K Number
K101692

Validate with FDA (Live)

Device Name
PATIENT MONITOR
Date Cleared
2011-06-28

(377 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CMS8000 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG, Heart Rate (HR), Respiration Rate (RESP), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient. The Pulse Oxygen Saturation (SpO2) and Pulse Rate (PR) are intended use for adult and pediatric patient. The monitor is to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for helicopter transport or hospital ambulance.

Device Description

The proposed device, CMS8000 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters. It has the physical and technical alarming function with audio and visual alarming, The device can driven by AC or DC power supply.

AI/ML Overview

The provided 510(k) summary for the Contec Medical Systems CMS8000 Patient Monitor discusses performance testing but does not explicitly provide a table of acceptance criteria and reported device performance in the format requested. It also does not include information on sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, or training set details as typically expected for advanced AI/ML device submissions.

However, based on the provided text, here's what can be extracted and inferred:

1. Table of Acceptance Criteria and Reported Device Performance

The document mentions that performance testing was conducted to "validate and verify that the proposed device, the Patient Monitors met all design specifications." It specifically states clinical tests were performed for SpO2 and NIBP according to ISO 9919 and AAMI SP10 standards. These standards themselves contain acceptance criteria relevant to the performance of pulse oximeters and automated sphygmomanometers. While the document doesn't report the specific acceptance criteria or the numerical results to populate the table directly, it implicitly states that the device met these standards.

Inferred Table Structure (Based on common practice for similar devices and standards):

Performance MetricAcceptance Criteria (Implied by Standards)Reported Device Performance
SpO2 Accuracy (for different ranges)Defined by ISO 9919 (e.g., RMS difference between device and reference within a certain limit across various SpO2 levels)Met ISO 9919:2005 standard
NIBP Accuracy (systolic, diastolic, mean)Defined by AAMI SP10 (e.g., mean difference and standard deviation between device and reference measurements within certain limits)Met AAMI SP10:2002 standard
ECG PerformanceDefined by AAMI / ANSI EC13:2002/(R) 2007 (e.g., accuracy of heart rate detection, alarm performance)Met AAMI / ANSI EC13:2002/(R) 2007 standard
Respiration Rate (RESP) AccuracyNot explicitly stated but would be part of general design specificationsMet all design specifications (implied)
Invasive Blood Pressure (IBP) AccuracyDefined by IEC 60601-2-34:2000Met IEC 60601-2-34:2000 standard
CO2 Measurement AccuracyDefined by ISO 21647:2004Met ISO 21647:2004 standard
Temperature (TEMP) AccuracyNot explicitly stated but would be part of general design specificationsMet all design specifications (implied)

2. Sample size used for the test set and the data provenance

The document states: "Performance testing including clinical and bench testing was conducted to validate and verify that the proposed device, the Patient Monitors met all design specifications."

  • Sample Size for Test Set: Not specified in the provided text. The standards (ISO 9919, AAMI SP10) would dictate minimum sample sizes for their respective clinical validation components, but these numbers are not present here.
  • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. It only generically mentions "clinical test" for SpO2 and NIBP functions.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the given text. While clinical tests were conducted, details about who established the ground truth (e.g., reference measurements by qualified clinicians) are absent.

4. Adjudication method for the test set

This information is not provided in the given text.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study is not mentioned in the provided text. The device is a patient monitor, not an AI-assisted diagnostic imaging tool where MRMC studies are typically relevant. This submission predates the widespread regulatory focus on AI/ML in medical devices.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This concept is not applicable in the context of this traditional patient monitor device. The device inherently provides parameters for human interpretation and action. Its performance is measured directly against reference standards, not as an "algorithm only" component separate from human interaction in the way AI/ML algorithms often are.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Based on the mentioned standards for validation:

  • For SpO2 (ISO 9919): Ground truth is typically established using a reference oximeter in controlled desaturation studies or by comparing to arterial blood gas measurements (co-oximetry values).
  • For NIBP (AAMI SP10): Ground truth is typically established by simultaneous auscultatory measurements performed by trained observers using a mercury sphygmomanometer or another validated reference method.
  • For other physiological parameters (ECG, IBP, CO2, RESP, TEMP): Ground truth would be established by comparison to reference measurement devices or validated clinical methods.

The document does not explicitly state these methods but references to the standards imply adherence to their ground truth methodologies.

8. The sample size for the training set

This information is not applicable/provided. The CMS8000 Patient Monitor, as described (submitted in 2011), is a traditional physiological monitor, not an AI/ML device that requires a "training set" in the modern sense. Its algorithms are based on established physiological principles and signal processing, not statistical learning from large datasets.

9. How the ground truth for the training set was established

This information is not applicable/provided for the same reason as point 8.

{0}------------------------------------------------

JUN 28 2011

Premarket Notification
Section 510(k) Submission
Section III 510(k) Submission
Section III 510(k) Summary

Section III 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Date of SubmissionJun 8th, 2010
SponsorEstablishment Registration No.: 3006979678
Contec Medical Systems Co., Ltd, No. 24, West Huanghe Road,Qinhuangdao, Hebei, 066000, China
Contact Person: XueYong Li, Quality Manager
CorrespondentMs. Diana Hong / Mr. Tarzan. Wang
Shanghai Mid-Link Business Consulting Co., Ltd
Suite 5D, No.19, Lane 999, Zhongshan Road (S-2), Shanghai, 200030, China
T: +86-21-64684973;
F: 240-238-7587
Tarzan.wang@mid-link.net
Proposed DevicePatient Monitors, CMS8000
Common NamePatient Monitor
ClassificationMonitor, Physiological, Patient , MHX , 21 CFR 870.1025
PanelCardiovascular
Subsequent CodeDSI, MLD, DRT, DXN, DSK, DQA, BZQ, CCK, FLL
Device DescriptionThe proposed device, CMS8000 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters. It has the physical and technical alarming function with audio and visual alarming, The device can driven by AC or DC power supply.
TechnologicalCharacteristicsThe proposed device, CMS8000 Patient Monitor has no any new technological characteristics, all function and technological characteristic are same or similar as the predicate device.
Intended UseThe CMS8000 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG, Heart Rate (HR), Respiration Rate (RESP), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), carbon dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient. The Pulse Oxygen Saturation (SpO2) and Pulse Rate (PR) are intended use for adult and pediatric

III-I

.

{1}------------------------------------------------

10 1692 7/2

.

Premarket Notification
Section 510(k) Submission
Section III 510(k) Submission
III 510(k) Summary

patient.
The monitor is to be used in healthcare facilities by clinical physicians or appropriate medical staffunder the direction of physicians. It is not intended for helicopter transport or hospital ambulance.
TestingPerformance testing including clinical and bench testing was conducted to validate and verify thatthe proposed device, the Patient Monitors met all design specifications.
Clinical DiscussionThe proposed device, General Patient Monitors, has conducted clinical test into performanceverification of SpO2 and NIBP measure function according with the standard of ISO 9919 andAAMI SP10.
Non Clinical DiscussionThe proposed device, General Patient Monitors, has conducted the test according with thefollowing standard to demonstrate the effectiveness and safety performance of device.
IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988;
Amendment 1, 1991-11, Amendment 2, 1995.
IEC 60601-1-2:2007 Medical Electrical Equipment - Part 1-2: General Requirements for Safety -Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
AAMI / ANSI EC13:2002/(R) 2007, Cardiac monitors, heart rate meters, and alarms.
ISO 9919:2005: Medical electrical equipment - Particular requirements for the basic safety andessential performance of pulse oximeter equipment for medical use.
AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
IEC 60601-2-34:2000, Medical electrical equipment - Part 2-34: Particular requirements for thesafety, including essential performance, of invasive blood pressure monitoring equipment.
ISO 21647:2004, Medical electrical equipment - Particular requirements for the basic safety andessential performance of respiratory gas monitors.
SE ConclusionThe proposed device, General Patient Monitors, is Substantially Equivalent (SE) to the PredicateDevice, PM-7000 Patient Monitor, K072346

. . . .

.

·

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Contect Medical System Co., Ltd. c/o Diana Hong Suite 8D , Zhongxin Zhongshan Mansion, NO. 19, Lane 999, Zhong Shan, Shanghai China 200030

JUN 2 8 2011

Re: K101692

Trade/Device Name: CMS8000 Patient Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection and alarms) Regulatory Class: Class II (two) Product Code: MHX, DSI, MLD, DRT, DXN, DSK, DQA, BZQ, CCK, FLL Dated: June 14, 2011 Received: June 15, 2011

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Hong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. 601 Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Premarket Notification Section 510(k) Submission Section II Indication for Use Statement

Section II Indication for Use Statement

510(k) Number: Device Name: Patient Monitor Model: CMS8000

Indications for Use:

The CMS8000 Patient Monitor is intended to be used for monitoring, reviewing, storing and alaming of multiple physiological parameters including ECG, Heart Rate (HR), Respiration Rate (RESP), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), cation dioxide (CO2), and Temperature (TEMP) of adult, pediatric and neonatal patient.

The Pulse Oxygen Saturation (SpO2) and Pulse Rate (PR) are intended use for adult and pediatric patient. The monitor is to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for helicopter transport or hospital ambulance.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEBDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
(Division Sign-Off)Division of Cardiovascular Devices
510(k) Numberk10106

II-I


§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.