K Number
K101054

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-09-03

(141 days)

Product Code
Regulation Number
N/A
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RomiAPEX™ A-15 is indicated for patients who need to undergo root canal treatment, which requires precise determination of position of the dental file in the canal relatively to the root apex.

Device Description
  • The RomiApex™ A-15 is used to measure the distance to the apex during root canal procedures. A low frequency low volt AC signal is applied between the lip electrode and the endodontic file, which is inserted into the root canal. The impedance of the tissues between the electrodes change as the file advances toward the root apex and the measured signals are used to monitor the progress of the file in the tooth.
  • The RomiApex™ A-15 operates on the measurement of a weak electrical current flowing between two electrodes. One electrode is a metal hook that rests on the patient's lower lip and the other is the endodontic file that is attached to the file clasp and inserted into the canal.
  • The measurements in RomiAPEX™ A-15 are performed utilizing AC signals at two frequencies - 500 Hz and 8 kHz. The frequencies are alternated rather then mixed, as it is done in other apex locators, thus canceling the need for signal filtering and eliminating the noise caused by non-ideal filters.
  • The device consists of a main body incorporating the LCD display, a lip Clip, a file clip a cable and a touch prob.
  • The device is powered by one 1.5 AAA Alkaline battery.
AI/ML Overview

The provided text is a 510(k) summary for the RomiApex™ A-15 Electronic Apex Locator. This type of document is for demonstrating substantial equivalence to a predicate device, not typically for reporting on detailed acceptance criteria and a specific study proving those criteria were met in the same way a de novo device would. Therefore, much of the requested information (e.g., number of experts for ground truth, adjudication method, MRMC study, sample size for training set, how training ground truth was established) is not available in the provided text.

However, I can extract information related to the device's performance characteristics in comparison to its predicate.

1. Table of acceptance criteria and the reported device performance

The document presents a "Substantial Equivalence Comparison Table" (Table 5.6) which compares the RomiApex™ A-15 (Candidate Device) to the predicate device, BINGO-1020, focusing on characteristics rather than specific performance metrics and acceptance criteria in numerical terms. The acceptance criteria, in this context, appear to be that the candidate device's performance characteristics should be equivalent or better than the predicate's.

CharacteristicPredicate Device (Bingo-1020)Candidate Device (RomiApex™ A-15)Equivalence Claim / Performance Relative to Predicate
Device definitionElectronic apex locatorSame as in Bingo-1020Equivalent
Intended UsePrecise apex localization during root canal treatment.Same as in Bingo-1020Equivalent
Indications for useBingo-1020 is a modern device for precise apex localization during root channel treatment. Bingo-1020 is distinguished by increasing precision at 0.1mm on wet/dry, large graphic display where current position of endo file is reflected and other essential information is displayed.RomiAPEX™ A-15 is indicated for patients who need to undergo root canal treatment, which requires precise determination of position of the dental file in the canal relatively to the root apex.Equivalent (description slightly different, but core indication is the same)
Where to be used (clinics, home etc.)This product must only be used in hospital environments, clinics or dental offices by qualified dental personnel.Same as in Bingo-1020Equivalent
Device categoryActive, invasiveSame as in Bingo-1020Equivalent
Power Source2.4V NiMH rechargeable batteries1.5V alkaline primary batteryDifferent, but considered acceptable
External chargerInput: 230V/50-60Hz Output: 6Vdc @ 200mAExternal charger is not requiredDifferent, but considered acceptable
Current ConsumptionMaximum - 30 mA DC.Maximum – 13 mA DC.Lower (better)
Method of calculating location of root canal apexRMS functions of the measured signals at two frequencies are used to calculate the test scores which are compared to statistically predefined thresholds.Same as in Bingo-1020Equivalent
DisplayCustom monochrome LCD DisplayCustom color LCD DisplayDifferent, considered an improvement
ButtonsThree push buttons: 1. On / Off 2. Sound control 3. Demo MODE (training mode)Two push buttons: 1. Same as in Bingo-1020 2. Same as in Bingo-1020 3. "Demo MODE" (for training) is not an offered feature in RomiApex™ A-15.Slightly different feature set, considered acceptable
Sound indicationPiezo transducer with sound level control (high, medium, low, mute).Same as in Bingo-1020Equivalent
Adjustment before measurementNot requiredSame as in Bingo-1020Equivalent
CalibrationNot requiredSame as in Bingo-1020Equivalent
Measuring signal amplitudeNominal - doesn't exceed 25 mV AC. The maximum voltage applied to the patient doesn't exceed 300 mV DC.Same as in Bingo-1020Equivalent
Frequencies used for measurements500 Hz and 8 kHzSame as in Bingo-1020Equivalent
Weight430 Gr100 GrLighter (better)
DimensionsW160 x H95 x T35 mmW55 x H92 x T16mmSmaller (better)
Files to be used with the deviceAny kind of standard dental files may be usedSame as in Bingo-1020Equivalent
Safety featuresThe same connector on the device is used both for measuring cable connection and for battery charging. This safety feature prevents simultaneous connection of the device to the patient and to the mains via external charger.No external charger should be connected to the device. And due that such safety feature is not needed in RomiApex™ A-15 at all. A new type of connector is used - Micro USB. This type of connector was specially designed to provide particularly high mechanical strength and is guaranteed by the manufacturer for 10,000 mating cycles.Different (no external charger needed, improved connector strength)
Automatic on/off switchAutomatic turn-off after 5 min. of the idle state.Same as in Bingo-1020Equivalent

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a separate "test set" or provide details on sample size, data provenance, or whether it was retrospective or prospective. The submission is for substantial equivalence, which primarily relies on comparing device specifications and intended use to a legally marketed predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not available in the provided text.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not available in the provided text.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not available in the provided text. Electronic Apex Locators are not AI systems in the context of "human readers" interpreting data.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The RomiApex™ A-15 is a device for measuring the distance to the root apex during root canal procedures, it's not an "algorithm only" device. Its performance is measured directly by its ability to accurately determine the apex, which is inherent in its design and comparison to the predicate. No specific "standalone" performance study in the context of an AI algorithm is mentioned.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document does not explicitly state the type of ground truth used. For apex locators, "ground truth" for accuracy would typically be established by direct measurement with radiography or physical measurement after tooth extraction, compared to the device's reading. However, this level of detail for a performance study is not included in the 510(k) summary provided. The comparison relies on the predicate's established performance and the candidate device's similar scientific technology.

8. The sample size for the training set

This information is not available in the provided text, as this type of device does not have a "training set" in the context of machine learning. Its operation is based on established electrical impedance principles.

9. How the ground truth for the training set was established

This information is not available in the provided text, for same reasons as point 8.

{0}------------------------------------------------

K101054

RomidanLtd.
premium quality products

  1. Simcha Holzberg St., 55022 Kiryat Ono, Israel.

图 +972-3-7360099 export@romidan.com 록 +972-3-7360097 Web: www.romidan.com

Document No .: 05.00

"510(k) Summary" Section 5.:

C

SEP 3 · 2010

RomiApex™ A-15, Electronic Apex Locator

The following 510(k) Summary of Safety and Effectiveness has been prepared pursuant to requirements for 510(k) summaries specified in 21 CFR § 807.92(a)

Section's content

  • 807.92(a)(1) Owner & Submitter's Details 5.1.
  • 807.92(a)(2) Candidate Device Details. 5.2.
  • 5.3. 807.92(a)(3) - Predicate Device Identification:
  • 5.4. 807.92(a)(4) - Device Description:
  • 5.5. 807.92(a)(5) - Intended Use
  • 5.6. 807.92(a)(6) - Substantial Equivalence Comparison Table:
  • 807.92(b)(3) Conclusions 5.7.

5.1 Owner &Submitter's Details: [ 807.92(a)(1) ]

Owner & Submitter Name:Romidan LTD
Address:5 Simcha Holzberg St., 55022 Kiryat Ono, Israel
Phone:+972-3-7360099
Fax number:+972-3-7360097
E-maileitan@romidan.com
Name of Contact Person:Mr. Eitan Margalit

Establishment Registration No.: 3003518307

Date prepared:

July 18t, 2010

US Agent:

15.00 100701-6 51000 Summenr of Sulaty and Effe

NORMAN F.ESTRIN, PH.D., ESTRIN CONSULTING GROUP, INC. (ECG) 9109 Copenhaver Drive, Potomac, MD 20854 Phone: +001-301-279-2899 +001-301-294-0126 Fax: estrin@yourfdaconsultant.com Email:

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Romidan Ltd. The logo features a diamond shape on the left, followed by the company name in bold, sans-serif font. Below the company name, the words "premium quality products" are written in a smaller, italicized font.

5, Simcha Holzberg St., 55022 Kiryat Ono, Israel. 粵 +972-3-7360097 @ +972-3-7360099 export@romidan.com Web: www.romidan.com

Document No .: 05.00

5.2 Candidate Device Details:[ 807.92(a)(2) ]
Trade Name:RomiApex™ A15 – Electronic Apex Locator
Common Name:Apex Locator
Classification Name:Locator, Root Apex
Product Code:LQY – Locator, ROOT APEX
Review Panel:Dental
Device Class:Unclassified

5.3 Predicate Device Identification: [ 807.92(a)(3) ]

Devices to which substantial equivalence is claimed:

Table 5.3: Predicate Device Identification

Predicate MedicalDevice NameApplicant Name510(k)Number(*)
BINGO-1020DENT CORP, Research & DevelopmentK992233

(4) See "510(k) Premarket Notification Database Search for more details.

Additional Substantial Equivalence Information is provided in the attached Substantial Equivalence Comparison Table. (see Para 807.92(a)(6)).

5.4 Device Description: [ 807.92(a)(4) ]

  • The RomiApex™ A-15 is used to measure the distance to the apex during root � canal procedures. A low frequency low volt AC signal is applied between the lip electrode and the endodontic file, which is inserted into the root canal. The impedance of the tissues between the electrodes change as the file advances toward the root apex and the measured signals are used to monitor the progress of the file in the tooth.
  • The RomiApex™ A-15 operates on the measurement of a weak electrical current . flowing between two electrodes. One electrode is a metal hook that rests on the patient's lower lip and the other is the endodontic file that is attached to the file clasp and inserted into the canal.
  • The measurements in RomiAPEX™ A-15 are performed utilizing AC signals at . two frequencies - 500 Hz and 8 kHz. The frequencies are alternated rather then mixed, as it is done in other apex locators, thus canceling the need for signal filtering and eliminating the noise caused by non-ideal filters.
  • The device consists of a main body incorporating the LCD display, a lip Clip, a file . clip a cable and a touch prob.
  • The device is powered by one 1.5 AAA Alkaline battery. .

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo for Romidan LTD. The logo features a diamond shape on the left, followed by the word "Romidan" in bold, sans-serif font. To the right of "Romidan" is "LTD." in a smaller font size. Below "Romidan" is the phrase "premium quality products" in a smaller, italicized font, underlined with a solid line.

5, Simcha Holzberg St., 55022 Kiryat Ono, Israel. 图 +972-3-7360099 톰 +972-3-7360097

export@romidan.com (a)

Web: www.romidan.com

Document No .: 05.00

5.5 Intended Use : [ 807.92(a)(5) ]

RomiAPEX™ A-15 is indicated for patients who need to undergo root canal treatment, which requires precise determination of position of the dental file in the canal relatively to the root apex.

5.6 Substantial Equivalence Comparison Table: [ 807.92(a)(6) ]

LineNo.DeviceCharacteristicsPredicate DeviceBingo-1020Candidate DeviceRomiApex™ A-15
1.Device definitionElectronic apex locatorSame as in Bingo-1020
2.Intended UsePrecise apex localization duringroot canal treatment.Same as in Bingo-1020
3.Indications for useBingo-1020 is a modern devicefor precise apex localizationduring root channel treatment.Bingo-1020 is distinguished byincreasing precision at 0.1mmon wet/dry, large graphicdisplay where current positionof endo file is reflected andother essential information isdisplayed.RomiAPEX™ A-15 is indicatedfor patients who need toundergo root canal treatment,which requires precisedetermination of position of thedental file in the canal relativelyto the root apex.
4.Where to be used(clinics, home etc.)This product must only be usedin hospital environments, clinicsor dental offices by qualifieddental personnel.Same as in Bingo-1020
5.Device categoryActive, invasiveSame as in Bingo-1020
6.Power Source2.4V NiMH rechargeablebatteries1.5V alkaline primary battery
7.External chargerInput: 230V/50-60HzOutput: 6Vdc @ 200mAExternal charger is not required
8.Current ConsumptionMaximum - 30 mA DC.Maximum – 13 mA DC.
9.Method of calculatinglocation of root canalapexRMS functions of the measuredsignals at two frequencies areused to calculate the testscores which are compared tostatistically predefinedthresholds.Same as in Bingo-1020
10.DisplayCustom monochrome LCDDisplayCustom color LCD Display
LineDevicePredicate DeviceCandidate Device
No.CharacteristicsBingo-1020RomiApex™ A-15
11.ButtonsThree push buttons:1. On / Off2. Sound control3. Demo MODE(training mode)Two push buttons:1. Same as in Bingo-10202. Same as in Bingo-10203. "Demo MODE" (for training)is not an offered feature inRomiApex™ A-15.
12.Sound indicationPiezo transducer with soundlevel control (high, medium,low, mute).Same as in Bingo-1020
13.Adjustment beforemeasurementNot requiredSame as in Bingo-1020
14.CalibrationNot requiredSame as in Bingo-1020
15a.Measuring signalNominal - doesn't exceed 25mV AC.Same as in Bingo-1020
15.bamplitudeThe maximum voltage appliedto the patient doesn't exceed300 mV DC.Same as in Bingo-1020
16.Frequencies used formeasurements500 Hz and 8 kHzSame as in Bingo-1020
17.Weight430 Gr100 Gr
18.DimensionsW160 x H95 x T35 mmW55 x H92 x T16mm
19.Files to be used withthe deviceAny kind of standarddentalfiles may be usedSame as in Bingo-1020
20a.Safety featuresThe same connector on thedevice is used both formeasuring cable connectionand for battery charging.This safety feature preventssimultaneous connection of thedevice to the patient and to themains via external charger.No external charger should beconnected to the device. Anddue that such safety feature isnot needed in RomiApex™ A-15 at all.
20b.Safety features(Continue)The type of connector used -Mini DINA new type of connector isused - Micro USB.This type of connector wasspecially designed to provideparticularly high mechanicalstrength and is guaranteed bythe manufacturer for 10,000mating cycles.
21.Automatic on/offswitchAutomatic turn-off after 5 min.of the idle state.Same as in Bingo-1020

Table 5.6: Substantial Equivalence Comparison

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for Romidan Ltd. The logo features a diamond shape on the left, followed by the word "Romidan" in bold, and "Ltd." in smaller font. Below the company name, the words "premium quality products" are printed in a smaller font with a line underneath.

5, Simcha Holzberg St., 55022 Kiryat Ono, Israel.

图 +972-3-7360099 | | |

惠 +972-3-7360097

Document No.: 05.00

@ export@romidan.com

Web: www.romidan.com

End of Table 5.6

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Romidan Ltd. The logo features a diamond shape on the left, followed by the company name in bold, black letters. Below the company name, the words "premium quality products" are written in a smaller font.

5, Simcha Holzberg St., 55022 Kiryat Ono, Israel.

图 +972-3-7360099

g

export@romidan.com

鳥 +972-3-7360097 Web: www.romidan.com

Document No .: 05.00

5.7 Conclusions: [ 807.92(b)(3) ]

  • Romidan's RomiApex™ A-15, subject of this submission, constitutes a safe, . reliable, and effective medical device, meeting all the declared requirements of its intended use. Device presents no adverse health effects or safety risks to patients when use as intended.
  • The RomiApex™ A-15 has the same intended use and fundamental scientific . technology as its predicate device - Bingo 1020 (K992233)
  • The RomiApex™ A-15 was evaluated against it's predicate, and was found to be . Substantial Equivalent.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like extensions. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Eitan Margalit General Manager Romidan, Limited 5 Simcha Holzberg Street Kirvat Ono Israel 55022

2010 SEP 3

Re: K101054

Trade/Device Name: RomiApex™ A-15, (an Apex Locator) Regulation Number: Unclassified Regulation Name: None Regulatory Class: None Product Code: LOY Dated: August 26, 2010 Received: August 30, 2010

Dear Mr. Margalit:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2- Mr. Margalit

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Antony D. Anatan

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

K101054

Romidan Ltd.
5, Simcha Holzberg St., 55022 Kiryat Ono, Israel.
+972-3-7360099+972-3-7360097
eitan@romidan.comWeb: www.romidan.com
Device Name:RomiApex™ A-15Valid from: 4/9/2010
Document No.:04.00Section 4; Page 1 (of 1)
Document Name:Indication for Use Statement

2010 SEP 3

Indication for Use Statement Section 4 .:

510(k) Number (if known):

RomiApex™ A-15, (an Apex Locator) Device Name:

Indications for Use:

  • RomiAPEX™ A-15 is indicated for patients who need to undergo root canal . treatment, which requires precise determination of position of the dental file in the canal relatively to the root apex.
    Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT.WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

&oncurrence of CDRH, Office of Device Evaluation (ODE)

Susan Renner

Division Sign-OH) livision of Anesthesiology, General Hospital nfection Control, Dental Devices

510(k) Number: K110154

Page 1 of 1

4-1 (Page 1 of 1)

N/A