K Number
K100831

Validate with FDA (Live)

Date Cleared
2011-05-26

(428 days)

Product Code
Regulation Number
866.6010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CYFRA 21-1 EIA kit is intended for the quantitative determination of soluble cytokeratin 19 fragments in human serum. The assay is to be used as an aid in monitoring disease progression during the course of disease and treatment in lung cancer patients. Serial testing for patient CYFRA 21-1 assay values should be used in conjunction with other clinical methods used for monitoring lung cancer.

Device Description

The CYFRA 21-1 EIA is a solid phase, non-competitive immunoassay based on two monoclonal antibodies (derived from mice) directed against two separate antigenic determinants of soluble fragments of cytokeratin 19. Calibrators, controls and patient samples are incubated together with biotinylated Anti-CYFRA 21-1 MAb and horseradish peroxidase (HRP) labeled Anti-CYFRA 21-1 MAb in streptavidin coated micro strips. After washing, buffered Substrate/Chromogen reagent (hydrogen peroxide and 3, 3', 5, 5' tetramethylbenzidine) is added to each well and the enzyme reaction is allowed to proceed. During the enzyme reaction a blue color will develop if antigen is present. The intensity of the color development is proportional to the amount of CYFRA 21-1 present in the samples. The color intensity is determined in a microplate spectrophotometer at 620 nm (or optionally at 405 nm after addition of Stop Solution). Calibration curves are constructed for each assay by plotting absorbance value versus the concentration for each calibrator. The CYFRA 21-1 concentrations of patient samples are then read from the calibration curve.

AI/ML Overview

Here's an analysis of the provided text regarding the CYFRA 21-1 EIA Kit, focusing on acceptance criteria and the supporting study:

Acceptance Criteria and Device Performance

The document does not present explicit, pre-defined "acceptance criteria" in a typical numerical format (e.g., "Sensitivity must be >X%"). Instead, it describes performance characteristics, and the "acceptance" is implied by the FDA's clearance based on demonstrated substantial equivalence to a predicate device and adherence to relevant guidelines.

However, based on the studies presented, we can infer performance metrics that were deemed acceptable for clearance. The key study provided is a clinical study for monitoring disease progression in lung cancer.

Table of Performance Metrics

Performance CharacteristicReported Device PerformanceComments / Context
Precision< 8.6 %Total precision based on multi-site, multi-lot testing.
Linearity100 ± 20% mean dilution linearity; ≤ 10% non-linearity across 0.5-50.0 ng/mLBased on CLSI guideline EP6-A.
Limit of Detection (LoD)0.12 ng/mLDetermined using CLSI guideline EP17-A.
Limit of Quantitation (LoQ)0.21 ng/mLCorresponds to highest allowable imprecision of 17.78%.
Interference (Endogenous)Mean assay specificity 100 ± 15% (for tested substances)No interference found for Triglycerides (30 mg/mL), Bilirubin (0.2 mg/mL), Hemoglobin (5 mg/mL), Total Protein (120 mg/mL).
Interference (Chemotherapeutic drugs)No interference (for tested substances)No interference found for Carboplatin (500 µg/mL), Cisplatin (165 µg/mL), Dexamethasone (10 µg/mL), Doxorubicin (1.16 µg/mL), Leucovorin (2.68 µg/mL), Methotrexate (45 µg/mL), Paclitaxel (3.5 ng/mL).
Interference (Clinical Conditions)Mean % recovery: HAMA 98%, RF 101%For CYFRA 21-1 spiked at ~5 and 25 ng/mL.
Clinical Study Performance (Monitoring Lung Cancer Progression)(See details below for specific percentages at a >50% change threshold)
- Concordance76% (239/314)Total concordance between CYFRA 21-1 change and disease status.
- Sensitivity (>50% increase)45.9%Proportion of patients with disease progression that had elevated CYFRA 21-1.
- Specificity (>50% increase)87.3%Proportion of patients without disease progression that had no elevation in CYFRA 21-1.
- Positive Predictive Value (PPV) (>50% increase)57.4%Proportion of patients with elevated CYFRA 21-1 that had disease progression.
- Negative Predictive Value (NPV) (>50% increase)81.3%Proportion of patients with no elevation in CYFRA 21-1 that did not have disease progression.

Study Details

The primary study mentioned proving the device meets its intended use is a retrospective clinical study focused on monitoring the course of disease in lung cancer patients.

  1. Sample size used for the test set and the data provenance:

    • Sample Size: 100 lung cancer patients, yielding a total of 314 observation pairs of serial serum samples. An average of 4.1 observations per patient.
    • Data Provenance: Retrospective clinical study. Patient samples were collected from a tertiary cancer center. The country of origin is not explicitly stated, but the submission is to the US FDA, implying the study was likely conducted in the US or under comparable standards.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not specify the number of experts or their qualifications used to establish the ground truth for disease progression/no progression. It states that "Changes in CYFRA 21-1 levels in serial serum samples... were compared to changes in disease status." This implies clinical assessment of disease status was the ground truth, likely determined by treating physicians or a review board at the tertiary cancer center, but no details are provided.
  3. Adjudication method for the test set:

    • The document does not describe an explicit adjudication method (like 2+1 or 3+1) for establishing disease status. The "changes in disease status" used for ground truth would typically be based on standard clinical practices (imaging, biopsy, clinical examination), but the specific process for confirming or adjudicating these statuses for the study is not outlined.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not performed. This device is an in vitro diagnostic (IVD) assay (a blood test), not an imaging-based AI system that would typically involve human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance was done in the sense that the assay results (CYFRA 21-1 concentrations) were directly compared to the established disease status without an intermediate "human-in-the-loop" interpretation of the assay values themselves for the purpose of the study. The device provides a quantitative measurement, and thresholds were applied to this measurement to determine its correlation with disease progression. However, the intended use states it is an "aid" in monitoring and "should be used in conjunction with other clinical methods," implying human interpretation within a broader clinical context. The performance metrics (sensitivity, specificity etc.) presented in the tables are for the assay's output as a standalone indicator of change.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for the clinical study was based on "changes in disease status." This likely encompasses a combination of clinical assessments, imaging results, and possibly pathology reports over time, interpreted by clinicians at the tertiary cancer center. The specific methods used to define "disease progression" and "no progression" within the study are not detailed beyond "changes in disease status."
  7. The sample size for the training set:

    • The document does not provide information on a specific training set size. The clinical study described is the test set for assessing the final device's performance for monitoring. For an immunoassay, the "training" analogous to machine learning would be the development and optimization of the assay itself (antibody selection, reagent concentrations, calibration methods), which doesn't typically rely on a "training set" of patient data in the same way an AI algorithm does.
  8. How the ground truth for the training set was established:

    • As there's no explicitly defined "training set" in the context of an AI/ML algorithm development, this question is not directly applicable. The assay's analytical performance (precision, linearity, LoD, LoQ, interference) and the clinical utility (monitoring study) are reported, which are evaluated against established analytical standards and clinical outcomes, respectively.

{0}------------------------------------------------

Premarket Notification (510(k)) CYFRA 21-1 EIA

Image /page/0/Picture/2 description: The image shows the logo for "FUJIREBIO Diagnostics, Inc.". The logo consists of a stylized graphic on the left, followed by the company name in bold, sans-serif font. The word "Diagnostics, Inc." is placed below "FUJIREBIO" in a smaller font size. The logo appears to be designed for brand recognition and corporate identity.

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K100831.

Submitter Information

Address: .

Fuiirebio Diagnostics, Inc. 201 Great Valley Parkway Malvern, PA 19355

  • Diana Dickson Contact person: (610) 240-3917 dicksond@fdi.com
    Summary preparation date: May 26, 2011

Name of Device

Trade/Proprietary Name:CYFRA 21-1 EIA Kit
---------------------------------------------

Common/Usual Name: CYFRA 21-1 EIA Test Kit

Regulation Number: 21 CFR 866.6010

Regulatory Class: Class II

Product Code: OVK

Predicate Device

ARCHITECT CEA (K990774)

Device Description

The CYFRA 21-1 EIA is a solid phase, non-competitive immunoassay based on two monoclonal antibodies (derived from mice) directed against two separate antigenic determinants of soluble fragments of cytokeratin 19. Calibrators, controls and patient samples are incubated together with biotinylated Anti-CYFRA 21-1 MAb and horseradish peroxidase (HRP) labeled Anti-CYFRA 21-1 MAb in streptavidin coated micro strips. After washing, buffered Substrate/Chromogen reagent (hydrogen peroxide and 3, 3', 5, 5' tetramethylbenzidine) is added to each well and the enzyme reaction is allowed to proceed. During the enzyme reaction a blue color will develop if antigen is present. The intensity of the color development is proportional to the amount of CYFRA 21-1 present in the samples.

1

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic to the left of the company name. The graphic is a three-part design with black, white, and striped sections. The text "FUJIREBIO Diagnostics, Inc." is to the right of the graphic.

The color intensity is determined in a microplate spectrophotometer at 620 nm (or optionally at 405 nm after addition of Stop Solution).

Calibration curves are constructed for each assay by plotting absorbance value versus the concentration for each calibrator. The CYFRA 21-1 concentrations of patient samples are then read from the calibration curve.

Reportable Range

The CYFRA 21-1 ElA measures concentrations between 0.5 and 50 ng/mL.

Intended Use

The CYFRA 21-1 EIA kit is intended for the quantitative determination of soluble cytokeratin 19 fragments in human serum. The assay is to be used as an aid in monitoring disease progression during the course of disease and treatment in lung cancer patients. Serial testing for patient CYFRA 21-1 assay values should be used in conjunction with other clinical methods used for monitoring lung cancer.

Statement of Substantial Equivalence

The CYFRA 21-1 EIA kit is intended for the quantitative determination of soluble cytokeratin 19 fragments in human serum.

The assay is to be used as an aid in monitoring disease progression during the course of disease and treatment in lung cancer patients. Serial testing for patient CYFRA 21-1 assay values should be used in conjunction with other clinical methods used for monitoring lung cancer.

As there is no FDA cleared or approved device for the CYFRA 21-1 EIA Kit, substantial equivalence for the CYFRA 21-1 EIA test kit was determined by comparing the performance characteristics obtained with the CYFRA 21-1 EIA to the package insert claims in the ARCHITECT CEA assay.

The requlatory submission will be prepared pursuant to Title 21CFR § 866.6010(b) which states Tumor Markers must comply with the following special controls;

    1. Guidance document entitled "Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA".
    1. Voluntary assay performance standards issued by the Clinical Laboratory Standards Institute (CLSI), formally NCCLS.

A comparison of the features of the CYFRA 21-1 EIA Kit and the ARCHITECT CEA assay are as follows:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains the logo for "FUJIREBIO Diagnostics, Inc.". The logo consists of a stylized graphic on the left, followed by the company name in bold, sans-serif font. The word "Diagnostics, Inc." is written in a smaller font size below the company name.

Similarities
CYFRA 21-1 EIA Kit(Proposed Device)ARCHITECT CEA(Predicate Device)K990774
Device TypeIn vitro diagnosticIn vitro diagnostic
ClassificationClass IIClass II
Regulation Number21 CFR 866.601021 CFR 866.6010
Product UsageClinical and Hospital laboratoriesClinical and Hospital laboratories
Specimen CollectionMethodRoutine Phlebotomy TechniquesRoutine Phlebotomy Techniques
Interpretation of ResultsStandard CurveStandard Curve
Differences
CYFRA 21-1 EIA Kit(Proposed Device)ARCHITECT CEA(Predicate Device)K042731
Product CodeTo be determinedDHX
Type of SpecimenHuman Serum OnlyHuman Serum or Plasma
Antigen DetectedCYFRA 21-1CEA
CalibratorsSupplied with KitSupplied as separate Kit
ControlsSupplied with KitSupplied as separate Kit
Principle of OperationManualEnzymatic Immunoassay (EIA)AutomatedChemiluminescent MicroparticleImmunoassay (CMIA)
Intended UseAid in monitoring diseaseprogression during the course ofdisease and treatment in lungcancer patients.Aid in the prognosis andmanagement of cancer patients inwhom changing concentrations ofCEA are observed.

{3}------------------------------------------------

Image /page/3/Figure/1 description: The image shows the logo for FUJIREBIO Diagnostics, Inc. The logo consists of a stylized graphic on the left, followed by the text "FUJIREBIO" in bold, uppercase letters. Below the company name, the text "Diagnostics, Inc." is printed in a smaller, regular font. The graphic appears to be an abstract design, possibly representing a molecule or a stylized flower.

Performance Characteristics

Precision:

Two (2) sites tested two (2) CYFRA 21-1 EIA Kit lots for the precision study. For each lot, the two (2) sites repeated the assay for twenty (20) days, (twenty (20) days each lot), and performed two (2) runs per day with four (4) serum based panel samples tested as unknowns. One (1) site tested three (3) CYFRA 21-1 EIA Kit lots for the precision study. This site repeated the assay for thirty (30) days (ten (10) days each lot), performing two (2) runs per day with eight (8) serum based panel samples tested as unknowns. The total precision of the CYFRA 21-1 EIA was found to be < 8.6 %.

Linearity:

The CYFRA 21-1 ElA assay mean dilution linearity is 100 ± 20%. A study was conducted for the CYFRA 21-1 EIA Kit modeled after the CLSI quideline EP6-A. Serum samples with elevated CYFRA 21-1 values were diluted with CYFRA 21-1 Calibrator A (zero). The CYFRA 21-1 concentration was determined for each dilution and the percent (%) recovery was calculated. The nonlinearity calculated by weighted polynomial regression is ≤ 10% across the measurement range of 0.5 to 50.0 ng/mL.

Detection and Quantitation Limit:

The Limit of Detection of the CYFRA 21-1 EIA Kit was determined to be 0.12 ng/mL. The CLSI quideline EP17-A (22) was used to design the LoD experiments. The limit of detection (LoD) corresponds to the upper limit of the 95% confidence interval and represents the lowest concentration of CYFRA 21-1 antigen that can be distinquished from zero.

The Limit of Quantitation of the CYFRA 21-1 EIA Kit was determined to be 0.21 ng/mL. The limit of quantitation (LoQ) corresponds to the lowest amount of analyte in a sample that can be accurately quantitated with the highest allowable imprecision of 17.78%.

Interference:

The CYFRA 21-1 EIA Kit mean assay specificity is 100 ± 15%. Recovery studies were performed to compare sera containing the following compounds at the indicated concentrations with control sera. The CLSI guideline EP7-A was used to design the interference experiments. The following substances and concentrations were tested and found not to interfere with the test.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for FUJIREBIO Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a three-part design with a white center and textured black outer sections. The text "FUJIREBIO" is in bold, sans-serif font, with "Diagnostics, Inc." in a smaller font size below.

Endogenous serum interferencesTest Concentration
Triglycerides30 mg/mL
Billirubin0.2 mg/mL
Hemoglobin5 mg/mL
Total Protein120 mg/mL
Chemotherapeutic drug interferencesTest Concentration
Carboplatin500 µg/mL
Cisplatin165 µg/mL
Dexamethasone10 µg/mL
Doxorubicin1.16 µg/mL
Leucovorin2.68 µg/mL
Methotrexate45 µg/mL
Paclitaxel3.5 ng/mL

Potentially interfering clinical conditions

The CYFRA 21-1 EIA assay was evaluated using specimens with HAMA and Rheumatoid Factor (RF) to further assess the assay specificity. Six specimens positive for HAMA and five specimens positive for RF were evaluated for % recovery with CYFRA 21-1 antigen spiked into each specimen at approximately 5 and 25 ng/mL.

Clinical conditionNumber of specimensMean % recovery
HAMA698
RF5101

MONITORING THE COURSE OF DISEASE IN PATIENTS DIAGNOSED WITH LUNG CANCER

The effectiveness of the CYFRA 21-1 EIA as an aid in monitoring the course of disease in lung cancer patients was determined through a retrospective clinical study. Changes in CYFRA 21-1 levels in serial serum samples collected from a tertiary cancer center were compared to changes in disease status. A study involving 100 patients was undertaken with a total of 314 observation pairs with an average number of 4.1 observations per patient. For the 100 lung cancer cases, 95 were classified as non-small cell lung cancer (NCSLC) and 5 were classified as small cell lung cancer (SCLC). 90 of the NSCLC were further classified as adenocarcinoma (68), squamous cell carcinoma (19), and large cell carcinoma (3). 79 of these the 100 lung cancer cases had staging information as shown in the following table. In this study, only 5 patients had Stage I or II disease, the performance of CYFRA 21-1 has not been adequately assessed in this subpopulation.

{5}------------------------------------------------

Premarket Notification (510(k)) CYFRA 21-1 EIA

Image /page/5/Figure/1 description: The image shows the logo for "FUJIREBIO Diagnostics, Inc." The logo consists of a stylized graphic on the left, followed by the company name in bold, sans-serif font. The graphic is a circular shape with a combination of solid black and white areas, creating a visually interesting design. The text "Diagnostics, Inc." is placed below the main company name in a smaller font size.

Number ofPatients
Stage I2
Stage II3
Stage III36
Stage IV38
Total (with Stage Information)79
Unknown17
Unstaged4
Total Lung Cancer Cases100

A positive change in CYFRA 21-1 was defined as a measurable increase in the value that was at least 50% greater than the previous value of the test. Observation pairs with both values below the normal reference range of 1.8 ng/mL were defined as no significant change. This level of change takes into account the variability of the assay and the biological variability.

Forty-six percent (46%) or 39/85 of the patient samples with a positive change correlated with the disease progression while eighty-seven percent (87%) or 200/229 of the patient serial samples with no significant change in CYFRA 21-1 value correlated with no progression. The total concordance was seventy-six percent (76%) or 239/314. The following table presents the data in a 2 x 2 format.

Change in Disease State per Sequential Pair
Increase in CYFRA21-1 concentrationProgressionNo ProgressionTotal
>50%392968
≤50%46200246
Total85229314

Clinicians may wish to use other percent changes in CYFRA 21-1 concentration to reflect their preferences in the trade-off between sensitivity and specificity. The following table shows the resulting sensitivities and specificities of the CYFRA 21-1 EIA at various percent changes in CYFRA 21-1 EIA concentrations, together with the positive values (PPV) and negative predictive values (NPV) for the population tested (85 sequential pairs from patients with disease progression and 229 sequential pairs from patients with no progression.)

  • . Sensitivity is represented as the proportion of patients with disease progression that had elevated CYFRA 21-1
  • Specificity is represented as the proportion of patients without disease progression that . had no elevation in CYFRA 21-1
  • PPV is represented as the proportion of patients with elevated CYFRA 21-1 that had . disease progression
  • NPV is represented as the proportion of patients with no elevation in CYFRA 21-1 that . did not have disease progression

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for FUJIREBIO Diagnostics, Inc. The logo consists of a stylized graphic on the left, followed by the company name in bold, sans-serif font. The word "Diagnostics, Inc." is placed below the company name in a smaller font size.

Percent Increase inCYFRA 21-1Concentration (%)Sensitivity(%)Specificity(%)PPV(%)NPV(%)
3052.984.355.682.8
4048.285.655.481.7
5045.987.357.481.3
6044.788.258.581.1
7043.589.560.781.0

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Fujirebio Diagnostics, Inc. c/o Ms. Diana Dickson Regulatory Affairs Manager 201 Great Valley Parkway Malvern, PA 19355

MAY 2 6 2011

Re: K100831

Trade/Device Name: CYFRA 21-1 EIA Kit Regulation Number: 21 CFR §866.6010 Regulation Name: Tumor-associated antigen immunological test system Regulatory Class: Class II Product Codes: OVK Dated: May 18, 2011 Received: May 19, 2011

Dear Ms. Dickson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice

{8}------------------------------------------------

Page 2 - Ms. Diana Dickson

requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Maria Chan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K100831

Device Name: _CYFRA 21-1 EIA

Indications for Use:

The CYFRA 21-1 EIA kit is intended for the quantitative determination of soluble cytokeratin 19 fragments in human serum. The assay is to be used as an aid in monitoring disease progression during the course of disease and treatment in lung cancer patients. Serial testing for patient CYFRA 21-1 assay values should be used in conjunction with other clinical methods used for monitoring lung cancer.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Deena Philip

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K100831

Page 1 of 1 .

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.