(15 days)
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
Here's an analysis of the provided information regarding the device's acceptance criteria and the study demonstrating its compliance:
1. Table of Acceptance Criteria and Reported Device Performance:
| Characteristics | Standard (Acceptance Criteria) | Device Performance (Reported) |
|---|---|---|
| Dimension | ASTM standard D 5250-06 | Meets |
| Physical Properties | ASTM standard D 5250-06 | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Amount | ASTM standard D 5250-06 | <10mg/dm2 |
| Biocompatibility | Primary Skin Irritation in rabbits | Passes (Not a Primary Skin Irritation) |
| Dermal sensitization in the guinea pig | Passes (Not a Dermal sensitization) |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample sizes used for testing the various characteristics. However, for the biocompatibility tests:
- Primary Skin Irritation: "rabbits" (plural, but exact number not specified).
- Dermal Sensitization: "guinea pig" (singular or plural, but exact number not specified).
The data provenance is not explicitly mentioned (e.g., country of origin, retrospective or prospective).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
This information is not applicable to the type of device and testing described. The "ground truth" for glove performance refers to meeting established engineering and biological standards, not expert interpretation of medical images or conditions.
4. Adjudication Method for the Test Set:
Not applicable. The testing involves objective measurements against established standards, not interpretation that would require adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No MRMC study was done, nor is it applicable to this type of medical device. This device is an examination glove, not an AI-powered diagnostic tool.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
Not applicable. This device is a physical product (gloves), not an algorithm or AI system.
7. Type of Ground Truth Used:
The ground truth for the device's performance is based on established industry standards and regulatory requirements:
- ASTM standard D 5250-06 (for dimension, physical properties, powder amount)
- 21 CFR 800.20 (for freedom from pinholes)
- ISO 10993-10 (for biocompatibility, referred to generally in the nonclinical discussion)
8. Sample Size for the Training Set:
Not applicable. There is no mention or indication of a "training set" as this is not an AI/ML device. The testing described is for a manufactured product against specified performance standards.
9. How the Ground Truth for the Training Set was Established:
Not applicable, as there is no training set for this device.
{0}------------------------------------------------
Summary
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
JUL 1 6 2010
"The assigned 510(k) number is: |</0/84 | .. " (applicant leave blank)
Premarket Notification [510(k)] Summary
[{a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :
| Submitter's name : | SHIJIAZHUANG SAIL PLASTIC CO., LTD |
|---|---|
| Submitter's address : | RM 201, UNIT 7, #1 BUILDING, #136 SHIFAN ST.SHIJIAZHUANG, 050000, CHINA |
| Phone number : | (86)- 311-83986806 |
| Fax number : | (86)- 311-83986358 |
| Name of contact person: | Mr. Frank Zhang |
| Date the summary was prepared: | June 28th, 2010 |
[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
| Device Name: | Powdered Vinyl Patient Examination Gloves,Clear (Non-colored) |
|---|---|
| Proprietary/Trade name: | Powdered Vinyl Patient Examination GlovesOther clients private labeling |
| Common Name: | Patient examination glove |
| Classification Name: | Patient examination glove |
| Device Classification: | I |
| Regulation Number: | 21 CFR 880.6250 |
| Panel: | General Hospital (80) |
| Product Code: | LYZ |
[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .
Class I* powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
Predicate device : FUGUAN (Brand) Powdered Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032907.
[(a)(4)] A description of the device
{1}------------------------------------------------
K101841$\rho$2$\beta$2
Device Description : powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
{(a)(5)] The summary describes the intended use of the device
Device Intended Use: powdered vinyl patient examination glove, Clear(non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
The powdered vinyl patient examination gloves, Clear (non-colored) non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.
| Characteristics | Standard | Device performance |
|---|---|---|
| Dimension | ASTM standard D 5250-06 | Meets |
| Physical Properties | ASTM standard D 5250-06 | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Amount | ASTM standard D 5250-06 | <10mg/dm2 |
| Biocompatability | Primary Skin Irritation inrabbits | Passes |
| Not a Primary Skin Irritation | ||
| Dermal sensitization in theguinea pig | Passes | |
| Not a Dermal sensitization |
[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Powdered vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-06, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10.
[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
[(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).
It can be concluded that the Powdered Vinyl Patient Examination Gloves, Clear(non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL, meet labeling claims .
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. The logo is accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the emblem.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 1 3 2010
Shijiazhuang Sail Plastic Company, Limited C/O Mr. Chu Xiaoan Jianxiang Yuan #209 Bei Si Huan Zhong Road Building 1, RM 1606 Beijing China 100083
Re: K101841
Trade/Device Name: Powdered Vinyl Patient Examination Gloves, Clear (Non-colored) Regulation Number: 21CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: June 28, 2010 Received: July 1, 2010
Dear Mr. Xiaoan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
Applicant: SHIJIAZHUANG SAIL PLASTIC CO., LTD
510(k) Number (if known):_* K101841
Device Name:_Powdered Vinyl Patient Examination Gloves, Clear (Non-colored)
Indications For Use:
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---|---|
| Division of Anesthesiology, General Hospital | |
| Infection Control, Dental Devices | |
| 510(k) Number: | K101841 |
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.