K Number
K101847

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-10-21

(112 days)

Product Code
Regulation Number
876.5310
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Urgent PC Neuromodulation System is intended to treat patients with Overactive Bladder (OAB) and associated symptoms of urinary urgency, urinary frequency, and urge incontinence.

Device Description

The Urgent® PC Neuromodulation System is a minimally invasive neuromodulation system designed to deliver retrograde access to the sacral nerve through percutaneous electrical stimulation of the tibial nerve. The method of treatment is referred to as Percutaneous Tibial Nerve Stimulation (PTNS). The Urgent PC Neuromodulation System is a combination of the Urgent PC Stimulator and the Urgent PC Stimulation Lead Set. The Urgent PC Stimulator is a battery-operated external pulse generator and is designed, constructed, and manufactured for multiple use, only in conjunction with the Urgent PC Stimulation Lead Set. The Urgent PC Stimulation Lead Set transfers the electrical current from the Urgent PC Stimulator to the tibial nerve via the Needle Electrode. The entire Stimulation Lead Set is intended for single use only and is not to be reused.

AI/ML Overview

The provided text does not contain a study that proves the device meets specific acceptance criteria in the manner requested. The document is a 510(k) premarket notification for the Urgent® PC Neuromodulation System. It highlights that the "new device" is substantially equivalent to a previously cleared predicate device (K071822). The primary difference noted is a change in the wording of the indications statement to incorporate "Overactive Bladder (OAB)" terminology, which is described as an "update in wording rather than a change in the device patient population."

Therefore, the submission relies on the established performance and safety of the predicate device, rather than presenting new clinical study data with specific acceptance criteria and detailed performance metrics for the updated device itself.

Based on the provided information, I cannot complete the requested tables and details as the document does not include a new study with explicit acceptance criteria and corresponding performance data for the device.

Here's an explanation of why the requested information cannot be extracted:

  • No explicit acceptance criteria or measured performance: The document states that the "Urgent PC Neuromodulation System and Stimulation Lead Set allows for the successful performance of the product's intended use," but it does not provide any specific quantitative acceptance criteria (e.g., sensitivity, specificity, accuracy, or clinical efficacy rates) or reported performance values against those criteria.
  • No new study details: The submission is a 510(k) for substantial equivalence, not a new clinical trial reporting primary endpoints and statistical analysis. It does not describe a study involving test sets, ground truth establishment, expert adjudication, or MRMC studies.

In summary, the provided document is a regulatory submission asserting substantial equivalence to a predicate device, and thus does not contain the detailed study information (acceptance criteria, sample sizes, ground truth methodology, etc.) that would typically be found in a clinical study report.

{0}------------------------------------------------

Uroplasty, Inc. Premarket Notification [510(k)] Submission Urgent® PC Neuromodulation System

Page 7 of 36

K101847

Section 5: 510(k) Summary

Date PreparedOctober 8, 2010
New Device NameUrgent® PC Neuromodulation SystemOCT 2 1 2010
Predicate DeviceUrgent® PC Neuromodulation System (K071822)
ContactUroplasty, Inc.5420 Feltl RoadMinnetonka, MN 55343 USATel: 952.426.6140; Fax: 952.426.6199

Intended Use

i

The Urgent PC Neuromodulation System is intended to treat patients with Overactive Bladder (OAB) and associated symptoms of urinary urgency, urinary frequency, and urge incontinence.

Device Description

The Urgent® PC Neuromodulation System is a minimally invasive neuromodulation system designed to deliver retrograde access to the sacral nerve through percutaneous electrical stimulation of the tibial nerve. The method of treatment is referred to as Percutaneous Tibial Nerve Stimulation (PTNS).

The Urgent PC Neuromodulation System is a combination of the Urgent PC Stimulator and the Urgent PC Stimulation Lead Set. The Urgent PC Stimulator is a battery-operated external pulse generator and is designed, constructed, and manufactured for multiple use, only in conjunction with the Urgent PC Stimulation Lead Set. The Urgent PC Stimulation Lead Set transfers the electrical current from the Urgent PC Stimulator to the tibial nerve via the Needle Electrode. The entire Stimulation Lead Set is intended for single use only and is not to be reused.

Indications Statement

The primary difference between the new and predicate devices is the wording of indications statement. The new device indications statement incorporates the overactive bladder terminology listed in the intended use. This difference in indication statement represents an update in wording rather than a change in the device patient population.

Technological Characteristics

The new and predicate devices are technologically the same; they are percutaneous tibial nerve stimulator devices with lead sets intended to deliver retrograde access to the sacral nerve for the overactive bladder symptoms of urinary urgency, urinary frequency, and urge incontinence. Both devices have the same intended use and principles of action. Only the wording of the indications statement differs between this device and the predicate.

{1}------------------------------------------------

Uroplasty, Inc.

Premarket Notification [510(k)] Submission Urgent® PC Neuromodulation System

Performance

The Urgent PC Neuromodulation System and Stimulation Lead Set allows for the successful performance of the product's intended use.

Conclusion

The subject device of the 510(k) submission is substantially equivalent to the previously cleared Urgent PC Neuromodulation System by Uroplasty, Inc. (K071822).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo features a stylized representation of a human figure, composed of three curved lines, positioned to the right of the department's name. The text "DEPARTMENT OF HEALTH & HUMAN" is arranged vertically along the left side of the logo, with "SERVICES" completing the circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G60 Silver Spring, MD 20993-0002

Mr. Michael Morrell, RAC Director of Regulatory Affairs and Quality Assurance Uroplasty. Inc., 5420 Feltl Road MINNETONKA MN 55343

OCT 2 1 2010

Re: K101847

Trade/Device Name: Urgent® PC Neuromodulation System Regulation Number: 21 CFR 8876,5310 Regulation Name: Non-implanted, peripheral electrical continence device Regulatory Class: II Product Code: NAM Dated: October 8, 2010 Received: October 14, 2010

Dear Mr. Morrell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate . commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{3}------------------------------------------------

Page 2

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Hubert Lemmer MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Uroplasty, Inc.

Premarket Notification [510(k)] Submission Urgent® PC Neuromodulation System

Page 6 of 36

K101847

Section 4: Indications for Use Statement

510(k) Number (if known): K101847

OCT 2 1 2010

Device Name:

Urgent® PC Neuromodulation System

The Urgent PC Neuromodulation System is intended to treat patients Indications for Use: with Overactive Bladder (OAB) and associated symptoms of urinary urgency, urinary frequency, and urge incontinence.

Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

HReinu

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K101847

CONFIDENTIAL

§ 876.5310 Nonimplanted, peripheral electrical continence device.

(a)
Identification. A nonimplanted, peripheral electrical continence device is a device that consists of an electrode that is connected by an electrical cable to a battery-powered pulse source. The electrode is placed onto or inserted into the body at a peripheral location and used to stimulate the nerves associated with pelvic floor function to maintain urinary continence. When necessary, the electrode may be removed by the user.(b)
Classification. Class II, subject to the following special controls:(1) That sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter.
(2) That the labeling must bear all information required for the safe and effective use of the device as outlined in § 801.109(c) of this chapter, including a detailed summary of the clinical information upon which the instructions are based.