The sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps.
The Sharps Container is a piece of equipment used in hospitals, physicians' offices, dental offices, laboratories, home health areas, patient rooms and other locations of medical waste. The sharps container is designed to safely contain used needles or syringes and prevent accidental needle sticks from used syringes or needles.
The device is made of injected molded polypropylene material. The material . is made of molded plastic into a specific shape. The actual shape is designed to sit on the floor or a desk.
On the outer casing of all the sharps containers is the Biohazard Warning label in Orange, noticeably displayed.
The molded plastic container is designed and manufactured such that there . will be no leakage during storage, handling or transport.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Device: Sharps Container, Model Lns-T1
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Impact Resistance | Passed (according to testing against BS7230:1990, ASTM F2132:2008, and OSHA 29CFR1910) |
| Puncture Resistance | Passed (according to testing against BS7230:1990, ASTM F2132:2008, and OSHA 29CFR1910) |
| Leak Resistance (leak-proof on sides/bottom) | Passed (according to testing against BS7230:1990, ASTM F2132:2008, and OSHA 29CFR1910) |
| Mechanical Safety | Passed (according to testing against BS7230:1990, ASTM F2132:2008, and OSHA 29CFR1910) |
| Environmental Safety | Passed (according to testing against BS7230:1990, ASTM F2132:2008, and OSHA 29CFR1910) |
| Performance | Passed (according to testing against BS7230:1990, ASTM F2132:2008, and OSHA 29CFR1910) |
| Material (Polypropylene) | Meets, device is made of injected molded polypropylene material. |
| Sharps closure (flaps closed and locked) | Meets, designed for flaps to be closed and locked in place for removal. |
| Single Use | Meets, device is indicated as single-use. |
The acceptance criteria are implied by the comparison to the predicate device and the standards used for performance testing (BS7230:1990, ASTM F2132:2008, OSHA 29CFR1910). The reported device performance is that the "product passed all tests."
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state the sample size for the test set or the exact data provenance (e.g., country of origin, retrospective/prospective). It only mentions that "Mechanical, environmental safety and performance testing have been accomplished according to BS7230:1990, ASTM F2132:2008 and OSHA 29CFR1910." and that the "product passed all tests." This suggests that the testing was conducted in a laboratory setting to meet these specific standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable and is not provided in the document. The device is a physical sharps container, and its performance (e.g., impact, puncture, leak resistance) is evaluated through standardized mechanical and environmental testing against established engineering standards, not through human expert interpretation of data.
4. Adjudication Method for the Test Set
This information is not applicable and is not provided in the document. As mentioned above, the evaluation is based on objective measurements against engineering standards, not expert consensus requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic or AI-assisted devices where human interpretation is involved. This device is a physical container, and its effectiveness is determined by its physical properties.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This is not applicable to a physical device like a sharps container. "Standalone" performance usually refers to an algorithm's ability to perform a task without human intervention, which doesn't fit the context of this product. The device's performance is inherently "standalone" in that it performs its function (containing sharps) without human "in-the-loop" interaction for its core function evaluation beyond its initial use.
7. The Type of Ground Truth Used
The ground truth for the performance testing was established by recognized industry and safety standards: BS7230:1990, ASTM F2132:2008, and OSHA 29CFR1910. These standards define the parameters and methodologies for evaluating the safety and effectiveness of sharps containers (e.g., impact, puncture, and leak resistance).
8. The Sample Size for the Training Set
This information is not applicable and is not provided. The development and testing of a physical product like a sharps container do not typically involve "training sets" in the context of algorithms or machine learning. The design implicitly relies on engineering principles and material science, which are validated through physical testing.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable. There is no "training set" in the context of this device. The design principles and manufacturing processes are likely informed by engineering best practices, prior product development, and regulatory requirements (e.g., those found in the referenced standards).
{0}------------------------------------------------
K 101567 2 2010 SEP
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: June 7, 2009
- Company and Correspondent making the submission:
Name - Cixi Leinuo Plastic Co., Ltd
Address - No. 56, Luoming South Road, Guanhaiwei Town, Cixi , Ningbo,
Zhejiang, China,315316
Telephone - +86-574-63673001
Fax - +86-574-63671001
Contact - Mr. Leiping Wang
Email – cxleinuo@163.com
- Device :
Trade/proprietary name: Sharps Container Lns-T1
Common Name : Sharps Container Classification Name : Container, Sharps
Predicate Devices:
| Model | Manufacturer | K Number | Submitted Device |
|---|---|---|---|
| Sharps AwayDisposableContainers | Solutions, Inc. | K072667 | Lns-T1 |
- Classifications Names & Citations :
21CFR 884.6120, MMK, Sharps Container, Class II
15 of 58
{1}------------------------------------------------
4. Description :
4.1 General
The Sharps Container is a piece of equipment used in hospitals, physicians' offices, dental offices, laboratories, home health areas, patient rooms and other locations of medical waste. The sharps container is designed to safely contain used needles or syringes and prevent accidental needle sticks from used syringes or needles.
Structure
- The device is made of injected molded polypropylene material. The material . is made of molded plastic into a specific shape. The actual shape is designed to sit on the floor or a desk.
- · On the outer casing of all the sharps containers is the Biohazard Warning label in Orange, noticeably displayed.
- The molded plastic container is designed and manufactured such that there . will be no leakage during storage, handling or transport.
for the daving thinker
| Please refer to the table below for the device thickness. | ||||
|---|---|---|---|---|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Comments of the comments of the comments of the comments of | - A L L L L L L L L L L L L L L L L B L B R L B R B R B R B R B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L | ||
| Model Name | Size (cm) | Volume (L) | EmptyWeight(grams) | Thickness of material |
|---|---|---|---|---|
| LNS-T1 | 10.5818 | 1.0 | 80g | body:1.2mm lid:1.3mm |
- Indication for use : Device Name:
Lns-T1
Indications for Use:
The sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps.
{2}------------------------------------------------
- Comparison with predicate device :- Please see the next page
:
. . . . . .
.
. . . .
and the same of the same of the same
:
:
:
.
- September 2017
and the comments of the comments of the comments of
{3}------------------------------------------------
| Element of comparison | Subject Device | Claimed SE Device |
|---|---|---|
| Company | Cixi Leinuo Plastic Co., Ltd. | Solutions Inc. |
| FDA510(K) Number | N/A | K072667 |
| Device Name | Sharps Container | Sharps Away Disposable Container |
| Intended use(s) | Leinuo sharps containers areintended to be used for the safedisposal of hazardous sharps. | Sharps Away Disposable Containers areintended to be used for the safe disposalof hazardous sharps. |
| Target Population | Healthcare professional | Healthcare professional |
| where used: Hospital, PhysiciansOffices, Dental offices, Laboratories,Home health, Patient room, etc. | Yes | Yes |
| Material | Polypropylene | Polypropylene |
| Sharps closure | Flaps are closed and locked in placefor removal | Flaps are closed and locked in place forremoval |
| Impact Resistance | Yes | Yes |
| Puncture Resistance | Yes | Yes |
| Leak Resistance | Leak-proof on the sides and bottom | Leak-proof on the sides and bottom |
| Single Use | Yes | Yes |
18 of 58
:
{4}------------------------------------------------
-
- Performance Testing :
Mechanical, environmental safety and performance testing have been accomplished according to BS7230:1990, ASTM F2132:2008 and OSHA 29CFR1910. The product passed all tests. These tests demonstrate this product is substantially equivalent to the predicate device.
- Performance Testing :
8. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification The Cixi Leinuo Plastic Co., Ltd. Sharps Container, Model Lns-T1 is substantially equivalent to the predicate devices as described herein.
-
- Cixi Leinuo Plastic Co., Ltd will update and include in a summary any other information deemed Yeasonably necessary by the FDA.
END
- Cixi Leinuo Plastic Co., Ltd will update and include in a summary any other information deemed Yeasonably necessary by the FDA.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Cixi Leinuo Plastic Company, Limited C/O Mr. Charles Mack IRC 77325 Joyce Way Echo, Oregon 97826
S.. 2 2010
Re: K101567
Trade/Device Name: Sharps Container, Models Lns-T1 Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK Dated: March 25, 2010 Received: June 4, 2010
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Mr. Mack
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
Device Name: Sharps Container, Models Lns-T1 Indications For Use:
The sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps.
Prescription Use ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth T. Clavie-Will
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital Division Control, Dental Devices
510(k) Number: KI01567
§ 880.5570 Hypodermic single lumen needle.
(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).