K Number
K101658

Validate with FDA (Live)

Date Cleared
2010-08-11

(61 days)

Product Code
Regulation Number
878.4810
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

General Indications:

  • Class I, II, III, IV and V cavity preparation
  • Caries removal
  • Hard tissue surface roughening or etching
  • Enameloplasty, excavation of pits and fissures for placement of sealants
    • For use on adult and pediatric patients

Root Canal Hard Tissue Indications:

  • Tooth preparation to obtain access to root canal
  • Root canal preparation including enlargement
  • Root canal debridement and cleaning

Root Canal Disinfection:

  • Laser root canal disinfection after endodontic treatment

Endodontic Surgery (Root Amputation) Indications:

  • Flap preparation incision of soft tissue to prepare a flap and expose the bone.
  • Cutting bone to prepare a window access to the apex (apices) of the root(s).
  • Apicoectomy - amputation of the root end.
  • Root end preparation for retrofill amalgam or composite.
  • Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex
    NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.

Bone Surgical Indications:

  • Cutting, shaving, contouring and resection of oral osseous tissues (bone)
  • Osteotomy

Soft Tissue Indications including Pulpal Tissues*:
Incision, excision, vaporization, ablation and coagulation of oral soft tissues, including:

  • Excisional and incisional biopsies
  • Exposure of unerupted teeth
  • Fibroma removal
  • Flap preparation incision of soft tissue to prepare a flap and expose the bone.
  • Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions)
  • Frenectomy and frenotomy
  • Gingival troughing for crown impressions
  • Gingivectomy
  • Gingivoplasty
  • Gingival incision and excision
  • Hemostasis
  • Implant recovery
  • Incision and drainage of abscesses
  • Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery
  • Leukoplakia
  • Operculectomy
  • Oral papillectomies
  • Operculectomy
  • Oral papillectomies
  • Pulpotomy
  • Pulp extirpation
  • Pulpotomy as an adjunct to root canal therapy
  • Root canal debridement and cleaning
  • Reduction of gingival hypertrophy
  • Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex
    NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
  • Soft tissue crown lengthening
  • Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
  • Vestibuloplasty
  • For use on adult and pediatric patient

Laser Periodontal Procedures:

  • Full thickness flap
  • Partial thickness flap
  • Split thickness flap
  • Laser soft tissue curettage
  • Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
  • Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining junctional epithelium
  • Removal of granulation tissue from bony defects
  • Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
  • Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)
  • Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)
  • Osseous crown lengthening
  • Removal of subgingival calculi in periodontal pockets with periodontitis by closed or open curettage
  • Waterlase MD Er,Cr:YSGG assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium)
Device Description

The Waterlase MD Turbo Plus is a dental laser device indicated for incision, excision, removal, and specific dental therapies, such as root canal therapy and periodontal therapy. All indications for use submitted for the Waterlase® MD Turbo Plus have been previously cleared by the FDA and there are no new indications included for this device.

AI/ML Overview

This K101658 510(k) submission for the Waterlase MD Turbo Plus does not contain specific acceptance criteria or a detailed clinical study demonstrating the device's performance against such criteria. Instead, it relies on substantial equivalence to previously cleared predicate devices.

Here's a breakdown based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not specify quantitative acceptance criteria or report device performance in a formal table with metrics like sensitivity, specificity, accuracy, etc. The "performance" described is largely qualitative and comparative to predicate devices.

Acceptance Criteria (Not explicitly stated as numeric targets)Reported Device Performance (as described in the "Discussion")
Equivalency for higher energy per pulse (600 mJ)Equivalent to Waterlase® MD Turbo (predicate K090181) at 600 mJ per pulse for preparation of Class II cavities
Equivalency for higher energy per pulse (600 mJ) to other predicatesEquivalent to predicate devices cleared under K030146 and K070355 for 600 mJ per pulse
Safety and Effectiveness for additional pulse frequencies (75 and 100 Hz)Determined safe and effective through performance data for gingivoplasty on pig jaw gingiva at 75 and 100 Hz, demonstrating equivalency to cleared indication for predicate devices.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: Not explicitly stated. For the higher energy per pulse, it mentions "comparison of class II cavity preparations prepared at 600 mJ per pulse." For the new frequencies, it mentions "Gingivoplasties performed at the two new frequencies on pig jaw gingiva." No specific numbers of preparations or gingivoplasties are provided.
  • Data Provenance: The document does not specify the country of origin. The test for gingivoplasty was performed on "pig jaw gingiva," indicating an in vitro or ex vivo animal model rather than human clinical data. The comparison of Class II cavity preparations suggests an in vitro or ex vivo setting as well, given the lack of clinical study details. The data is likely retrospective in the sense that it's a comparison to established performance of predicate devices, or bench/laboratory for the specific tests mentioned.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

  • Number of Experts: Not mentioned.
  • Qualifications of Experts: Not mentioned.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not performed. The submission explicitly states: "No clinical performance data is required for this submission." The equivalency is established through comparison to predicate devices and limited performance data (Class II cavity preparations and gingivoplasties on pig jaw gingiva), not through human reader studies.

6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study:

The Waterlase MD Turbo Plus is a physical dental laser device, not an algorithm. Therefore, a "standalone algorithm performance" study is not applicable. The performance data mentioned (cavity preparations, gingivoplasties) represents the device's direct physical interaction with tissue.

7. Type of Ground Truth Used:

  • For the Class II cavity preparations, the "ground truth" implicitly refers to the established effectiveness and safety of the predicate device (Waterlase® MD Turbo) for this application. The new device's performance is compared to this established benchmark.
  • For the gingivoplasties on pig jaw gingiva, the "ground truth" would be the observed clinical effect (e.g., successful tissue removal, absence of adverse effects) as assessed by technical or dental experts, compared to the expected outcome from predicate devices for a similar indication. This is a technical assessment rather than a diagnostic "ground truth" like pathology or clinical outcomes in a human patient.

8. Sample Size for the Training Set:

This is irrelevant as the Waterlase MD Turbo Plus is a physical device and not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

This is irrelevant as the Waterlase MD Turbo Plus is a physical device and not an AI/ML algorithm.

{0}------------------------------------------------

K 101658

(

510k Summary of Safety and Effectiveness Waterlase® MD Turbo Plus July 13, 2010

ATTACHMENT 7 510(k) Summary of Safety and Effectiveness Information

AUG 1 1 2010

Date Prepared:July 13, 2010
Regulatory Authority:Safe Medical Devices Act of 1990, 21 CFR 807.92
Company:Biolase Technology, Inc.4 CromwellIrvine, CA 92618
Contact:Ms. Ioana RizoiuBiolase Technology, Inc.4 CromwellIrvine, CA 92618Tel: (949) 226-8144Fax: (949) 273-6680
Trade Name:Waterlase® MD Turbo Plus
Common Name:Er,Cr:YSGG laser
Classification Name:-Powered laser surgical instrument-System, dental, hydrokinetic, caries removal & cavity preparation-Drill, bone, powered
Classification Code(s):79 GEX, MXF, DZI, a Class II device
Equivalent Devices:Waterlase® MDBiolase Technology, Inc.K031140, July 7, 2004; K071363, February 12, 2008
Waterlase® MD TurboBiolase Technology, Inc.K090181, February 11, 2009; K083927, October 1, 2009;K091746, December 7, 2009

Device Description:

ﺎﻳ

The Waterlase MD Turbo Plus is a dental laser device indicated for incision, excision, removal, and specific dental therapies, such as root canal therapy and periodontal therapy. All indications for use submitted for the Waterlase® MD Turbo Plus have been previously cleared by the FDA and there are no new indications included for this device.

{1}------------------------------------------------

K 1016)8

Indications for Use:

General Indications

  • 트 Class I, II, III, IV and V cavity preparation
  • 트 Caries removal
  • . Hard tissue surface roughening or etching
  • Enameloplasty, excavation of pits and fissures for placement of sealants 트
      • For use on adult and pediatric patients

Root Canal Hard Tissue Indications

  • I Tooth preparation to obtain access to root canal
  • 트 Root canal preparation including enlargement
  • I Root canal debridement and cleaning

Root Canal Disinfection

  • Laser root canal disinfection after endodontic treatment ■

Endodontic Surgery (Root Amputation) Indications

  • Flap preparation incision of soft tissue to prepare a flap and expose the bone. 트
  • Cutting bone to prepare a window access to the apex (apices) of the root(s). 트
  • 트 Apicoectomy - amputation of the root end.
  • Root end preparation for retrofill amalgam or composite. 트
  • Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues . (i.e., granulation tissue) from around the apex

NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.

Bone Surgical Indications

  • Cutting, shaving, contouring and resection of oral osseous tissues (bone) l
  • I Osteotomy

Soft Tissue Indications including Pulpal Tissues*

Incision, excision, vaporization, ablation and coagulation of oral soft tissues, including:

  • 트 Excisional and incisional biopsies
  • Exposure of unerupted teeth
  • l Fibroma removal
  • Flap preparation incision of soft tissue to prepare a flap and expose the bone. ■
  • Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth I (hard and soft tissue impactions)
  • Frenectomy and frenotomy 트
  • Gingival troughing for crown impressions 트
  • Gingivectomy 메
  • 트 Gingivoplasty

{2}------------------------------------------------

jK161658

  • 트 Gingival incision and excision
  • I Hemostasis
  • 이 Implant recovery
  • 트 Incision and drainage of abscesses
  • .. Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery
  • 트 Leukoplakia
  • 피 Operculectomy
  • Oral papillectomies 메
  • Operculectomy 트
  • 트 Oral papillectomies
  • 트 Pulpotomy
  • 피 Pulp extirpation
  • I Pulpotomy as an adjunct to root canal therapy
  • 트 Root canal debridement and cleaning
  • · · Reduction of gingival hypertrophy
  • Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic . tissues (i.e., granulation tissue) from around the apex

NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.

  • 트 Soft tissue crown lengthening
  • Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa .
  • . Vestibuloplasty

*For use on adult and pediatric patient

Laser Periodontal Procedures

  • 트 Full thickness flap
  • 트 Partial thickness flap
  • Split thickness flap 1
  • Laser soft tissue curettage 1
  • Laser removal of diseased, inflamed and necrosed soft tissue within the 트 periodontal pocket
  • Removal of highly inflamed edematous tissue affected by bacteria penetration of the 트 pocket lining junctional epithelium
  • Removal of granulation tissue from bony defects 제
  • Sulcular debridement (removal of diseased, inflamed or necrosed soft 피 tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
  • Osteoplasty and osseous recontouring (removal of bone to correct osseous defects 미 and create physiologic osseous contours)

510k Summary of Safety and Effectiveness Waterlase® MD Turbo Plus June 7, 2010

{3}------------------------------------------------

K101658

r
Franchise

ﺎ ﺳﮯ ﺍ

S 10k Summary of Safety and Effectiveness Waterlase® MD Turbo Plus June 7, 2010

  • . Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)
  • 트 Osseous crown lengthening
  • .. Removal of subgingival calculi in periodontal pockets with peridonditis by closed or open curettage
  • . Waterlase MD Er.Cr.YSGG assisted new attachment procedure (cementummediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium)

Contraindications:

All clinical procedures performed with the Waterlase MD Turbo Plus must be subjected to the same clinical judgment and care as with traditional techniques. Patient risk must always be considered and fully understood before clinical treatment. The clinician must completely understand the patient's medical history prior to treatment. Exercise caution for general medical conditions, which might contraindicate a local procedure. Such conditions may include, but are not limited to, allergy to local or topical anesthetics, heart disease, lung disease, bleeding disorders, or an immune system deficiency. Medical clearance from the patient's physician is advisable when doubt exists regarding treatment.

Discussion:

The Waterlase MD Turbo Plus has been compared and determined to be equivalent to all predicated devices presented as part of the Substantial Equivalency discussion. In summary, for changes to the higher energy per pulse of 600 mJ, Waterlase MD Turbo Plus is equivalent to Waterlase® MD Turbo (predicate). The two devices have the same energy per square cm when compared at settings related to the preparation of class II cavities. Equivalency of the Waterlase MD Turbo Plus was also established through performance data in a comparison of class II cavity preparations prepared at 600 mJ per pulse with the Waterlase MD Turbo Plus to the same prepared with the Waterlase® MD Turbo (predicate, K090181). In addition, the Waterlase MD Turbo Plus operated at 600 mJ per pulse is also equivalent to the other predicated devices cleared by the FDA under K030146 and K070355. For the additional pulse frequencies of 75 and 100Hz, the Waterlase MD Turbo Plus device was determined safe and effective through performance data. Gingivoplasties performed at the two new frequencies on pig jaw gingiva demonstrated that the Waterlase® MD Turbo Plus is a safe and effective device for the gingivoplasty indication. Gingivoplasty is an indication already cleared for the listed predicate devices.

Conclusion:

The indications included herein are the same as the indications that have been previously cleared by the FDA for the Waterlase MD predicates under K031140, K071363, K090181, K083927, and K091746, as well as other dental laser predicates, such as K001527, K070355, K983100, K000805, K030146, and K073074 . No clinical performance data is required for this submission. Therefore, substantial equivalency for the Waterlase® MD Turbo Plus has been determined through comparison to these previously cleared dental laser devices.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wavy lines extending from its head, all enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged around the upper portion of the circle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Biolase Technology, Inc. % Ms. Ioana Rizoiu Vice President of Clinical Research and Development 4 Cromwell Irvine, California 92618

AUG 1 1 2010

Re: K101658

Trade/Device Name: Waterlase® MD Turbo Plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, MXF Dated: June 07, 2010 Received: June 11, 2010

Dear Ms. Rizoiu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Ioana Rizoiu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Waterlase MD Turbo Plus

Indications for Use:

AUG11 2010

General Indications*

  • 트 Class I, II, III, IV and V cavity preparation
  • l Caries removal
  • 트 Hard tissue surface roughening or etching
  • 지 Enameloplasty, excavation of pits and fissures for placement of sealants
    • For use on adult and pediatric patients

Root Canal Hard Tissue Indications

  • 트 Tooth preparation to obtain access to root canal
  • Root canal preparation including enlargement
  • Root canal debridement and cleaning

Root Canal Disinfection

  • I Laser root canal disinfection after endodontic treatment

Endodontic Surgery (Root Amputation) Indications

  • Flap preparation - incision of soft tissue to prepare a flap and expose the bone.
  • Cutting bone to prepare a window access to the apex (apices) of the root(s). 트
  • Apicoectomy - amputation of the root end.
  • 포 Root end preparation for retrofill amalgam or composite. .
  • Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation I tissue) from around the apex

NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.

Bone Surgical Indications

  • 트 Cutting, shaving, contouring and resection of oral osseous tissues (bone)
  • Osteotomy

Soft Tissue Indications including Pulpal Tissues*

Incision, excision, vaporization, ablation and coagulation of oral soft tiss

  • 트 Excisional and incisional biopsies
  • Exposure of unerupted teeth
  • Fibroma removal

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K101658

{7}------------------------------------------------

  • " Flap preparation - incision of soft tissue to prepare a flap and expose the bone.
  • . Flap preparation - incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions)
  • 트 Frenectomy and frenotomy
  • . Gingival troughing for crown impressions
  • . Gingivectomy
  • I Gingivoplasty
  • 페 Gingival incision and excision
  • Hemostasis ■
  • implant recovery
  • 트 Incision and drainage of abscesses
  • Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during ■ apical surgery
  • 사 Leukoplakia
  • 트 Operculectomy
  • 트 Oral papillectomies
  • 트 Operculectomy
  • ' Oral papillectomies
  • 트 Pulpotomy
  • 트 Pulp extirpation
  • E Pulpotomy as an adjunct to root canal therapy
  • . Root canal debridement and cleaning
  • 8 Reduction of gingival hypertrophy
  • Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues
    • (i.e., granulation tissue) from around the apex

NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.

  • 발 Soft tissue crown lengthening
  • .. Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
  • 지 Vestibuloplasty
  • For use on adult and pediatric patient

Laser Periodontal Procedures

  • 트 Full thickness flap
  • 트 Partial thickness flap
  • 트 Split thickness flap
  • . Laser soft tissue curettage
  • . Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
  • . Removal of highly inflamed edematous tissue affected by bacteria pen junctional epithelium
  • 트 Removal of granulation tissue from bony defects

(Division Sign Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K101658

{8}------------------------------------------------

  • Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the l periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
  • 트 Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)
  • 피 Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)
  • 8 Osseous crown lengthening
  • Removal of subgingival calculi in periodontal pockets with periodontitis by closed or open curettage 에
  • Waterlase MD Er,Cr:YSGG assisted new attachment procedure (cementum-mediated periodontal 트 ligament new-attachment to the root surface in the absence of long junctional epithelium)

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Niel R.P. Dyben for mxm
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K101658

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.