K Number
K101535

Validate with FDA (Live)

Device Name
BT-CATH
Date Cleared
2011-01-11

(222 days)

Product Code
Regulation Number
884.4530
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BT-Cath (BTC-100) is intended to provide temporary control or reduction of uterine bleeding during postpartum hemorrhage that is unresponsive to standard therapy including massage and oxytocin administration.

Device Description

The BT-Cath consists of an extruded silicone dual lumen catheter with an inflatable molded silicone tamponade balloon attached on the distal end. One lumen of the catheter serves as the conduit through which saline is infused into the balloon after insertion into the bleeding uterus. When inflated properly to press against the uterine wall, the balloon provides compression against multiple sites of torn arteries, ruptured sinuses, lacerations and other tissue evoking hemorrhage. The second lumen, which is open at the distal end of the catheter, provides a drainage port for blood and other bodily fluids. At the proximal end of the catheter, a stopcock retains the saline in the balloon until a clinician gradually drains it allowing the uterus to contract. For ease of filling the balloon with multiple syringes, a check valve is also provided at the proximal end. The device is latex-free and provided sterile.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the BT-Cath device:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Material/Design Equivalence"insignificant differences in the physical dimensions and physical configuration of the balloon tamponade catheters. The technological characteristics, methods of use, materials, intended use, indications for use and contraindications are the same as the predicate devices."The BT-Cath is described as an "extruded silicone dual lumen catheter with an inflatable molded silicone tamponade balloon attached on the distal end." This matches the general description of the predicate device (Bakri Uterine Balloon Tamponade) in terms of function and materials (silicone). The submission states "insignificant differences" between the BT-Cath and predicate.
BiocompatibilityCompatibility with biological systems."Biocompatibility...testing were also assessed to demonstrate substantial equivalence." (Implies satisfactory results to meet equivalence)
Shelf-LifeAbility to maintain functionality and sterility over an intended storage period."shelf-life testing were also assessed to demonstrate substantial equivalence." (Implies satisfactory results to meet equivalence)
Mechanical Integrity - BalloonStructural integrity of the balloon, including: - Relationship of volume of saline infused into the balloon and resulting balloon diameter. - Burst volume of the balloon. - Ability of the balloon to hold its shape and not leak during repeated and prolonged inflations."Both the BT-Cath and the Bakri device were tested and found to achieve structural integrity with a reasonable safety factor." (This statement confirms that the BT-Cath met the mechanical integrity requirements for the balloon, matching the predicate.)
Mechanical Integrity - CatheterStructural integrity of the catheter using leak tests."Both the BT-Cath and the Bakri device were tested and found to achieve structural integrity with a reasonable safety factor." (This statement confirms that the BT-Cath met the mechanical integrity requirements for the catheter, matching the predicate.)
Tensile StrengthTensile strength in each joint in the devices."Both the BT-Cath and the Bakri device were tested and found to achieve structural integrity with a reasonable safety factor." (This statement confirms that the BT-Cath met the tensile strength requirements for joints, matching the predicate.)
Clinical Performance/SafetyDevice consistently achieving its intended purpose without reported likelihood of patient injury."Clinical feedback was obtained after about 840 uses of BT-Cath by U.S. clinicians that indicates the device consistently achieved its intended purpose without any reported likelihood of patient injury." (This is presented as evidence of field performance meeting the safety and efficacy expectations.)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: "about 840 uses"
  • Data Provenance: "U.S. clinicians" (prospective, real-world clinical feedback after deployment, rather than a controlled test set from a formal clinical trial for pre-market submission directly).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The document mentions "U.S. clinicians" providing feedback. It does not specify a number of experts or their qualifications for establishing a ground truth in the context of a controlled test set with expert consensus. The feedback provided is general clinical observation rather than a structured expert review of specific cases.

4. Adjudication Method for the Test Set

  • No formal adjudication method is described. The "clinical feedback" appears to be an aggregation of observations from multiple clinicians.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study was done, nor is AI involved with this device. This is a medical device, specifically a balloon tamponade catheter, not an AI/software-as-a-medical-device (SaMD).

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable as this is a physical medical device and does not involve an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" for the clinical performance assessment seems to be real-world clinical outcomes and observations as reported by clinicians using the device. This is more akin to post-market surveillance data contributing to a pre-market submission, rather than a pre-defined ground truth for a structured clinical trial.

8. The Sample Size for the Training Set

  • Not applicable. This is a physical medical device, not a machine learning algorithm requiring a training set.

9. How the ground truth for the training set was established

  • Not applicable as this is a physical medical device.

{0}------------------------------------------------

K101535...
Page 1/2

Image /page/0/Picture/1 description: The image shows the logo for Utah Medical Products Inc. The logo consists of the text "UTAH MEDICAL PRODUCTS INC." in a sans-serif font, with a line underneath the text. Below the text is a stylized symbol that resembles the letter "U" or a horseshoe shape, with a vertical line extending upwards from the center of the "U" shape.

JAN 1 1 2011

SECTION 9: 510(k) SUMMARY

and the state of the states and

510(k) Owner:Address:Utah Medical Products, Inc.7043 South 300 WestMidvale, UT 84047 USA
Telephone No .:Fax No .:(801) 566-1200(801) 566-2062
Contact:Kevin L. Cornwell, CEO
Date:3 January 2011
Trade Name:BT-Cath
Common Name:Intrauterine Balloon Tamponade Catheter
Classification:Obstetric-gynecologic specialized manual instrument,21 CFR 884.4530; Product Code OQY
Predicate Devices:Bakri Uterine Balloon Tamponade, (K013597, K062438)Cook Ob/Gyn
Indications for Use:BT-Cath is intended to provide temporary control or reduction ofuterine bleeding during postpartum hemorrhage that is unresponsiveto standard conservative therapy including uterine massage andoxytocin administration.

{1}------------------------------------------------

7080 2

Device Description:

Comparison to predicate devices:

Performance Data:

The BT-Cath consists of an extruded silicone dual lumen catheter with an inflatable molded silicone tamponade balloon attached on the distal end. One lumen of the catheter serves as the conduit through which saline is infused into the balloon after insertion into the bleeding uterus. When inflated properly to press against the uterine wall, the balloon provides compression against multiple sites of torn arteries, ruptured sinuses, lacerations and other tissue evoking hemorrhage. The second lumen, which is open at the distal end of the catheter, provides a drainage port for blood and other bodily fluids. At the proximal end of the catheter, a stopcock retains the saline in the balloon until a clinician gradually drains it allowing the uterus to contract. For ease of filling the balloon with multiple syringes, a check valve is also provided at the proximal end. The device is latex-free and provided sterile.

There are insignificant differences in the physical dimensions and physical configuration of the balloon tamponade catheters. The technological characteristics, methods of use, materials, intended use, indications for use and contraindications are the same as the predicate devices. Biocompatibility and shelf-life testing were also assessed to demonstrate substantial equivalence.

Substantial equivalence is also based on an assessment of mechanical tests where the structural integrity of the balloon, including the relationship of volume of saline infused into the balloon and resulting balloon diameter, the burst volume of the balloon, and the ability of the balloon to hold its shape and not leak during repeated and prolonged inflations; the structural integrity of the catheter using leak tests; and the tensile strength in each joint in the devices were evaluated. Both the BT-Cath and the Bakri device were tested and found to achieve structural integrity with a reasonable safety factor. Clinical feedback was obtained after about 840 uses of BT-Cath by U.S. clinicians that indicates the device consistently achieved its intended purpose without any reported likelihood of patient injury.

K. Howell
Kevin L. Cornwell

Kevin L. Cornwell Chairman & CEO Utah Medical Products, Inc.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN USA" around the edge. Inside the circle is a stylized eagle symbol. The eagle is depicted with three thick, curved lines representing its body and wings. The image is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Mr. Kevin L. Cornwell Chairman & CEO Utah Medical Products, Inc. 7043 South 300 West MIDVALE UT 84047

JAN 1 1 201

Re: K101535 Trade Name: BT-Cath® Regulation Number: 21 CFR §884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: OQY Dated: January 3, 2011 Received: January 3, 2011

Dear Mr. Cornwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{3}------------------------------------------------

Page 2

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Herbert Semler MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

SECTION 3: INTENDED USE / INDICATIONS for USE

510(k) Number:K101535JAN 1 1 2011
---------------------------------------

Device Name: UTMD BT-Cath - Balloon Tamponade Catheter for Postpartum Uterine Hemorrhage (BTC-100)

Indications for Use:

BT-Cath (BTC-100) is intended to provide temporary control or reduction of uterine bleeding during postpartum hemorrhage that is unresponsive to standard therapy including massage and oxytocin administration.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use _ X (Per 21 CFR 801.109) OR

Lori K. Wertz
Visible Sign Off

10

Over-the-Counter Use

§ 884.4530 Obstetric-gynecologic specialized manual instrument.

(a)
Identification. An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes.
(2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant.
(3) An umbilical clamp is an instrument used to compress the umbilical cord.
(4) A uterine curette is an instrument used to scrape and remove material from the uterus.
(5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix.
(6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus.
(7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination.
(8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix.
(9) A gynecological cerclage needle is a looplike instrument used to suture the cervix.
(10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus.
(11) A gynecological fibroid screw is an instrument used to hold onto a fibroid.
(12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity.
(13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina.
(14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures.
(15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus.
(16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis.
(17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina.
(18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
(b)
Classification. (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.