K Number
K101194

Validate with FDA (Live)

Date Cleared
2011-07-11

(439 days)

Product Code
Regulation Number
880.5725
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Navigator™ Delivery System (Navigator DS) is intended for use in the delivery of medication and/or fluids in a controlled manner. The Navigator DS is indicated for use in the intermittent delivery of medications and other fluids in intra-articular applications.

Device Description

The device consists of a fluid delivery module (motor driven piston syringe pump), a daily disposable cassette, a per-patient disposable handpiece and tubing set. and a wired foot pedal. The fluid delivery module delivers the medications or fluids from off-the-shelf vials via aspirated dispense. The user interface is via touch screen that allows the user to define per-delivery treatment volumes and regimens of off-the-shelf medications or fluids in accordance with their labeled indication and proposed dosage. Delivery of the user-defined regimen occurs through depression of the foot pedal that actuates dispense of the fluids through three individual fluid lines within the cassette. The fluid is then transported through the per-patient tubing and handpiece set where off-the-shelf needles are attached.

AI/ML Overview

The provided text describes the Navigator™ Delivery System (Navigator DS), an infusion pump, and its regulatory submission. It does not contain information about acceptance criteria and a study proving the device meets these criteria in the typical format of a clinical or performance study with specific metrics and results. The document focuses on regulatory approval based on substantial equivalence to a predicate device and various engineering and safety tests.

Therefore, most of the requested information cannot be extracted from this text.

Here's what can be gathered based on the provided input:

  1. A table of acceptance criteria and the reported device performance

    • The document states: "The test results demonstrated that the Navigator DS functions as designed and can be operated by the user as intended through the user interface and instructions provided. The results demonstrate that the Navigator DS is as safe and effective and performs as well as the predicate device."
    • Specific quantitative acceptance criteria (e.g., flow rate accuracy within X%, volume accuracy within Y%) and corresponding reported performance values are not provided in this summary.
    • The tests mentioned include:
      • Characterization of system functionality over viscosity, temperature, and pressure extremes.
      • Flow rate characterization.
      • Accuracy of volumes dispensed.
      • Flow profiles and rate accuracy.
      • Occlusion testing.
      • Integrity of fluid pathway components.
      • Software verification and validation.
      • Design verification and validation.
      • Sterilization validation.
      • Microbial ingress testing.
      • Viral ingress testing.
      • Dye ingress testing.
      • Cleaning validation.
      • Disinfection validation.
      • Biocompatibility.
      • IPA residuals assessment.
      • Chemical compatibility and extractables/leachables testing.
      • Simulated use/human factors studies.
      • Electrical safety testing.
      • Shipping validations.
  2. Sample sized used for the test set and the data provenance

    • The document mentions "simulated use study of human factors" and "performance testing" but does not specify sample sizes for any of these tests.
    • Data provenance (e.g., country of origin, retrospective/prospective) is not mentioned.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Ground truth or expert review in the context of device performance metrics is not explicitly discussed. The "simulated use study" involved "intended users," but their qualifications for establishing a "ground truth" are not detailed.
  4. Adjudication method for the test set

    • An adjudication method (e.g., 2+1, 3+1) is not mentioned as this type of study was not conducted.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • An MRMC study is relevant for imaging devices or AI-assisted diagnostic tools. The Navigator DS is an infusion pump. Therefore, an MRMC comparative effectiveness study was not conducted, and AI assistance is not applicable to this device's stated function.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is typical for AI/algorithmic devices. The Navigator DS is a mechanical infusion pump with software control. Its "standalone performance" is implicitly covered by the various engineering and functional tests mentioned (flow rate accuracy, volume dispensed, etc.), which assess the device's inherent capabilities separate from a human operator's specific actions, although the human factors study evaluates device operation by users.
  7. The type of ground truth used

    • For an infusion pump, "ground truth" would typically refer to established engineering standards, physical measurements (e.g., actual vs. desired fluid volume, actual vs. desired flow rate), and safety specifications. The document implies these were used as the basis for performance evaluation, stating tests were "carried out to meet all elements of FDA's Draft Guidance 'Total Product Life Cycle: Infusion Pump - Premarket Notification [510(k)] Submissions'." However, specific details are not provided.
  8. The sample size for the training set

    • The concept of a "training set" is relevant for machine learning models. This device is an electromechanical infusion pump, not an AI/ML device. Therefore, a "training set" in this context is not applicable. The device's software was subject to "software verification and validation," which involves testing against requirements, but not "training" in the ML sense.
  9. How the ground truth for the training set was established

    • As a "training set" is not applicable, the establishment of its ground truth is also not applicable.

{0}------------------------------------------------

K101194/S002 Additional Information on Navigator DS

K101194
JUL 11 2011

510(k) Summary

SUBMITTER'S INFORMATION

Owner:Carticept Medical, Inc.
Address:6120 Windward Parkway, Suite 220, Alpharetta, GA 30005
Phone:770-754-3800
Fax Numbers:770-754-3808
Contact Person:Tanya Eberle, Director, Regulatory Affairs
Date Summary Prepared:May 23, 2011

DEVICE INFORMATION

Name of Device:Navigator™ Delivery System (Navigator DS)
Common/Usual Name:Infusion Pump, External
Classification Name:Infusion Pump, Class II, 21 CFR 880.5725 (Product Code FRN)
Predicate Device(s):Milestone Scientific's CompuFlo™ Infusion Pump (K053554)
Device Description:The device consists of a fluid delivery module (motor drivenpiston syringe pump), a daily disposable cassette, a per-patientdisposable handpiece and tubing set. and a wired foot pedal.The fluid delivery module delivers the medications or fluidsfrom off-the-shelf vials via aspirated dispense. The userinterface is via touch screen that allows the user to define per-delivery treatment volumes and regimens of off-the-shelfmedications or fluids in accordance with their labeled indicationand proposed dosage. Delivery of the user-defined regimenoccurs through depression of the foot pedal that actuatesdispense of the fluids through three individual fluid lines withinthe cassette. The fluid is then transported through the per-patient tubing and handpiece set where off-the-shelf needlesare attached.
Indication for Use:The Navigator™ Delivery System (Navigator DS) is intendedfor use in the delivery of medication and/or fluids in a controlledmanner. The Navigator DS is indicated for use in theintermittent delivery of medications and other fluids in intra-articular applications.
TechnologicalCharacteristics:The Navigator DS fluid delivery module is a motor-driven pistonsyringe pump that consists of three separate drives andindependently calibrated force sensors. The fluid deliverymodule is a software-driven, microprocessor controlledelectromechanical system that meters fluids through anadministration set to the patient. Delivery volumes and rates

Confidential and Proprietary to Carticept Medical

Page 1-169

June 2011

.

{1}------------------------------------------------

K101194/S002 Additional Information on Navigator DS

are programmed by the operator on an LCD touch screen display.

All aspects of the fluid delivery module are contained within a plastic housing that receives the system-dedicated, sterile, disposable cassette. Once installed, the cassette is the point of attachment for the sterile, per-patient tubing and handpiece set. The wired foot pedal provides control of delivery. The system operates with user-provided disposable supplies, such as off-the-shelf vials and off-the-shelf needles.

The Navigator DS is substantially equivalent to the Milestone CompuFlo Infusion Pump (K053554). The Navigator DS raises no new questions of safety or effectiveness as compared to the CompuFlo infusion pump. The Navigator DS has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Navigator DS and its predicate devices raise no new issues of safety or effectiveness.

Testing of the Navigator DS device was carried out to meet all elements of FDA's Draft Guidance "Total Product Life Cycle: Infusion Pump - Premarket Notification [5 10(k)] Submissions", including an Assurance Case Report and Hazard Analysis. A Risk Management Report demonstrated an acceptable risk profile based on design-based risk mitigation and satisfactory performance testing, including characterization of the system functionality over viscosity, temperature, and pressure extremes, flow rate characterization, accuracy of volumes dispensed, flow profiles and rate accuracy, occlusion testing, integrity of fluid pathway components, and software verification and validation. Additional safety evaluations related to performance included design verification and validation, sterilization validation, software validation, microbial ingress testing, viral ingress testing, dye ingress testing, cleaning validation, disinfection validation, biocompatibility, IPA residuals assessment, chemical compatibility and extractables/leachables testing with varying pH and polarity, simulated use/human factors studies, electrical safety testing, and shipping validations. A Clinical Evaluation was determined not to be required for the Navigator DS. A simulated use study of human factors was conducted with intended users in the intended use environment that evaluated device performance, possible use error and user perception of difficulties with pump use. The study assessed the critical tasks or use scenarios where use related errors are most likely to occur.

The test results demonstrated that the Navigator DS functions as designed and can be operated by the user as intended through the user interface and instructions provided. The results demonstrate that the Navigator DS is as safe and effective and performs as well as the predicate device.

Comparison to Predicate Devices:

Performance Data:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and head. The bird is positioned to the right of the text, which is arranged in a circular pattern around the bird. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Ms. Tanya Eberle Director of Regulatory Affairs Carticept Medical, Inc. 120 Windward Parkway. Suite 220 Alpharetta, Georgia 30005

JUL_ 1 1 2011

Re: K101194

Trade/Device Name: Navigator Delivery System (or Navigator DS) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: June 2, 2011 Received: June 3, 2011

Dear Ms. Eberle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Eberle

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Anthen ilman

Anthony D. Watson. B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K101194/S002 Additional Information on Navigator DS

Indications for Use Statement

510(k) No. (if known):
K161194
Device Name: Navigator™ Delivery System (Navigator DS)
Indications for Use: The Navigator™ Delivery System (Navigator DS) is intended for
use in the delivery of medication and/or fluids in a controlled
manner. The Navigator DS is indicated for use in the intermittent
delivery of medications and other fluids in intra-articular
applications.

LabelsValues
Prescription Use:X
(Part 21 CFR 801 Subpart D)

AND/OR Over-the-Counter Use
(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rill C. Chaye 7/8/11

Page of

Tivision Sign-Off) vision of Anesthesiology, General Hospital fection Control, Dental Devices

10(k) Number: K101194

Page 1-172

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).