K Number
K101153

Validate with FDA (Live)

Date Cleared
2010-12-08

(229 days)

Product Code
Regulation Number
870.4350
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QUADROX-iD Adult diffusion membrane oxygenator is intended for use in an extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide from the blood and regulates the blood temperature. The application duration is limited to 6 hours.

Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.

Device Description

The QUADROX-iD Adult diffusion membrane oxygenator with BIOLINE Coating and with SOFTLINE Coating is a blood-gas exchanger with integrated heat exchanger.

AI/ML Overview

This document is a 510(k) summary for the MAQUET QUADROX-iD Adult diffusion membrane oxygenator. It outlines the device's technical comparison to predicate devices and the non-clinical testing performed to establish substantial equivalence.

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state acceptance criteria in a quantitative format with corresponding reported device performance values. Instead, it refers to compliance with established international standards for medical devices.

Acceptance Criteria CategoryReported Device Performance (Compliance)
IntegrityTested and / or evaluated
PerformanceTested and / or evaluated
BiocompatibilityMet the requirements of ISO 10993-1
SterilityTested and / or evaluated
General Cardiovascular Implants and Artificial Organs StandardsMet the requirements of ISO 7199 (Blood-gas exchangers)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample size used for the test set or the data provenance (country of origin, retrospective/prospective). The testing appears to be non-clinical (laboratory/bench testing) as opposed to human subject trials.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a non-clinical evaluation of a medical device (oxygenator) against performance standards, not an AI/diagnostic device requiring expert consensus for ground truth.

4. Adjudication method for the test set

Not applicable. This is a non-clinical evaluation of a medical device against performance standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (oxygenator) and not an AI or diagnostic imaging device involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a medical device (oxygenator), not an algorithm. The testing described is for the physical device's performance characteristics.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is compliance with established international standards (ISO 10993-1 and ISO 7199) and the performance characteristics demonstrated during testing (e.g., oxygenation, CO2 removal, heat exchange). This is based on objective measurements against engineering and biological standards, not expert consensus or pathology in the diagnostic sense.

8. The sample size for the training set

Not applicable. This is a medical device, not a machine learning model. There is no concept of a "training set" in the context of this 510(k) submission.

9. How the ground truth for the training set was established

Not applicable. See point 8.

{0}------------------------------------------------

MAQUET
GETINGE GROUP

101123

· DEC - 8 2010

510(k) Summarv

[as required by 21 CFR 807.92(c)]

Submitter

MAQUET Cardiopulmonary AG Hechinger Strasse 38 72145 Hirrlingen Germany

Contact Person

Katrin Schwenkglenks Phone: 011 49 7478 921 151 Fax: 011 49 7478 921 8667

Date Prepared

April 19, 2010

Device Trade Name

QUADROX-iD Adult diffusion membrane oxygenator with BIOLINE Coating and with SOFTLINE Coating

Common/Usual Name

Classification Names

Oxygenator with integrated heat exchanger

Cardiopulmonary bypass oxygenator (21 CFR 870.4350 - Product Code: DTZ) Cardiopulmonary bypass heat exchanger (21 CFR 870.4240 - Product Code: DTR)

Legally Marketed Devices

QUADROX-i Adult microporous membrane Oxygenator with SOFTLINE Coating (K082117),

QUADROX-i Adult microporous membrane Oxygenator with BIOLINE Coating (K090511), QUADROX-D diffusion membrane oxygenator with SAFELINE Coating (K061628) and with BIOLINE Coating (K071774).

Page 1 of 3

{1}------------------------------------------------

Device Description

The QUADROX-iD Adult diffusion membrane oxygenator with BIOLINE Coating and with SOFTLINE Coating is a blood-gas exchanger with integrated heat exchanger.

Indications for Use

The QUADROX-iD Adult diffusion membrane oxygenator is intended for use in an extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide from the blood and regulates the blood temperature. The application duration is limited to 6 hours.

Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.

Statement of Technical Comparison

The QUADROX-iD Adult diffusion membrane oxygenator with BIOLINE Coating and with SOFTLINE Coating is comparable to the QUADROX-i Adult micropyrons membrane oxygenator with BIOLINE Coating and with Softline Coating with the only exception that the QUADROX-iD Adult diffusion membrane oxygenator has a diffusive oxygenation membrane instead of a microporous oxygenation membrane. However, the diffusion membrane is the same as with the QUADROX-D diffiusion membrane oxygenator. Besides this difference the QUADROX iD Adult diffusion membrane oxygenator with SOFTLINE Coating and with BIOLINE Coating is comparable in design, method of operation, components, packaging, and fundamental scientific technology as compared to the QUADROX, Adult microporous membrane oxygenator with Softline Coating and with BIOLINE Coating.

Non-clinical Testing

The QUADROX-iD Adult Diffusion Membrane Oxygenator with BIOLINE COATING and SOFLTINE Coating has been tested to and met the requirements of ISO 10993-1 Biological evaluation of medical devices · Padt 1: Evaluation and testing as well as the requirements of ISO 7199 Cardiovascular implants and artificial organs - Bloodgas exchangers (oxygenators).

Determination of Substantial Equivalence

Evaluation and testing on safety and effectiveness was executed to demonstrate that the QUADROX-ID Adult diffusion membrane oxygenator with SOFTLINE Coating and with BIOLINE Coating described in this submission is substantially equivalent to the QUADROX-i Adult microporous membrane Oxygenator with Softline Coating and with BIOLINE Coating.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Maquet Getinge Group. The word "MAQUET" is in large, bold, sans-serif font on the top line. Below that, "GETINGE GROUP" is in a smaller, sans-serif font.

The following areas have been tested and / or evaluated:

  • Integrity .
    • Performance .

.

  • Biocompatibility -
    .

  • . Sterility

Conclusion

.

.

.

.

.

The data given demonstrate that the QUADROX-iD Adult diffusion membrane oxygenator with SOFLTINE Coating and with BIOLINE Coating is substantially equivalent to the named predicate devices which currently hold market clearance.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with healthcare, with three intertwined snakes and a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue i Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Maquet Cardiopulmonary AG c/o Ms. Katrin Schwenkglenks Hechinger Straße 38 D-72145 Hirrlingen, Germany

DEC - 8 2010

Re: K101153

Trade/Device Name: QUADROX-iD Adult diffusion membrane oxygenator with BIOLINE Coating, art. codes: BE-HMOD 70000-USA, BEQ-HMOD 70000-USA; and QUADROX-iD Adult diffusion membrane oxygenator with SOFTLINE Coating, art. codes: HMOD 70000-USA

Regulation Number: 21 CFR 870.4350

Regulation Name: Cardiopulmonary bypass oxygenator

Regulatory Class: II

Product Code: DTZ, DTR

Dated: November 30, 2010

Received: December 2, 2010

Dear Ms. Schwenkglenks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean

{4}------------------------------------------------

Page 2 - Ms. Katrin Schwenkglenks

that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

smma R.V.hmn

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

DEC - 8 2010

510(k) Number (if known): _ K10 |153

Device Name: QUADROX-iD Adult diffusion membrane oxygenator with BIOLINE Coating and with SOFTLINE Coating

Indications for Use:

The QUADROX-iD Adult diffusion membrane oxygenator is intended for use in an extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide from the blood and regulates the blood temperature. The application duration is limited to 6 hours.

Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.

Prescription Use × (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

DMAR R. Vic/m
(Division Sign-Off)

Page 1 of 1

Division of Cardiovascular Devices

510(k) Number K101153

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”