K Number
K101827

Validate with FDA (Live)

Device Name
OSTEOSCAF
Date Cleared
2011-04-20

(293 days)

Product Code
Regulation Number
872.3930
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

OsteoScaf™ bone grafting material is indicated for filling and/or augmenting intraoral/maxillofacial osseous defects, such as intrabony periodontal osseous defects, furcation defects, augmentation of bony defects of the alveolar ridge, filling of tooth extraction sites, and sinus elevation grafting.

Device Description

OsteoScaf™ is a synthetic osteoconductive and bioresorbable bone grafting material composed of 3-phase biodegradable polymer and calcium phosphate composite, intended to fill and/or augment intraoral/maxillofacial osseous defects. It functions in the same manner as the predicate devices. The product is available as a particulate, cylinder, or block and is provided in sterile packaging in various dosage volumes.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the OsteoScaf™ device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document describes the device's characteristics and performance relative to predicate devices, rather than explicit numerical acceptance criteria. The criteria are framed around demonstrating equivalence or superiority to predicate devices in various properties.

Acceptance Criteria CategoryReported Device Performance (OsteoScaf™)
Chemical Composition3-phase composite of 22% (wt) PLGA, 66% Calcium phosphates, and 12% hydroxyapatite. (Implicitly, this composition is acceptable and comparable to predicates).
ImpurityTesting demonstrates compliance with ASTM 1185 requirements. (Acceptance: Meets ASTM 1185).
Pore StructureHighly interconnected macroporous structure of larger than 80% porosity and 0.25~1.2 mm pore size similar to human cancellous bone. (Acceptance: Similar to human cancellous bone and predicates).
Mechanical StrengthProvides adequate handling properties and strength to ensure dimensional integrity when handled by the surgeon and delivered to bony application sites. (Acceptance: Adequate handling and dimensional integrity, comparable to predicates).
DegradationAll three phases of the composite can be resorbed over time through hydrolytic and/or cell-mediated degradation. (Acceptance: Resorbable over time, comparable to predicates).
OsteoconductivityIn vivo studies demonstrate that when OsteoScaf™ is placed in direct contact with viable bone, new bone forms in apposition to its surfaces and within its interstices. (Acceptance: Osteoconductive and comparable to predicates).
Bioresorption & Bone Formation (Animal Studies)1) More new bone formed with OsteoScaf™ compared to coagulum control in a rat calvaria defect model after 4 weeks. 2) Gave rise to more new bone formation and faster resorption of the material than the predicate in a rat model. 3) Biomechanical testing in a rat model demonstrates that OsteoScaf™ facilitates replacement with native bone. (Acceptance: Comparable to or better than predicate devices in promoting new bone formation and appropriate resorption rate).
Material EquivalenceThe device is substantially equivalent to predicate devices in terms of material composition, geometry, and technological characteristics. (Overall Acceptance: Demonstrates substantial equivalence to predicates in key characteristics).
Safety and EffectivenessThe material is at least as safe and effective as predicate devices. (Overall Acceptance: At least as safe and effective as predicates).

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not explicitly stated as a number of individual animals. The studies are described as using "a rat calvaria model" and "a rat model." This suggests multiple rats were involved, but the exact count is not provided.
  • Data Provenance: Retrospective, as the study has already been conducted and the results are being reported in the 510(k) summary. The country of origin is not specified, but given the applicant's address in San Antonio, Texas, USA, it is likely the studies were conducted domestically or by contractors overseen by the US applicant.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. This device is a bone grafting material, and its performance is evaluated through biological and mechanical studies (animal models, physical/chemical tests), not through expert interpretation of images or clinical assessments requiring adjudicated ground truth. The "ground truth" for the animal studies would be objective measurements of new bone formation, material resorption, and biomechanical strength.

4. Adjudication Method for the Test Set:

Not applicable for this type of device and study. Adjudication methods like "2+1" or "3+1" are typically used in clinical imaging studies where expert consensus is needed to establish a diagnostic ground truth. The animal studies involve objective measurements, not subjective expert interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size:

No. An MRMC study is not relevant for this type of medical device (bone graft material). MRMC studies evaluate the performance of diagnostic devices or AI algorithms when interpreted by multiple human readers, often with and without AI assistance. This device is a therapeutic/reconstructive material, not a diagnostic tool.

6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a material for grafting, not a software algorithm.

7. The Type of Ground Truth Used:

For the animal studies, the ground truth was based on:

  • Histological and Morphological Analysis: Measurement of new bone formation and material resorption (e.g., "more new bone formed," "faster resorption").
  • Biomechanical Testing: Objective measurement of the mechanical properties of the reconstructed bone ("facilitates replacement with native bone").

For the bench tests, the ground truth was based on:

  • Analytical Chemistry and Material Science Techniques: Measuring chemical composition, impurity levels, pore structure, and mechanical strength against established standards (e.g., "compliance with ASTM 1185 requirements").

8. The Sample Size for the Training Set:

Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device. The studies described are performance validation tests.

9. How the Ground Truth for the Training Set was Established:

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K101827

K081432

Section 5

APR 2 0 2011

510(k) Summary

Applicant Nameand AddressTexas Innovative Medical Devices (DBA)Skeletal Regeneration, Ltd.14785 Omicron DrBuilding 100, Suite 101San Antonio, Texas 78245
Contact Person:Jerry ChangSenior Director of Quality Assurance/Regulatory Affairs(210) 696-7616
Date of Summary:January 18, 2011 (edits made) Original Submission madeJune 30, 2010
Name of Device:OsteoScaf™
Trade/Proprietary Name:OsteoScaf™
Common Name:Bone Grafting Materials, Synthetic
Establishment Name:OsteoScaf™
Establishment Number:21 CFR 872.3930
Proposed Classification:Class II
Product Code:LYC
Panel:Dental
Predicate Devices:
ReOSS™Intra-Lock International, IncK082419
Bio-OSS™Geistlich Biomaterials, IncK033815
Norian PDC™Norian CorpK983104
Vitoss™Orthovita, IncK083033

Depuy Spine, Inc

HEALOS™

{1}------------------------------------------------

Intended Use:

OsteosScaf™ bone grafting material is indicated for filling and/or augmenting intraoral/maxillofacial osseous defects, such as intrabony periodontal osseous defects, furcation defects, augmentation of bony defects of the alveolar ridge, filling of tooth extraction sites, and sinus elevation grafting.

Device Description:

OsteoScaf™ is a synthetic osteoconductive and bioresorbable bone grafting material composed of 3-phase biodegradable polymer and calcium phosphate composite, intended to fill and/or augment intraoral/maxillofacial osseous defects. It functions in the same manner as the predicate devices. The product is available as a particulate, cylinder, or block and is provided in sterile packaging in various dosage volumes.

Summary of Technological Characteristics:

The OsteoScaf™ bone grafting material is comprised of a 3-phase composite scaffold comprising poly (lactide-co-glycolide) (PLGA) and 2 osteoclast-resorbable calcium phosphate (CaP) phases that provide the same physical characteristics as the listed predicate devices. The latter are particulates synthesized at physiological temperature and blended with the PLGA phase, and a thin layer of hydroxyapatite surface coating generated in a simulated body fluid at physiological temperature. The Device is configured as a particulate, cvlinder or in a block shape and is packaged in a tray and a chevron type peel pouch for easy use. The product is irradiated to ensure sterility. The Device has a trabecular structure that is the same as the predicate devices and mimics the multidirectional interconnected porosity of human cancellous bone. The Device resorbs and is replaced with natural bone, which remodels during the healing process in the same fashion as that of the predicate devices.

Summary of Performance Testing:

Bench and animal studies were conducted to ensure that OsteoScaf™ bone grafting material met the predetermined design specifications. OsteoScaf™ has been tested for a number of chemical and physical characteristics: chemical composition, phase composition, trace of impurities, density and porosity, particle size, morphology, compressive strength and elastic modulus. All these characteristics are equivalent to those of the predicate device. OsteoScaf™ is osteoconductive and bioresorbable based on in vivo studies. When OsteoScaf™ is placed in direct contact with viable bone, new bone forms in apposition to the OsteoScaf™ surfaces and within the interstices of the implant. As the material resorbs, bone tissue grows into the space previously occupied by the implant. The testing performed included the following:

{2}------------------------------------------------

Test ItemsSummarized Results
Chemical composition3-phase composite of 22% (wt) PLGA, 66% Calcium phosphatesand 12% hydroxyapatite.
ImpurityTesting demonstrates compliance with ASTM 1185 requirements.
Pore structureHighly interconnected macroporous structure of larger than 80%porosity and 0.25~1.2 mm pore size similar to human cancellousbone.
Mechanical strengthThe devices provides adequate handling properties and strength toensure dimensional integrity when handled by the surgeon anddelivered to bony application sites.
DegradationAll three phases of the composite can be resorbed over time throughhydrolytic and/or cell-mediated degradation.
Animal studies1) OsteoScaf™ was used to reconstruct the critical-sized defect in arat calveria model. The result demonstrated that more new boneformed with OsteoScaf™, compared to coagulum control, after 4weeks.2) OsteoScaf™ gave rise to more new bone formation and fasterresorption of the material than the predicate in a rat model.3) The biomechanical testing in rat model demonstrates thatOsteoScaf™bone grafting material facilitates replacement withnative bone.

Conclusion:

The Company Device (OsteoScaf™) is substantially equivalent to the predicate devices for filling and/or augmenting intraoral/maxillofacial osseous defects, such as intrabony periodontal osseous defects, furcation defects, augmentation of bony defects of the alveolar ridge, filling of tooth extraction sites, and sinus elevation grafting. The Device is also substantially equivalent to the predicate devices in terms of material composition, geometry, and technological characteristics. The material is at least as safe and effective as predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Jerry Chang Senior Director of Quality Assurance and Regulatory Affairs Texas Innovative Medical Devices (DBA) Skeletal 4808 Research Drive San Antonio, Texas 78240

Re: K101827

Trade/Device Name: OsteoScafTM Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: March 31, 2011 Received: April 1, 2011

APR 2 0 2011

Dear Mr. Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act'(Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Chang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section 4 Indications for Use

510(k) Number (if known): K101827

Device Name: OsteoScaf™

Indications for Use:

OsteoScaf™ bone grafting material is indicated for filling and/or augmenting intraoral/maxillofacial osseous defects, such as intrabony periodontal osseous defects, furcation defects, augmentation of bony defects of the alveolar ridge, filling of tooth extraction sites, and sinus elevation grafting.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use __ (21 CFR 801 Subpart C)

· (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K. Alberons for Dr. Susan Runner
Division Sign-Off

ivision Sign-K

Office of Device Evaluation (ODE)

510(k) K101827

Page 1 of 1

§ 872.3930 Bone grafting material.

(a)
Identification. Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.(b)
Classification. (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.