K Number
K101307

Validate with FDA (Live)

Date Cleared
2011-08-09

(456 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AG-6081 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6081 Single Blood Glucose Monitoring System is to be used by a single person and should not be shared.

The AG-6081 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-6081 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 Single Blood Glucose Test Strips are for use with the AG-6081 Single Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6081 Single meters and AGS-1100 Single test strips.

The AG-6951 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6951 Single Blood Glucose Monitoring System is to be used by a single person and should not be shared.

The AG-6951 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-6951 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 Single Blood Glucose Test Strips are for use with the AG-6951 Single Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6951 Single meters and AGS-1100 Single test strips.

This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a glucose test.

The AG-6081 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6081 MULTI Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of a diabetes control program.

The AG-6081 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, upper arm, calf and thigh should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 MULTI Blood Glucose Test Strips are for use with the AG-6081 MULTI Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6081 MULTI meters and AGS-1100 MULTI test strips.

The AG-6951 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6951 MULTI Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of a diabetes control program.

The AG-6951 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, upper arm, calf and thigh should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 MULTI Blood Glucose Test Strips are for use with the AG-6951 MULTI Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6951 MULTI meters and AGS-1100 MULTI test strips.

This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a glucose test.

Device Description

AG-6081 and AG-6951 Single Blood Glucose Monitoring System (BGMS) consist of blood glucose meter, single use test strips, sterile lancets, lancing device and the control solutions.

While AG-6081 and AG-6951 MULTI Blood Glucose Monitoring System (BGMS) consist of blood glucose meter, MULT! use test strips, Auto-disabling Lancing device and the control solutions.

The four blood Glucose Monitoring system AG-6081 Single, AG-6951 Single, AG-6081 MULTI and AG-6951 MULTI are all based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 5 seconds. The control solution available is used to test the performance of the device. All of them use the same technological characteristics for testing with their predicate device. Only the appearance is different from their predicate device, and the new devices can test the blood glucose at the alternative site: the palm, the forearm, the upper arm, the calf and the thigh.

More over, AG-6951 Single and AG-6951 MULTI Blood Glucose Monitoring Systems have a voice function, which is also different from their predicate device.

AI/ML Overview

This document describes the AG-6081 Single, AG-6081 MULTI, AG-6951 Single, and AG-6951 MULTI Blood Glucose Monitoring Systems (BGMS). These devices are intended for the quantitative measurement of glucose in fresh capillary whole blood samples. The document focuses on comparing the new devices to a predicate device (AG-606 Blood Glucose Monitoring System, K073030) and claims conformity to relevant standards, but provides very limited details about the specific studies.

Here's an attempt to answer your questions based on the provided text, while acknowledging the severe limitations in the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that the devices conform to the following standards:

  • ISO 15197: In vitro diagnostic test systems- Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
  • FDA Draft Guidance Document-Total Product Life Cycle for Portable Invasive Blood Glucose Monitoring Systems: October 24, 2006
  • CLSI/NCCLS Guideline, EP07-A2 Interference Testing in Clinical Chemistry; Approved Guideline-Second edition

However, the specific "acceptance criteria" (e.g., accuracy percentages, bias limits) from these standards and the "reported device performance" in relation to these criteria are not explicitly provided in the text. The document only states that "the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness," implying that the devices met the performance requirements referenced in these standards without detailing the actual results.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document. There is no mention of the number of samples, participants, or the type of study (retrospective or prospective) used for performance evaluation. The country of origin of the data is also not specified, though the submitter is based in Tianjin, China.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The text does not detail any expert involvement in establishing ground truth.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study

A multi-reader, multi-case comparative effectiveness study with human readers and AI assistance is not applicable to this type of medical device (blood glucose monitoring system). These devices are standalone measurement tools and do not involve human interpretation of complex images or data that AI would assist with in the usual sense of an MRMC study.

6. Standalone Performance Study

Based on the claim of conformity to ISO 15197 and FDA guidance for blood glucose monitoring systems, a standalone performance study would have been conducted. ISO 15197 specifies accuracy requirements for glucose meters, which are assessed by comparing meter readings to a reference method (e.g., laboratory analyzer). While the document states "the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness," the detailed results of such a standalone study are not explicitly presented.

7. Type of Ground Truth Used

For blood glucose monitoring systems, the ground truth is typically established using a laboratory reference method (e.g., hexokinase method on a clinical chemistry analyzer) to measure blood glucose concentrations. The device's readings are then compared against these highly accurate reference values. This is implied by the nature of such devices and the standards cited, but not directly stated in the provided text.

8. Sample Size for the Training Set

This information is not applicable as blood glucose monitoring systems of this type (electrochemical biosensors) are not typically "trained" in the machine learning sense. Their performance is inherent to their design and manufacturing, rather than learned from a dataset.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated above.

{0}------------------------------------------------

Additional information for K101307 柯 顿(天 津) 电 子 医 疗 器 械 有 限 公 司 Andon Medical Co., Ltd. 地址:天津市,天津空港经济开发区,航宇路26号 AUG - 9 2011 Add: No.26 HangYu Road, Tianjin Airport Economic Area, Tianjin, P.R. China 邮编(P.C.) 300381

传真(Fax): 86-22-6052 6162 电话(Tel): 86-22-87612426 E-mail : andonmedical@yahoo.com.cn

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1.0 submitter's information_

Name:Andon Medical Co., Ltd.
Address:No.04-23-3 AIRPORT INDUSTRIAL PARK, TIANJIN
Phone number:86-22-8761 2426
Fax number:86-22-6052 6162
Contact:Yi Liu
Date of Application:5/9/2010

2.0 Device name

AG-6081 Single Blood Glucose Monitoring System AG-6951 Single Blood Glucose Monitoring System AG-6081 MULTI Blood Glucose Monitoring System AG-6951 MULTI Blood Glucose Monitoring System

3.0 Classification

Production code: NBW- Blood Glucose Monitoring System. Regulation number: 862.1345 Classification: II Panel: Clinical Chemistry

4.0 Predict device information

Manufacturer: Andon Health Co., Ltd. Device: AG-606 Blood Glucose Monitoring System 510(k) number: K073030

5.0 Intended use

5.1 AG-6081 Single and AG-6951 Single BGMS

The AG-6081 Single Blood Glucose Monitoring System is intended to be used

{1}------------------------------------------------

for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6081 Single Blood Glucose Monitoring System is to be used by a single person and should not be shared.

The AG-6081 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-6081 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 Single Blood Glucose Test Strips are for use with the AG-6081 Single Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6081 Single meters and AGS-1100 Single test strips.

The AG-6951 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6951 Single Blood Glucose Monitoring System is to be used by a single person and should not be shared.

The AG-6951 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-6951 Single Blood Glucose Monitoring System should not be used for the diagnosis

{2}------------------------------------------------

Additional information for K101307 柯 顿 (天 津) 电 子 医 疗 器 械 有 限 公 司 Andon Medical Co., Ltd. 地址:天津市,天津空港经济开发区,航宇路26号 Add: No.26 HangYu Road, Tianjin Airport Economic Area, Tianjin, P.R. China 邮编(P.C.): 300381 电话(Tel): 86-22-87612426 传真(Fax): 86-22-6052 6162 E-mail: andonmedical@yahoo.com.cn

of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 Single Blood Glucose Test Strips are for use with the AG-6951 Single Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6951 Single meters and AGS-1100 Single test strips.

This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a glucose test.

5.2 AG-6081 MULTI and AG-6951 MULTI BGMS

The AG-6081 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6081 MULTI Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of a diabetes control program.

The AG-6081 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, upper arm, calf and thigh should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 MULTI Blood Glucose Test Strips are for use with the AG-6081 MULTI Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper

{3}------------------------------------------------

Additional information for K101307 柯 顿(天 津)电子 医疗 器 械 有 限 公 司 Andon Medical Co., Ltd. 地址:天津市,天津空港经济开发区,航宇路26号 Add: No.26 HangYu Road, Tianjin Airport Economic Area, Tianjin, P.R. China 邮编(P.C.): 300381 电话(Tel): 86-22-87612426 传真(Fax): 86-22-6052 6162 E-mail: andonmedical@yahoo.com.cn

arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6081 MULTI meters and AGS-1100 MULTI test strips.

The AG-6951 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6951 MULTI Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of a diabetes control program.

The AG-6951 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, upper arm, calf and thigh should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 MULTI Blood Glucose Test Strips are for use with the AG-6951 MULTI Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6951 MULTI meters and AGS-1100 MULTI test strips.

This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a glucose test.

{4}------------------------------------------------

Additional information for K101307 柯顿(天津)电子医疗器械有限公司 Andon Medical Co., Ltd. 地址:天津市,天津空港经济开发区,航宇路26号 Add: No.26 HangYu Road, Tianjin Airport Economic Area, Tianjin, P.R. China 邮编(P.C.): 300381 传真(Fax): 86-22-6052 6162 电话(Tel): 86-22-87612426 E-mail: andonmedical@yahoo.com.cn

6.0 Device description

AG-6081 and AG-6951 Single Blood Glucose Monitoring System (BGMS) consist of blood glucose meter, single use test strips, sterile lancets, lancing device and the control solutions.

While AG-6081 and AG-6951 MULTI Blood Glucose Monitoring System (BGMS) consist of blood glucose meter, MULT! use test strips, Auto-disabling Lancing device and the control solutions.

The four blood Glucose Monitoring system AG-6081 Single, AG-6951 Single, AG-6081 MULTI and AG-6951 MULTI are all based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 5 seconds. The control solution available is used to test the performance of the device. All of them use the same technological characteristics for testing with their predicate device. Only the appearance is different from their predicate device, and the new devices can test the blood glucose at the alternative site: the palm, the forearm, the upper arm, the calf and the thigh.

More over, AG-6951 Single and AG-6951 MULTI Blood Glucose Monitoring Systems have a voice function, which is also different from their predicate device.

{5}------------------------------------------------

柯 顿 (天 津) 电 子 医 疗 器 械 有 限 公 司

Andon Medical Co., Ltd. 地址:天津市,天津空港经济开发区,航宇路 26号 Add: No.26 HangYu Road, Tianjin Airport Economic Area, Tianjin, P.R. China 邮编(P.C.): 300381 传真(Fax): 86-22-6052 6162 电话(Tel): 86-22-87612426 E-mail: andonmedical@yahoo.com.cn

7.0 Summary_comparing technological characteristics with predicate device

7.1 AG-6081 Single and AG-6081 MULTI BGMS

Similarities
NEW DEVICE:PREDICATE:
CHARACTERISTICSAG-6081 Single andAG-606 Blood Glucose
AG-6081 MULTI BloodMonitoring System
Glucose Monitoring System(K073030)
Detection MethodAmperometryAmperometry
EnzymeGlucose OxidaseGlucose Oxidase
Type of MeterBiosensor (Electrode)Biosensor (Electrode)
OperatingTemperature10°C~40°C(50°-104°F)10°C ~40°C(50°-104°F)
RangeLCDLCD
Result Presentationmg/dL or mmol/Lmg/dL or mmol/L
Test StartAutomaticAutomatic
Test Time5 seconds5 seconds
Battery LifeApprox. 500 normal testsApprox. 1000 normal tests
Measurement Range20mg/dL-600mg/dL20mg/dL-600mg/dL
(1.1mmol/L~33.3mmol/L)(1.1mmol/L~33.3mmol/L)
Differences
AG-6081 Single andAG-606 Blood Glucose
CHARACTERISTICSAG-6081 MU MULTI BloodMonitoring System
Glucose Monitoring System(K073030)
Sample SourceAST(Alternative site testing)Capillary whole blood
Hematocrit Range20-60%30-55%
Dimensions87mmx 53mmx 9.9mm82mmx 59mmx 20mm
Memory Capabilities500 times with time and date350 times with time and date
displayingdisplaying
Power SourceDC 3V (CR2032)DC 3V (2 AAA)
Qualified Test StripAGS-1100 Single Test strip forAGS-600 Test Strip
AG-6081 Single BGMS
and AGS-1100 MULTI Test
Strip for AG-6081MULTI
BGMS
Sample VolumeMinimum 0.7 micro literMinimum 1 micro liter
Other functionUSB functionN/A
Intended UseTo quantitatively measure glucose.To quantitatively measure glucose in fresh capillary whole blood.
Sample ApplicationBlood sample is placed directly to the test strip after finger or other part of the body is lanced.Blood sample is placed directly to the test strip after finger is lanced.
Similarities
CHARACTERISTICSNEW DEVICE:AG-6951 Single andAG-6951 MULTI BloodGlucose Monitoring SystemPREDICATE:AG-606 Blood GlucoseMonitoring System(K073030)
Detection MethodAmperometryAmperometry
EnzymeGlucose OxidaseGlucose Oxidase
Type of MeterBiosensor (Electrode)Biosensor (Electrode)
Operating Temperature10℃~40℃(50°-104°F)10℃~40℃(50°-104°F)
DisplayLCDLCD
Result Presentationmg/dL or mmol/Lmg/dL or mmol/L
Test StartAutomaticAutomatic
Test Time5 second5 second
Power SourceDC 3V (2 AAA)DC 3V (2 AAA)
Measurement Range20mg/dL-600mg/dL(1.1mmol/L~33.3mmol/L)20mg/dL-600mg/dL(1.1mmol/L~33.3mmol/L)
Differences
CHARACTERISTICSNEW DEVICE:AG-6951 Blood GlucoseMonitoring SystemPREDICATE:AG-606 Blood GlucoseMonitoring System(K073030)
Intended UseTo quantitatively measureglucose.To quantitatively measureglucose in fresh capillarywhole blood.
Sample SourceAST(Alternative site testing)Capillary whole blood
Sample ApplicationBlood sample is placeddirectly to the test strip afterfinger or other part of the bodyis lanced.Blood sample is placeddirectly to the test strip afterfinger is lanced.
Hematocrit Range20-60%30-55%
Dimensions52mmx 92mmx 21mm82mmx 59mmx 20mm
Memory Capabilities500 times with time and datedisplaying350 times with time and datedisplaying
Battery LifeApprox. 500 normal testsApprox. 1000 normal tests
Qualified Test StripAGS-1100 Single Test strip forAG-6951 Single BGMSAGS-600 Test Strip
and AGS-1100 MULTI TestStrip for AG-6951 MULTIBGMS
Sample VolumeMinimum 0.7 microliterMinimum 1 microliter
Other functionVoice functionN/A

{6}------------------------------------------------

柯 顿(天 津)电 子 医 疗 器 械 有 限 公 司

Andon Medical Co., Ltd.

地址:天津市,天津空港经济开发区,航宇路 26 号 Add: No.26 HangYu Road, Tianjin Airport Economic Area, Tianjin, P.R. China 邮编(P.C.): 300381

传真(Fax): 86-22-6052 6162 电话(Tel): 86-22-87612426 E-mail: andonmedical@yahoo.com.cn

{7}------------------------------------------------

柯 顿 (天 津) 电 子 医 疗 器 械 有 限 公 司

Andon Medical Co., Ltd. 地址:天津市,天津空港经济开发区,航宇路 26 号

Add: No.26 HangYu Road, Tianjin Airport Economic Area, Tianjin, P.R. China 邮编(P.C.): 300381 电话(Tel): 86-22-87612426 E-mail: andonmedical@yahoo.com.cn

7.2 AG-6951 Single and AG-6951 MULTI BGMS

{8}------------------------------------------------

柯 顿 (天 津) 电 子 医 疗 器 械 有 限 公 司

Andon Medical Co., Ltd.

地址:天津市,天津空港经济开发区,航宇路 26 号 Add: No.26 HangYu Road, Tianjin Airport Economic Area, Tianjin, P.R. China 邮编(P.C.): 300381

电话(Tel): 86-22-87612426 传真 (Fax): 86-22-6052 6162 E-mail: andonmedical@yahoo.com.cn

{9}------------------------------------------------

Additional information for K101307 柯 顿 (天 津) 电 子 医 疗 器 械 有 限 公 司 Andon Medical Co., Ltd. 地址:天津市,天津空港经济开发区,航宇路26号 Add: No.26 HangYu Road, Tianjin Airport Economic Area, Tianjin, P.R. China 邮编(P.C.): 300381 传真(Fax): 86-22-6052 6162 电话(Tel): 86-22-87612426 E-mail: andonmedical@yahoo.com.cn

8.0 Performance summary

AG-6081 Single BGMS, AG-6081 MULTI BGMS, AG-6951 Single BGMS and AG-6951 MULTI BGMS conform to the following standards:

  • · ISO 15197: In vitro diagnostic test systems- Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
  • · FDA Draft Guidance Document-Total Product Life Cycle for Portable Invasive Blood Glucose Monitoring Systems: October 24, 2006
  • · CLSI/NCCLS Guideline, EP07-A2 Interference Testing in Clinical Chemistry; Approved Guideline-Second edition

9.0 Comparison to the predict device and the conclusion

The four blood glucose monitor AG-6081 Single, AG-6081 MULTI, AG-6951 Single and AG-6951 MULTI are very similar with the predicate device AG-606. Their appearance is different from AG-606, they use the different test strips, and the intended use of AG-6081 MULTI and AG-6951 MULTI Blood Glucose Monitoring Systems are also different from their predicate device AG-606. Besides, the four new devices can test the blood glucose at the alternative site other than the finger, and they have the USB function as well. More over, AG-6951 Single and AG-6951 MULTI BGMS have a new voice function.

However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is an emblem of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Andon Medical Co.. Ltd. c/o Yi Liu No. 04-23-3 Airport Industrial Park Tianjin, China 300381

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

AUG 0 9 2011

Re: K101307

Trade name: AG-6081 Single Blood Glucose Monitoring System AG-6081 Multi Blood Glucose Monitoring System AG-6951 Single Blood Glucose Monitoring System AG-6951 Multi Blood Glucose Monitoring System AGS-1100 Control Solution Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System

Regulatory Class: Class II

Product Code: NBW. CGA. JJX Dated: July 25, 2011

Received: July 28, 2011

Dear Yi Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{11}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to paremarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

C.C.

Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

510(k) Number (if known): K 101307

Device Name: AG-6081 Single Blood Glucose Monitoring System

Indication for Use:

The AG-6081 Sinale Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6081 Single Blood Glucose Monitoring System is to be used by a single person and should not be shared.

The AG-6081 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-6081 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 Single Blood Glucose Test Strips are for use with the AG-6081 Single Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6081 Single meters and AGS-1100 Single test strips.

Prescription Use J J (Part 21 CFR 801 Subpart D) And/Or

Over the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device (OIVD)

Rita charles

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) k 101307

t Page 1 of

{13}------------------------------------------------

510(k) Number (if known): K /0 / 30'7

Device Name: AG-6081 MULTI Blood Glucose Monitoring System

Indication For Use:

The AG-6081 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6081 MULTI Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of a diabetes control program.

The AG-6081 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, upper arm, calf and thigh should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 MULTI Blood Glucose Test Strips are for use with the AG-6081 MULTI Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6081 MULTI meters and AGS-1100 MULTI test strips.

Prescription Use And/Or (Part 21 CFR 801 Subpart D)

Over the Counter Use √ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device (OIVD)

Ruth Chesler

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety k 10 1307 510(k)

Page: 1 of

{14}------------------------------------------------

K101307 510(k) Number (if known):

Device Name: AG-6951 Single Blood Glucose Monitoring System

Indication For Use:

The AG-6951 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6951 Single Blood Glucose Monitoring System is to be used by a single person and should not be shared.

The AG-6951 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-6951 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady – state times (when glucose is not changing rapidly).

The AGS-1100 Single Blood Glucose Test Strips are for use with the AG-6951 Single Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6951 Single meters and AGS-1100 Single test strips.

This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a glucose test.

Over the Counter Use √ Prescription Use And/Or (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device (OIVD) Suite Chiller

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

101307 510(k)

Page 3 of 4

{15}------------------------------------------------

510(k) Number (if known):

Device Name: AG-6951 MULTI Blood Glucose Monitoring System

Indication For Use:

The AG-6951 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf and thigh. The AG-6951 MULTI Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of a diabetes control program.

The AG-6951 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, upper arm, calf and thigh should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-1100 MULTI Blood Glucose Test Strips are for use with the AG-6951 MULTI Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AGS-1100 Glucose Controls are intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the AG-6951 MULTI meters and AGS-1100 MULTI test strips.

This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a glucose test.

Prescription Use Over the Counter Use And/Or (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device (OIVD)

Ruta Club

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K101307

Page 4 of

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.