K Number
K101455

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-07-19

(55 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EKO 7 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal, Abdominal, Pediatric, Small Organs, Neonatal Cephalic, Adult Cephalic, Trans-esophageal (non-Cardiac, Cardiac), Muscular-Skeletal (conventional, superficial), Cardiac Adult, Cardiac Pediatric and Peripheral-vessel.

Device Description

The EKO 7 is a dedicated cardiac purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, M mode, color Doppler imaging, power Doppler imaging, Harmonic imaging, PW spectral Doppler mode, CW spectral Doppler mode or as a combination of these modes. The EKO 7 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The EKO 7 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

AI/ML Overview

The provided text is a 510(k) Premarket Notification for the EKO 7 Diagnostic Ultrasound System. This document focuses on establishing substantial equivalence to previously cleared predicate devices rather than proving performance against specific acceptance criteria through a dedicated study.

Therefore, the requested information about acceptance criteria, device performance, sample sizes for test and training sets, data provenance, expert ground truth establishment, adjudication methods, and MRMC comparative effectiveness studies is not available in the provided text for the EKO 7 Diagnostic Ultrasound System itself.

The document primarily outlines:

  • Device Description: The EKO 7 is a mobile, software-controlled diagnostic ultrasound system for cardiac purposes and other applications. It acquires and displays ultrasound data in various modes (B-mode, M-mode, Doppler, Harmonic imaging, etc.).
  • Intended Uses: Diagnostic ultrasound imaging and fluid analysis of the human body for various clinical applications include Fetal, Abdominal, Pediatric, Small Organs, Adult Cephalic, Trans-esophageal, Muscular-Skeletal, Cardiac Adult, Cardiac Pediatric, and Peripheral-vessel.
  • Technological Characteristics: The device is stated to be substantially equivalent to previously cleared ACCUVIX V20 and V10 systems.
  • Safety Standards: The EKO 7 has been designed to meet several product safety standards (e.g., UL 60601-1, IEC60601-2-37, NEMA UD-2).
  • Transducers: A list of transducers compatible with the EKO 7 system is provided, along with their respective cleared indications for use, indicating whether they are new ("N") or previously cleared ("P") for specific modes of operation and clinical applications.

Note on "P" and "N" in Transducer Tables:
The tables for each transducer (e.g., C1-4EC, L3-8, P2-4BA) indicate "N" for new indications and "P" for previously cleared indications. This "P" refers to whether that specific transducer model had its indications cleared under a previous 510(k) submission for a different device, or if the indication was previously cleared for the EKO 7 in an earlier submission (which is unlikely in a new 510(k) for the system itself). It does not imply a performance study comparing the EKO 7's performance against pre-defined acceptance criteria.

In summary: This 510(k) submission establishes substantial equivalence based on technological characteristics and intended uses aligning with predicate devices, and adherence to recognized safety standards. It does not contain the detailed performance study information requested.

{0}------------------------------------------------

K101455

510(k) Premarket Notification

19 2010

EKO 7 Diagnostic Ultrasound System

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

1. Submitter's Information: 21 CFR 807.92(a)(1)

MEDISON CO., LTD. 1003, Daechi-dong, Gangnam-gu, Seoul 135-280, Korea

Contact Person: Mr. Kyung-Am, Shim Regulatory Affairs Manager

Telephone:82.2.2194.1381
Facsimile:82.2.2194.1399
Email:kashim@medison.com

Data Prepared: March 25, 2010

2. Name of the device:

Common/Usual Name:

Diagnostic Ultrasound System and Accessories

Proprietary Name:

EKO 7 Diagnostic Ultrasound System

Classification Names:FR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.1550IYN
Ultrasound Pulsed Echo Imaging System892.1560IYO
Diagnostic Ultrasound Transducer892.1570ITX

3. Identification of the predicate or legally marketed device:

K092159, ACCUVIX V20 Diagnostic Ultrasound System K093849, ACCUVIX V10 Diagnostic Ultrasound System K020789, GE Vivid 3 Expert/Pro Diagnostic Ultrasound System

510(k) Summary / Statement Certification

ATTACHMENT 1

{1}------------------------------------------------

4. Device Description:

The EKO 7 is a dedicated cardiac purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, M mode, color Doppler imaging, power Doppler imaging, Harmonic imaging, PW spectral Doppler mode, CW spectral Doppler mode or as a combination of these modes. The EKO 7 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The EKO 7 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

The EKO 7 has been designed to meet the following product safety standards:

  • UL 60601-1, Safety requirements for Medical Equipment
  • CSA C22.2 No. 601.1, Safety requirements for Medical Equipment
  • IEC60601-2-37, Diagnostic Ultrasound Safety Standards
  • EN/IEC60601-1, Safety requirements for Medical Equipment
  • EN/IEC60601-1-2, EMC requirements for Medical Equipment
  • NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
  • NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
  • IEC 61157, Declaration of the acoustic output
  • ISO10993-1, Biocompatibility

છે. Intended Uses:

The EKO 7 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal, Abdominal, Pediatric, Small Organs, Adult Cephalic, Trans-esophageal (non-Cardiac, Cardiac), Muscular-Skeletal (conventional. superficial), Cardiac Adult, Cardiac Pediatric and Peripheral-vessel.

Technological Characteristics: 6.

The EKO 7 is substantially equivalent to the ACCUVIX V20 Diagnostic Ultrasound System, cleared via K092159, and the ACCUVIX V10 Diagnostic Ultrasound System, cleared via K094623. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations.

END of 510(K) Summary

510(k) Summary / Statement Certification

ATTACHMENT 1

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the eagle. The eagle is facing to the left. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

JUL 1 9 2010

Medison Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K101455

Trade/Device Name: EKO 7 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: June 23, 2010 Received: June 24, 2010

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the EKO 7 Diagnostic Ultrasound System, as described in your premarket notification:

Transducer Model Number

C1-4ECP2-4BA
C2-6ICP3-8CA
L3-8P4-12
L5-13ISCW2.0
MPT3-7CW4.0

{3}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your de subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.

Sincerely vours, -

Mahal D.O.K. Kim

Donald St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: EKO 7 Diagnostic Ultrasound System

Indications for Use:

The EKO 7 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal, Abdominal, Pediatric, Small Organs, Neonatal Cephalic, Adult Cephalic, Trans-esophageal (non-Cardiac, Cardiac), Muscular-Skeletal (conventional, superficial), Cardiac Adult, Cardiac Pediatric and Peripheral-vessel.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR .

· Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Michael P. O'Keefe

Page 1 of 12

(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K K101455

Indications For Use

{5}------------------------------------------------

510(k) No.:

Device Name: EKO 7 Diagnostic Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal (See Note 3)NNNNNNote 1Notes 4, 7
AbdominalNNNNNNote 1Notes 4, 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNNote 1Note 5,6,7,9
Small Organ (See Note 5)NNNNNNote 1Note 5,6,7,9
Neonatal CephalicNNNNNNote 1Note 4, 7
Adult CephalicNNNNNNote 1Note 4,7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)NNNNNNote 1Note 7
Musculo-skel. (Convent.)NNNNNNote 1Note 5,6,7,9
Musculo-skel. (Superfic.)NNNNNNote 1Note 5,6,7,9
Intra-luminal
Other (spec.)
Cardiac AdultNNNNNNote 1Note 4,7
CardiacCardiac PediatricNNNNNNote 1Note 4,7
Trans-esophageal (Cardiac)NNNNNNote 1Note 7
Other (spec.)
PeripheralVesselPeripheral vesselNNNNNNote 1Note 5,6,7,9
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

510K

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)


(Division Sign-Off)

Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety.

16-1545

{6}------------------------------------------------

Section 1.3, Page 3

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: C1-4EC for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal (See Note 3)PPPPPNote 1Notes 7
AbdominalPPPPPNote 1Notes 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPPPNote 1Notes 7
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared under K060087; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

510F

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

Nihil D'K
(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

{7}------------------------------------------------

510(k) No.:

Device Name: C2-6IC for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)PPPPPNote 1Notes 7
AbdominalPPPPPNote 1Notes 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatricPPPPNote 1Notes 7
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared under K092159; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

ation and Safety

Imgur

(Division Sign-Off)

Indications For Use

Office of In Vitro Diagnostic Device Evaluation and Sa
510K K101455

Division of Radiological Devices

{8}------------------------------------------------

510(k) No.:

Device Name: L3-8 for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote 1Note 5, 6, 7, 9
Small Organ (See Note 5)NNNNNote 1Note 5, 6, 7, 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNote 1Note 5, 6, 7, 9
Musculo-skel. (Superfic.)NNNNNote 1Note 5, 6, 7, 9
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselNNNNNote 1Note 5, 6, 7, 9
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)

Indications For Use

Office of In Vitro Diagnostic Device Evaluation and Safety

510K. K101455

Division of Radiological Devices

{9}------------------------------------------------

510(k) No.:

Device Name: L5-13IS for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Note 5, 6, 7, 9
Small Organ (See Note 5)PPPPNote 1Note 5, 6, 7, 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 5, 6, 7, 9
Musculo-skel. (Superfic.)PPPPNote 1Note 5, 6, 7, 9
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 5, 6, 7, 9
Other (spec.)

N= new indication; P= previously cleared under K092159; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)

Division of Radiological Devices

Indications For Use

Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K101455

{10}------------------------------------------------

510(k) No.:

Device Name: MPT3-7 for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorCombined*Other
(Track I only)(Tracks I & III)Doppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)NNNNNote 1Note 7
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)ﻤﺴ
Cardiac Adult1
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)NNNNNote 1Note 7
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

Indications For Use

{11}------------------------------------------------

510(k) No.:

Device Name: P2-4BA for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
AbdominalPPPPPNote 1Note 4, 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 4, 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared under K092159; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

510k

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)

Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

{12}------------------------------------------------

510(k) No.:

Device Name: P3-8CA for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
AbdominalPPPPPNote 1Note 4, 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging & OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 4, 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
Peripheral VesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared under K092159; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

510k

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

Michael D. Kirk

(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

{13}------------------------------------------------

510(k) No.:

Device Name: P4-12 for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
AbdominalNNNNNNote 1Note 4, 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNNote 1Note 4, 7
Small Organ (See Note 5)
Neonatal CephalicNNNNNNote 1Note 4, 7
Adult CephalicNNNNNNote 1Note 4, 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac AdultNNNNNNote 1Note 4, 7
CardiacCardiac PediatricNNNNNNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

Indications For Use

510K K101455

{14}------------------------------------------------

510(k) No.:

Device Name: CW2.0 for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac AdultP
Cardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselP
Other (spec.)

N= new indication; P= previously cleared under K092159; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)


(Division Sign-Off)

Division of Radiological Devices

Indications For Use

Office of In Vitro Diagnostic Device Evaluation and Safety

510K K101455

{15}------------------------------------------------

Section 1.3, Page 12

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: CW4.0 for use with EKO 7

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging&OtherFetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricP
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac AdultP
CardiacCardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselP
Other (spec.)

N= new indication; P= previously cleared under K092159; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+C+PW, B+PD+PW, B+C+CW, B+PD+CW, B+C+M, Dual B, Dual B+C, Dual B+PD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI) / Pulse Inversion Harmonic Imaging

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)

Indications For Use

Division of Radiological Devices tion and Safety Office of Ir

N/A