K Number
K100856

Validate with FDA (Live)

Date Cleared
2010-05-19

(54 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powdered Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powdered Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and the study proving the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 5250-06 e1.Meets
Physical PropertiesASTM standard D 5250-06 e1.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder AmountASTM standard D 5250-06 e1 and D6124-06Meets (<10mg/dm²)
BiocompatibilityPrimary Skin Irritation in rabbits (AAMI / ANSI / ISO 10993-10)Passes (Not a Primary Skin Irritation)
Dermal sensitization in guinea pig (AAMI / ANSI / ISO 10993-10)Passes (Not a Dermal sensitization)

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample size used for the test set for each characteristic. It broadly states that the device "meets" the requirements, implying testing was performed to the standards. The data provenance is not explicitly stated in terms of country of origin but is implicitly from the manufacturer's testing, specifically Suzhou Shengyuda Plastic Products Co.,Ltd. The studies were nonclinical.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as this device (patient examination gloves) does not involve human interpretation or subjective assessment by experts. The ground truth is established by objective physical and chemical testing against established industry standards and regulations.

4. Adjudication method for the test set

This information is not applicable as there is no subjective assessment or multiple interpretations requiring adjudication. Performance is determined by objective measurements against defined pass/fail criteria in the standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a medical glove, not an AI-powered diagnostic tool. Therefore, MRMC studies and the concept of human readers improving with AI are irrelevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a medical glove, not an algorithm or AI system.

7. The type of ground truth used

The ground truth used for this device's performance assessment is based on established industry standards and regulations. These include:

  • ASTM standard D 5250-06 e1 (for dimensions and physical properties)
  • 21 CFR 800.20 (for freedom from pinholes)
  • ASTM D6124-06 (for powder amount)
  • AAMI / ANSI / ISO 10993-10:2002 (for biocompatibility regarding primary skin irritation and dermal sensitization)

These standards define objective and measurable criteria for product performance.

8. The sample size for the training set

This information is not applicable. The device is a physical product (medical gloves), not a machine learning model. Therefore, there is no "training set" in the context of an algorithm.

9. How the ground truth for the training set was established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

MAY 1 9 2010

510(k) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: K 100850" (applicant leave blank)

Premarket Notification [510(k)] Summary

.

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name : Suzhou Shengyuda Plastic Products Co.,Ltd. Submitter's address : Zhitang Industrial Park,Zhitang Town Changshu City Jiangsu Province, 215531, China Phone number : 86-512-52553368 Fax number : 86-512-52515875 Name of contact person: Ms.Chen Su Date the summary was prepared: Mar. 10, 2010

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powdered Vinyl Patient Examination Gloves,Clear (non-colored)
Proprietary/Trade name:Powdered Vinyl Patient Examination Gloves,Clear (non-colored)Other clients private labeling
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

[(a)(3)). An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* Powdered Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.

Section C

{1}------------------------------------------------

[(a)(4)] A description of the device

Device Description: Powdered Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: Powdered Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The Powdered Vinyl Patient Examination Gloves, Clear (non-colored), non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-06 el.Meets
Physical PropertiesASTM standard D 5250-06 el.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder AmountASTM standard D 5250-06 e1and D6124-06Meets<10mg/dm²
BiocompatabilityPrimary Skin Irritation in rabbitsAAMI / ANSI / ISO 10993-10PassesNot a Primary Skin Irritation
Dermal sensitization in the guinea pigAAMI / ANSI / ISO 10993-10PassesNot a Dermal sensitization

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powdered Vinyl Patient Examination Gloves, Clear (non-colored), meet requirements per ASTM D5250-06 e1, per ASTM D6124-06, per 21 CFR 800.20 and FDA recognition number 2-87: AAMI / ANSI / ISO 10993-10:2002.

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powdered Vinyl Patient Examination Gloves, Clear (non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .

It can be concluded that the Powdered Vinyl Patient Examination Gloves, Clear (non-colored) is as safe, as effective, and performs as well as the predicate device, Powdered Vinyl Patient Examination Gloves, Clear(Non-colored) , ZHANG JIA GANG FENGYUAN PLASTIC PRODUCTS CO LTD. K091662.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract shapes stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Suzhou Shengyuda Plastic Products Company Limited C/O Mr. Chu Xiaoan Beijing Easy - Link Company, Limited Room 1606 Building 1 Jian Xiang Yuan Number 209 Beijing, China 100083

MAY 1 9 7010

Re: K100856

Trade/Device Name: Powdered Vinyl Patient Examination Gloves, Clear (Non-Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove

Regulatory Class: I Product Code: LYZ Dated: March 19, 2010

Received: March 26, 2010

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Chu Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Eyaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Applicant: Suzhou Shengyuda Plastic Products Co.,Ltd.

100856 510(k) Number (if known): *

Device Name: Vinyl Patient Examination Gloves, Clear (non-colored) Powdered

Indications For Use:

Powdered Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Se

(Division Sign-Off) Division of Anesthesiology, General Hospital

Infection Control, Dental Devices

510(k) Number: __ K i00856

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.