K Number
K101560

Validate with FDA (Live)

Date Cleared
2011-03-25

(294 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medline Retractable Safety Syringe is used to inject fluids into, or withdraw fluids from, the body. In addition, this safety syringe is designed to aid in the prevention of needle stick injuries and reduce the potential for syringe reuse.

Device Description

The Medline Retractable Safety Syringe is a retractable type piston syringe, designed to aid in the prevention of needle stick injuries. This single-use, disposable syringe consists of the following components:

  1. Barrel - The barrel has a scale showing the capacity of the syringe. In addition, the tip of the barrel has a luer lock fitting for the user in affach a needle.
  2. Plunger - Once the plunger is fully depressed, it engages the needle assembly. As the plunger is retracted, the needle assembly is retracted into the barrel. Once the plunger is fully retracted and locked in place, the plunger is snapped off leaving the needle in the barrel of the syringe.
  3. O-Ring - The O-ring minimizes the risk of leakage around the adapter.
  4. Gasket - The gasket maintains the fluid in the barrel between the adapter and plunger.
  5. Luer Lock or Needle Holder - The luer lock facilitates passage of the fluid, while holding the needle in place.
  6. Hub Holder - Engages the plunger after the injection.
  7. Locking Ring - A plastic insert at the top of the barrel. After the injection, the health care professional retracts the plunger with the needle into the barrel. Once the plunger is fully retracted, the plunger is locked into position at the top of the barrel. This safety mechanism makes sure the needle cannot be pushed back out the tip of the barrel.
  8. Needle - Pre-attached single lumen stainless steel needle which penetrates the patient's skin to inject/withdraw fluid. (1ml syringe only)
  9. Short Cap - Protective cover for the needle tip. Removed upon use. (1ml syringe only)
  10. Big Cap - Protective cover for the plunger. (1ml syringe only)
    The Medline Retractable Safety Syringe functions in a manner similar to standard syringes for fluid injection/withdrawal. After use, the health care professional fully depresses the plunger to engage the luer assembly. Once the luer assembly is engaged, pulling back the plunger causes the adapter and the attached needle to be withdrawn into the safety of the barrel. This retraction into the barrel of the syringe can be visually confirmed. In this position against the flange, single-handed lateral pressure on the plunger results in a controlled fracture of the plunger. Once this safety mechanism has been activated, the syringe is permanently disabled and the needle is completely secured within the barrel. Both the syringe and plunger are discarded in a Sharp's container.
AI/ML Overview

This document describes a 510(k) submission for the Medline Retractable Safety Syringe, a medical device designed to prevent needlestick injuries. The submission focuses on substantial equivalence to predicate devices rather than a complex clinical study proving novel performance. Therefore, many of the requested sections related to clinical study specifics (such as MRMC studies, training set details, and expert qualifications for ground truth) are not applicable or explicitly stated in this type of regulatory submission for this device.

Here's the breakdown of the available information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Compliance with ISO 7886-1:1993 (Sterile hypodermic syringes for single use - Part 1: Syringes for manual use)"Testing consisted of compliance to the applicable sections of... ISO 7886-1:1993."
Compliance with ISO 7886-4:2006 (Sterile hypodermic syringes for single use - Part 4: Syringes with an integral reuse prevention feature)"Testing consisted of compliance to the applicable sections of... ISO 7886-4:2006."
Compliance with ISO 9626 (Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods)"Testing consisted of compliance to the applicable sections of... ISO 9626."
Compliance with ISO 594 (Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment)"Testing consisted of compliance to the applicable sections of... ISO 594."
Compliance with ISO 10993 (Biological evaluation of medical devices)"Testing consisted of compliance to the applicable sections of... ISO 10993."
Prevention of needle stick injuriesDevice is "designed to aid in the prevention of needle stick injuries." The mechanism for achieving this is described: "After use, the health care professional fully depresses the plunger to engage the luer assembly... pulling back the plunger causes the adapter and the attached needle to be withdrawn into the safety of the barrel... Once this safety mechanism has been activated, the syringe is permanently disabled and the needle is completely secured within the barrel." This functional description serves as the performance claim against the acceptance criterion.
Reduction of potential for syringe reuseDevice is "designed to... reduce the potential for syringe reuse." The mechanism described (permanent disabling and securing the needle within the barrel) supports this claim.
Substantial Equivalence to predicate devices (Inviro Snap Safety Syringe / InviroSnap Safety Syringe - K032780 / K092430)"Information was submitted to demonstrate that there are no significant differences in technological characteristics between Medline's Retractable Safety Syringe and the cited predicate devices." The FDA concurred with this finding, stating the device is "substantially equivalent."

2. Sample size used for the test set and the data provenance

The document does not specify a separate "test set" in the context of clinical data for performance evaluation. The "testing" mentioned refers to compliance with international standards (ISO standards), which typically involve laboratory testing of device attributes on a batch-by-batch or statistically representative sample basis. The sample size for these engineering and materials tests is not provided. Data provenance is not applicable as this describes physical product testing, not patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. Ground truth as typically defined for clinical studies or AI algorithm evaluation (e.g., expert consensus on medical images) is not relevant for this type of device submission and its associated testing. The "ground truth" for compliance with ISO standards is the standard itself and its specified test methods and acceptance limits.

4. Adjudication method for the test set

Not applicable. No expert adjudication process for clinical data is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a physical medical device (syringe), not an AI algorithm for diagnostic interpretation. Therefore, MRMC studies and AI assistance metrics are not relevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the compliance testing against ISO standards, the "ground truth" is defined by the objective metrics and specifications within those standards (e.g., force required for retraction, leakage rates, material specifications). For the claim of substantial equivalence, the "ground truth" is the performance and technological characteristics of the predicate devices.

8. The sample size for the training set

Not applicable. There is no "training set" as this is not a machine learning or AI-based device.

9. How the ground truth for the training set was established

Not applicable. There is no "training set."

{0}------------------------------------------------

510(k) Summary

KIOISGO

Submitter Information

Medline Industries, Inc. One Medline Place Mundelein, IL 60060 Contact Person: Matt Clausen Phone Number: 847-643-4785 Date: March 15, 2011

Device Information

Classification Name: Syringe, Antistick Proprietary Name: Medline Retractable Safety Syringe Common Name: Piston or Safety Syringe

Predicate Information

Predicate Name: Inviro Snap Safety Syringe / InviroSnap Safety Syringe Predicate 510(k) #: K032780 / K092430 Predicate Product Code: MEG

Device Description

The Medline Retractable Safety Syringe is a retractable type piston syringe, designed to aid in the prevention of needle stick injuries. This single-use, disposable syringe consists of the following components:

  1. Barrel - The barrel has a scale showing the capacity of the syringe. In addition, the tip of the barrel has a luer lock fitting for the user in affach a needle. As a strenger and

  2. Plunger - Once the plunger is fully depressed, it engages the needle assembly. As the plunger is retracted, the needle assembly is retracted into the barrel. Once the plunger is fully retracted and locked in place, the plunger is snapped off leaving the needle in the barrel of the syringe.

  3. O-Ring - The O-ring minimizes the risk of leakage around the adapter.

  4. Gasket - The gasket maintains the fluid in the barrel between the adapter and plunger.

  5. Luer Lock or Needle Holder - The luer lock facilitates passage of the fluid, while holding the needle in place.

  6. Hub Holder - Engages the plunger after the injection.

  7. Locking Ring - A plastic insert at the top of the barrel. After the injection, the health care professional retracts the plunger with the needle into the barrel. Once the plunger is fully retracted, the plunger is locked into position at the top of the barrel. This safety mechanism makes sure the needle cannot be pushed back out the tip of the barrel.

MAR 2 5 2011

{1}------------------------------------------------

  1. Needle - Pre-attached single lumen stainless steel needle which penetrates the patient's skin to inject/withdraw fluid. (1ml syringe only)

  2. Short Cap - Protective cover for the needle tip. Removed upon use. (1ml syringe only)

  3. Big Cap - Protective cover for the plunger. (1ml syringe only)

The Medline Retractable Safety Syringe functions in a manner similar to standard syringes for fluid injection/withdrawal. After use, the health care professional fully depresses the plunger to engage the luer assembly. Once the luer assembly is engaged, pulling back the plunger causes the adapter and the attached needle to be withdrawn into the safety of the barrel. This retraction into the barrel of the syringe can be visually confirmed. In this position against the flange, single-handed lateral pressure on the plunger results in a controlled fracture of the plunger. Once this safety mechanism has been activated, the syringe is permanently disabled and the needle is completely secured within the barrel. Both the syringe and plunger are discarded in a Sharp's container.

Indications For Use Statement

The intended use of this device is to inject fluids into, or withdraw fluids from, the body. In addition, this retractable safety syringe is designed to aid in the prevention of needle stick injuries and reduce the potential for syringe reuse.

Comparison of Required Technological Characteristics

Information was submitted to demonstrate that there are no significant differences in technological characteristics between Medline's Retractable Safety Syringe and the cited predicate devices.

Summary and Conclusion of Testing

Testing consisted of compliance to the applicable sections of the following testing

  • ISO 7886-1:1993 .
  • ISO 7886-4:2006 .
  • ISO 9626 .
  • ISO 594 .
  • . ISO 10993

Conclusion

storidards .

In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that Medline's Retractable Safety Syringe is safe, effective and substantially equivalent as described herein.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle symbol. The logo is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Matt Clausen Regulatory Affairs Medline Industries, Incorporated One Medline Place Mundelein, Illinois 60060

MAR 2 5 2011

Re: K101560

Trade/Device Name: Medline Retractable Safety Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: March 7, 2011 Received: March 8, 2011

Dear Mr. Clausen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Clausen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/ default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours

fors 2

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K101560

Device Name:

Medline Retractable Safety Syringe

Indications For Use:

The Medline Retractable Safety Syringe is used to inject fluids into, or withdraw fluids from, the body. In addition, this safety syringe is designed to aid in the prevention of needle stick injuries and reduce the potential for syringe reuse.

Prescription Use X (Part 21 CFR 801 Subpart D) OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of DRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K101560

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).