K Number
K101792

Validate with FDA (Live)

Date Cleared
2011-03-18

(263 days)

Product Code
Regulation Number
868.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LIV is an inteqrated portable oxygen delivery system intended to provide supplemental oxygen to pediatrics and adults.. The device is MR-conditional (per ASTM standard 2503-05), and intended for use during MR imaging for MR1 systems up to 3.0T. Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.

Device Description

The Linde Integrated Valve® ("LIV") is a portable oxygen delivery system, consisting of a fully integrated cylinder, valve, requlator, flow meter, and shock-absorbing quard. A range of userselectable flow settings is available, inclúding low flows that may be clinically appropriate for certain classes of patients. An additional DISS-1240 connection provides standard 50 psig oxygen delivery, while an optional bed hanger allows the LV to be readily attached to a bed. The LV is suitable for use in all healthcare settings including, but not limited to, hospital, outpatient, imaging center, ambulatory, and home healthcare.

AI/ML Overview

The provided document, K101792, describes a premarket notification for the "Linde Integrated Valve - LIV," a portable oxygen delivery system. However, the document does not contain a study that proves the device meets specific acceptance criteria in the manner requested.

Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (Medcyl-E-lite K033897) based on technological characteristics and compliance with a series of recognized international and national standards for medical devices and gas cylinders. This type of submission (510(k)) primarily demonstrates that a new device is as safe and effective as a legally marketed predicate device, rather than providing detailed clinical study results with acceptance criteria.

Therefore, many of the requested elements for a study proving acceptance criteria cannot be extracted directly from this document. I will answer based on the information provided, explicitly stating when the requested information is not available.


Description of Acceptance Criteria and Proving Study

The Linde Integrated Valve (LIV) is a portable oxygen delivery system. The acceptance criteria for this device are primarily based on compliance with established safety and performance standards for similar medical devices and gas handling equipment. The "study" that proves the device meets these criteria is a series of non-clinical performance tests confirming adherence to these standards, as detailed in the "Non-Clinical Performance Summary" section of the 510(k) submission.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard / Parameter)Reported Device Performance (Compliance)
ISO 10524-3:2005 (Pressure regulators for use with medical gases -- Part 3: Pressure regulators integrated with cylinder valves)The new LIV valves have been revised to comply with this standard.
ISO 15996:2005 (Gas cylinders -- Residual pressure valves -- General requirements and type testing)The device conforms to this standard.
BS EN 1789:2007 (Medical vehicles and their equipment. Road ambulances)The device conforms to this standard.
ISO 9227:2006 (Corrosion tests in artificial atmospheres -- Salt spray tests)The device conforms to this standard.
DOT 49 CFR 178.46 (Specification 3AL seamless aluminum cylinder)The device conforms to this standard for aluminum cylinders.
CE marking; Council Directive 93/42/EEC (concerning medical devices)The device conforms to this directive.
ISO 11117:2008 (Gas cylinders-valve protection caps and valve guards for industrial and medical gas cylinder- Design, construction and tests)The device conforms to this standard.
MR Compatibility (ASTM standard 2503-05)The device is MR-conditional, tested up to 3.0T. This matches the predicate device.
Technological characteristicsFlow selector and flow outlet: Yes Fixed pressure outlet: Yes Cylinder On/Off: Yes (with optional open close indication) Contents Gauge: Active Excess Flow Device: No Cylinder Sizes: D, E Cylinder materials construction: Aluminum Guard colour: Green Guard hand grip: 2 grips

2. Sample size used for the test set and the data provenance

  • Sample size for the test set: Not explicitly stated for each individual test or standard listed. The document indicates "safety and performance testing" in general. It is typical for compliance testing to involve a representative sample of devices, but the specific numbers are not provided in this summary.
  • Data provenance: Non-clinical (laboratory/engineering) testing and design specifications. The country of origin for the data is not specified, but the standards are international (ISO, EN) and national (DOT, CFR, CE).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of experts: Not applicable/not specified. The "ground truth" here is compliance with published, recognized standards and engineering specifications for device performance. These standards are developed by expert committees, but the individual tests themselves typically involve engineering or technical personnel rather than clinical experts establishing a "ground truth" in the diagnostic sense.
  • Qualifications of experts: Not specified.

4. Adjudication method for the test set

  • Adjudication method: Not applicable/not specified. Compliance testing generally involves objective measurements against predefined criteria in the standards, rather than expert adjudication of subjective findings.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC study: No. This document describes a medical device (portable oxygen delivery system), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study related to human reader improvement with AI assistance is not relevant and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone performance: Not applicable. This is not an algorithm-based device. Performance testing for this device would assess its mechanical and functional specifications.

7. The type of ground truth used

  • Type of ground truth: Predetermined performance parameters and established safety requirements outlined in recognized medical device and gas cylinder standards (e.g., ISO, DOT, EN, CE Mark Directive) and engineering specifications. For MR compatibility, adherence to ASTM standard 2503-05 serves as the ground truth.

8. The sample size for the training set

  • Sample size for training set: Not applicable. This device is not an AI/machine learning product requiring a training set.

9. How the ground truth for the training set was established

  • Ground truth for training set: Not applicable.

{0}------------------------------------------------

K101792

Image /page/0/Picture/1 description: The image shows the logo for The Linde Group. The logo is white text on a black background. The words "THE LINDE GROUP" are in all caps and are located at the top left of the image. Below that is a white swoosh and the word "Linde" in cursive.

Section 6 - 510(k) Summary

MAR 1 8 2011

Contact Persons for this submission:

MICHAEL C. PIACENZA

Medical and FDA Compliance Manager Linde North America, Inc. 575 Mountain Avenue, Murray Hill, NJ 07974, USA Office: 908-508-2729 Fax: 908-771-1203 Mobile: 908-720-0200

Date prepared: June 25, 2010

Trade Names

LIV (Linde Integrated Valve)

    1. LIV Portable Oxygen System - with a 2000 psi service pressure rating
    1. LIV Portable Oxygen System - with a 3000 psi service pressure rating

Device Classification

Common NameClassificationNumberClassRegulation Number
Portable Oxygen Delivery Systemunclassified

Predicate Device Identification

Legally marketed devices to which equivalence is being claimed510(k) #
LIVK063354
Medicyl-E-liteK033897

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for The Linde Group. The words "THE LINDE GROUP" are in white text in the upper left corner of the image. The word "Linde" is in a stylized white font in the lower right corner of the image.

Device Description:

The Linde Integrated Valve® ("LIV") is a portable oxygen delivery system, consisting of a fully integrated cylinder, valve, requlator, flow meter, and shock-absorbing quard. A range of userselectable flow settings is available, inclúding low flows that may be clinically appropriate for certain classes of patients. An additional DISS-1240 connection provides standard 50 psig oxygen delivery, while an optional bed hanger allows the LV to be readily attached to a bed. The LV is suitable for use in all healthcare settings including, but not limited to, hospital, outpatient, imaging center, ambulatory, and home healthcare.

Intended Use of the Device:

The UV is an integrated portable oxygen delivery system intended to provide supplemental oxygen to pediatrics and adults.. The device is MR-conditional (per ASTM standard 2503-05), and intended for use during MR imaging for MRI systems up to 3.0T. Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling

Comparison to Predicate Devices:

A summary comparison of technological characteristics, including design and materials is provided in the table below.

ParameterMedicyl-E-lite (K033897)LIV(Linde Integrated Valve)
Flow selector and flow outletyesyes
Fixed pressure outletnoyes
Cylinder On/Offyesyes (with optional open close indication)
Contents Gaugenon-activeActive
Excess Flow DeviceNoNo
MR Compatibilityyes; tested up to 3.0Tyes; tested up to 3.0T
Cylinder SizesD,ED,E
Cylinder materials constructionAluminiumAluminium
Guard colourgreengreen
Guard hand grip2 grips2 grips

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for The Linde Group. The words "THE LINDE GROUP" are in white, block letters in the upper left corner of the image. The word "Linde" is written in a stylized, cursive font in the lower right corner of the image.

Based on the Medicyl-E-lite - K033897, the new LIV valves have been revised to comply with ISO 10524-3 (valid standard for Medical Valve Integrated Pressure Requiator - ratified April 2006). The modified devices also have the same intended use and indications for use as the aforementioned predicate device. Linde believes that the technological characteristics of the LIV (Linde Integrated Valve) is substantially similar to those of the predicate device.

Non-Clinical Performance Summary:

In terms of operating and safety specification the device conforms to applicable standards including

    1. ISO 10524-3:2005 Pressure requlators for use with medical gases -- Part 3: Pressure requlators integrated with cylinder valves
    1. 150 15996:2005 Gas cylinders -- Residual pressure valves -- General requirements and type testing
    1. BS EN 1789:2007 Medical vehicles and their equipment. Road ambulances
    1. ISO 9227:2006 Corrosion tests in artificial atmospheres -- Salt spray tests
    1. DOT 49 CFR 178.46. Specification SAL seamless aluminum cylinder (Aluminium cylinders)
    1. CE marking; Council Directive 93/42/EEC of 14.June 1993, concerning medical devices.
    1. ISO 11117:2008 Gas cylinders-valve protection caps and valve quards for industrial and medical gas cylinder- Design, construction and tests

Conclusion:

Based upon the safety and performance testing and compliance with voluntary standards, the manufacturer believes that the LIV (Linde Integrated Valve) is substantially equivalent to the predicate device, and does not raise any new questions of safety or effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, which is a traditional symbol of medicine and healthcare. The symbol is composed of three wavy lines that converge at the bottom and extend upwards.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Michael Piacenza Medical and FDA Compliance Manager Linde Gas North America LLC 575 Mountain Avenue Murray Hill, New Jersey 07974

MAR 1 8 201

Re: K101792

Trade/Device Name: Linde Integrated Valve - LIV Regulatory Class: II Product Code: ECX Dated: February 4, 2011 Received: February 7, 2011

Dear Mr. Piacenza:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Piacenza

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htmr.for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection control and Dental Devices

Office of Device Evaluation

Center for Devices and

Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo for The Linde Group. The words "THE LINDE GROUP" are in a sans-serif font at the top left of the image. Below that is a white streak that appears to be a stylized representation of a gas or vapor. The word "Linde" is written in a cursive font at the bottom right of the image.

Section 5 - Indications for Use Statement

510(k) Number: K101792

Device Name: Linde Integrated Valve - LIV

Indications for Use:

The LIV is an inteqrated portable oxygen delivery system intended to provide supplemental oxygen to pediatrics and adults.. The device is MR-conditional (per ASTM standard 2503-05), and intended for use during MR imaging for MR1 systems up to 3.0T. Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.

Prescription Use

AND / OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

510(k) Number:

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anest Stantal Devices Division of Ansol.
Infection Control, Dental Devices

W/01-

Page 13 of 47

§ 868.2700 Pressure regulator.

(a)
Identification. A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.