K Number
K101200

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-09-30

(154 days)

Product Code
Regulation Number
876.1300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The disposable PillCam® Express™ video capsule delivery device is indicated for the transendoscopic delivery of the PillCam® SB video capsule in patients age 8 and above who are either unable to ingest the PillCam capsule or are known to have slow gastric emptying time.

Device Description

The PillCam® Express Capsule Endoscopy Delivery System is comprised of three parts:

  • Catheter
  • Syringe
  • Capsule holder

The catheter is passed through the accessory channel of a standard endoscope and the capsule holder is snapped on to the distal end of the device.

The endoscope is used to guide the Capsule Endoscopy Delivery System to the proximal duodenum. The capsule is then released pneumatically, using an air-filled syringe attached to the proximal end of the catheter. Once the capsule is deployed, the endoscope is withdrawn and the capsule holder is cut off at the distal part of the device with scissors. The catheter is then retracted and discarded.

The PillCam® Express Capsule Endoscopy Delivery System is a single-use, disposable and latex-free product.

AI/ML Overview

The provided 510(k) summary (K101200) describes the PillCam® Express™ Video Capsule Delivery Device. Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria in a typical "pass/fail" format for the clinical study. Instead, it describes the outcome that serves as proof of acceptable performance.

Acceptance Criterion (Implicit)Reported Device Performance
Successful transendoscopic delivery of the PillCam® SB capsuleAll capsules were successfully placed into the duodenum (100% success rate).
Device safety (absence of serious adverse events)No serious adverse events were reported in the clinical study.
Strength of mechanical componentsNon-clinical bench testing included strength testing of the cup to the catheter, force to release the capsule, force to maintain the capsule in the cup, and strength of the catheter. (Implied successful completion, no specific metrics provided.)

2. Sample Size for the Test Set and Data Provenance

  • Sample Size for Test Set: Twenty-five (25) patients.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical study data, nor does it explicitly classify it as retrospective or prospective. However, the nature of a clinical validation study typically implies a prospective design for gathering new data. Given the submitter's location (Israel), the study likely originated there or in a collaborating country.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not provided in the document. The study focuses on the mechanical delivery of the capsule and patient safety, not on diagnostic accuracy requiring expert interpretation of images to establish ground truth for algorithm performance.

4. Adjudication Method for the Test Set

This information is not applicable/provided. The study evaluates successful delivery and adverse events, not diagnostic findings that typically require adjudication of interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The study described is a clinical validation of a delivery device, not an imaging device that requires human interpretation or comparison of AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

No, a standalone (algorithm only) study was not done. The device is a mechanical delivery system for a capsule endoscope, not an AI algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the clinical validation study was based on:

  • Direct observation of successful capsule placement: Clinicians would confirm the capsule's release into the duodenum during the procedure.
  • Clinical assessment of adverse events: Monitoring patients for any complications related to the device.
  • Bench testing results: Measurements of mechanical strength and forces for various components.

8. The Sample Size for the Training Set

This information is not applicable/provided. As the device is a mechanical delivery system and not an AI algorithm, there is no "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/provided for the reasons stated above.

{0}------------------------------------------------

K101200

Page 1 of 3

GlEN
IMAGING

Given Imaging Limited New Industrial Park PO Box 258, Yoqneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466

510(k) Summary

SEP 30 2010

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.

Submitter Name and Address:Given Imaging Ltd.Hermon BuildingNew Industrial ParkPO Box 258Yoqneam 20692IsraelTel.: 011-972-4-9097730Fax: 011-972-4-9938060
Contact Person:Tim Thomas RACVice PresidentRegulatory Affairs & Quality Assurance
Email: tim.thomas@givenimaging.com.
Phone Number:770-662-0870 ext. 1006
Fax Number:770-662-0510
Establishment Registration Number:9710107
Date Prepared:April 27, 2010
Device Trade Name(s):PillCam® Express™ Video Capsule Delivery Device
Device Common Name:Ingestible telemetric gastrointestinal capsule imaging system& Endoscope and/or accessories
Classification:Regulation No: 21 CFR 876.1300 & 876.1500Class: IIReview panel - Gastroenterology/UrologyNEZ - System, Imaging, Gastrointestinal, Wireless, Capsuleand KOG - Endoscope and/or accessories
Predicate Device(s):Capsule endoscopy delivery device (K040494)

Given Imaging Ltd. New Industrial Park. PO Box 258, Yoqncam 20692, I-racl

{1}------------------------------------------------

K101200 Page 2 of 3

Image /page/1/Picture/1 description: The image shows the logo for GI-EN Imaging. The logo is in black and white and features the words "GI-EN" in large, bold letters. Below the letters is an image of a medical device. To the right of the image is the word "IMAGING" in smaller letters.

Given Imaging Limited New Industrial Park PO Box 258, Yogneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466

General Device Description:

The PillCam® Express Capsule Endoscopy Delivery System is comprised of three parts:

  • Catheter
  • 트 Syringe
  • Capsule holder

The catheter is passed through the accessory channel of a standard endoscope and the capsule holder is snapped on to the distal end of the device.

The endoscope is used to guide the Capsule Endoscopy Delivery System to the proximal duodenum. The capsule is then released pneumatically, using an air-filled syringe attached to the proximal end of the catheter. Once the capsule is deployed, the endoscope is withdrawn and the capsule holder is cut off at the distal part of the device with scissors. The catheter is then retracted and discarded.

The PillCam® Express Capsule Endoscopy Delivery System is a single-use, disposable and latex-free product.

Indications for Use: Intended Use: The disposable PillCam® Express™ video capsule delivery device is indicated for the transendoscopic delivery of the PillCam® SB video capsule in patients age 8 and above who are either unable to ingest the PillCam capsule or are known to have slow gastric emptying time The technology characteristics are similar as the predicate Technological device. Characteristics: Performance data demonstrating safety and effectiveness of Performance Data: the PillCam® Express™ video capsule delivery device was obtained from non-clinical bench testing and from a clinical validation study. The non-clinical bench testing included strength testing of the cup to the catheter, force to release the capsule, force to maintain the capsule in the cup, and strength of the catheter. The clinical validation study included twenty five patients in the age range of 22-89 years old. All capsules were successfully placed into the duodenum and no serious adverse events were reported in the clinical study. Based on the technological characteristics and clinical Conclusion: performance of the device, Given Imaging Ltd. believes that

Given Imaging Ltd. New Industrial Park. P() Box 258, Yoqncam 20692, Israel

{2}------------------------------------------------

K101200

Page 3 of 3

Given Imaging Limited

New Industrial Park .PO Box 258, Yoqneam 20692 İsrael Voice 972 4 909 7777 Fax 972 4 959 2466

Given PillCam® Express™ video capsule delivery device and the predicate capsule delivery device selected are substantially equivalent and do not raise new issues of safety or effectiveness.

Image /page/2/Picture/5 description: The image shows the address of Given Imaging Ltd. The address is New Industrial Park, PO Box 258, Yoqneam 20692, Israel. The text is black and the background is white. The text is centered on the page.

Image /page/2/Picture/6 description: The image shows the logo for Glen Imaging. The logo is in black and white and features the words "Glen Imaging" in a stylized font. The word "Glen" is on the top line, and the word "Imaging" is on the bottom line. There is a graphic of a test tube between the two words.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol on the right side. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Mr. Tim Thomas VP Regulatory Affairs and Quality Assurance Given Imaging Limited Hermon Building, New Industrial Park P.O. Box 258 Yoqneam 20692 ISRAEL

SEP 3 0 2010

Re: K101200

Trade/Device Name: PillCam® Express™ Video Capsule Delivery Device Regulation Number: 21 CFR §876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: II Product Code: NEZ Dated: September 20, 2010 Received: September 21, 2010

Dear Mr. Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{4}------------------------------------------------

Page 2

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Hubert Lemire MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known):

Device Name: PillCam® Express™ Video Capsule Delivery Device

SEP 3 0 2010

Indications for Use:

The disposable PillCam® Express™ video capsule delivery device is indicated for the transendoscopic delivery of the PillCam® SB video capsule in patients age 8 and above who are either unable to ingest the PillCam capsule or are known to have slow gastric emptying time.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)


(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K10/200

Given Imaging Ltd. Traditional 510(k) Submission PillCam® Express™ Video Capsule Delivery Device

April 27, 2010

§ 876.1300 Ingestible telemetric gastrointestinal capsule imaging system.

(a)
Identification. An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.(b)
Classification. Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”