for the training set

Not applicable. No training set is relevant to this type of device modification.

9. How the ground truth for the training set was established

Not applicable.


Summary of Device Modification and How it "Meets" Acceptance (within the context of a 510(k) for modification):

The document explicitly states:

  • "The NovaBone Putty device of this submission has been modified from that of the predicate solely in terms of the device packaging; the device itself is unchanged." (Page 1, Section 6)
  • "The device action as a synthetic, inorganic, biocompatible and osteoconductive scaffold into which new bone will grow is unchanged." (Page 1, Section 6)
  • "This device modification does not result in a change in technological characteristics of the NovaBone Putty device." (Page 1, Section 7)
  • "NovaBone Putty continues to be safe and effective as a non-structural osteoconductive bone void filler for osseous defects." (Page 1, Section 7)

The "study" that proves the device meets the acceptance criteria (in this specific case) is the argument for Substantial Equivalence to the predicate device, K060728, K080009, and K082672.

The acceptance criteria are implicitly that the modified device (packaging change only) retains the same technological characteristics, intended use, safety, and effectiveness as the previously cleared predicate devices. The "proof" is the detailed comparison provided in Section 6, explicitly stating the device itself is unchanged and only the packaging for delivery has been modified to permit the attachment of an extension cannula. This demonstrates that the change does not introduce new questions of safety or effectiveness.

The FDA's letter (Page 2) confirms this by stating: "We have reviewed your Section 510(k) premarket notification... and have determined the device is substantially equivalent...". This determination by the FDA serves as the "proof" that the device modification meets the necessary regulatory acceptance criteria for substantial equivalence.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for "NOVABONE SYNTHETIC BONE GRAFT PRODUCTS". The word "NOVABONE" is written in a bold, sans-serif font, with the letters "NOVA" in a distressed, textured style and the letters "BONE" in a solid, black style. Below the word "NOVABONE" is the phrase "SYNTHETIC BONE GRAFT PRODUCTS" in a smaller, sans-serif font.

K101860

Special 510(k): Device Modification Summary NovaBone Putty - Bioactive Synthetic Bone Graft

06/30/10

1- 2010

Submitter Information: 1.

Name:NovaBone Products, LLCSEF
Address:13631 Progress Boulevard, #600Alachua, FL 32615
Telephone:(386) 462-7661
Facsimile:(386) 418-1636
Contact:David M. Gaisser

Name of Device: 2.

Trade Name:NovaBone Putty – Bioactive Synthetic Bone Graft
Common Name:Osteoconductive Bone Void FillerSynthetic Resorbable Bone Graft Material
Regulation Number:21 CFR 888.3045
Regulation Name:Bone Void Filler

Legally Marketed Predicate Device: 3.

NovaBone Putty - Bioactive Synthetic Graft Predicate #1: [K060728, K080009, K082672]

4. Device Description

NovaBone Putty is an osteoconductive, bioactive, bone void filler device. It is composed of a calcium-phosphorus-sodium-silicate (Bioglass) particulate mixed with a synthetic binder that acts as a temporary binding agent for the particulate. The particulate and binder are provided premixed as a pliable cohesive material. The mixed device is supplied sterile, packaged in a disposable plastic syringe with one or more attachable cannula allowing for placement of the putty at a predetermined distance from the syringe. On implantation, the binder is absorbed to permit tissue infiltration between the Bioglass particles. The particles then are slowly absorbed and replaced by new bone tissue during the healing process.

Intended Use ડ.

NovaBone Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. NovaBone Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. the extremities, posterolateral spine, and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

{1}------------------------------------------------

NovaBone Putty is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.

6. Technological Characteristics and Substantial Equivalence

The technological characteristics of the NovaBone Putty device are identical to the NovaBone Putty device cleared per K060728, K080009, and K082672. The device and predicate are designed as osteoconductive space-filling devices to be gently packed into defect sites and used as non-structural scaffolds for the body's natural healing and bone regeneration process. The device indications are the same as for the predicate. The device is intended to be used alone, or in combination with autogenous bone.

The primary component of NovaBone Putty is bioactive glass (45S5 Bioglass) particulate. This synthetic material is both biocompatible and osteoconductive. The NovaBone Putty includes a synthetic binder as an inert carrier for ease of handling and delivery, forming a premixed cohesive material. The binder is biocompatible and is absorbed after implantation, opening space between the bioactive glass particles for cell infiltration and bone formation. The bioactive glass particulate remains for a longer post-implantation period, acting as a scaffold for bone ingrowth. This particulate is absorbed and replaced by new bone tissue. Animal testing has demonstrated that the majority of the material is absorbed within six months of implantation, with >98% of the material being absorbed by 12 months. The timeframe for full absorption in humans has not been determined, but is expected to be at least 12 months.

The NovaBone Putty device of this submission has been modified from that of the predicate solely in terms of the device packaging; the device itself is unchanged. The previous syringe packaging has been modified to permit attachment of an extension cannula (included) to better access more remote graft sites. The device action as a synthetic, inorganic, biocompatible and osteoconductive scaffold into which new bone will grow is unchanged.

7. Conclusion

The NovaBone Putty device modification subject to this submission is to include a new syringe package format. The syringe applicator has been modified from that of prior devices to permit interfacing with a large-bore cannula. This device modification does not result in a change in technological characteristics of the NovaBone Putty device. NovaBone Putty continues to be safe and effective as a non-structural osteoconductive bone void filler for osseous defects.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

NovaBone Products, LLC % Mr. David M. Gaisser 13631 Progress Boulevard, #600 Alachua, Florida 32615

SEP 1 2010

Re: K101860 ·

Trade/Device Name: NovaBone Putty - Bioactive Synthetic Bone Graft Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MOV Dated: July 29, 2010 Received: August 2, 2010

Dear Mr. Gaisser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. David M. Gaisser

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Sincerely, yours,

Barbara Buchner

Mark N. Melker Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K101860

1 2010 SEP STATEMENT OF INDICATIONS FOR USE

K101860 510(k) Number (if known):

Device Name: __________________ NovaBone Putty – Bioactive Synthetic Bone Graft

Indications For Use:

NovaBone Putty - Bioactive Synthetic Bone Graft is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. NovaBone Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. the extremities, posterolateral spine, and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

NovaBone Putty is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.

Prescription Use XX

OR (Per 21 CFR 801.109) Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign Off

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K101860