K Number
K101012

Validate with FDA (Live)

Date Cleared
2010-12-15

(247 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The V-TRUST Handheld Pulse Oximeter is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adults.

This device with reusable SpO2 sensor accessories is indicated for non-invasive spot checking or continuous monitoring.

Device Description

The V-TRUST Handheld Pulse Oximeter is a light weight, portable and non-invasive pulse oximeter designed for use in measuring and displaying functional arterial oxygen saturation and pulse rate for adults during no motion based on the principle of spectrophotometry.

The V-TRUST Handheld Pulse Oximeter uses the same fundamental technology with the predicate device, 503 Pulse Oximeter. The SpO₂ and pulse rate software algorithms, interference-filtering software, and alarming system are identical to the software in the legally marketed predicate device cleared under K911124, with minor changes that do not raise new questions of safety or efficacy.

The V-TRUST Handheld Pulse Oximeter is allowed to link individual Solaris Compatible Reusable Adult SpO2 Finger Sensor (K061931) only. Solaris finger sensors are electro-optical sensors which function without skin penetration, electrical contact, or heat transfer. The sensors use optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the V-TRUST Handheld Pulse Oximeter:

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Proposed Device Specifications)Reported Device Performance (as tested)
SpO2 measurement range0% to 100%Not explicitly stated as meeting this, but accuracy tested within 70% to 100%.
SpO2 Accuracy70% to 100%: within $\pm$ 2%Equal to or less than 2%
Pulse Rate Measurement Range30 to 250 BPMNot explicitly stated as meeting this, but accuracy tested.
Pulse Rate Accuracy$\pm$ 1 bpmWithin $\pm$ 1 bpm
DisplayLED (light-emitting diode)LED (light-emitting diode) (Implicitly met as a design characteristic)
Power buttonYesYes (Implicitly met as a design characteristic)
SpO2 % displayYesYes (Implicitly met as a design characteristic)
Pulse amplitude indicatorYesYes (Implicitly met as a design characteristic)
Pulse rate displayYesYes (Implicitly met as a design characteristic)
Low battery indicatorYesYes (Implicitly met as a design characteristic)
Sensor error indicatorYesYes (Implicitly met as a design characteristic)
Sensor connector portYesYes (Implicitly met as a design characteristic)
AlarmYesYes (Implicitly met as a design characteristic)
Storage condition-4°F to 122°F (-20°C to 50°C), below 95%, non-condensing-4°F to 122°F (-20°C to 50°C), below 95%, non-condensing (Same as specification)
Operating condition32 °F to 113 °F (0 °C to 45 °C)32 °F to 113 °F (0 °C to 45 °C) (Same as specification)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 10 healthy subjects (5 males and 5 females).
  • Data Provenance: Not explicitly stated (e.g., country of origin), but the study was a "hypoxia study" conducted in adherence to ISO 9919:2005, implying a controlled clinical environment. It is a prospective study because subjects were tested during no motion to determine accuracy against an ABG analyzer.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the number of experts used or their qualifications for establishing the ground truth. It mentions that the ground truth for SpO2 and heart rate was determined "against the artery blood gas (ABG) analyzer in hospital," implying a gold standard measurement rather than expert consensus on visual interpretation.

4. Adjudication Method for the Test Set

Not applicable. The ground truth was established by an objective medical device (ABG analyzer) rather than expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This device is a standalone pulse oximeter, not an AI-assisted diagnostic tool that would typically involve human readers. The study compared the device's performance against a gold standard and a predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance study was done. The study evaluated the accuracy of the V-TRUST Handheld Pulse Oximeter's measurements (SpO2 and heart rate) directly against a clinical gold standard (ABG analyzer) and a predicate device, without involving human interpretation of the device's output.

7. The Type of Ground Truth Used

The ground truth used for SpO2 was established by an artery blood gas (ABG) analyzer in a hospital setting. For heart rate, the ground truth was also established through comparison against the ABG analyzer and the predicate device. This represents an objective, clinically accepted gold standard.

8. The Sample Size for the Training Set

The document does not provide information on a training set or its sample size. This is typical for a device like a pulse oximeter, where the fundamental technology (spectrophotometry) and algorithms are well-established and not based on a machine learning model that requires explicit "training data" in the same sense as an AI diagnostic tool. The software algorithms are stated to be "identical to the software in the legally marketed predicate device cleared under K911124."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no specific training set for a machine learning model is mentioned or implied. The device's software algorithms are inherited from the predicate device.

{0}------------------------------------------------

K101012

DEC 1 5 2010

Attachment C4

510(k) Summary

  1. Submitter Information Company name Contact person Address

Phone FAX E-mail Date Prepared

พละ แมร่เป็นตอนตรี และ เกม


  1. Name of Device .Trade Names Common Names Product Code Classification Panel Regulations

TaiDoc Technology Corporation Teling Hsu 3F, 5F, No. 127, Wugong 2nd Rd., Wugu Township, Taipei County, 24888, Taiwan (+886-2) 6625-8188 ext.1176 (+886-2) 6625-0288 teling.hsu@taidoc.com.tw April 2nd, 2010

V-TRUST Handheld Pulse Oximeter Oximeter DQA Anesthesiology Class II 21 CFR 870.2700

  1. Predicate Device Trade/Proprietary Name: Common/Usual Name: Submitter 510 (k) Number

503 PULSE OXIMETER Oximeter CRITICARE SYSTEMS, INC. K911124

  1. Device Description

The V-TRUST Handheld Pulse Oximeter is a light weight, portable and non-invasive pulse oximeter designed for use in measuring and displaying functional arterial oxygen saturation and pulse rate for adults during no motion based on the principle of spectrophotometry.

The V-TRUST Handheld Pulse Oximeter uses the same fundamental technology with the predicate device, 503 Pulse Oximeter. The SpO₂ and pulse rate software algorithms,

{1}------------------------------------------------

interference-filtering software, and alarming system are identical to the software in the legally marketed predicate device cleared under K911124, with minor changes that do not raise new questions of safety or efficacy.

The V-TRUST Handheld Pulse Oximeter is allowed to link individual Solaris Compatible Reusable Adult SpO2 Finger Sensor (K061931) only. Solaris finger sensors are electro-optical sensors which function without skin penetration, electrical contact, or heat transfer. The sensors use optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter device.

5. Intended Use

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

:

い・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


The V-TRUST Handheld Pulse Oximeter is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate for adults.

This device with reusable SpO2 sensor accessories is indicated for non-invasive spot checking or continuous monitoring.

6. Comparisons to the Predicate Device

The V-TRUST Handheld Pulse Oximeter has the following technological characteristics compared with the predicate device, 503 Pulse Oximeter (K911124):

DescriptionProposed DevicePredicate Device
V-TRUST Handheld Pulse Oximeter503 Pulse Oximeter
Intended patientpopulationAdultAdult
Intended application siteFingerFinger
PerformancespecificationsSubject who are well perfusedduring no motionSubject who are wellperfused during no motion
Detection modeContinuous monitoring or spotcheckingContinuous monitoring orspot checking
DisplayLED (light-emitting diode)LED (light-emitting diode)
SpO2 measurementrange0% to 100%1% to 99%
SpO2 accuracy70% to 100%: within $\pm$ 2%70% to 99%: within $\pm$ 2%
Pulse rate measurementrange30 to 250 BPM20 to 300 BPM

{2}------------------------------------------------

Pulse rate accuracy$\pm$ 1bpm$\pm$ 1bpm or $\pm$ 1%, whicheveris greater
Power buttonYesYes
SpO2 % displayYesYes
Pulse amplitude indicatorYesYes
Pulse rate displayYesYes
Low battery indicatorYesYes
Sensor error indicatorYesYes
Sensor connector portYesYes
AlarmYesYes
Storage condition-4°F to 122°F (-20°C to 50°C),below 95%, non-condensing-4 °F to 131 °F (-20°C to 55°C), below 95%,non-condensing
Operating condition32 °F to 113 °F(0 °C to 45 °C)50 °F to 104 °F(10 °C to 40 °C)

7. Performance Studies

ນ: ກິດສ່ວນພຣະເທດໄທແລນ ແລະ ບາງປີຕໍ່ໄດ້ຮັບສະໜາດການ, ການປະຊາທິປະເທດລາວ ແລະ ແລະ ແລະ ເພີ່ມເຕີຣ໌ວໄທແລນ ແລະ

היינגעינית י רטי יותר מומר מומרת מערכת מ

The clinical test included 10 healthy subjects (5 males and 5 females) aged 25-44 years old for the hypoxia study.SpO2 and heart rate accuracy were determined by testing on subjects during no motion over the range of 70% to 100% SpO2 against the artery blood gas (ABG) analyzer in hospital and the predicate device 503 Pulse Oximeter.

The study was performed in accordance with ISO 9919:2005. The SpO2 Arms of the study was equal to or less than the performance test criteria of 2% and the heart rate variation was within +/- 1 bpm as the specifications. The comparison between V-TRUST Handheld Pulse Oximeter and the predicate also showed 2% error limit for adults in the SpO2 range of 70% to 100%. Based on the results, V-TRUST pulse oximeter is substantial equivalent to the predicate device.

The consumer study was performed in both clinical and home environments and the device was used by healthcare professionals and lay users. The results demonstrate that the oximeter is suitable for use as intended SpO2 and heart rate accuracy.

The V-TRUST Handheld Pulse Oximeter meets the requirements of IEC/EN 60601-1:1995, IEC/EN 60601-1-2:2001, ISO 9919: 2005,ISO 10993-5: 1999 and ISO

{3}------------------------------------------------

10993-10: 2002. Software of the proposed device is validated.

The quality system for the Handheld Pulse Oximeter is well documented in the ` manufacturer's quality management system and conforms to ISO 13485 and ISO 9001.

  1. Conclusion

のお気になると、おすすめのお気になる。 このため、アイデントです。 このことです。 ここでは気になると、 ここでとても、 ここでとても、 ここでとても、 ここで


ﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮ

ディコスターになってる。

The V-TRUST Handheld Pulse Oximeter demonstrates satisfactory performance and is suitable for its intended use. The V-TRUST Handheld Pulse Oximeter is substantially equivalent to the predicate device 503 Pulse Oximeter.

{4}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Teling Hsu Regulatory Affairs Specialist TaiDoc Technology Corporation 3F, 5F, No. 127, Wugong 2nd Road Wugu Township, Taipei County Taiwan 24888

DEC 1 5 2010

Re: K101012

K101012
Trade/Device Name: V-TRUST Handheld Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: December 7, 2010 Received: December 10, 2010

Dear Mr. Hsu:

We have reviewed your Section 510(k) premarket notification of intent to market the device is We have reviewed your Section 310(x) promoted is substanially equivalent (for the referenced above and nave ucterimined the to her marketed predicate devices marketed in
indications for use stated in the enclosure and as a the Medical Device indications for use stated in the circlessity to the enactment date of the Medical Device interstate commerce prior to way 20, 1770, the sidential in accordance with the provisions of
Amendments, or to devices that have been received in accordance approval of a pr Amendments, or to devices that have been routes that do not require approval of a prematice of the Federal Food, Drug, and Cosment Act (rec) market the device, subject to the general
approval application (PMA). You may, therefore, market the device, subject to the gene approval application (PMA). Tou may, more of the Act include controls provisions of the Act. The general of devices, good manufacturing practice, requirements for annual registration, institution and adulteration. Please note: CDRH does
labeling, and prohibitions against mistrant libelity wereanies We remind you, h labeling, and prohibitions against institution of the remind you, however,
not evaluate information related to contract liability warranties. We remind you, however,
of the that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III
ng your device is a the same a trala - Evisting major regulations affecting you If your device is classified (See above) into this. Existing major regulations affecting your
(PMA), it may be subject to additional Controls. Tile 21, Parts 800 to 898, (PMA), it may be subject to adultions: Littlers, Title 21, Parts 800 to 898. In
device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In the device can be found in the Code of Federal Regulations, TMS = 7
addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Mr. Hsu

Please be advised that FDA's issuance of a substantial equivalence determination does not
ements of Please be advised that FDA's issuance of a substantials with other requirements
mean that FDA has made a determination that your devices with other requires, mean that FDA has made a determination interest by other Federal agencies.
of the Act or any Federal statutes and regulations and not timited to: registration of the Act or any Federal statutes and regulations administered of onlimited to: registration
You must comply with all the Act s requirements, including, but not limited to r You must comply with all the Act STequilements; mostumby one includes reporting
and listing (21 CFR Part 807); labeling (21 CFR Part) (21 CFR Party (21 CFF) Party (21 and listing (21 CFR Part 807); labeling (21 CFR 803); good manufacturing
(reporting of medical device-related adverse events) (21 CFR 803); good manufacturing (reporting of medical device-related adverse events) (2) ceplation (21 CFR Part 820);
practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 8 practice requirements as set forth in the quality systems (2) vegan of (1)
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm113809.html for http://www.fda.gov/AboutFDA/CentersOffices CDTFre of Compliance. Also,
the Center for Devices and Radiological Health's (CDRH's) of comerses notification" the Center for Devices and Radiological Treating by reference to premarket notification"
please note the regulation entitled, "Misbranding of environes events under the please note the regulation entitled, "Misolanding of reference of adverse events under the MDR regulation (21 CFR Part 803), please go to

MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Report/Problem/default.htmlfor the CDRH's
http://www.fda.gov/MedicalDevices/Safety/Rep http://www.ida.gov/MedicalDevieos/Division of Postmarket Surveillance.
Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the You may obtain other general Intornitures and Consumer Assistance at its toll-free
Division of Small Manufacturers, International and Consumer Assistance at its toll-free Division of Small Manufacturers, International its Internet address
number (800) 638-2041 or (301) 796-7100 or at its Internet school number (800) 038-2041 01 (301) 770 7100 et and 12 - 12 08 et and 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

KI010/2

Indications for Use

510(k) Number:

DEC 1 5 2010

Device Name:

V-TRUST Handheld Pulse Oximeter

Indications for Use:

The V-TRUST Handheld Pulse Oximeter is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adults.

This device with reusable SpO2 sensor accessories is indicated for non-invasive spot checking or continuous monitoring.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart D) AND/OR

Over the Counter Use _ (21 CFR Part 801 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

L. Shultheis

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K10/012

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Ancount of Antal Devices

Page 1 of -- 1-510(k) Number: _______________________________________________________________________________________________________________________________________________________________

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).