(184 days)
UROSTATION - 3D PROSTATE SUITE is a computer-based software application intended to process, visualize and record 3D digital ultrasound images of the prostate.
UROSTATION - 3D PROSTATE SUITE is intended to be used by physicians in the clinic or hospital for 2D and 3D visualization of ultrasound images of the prostate gland. Additional software features include patient data management, multi-planar reconstruction and 3D image registration.
UROSTATION - 3D PROSTATE SUITE is a computer-based software application designed to process, visualize and record 3D digital ultrasound images of the prostate, and to manage patient and clinical data in the context of transrectal prostate biopsy.
This 510(k) summary does not contain the acceptance criteria or a study proving that the device meets acceptance criteria in the traditional sense of a clinical performance study with statistical endpoints.
Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (3-D Imaging Workstation, K081093) based on shared intended use, technological characteristics, and functions. The "study" here is a comparison of features and capabilities.
Here's a breakdown of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance:
| Feature/Characteristic | Acceptance Criteria (Implied by Predicate) | Reported Device Performance (UROSTATION - 3D PROSTATE SUITE) |
|---|---|---|
| Intended Use | Process, visualize and record 3D digital ultrasound images of the prostate | Process, visualize and record 3D digital ultrasound images of the prostate |
| Function - Image Acquisition | 2D/3D image acquisition | 2D/3D image acquisition |
| Function - Image Viewing/Reviewing | 2D/3D image viewing/reviewing | 2D/3D image viewing/reviewing |
| Function - Image Processing | 2D/3D image processing | 2D/3D image processing |
| Function - Image Storage | 2D/3D image storage | 2D/3D image storage |
| Function - Multi-Planar Reformatting (MPR) | Multi-Planar Reformatting (MPR) (implied by predicate's "Volume rendering", "Segmentation") | Multi-Planar Reformatting (MPR) |
| Function - Patient and Clinical Data Management | Patient and clinical data management | Patient and clinical data management |
| Data Source | 3D TRUS scanners | 3D TRUS scanners |
| Physical Characterization | Software package, operates on off-the-shelf hardware, Windows OS | Software package, operates on off-the-shelf hardware, Windows OS |
| Differences from Predicate | Not applicable; differences are noted as not raising safety/effectiveness concerns. | • Linear workflow adapted to physician's practice. • Dedicated algorithm to visualize biopsies w.r.t. a single reference image. • No measurement features. |
Note on Acceptance Criteria: The acceptance criteria are implied by the features of the predicate device. The goal of this 510(k) is to show UROSTATION performs similarly or introduces no new risks despite differences. There are no explicit quantitative performance metrics or thresholds mentioned as "acceptance criteria" for a study to meet.
2. Sample size used for the test set and the data provenance:
- Sample Size: Not applicable. The document describes a comparison to a predicate device's functionalities, not a clinical test set for performance evaluation.
- Data Provenance: Not applicable. No specific data or images were used for performance evaluation in a clinical study context.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not applicable. No ground truth was established by experts for a test set in this submission.
- Qualifications of Experts: Not applicable.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication Method: Not applicable. No test set requiring expert adjudication was conducted.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No. This submission does not include an MRMC comparative effectiveness study. The device is image processing software, not an AI diagnostic tool requiring evaluation of human reader performance.
- Effect Size: Not applicable.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Standalone Study: No. This submission does not describe a standalone performance study. The device's function is to process and visualize images for physician use, not to make independent diagnoses.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Type of Ground Truth: Not applicable. No ground truth data (e.g., pathology, outcomes) was used for performance validation in this context. The "ground truth" for substantial equivalence is the predicate device's established safety and effectiveness.
8. The sample size for the training set:
- Sample Size: Not applicable. The document does not mention any machine learning or AI components that would require a "training set" in the context of performance evaluation. It describes software with image processing and display algorithms.
9. How the ground truth for the training set was established:
- How Ground Truth Was Established: Not applicable, as there is no mention of a training set.
{0}------------------------------------------------
K100793 - Page 1 of 3
510(k) SUMMARY 5.
SEP 2222010 510(k) Summary for UROSTATION - 3D PROSTATE SUITE
The 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92.
| 510(k) Owner: | KOELIS5, avenue du Grand Sablon 38700 La TroncheFRANCEPhone: +33 476 637 588 Fax: +33 476 549.561 |
|---|---|
| Contact Name | Cécile DESMULIEMail: cecile.desmulie@koelis.com |
| Date Prepared | 03/17/2010 |
Proposed Device:
| Trade Name: | UROSTATION - 3D PROSTATE SUITE |
|---|---|
| Common Name: | Medical Image Processing Software System |
| Classification Name: | System, Image processing, RadiologicalPicture archiving and communication system,21 CFR PART 892.2050 |
| Device Class | II |
| Product Code | LLZ |
Predicate Device:
The Urostation system is substantially equivalent to:
| 510(k) Number | K081093 |
|---|---|
| Device Name | 3-D Imaging Workstation |
Intended Use:
UROSTATION - 3D PROSTATE SUITE is a computer-based software application intended to process, visualize and record 3D digital ultrasound images of the prostate.
Indications for Use:
UROSTATION - 3D PROSTATE SUITE is intended to be used by physicians in the clinic or hospital for 2D and 3D visualization of ultrasound images of the prostate gland. Additional software features include patient data management, multi-planar reconstruction and 3D image registration.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{1}------------------------------------------------
K100793 - Page 2 of 3
Device Description:
UROSTATION - 3D PROSTATE SUITE is a computer-based software application designed to process, visualize and record 3D digital ultrasound images of the prostate, and to manage patient and clinical data in the context of transrectal prostate biopsy.
Hardware Platform and Operating System
The application runs on standard Intel PCs under Microsoft Windows® XP operating system
Peripheral and accessories
The application is controlled by a footswitch and a manual input device. It is designed to work with commercially available 3D ultrasound scanner through Ethernet connection, 3D transrectal ultrasound probe and needle guide.
Software Features
UROSTATION - 3D PROSTATE SUITE implements image fusion and display algorithms to provide 3D representation of prostate biopsies.
A typical workflow enables the physician to intraoperatively visualize the 3D mapping of biopsies with respect to a reference image of the patient's prostate.
For that purpose, 3D digital images may be transferred at any time from the 3D ultrasound scanner to the Urostation for process and display, while the physician keeps track of the organ using the usual 2D live ultrasound mode.
Alternatively, UROSTATION - 3D PROSTATE SUTE also provides a review mode that allows the mapping of histologic results on the said 3D representation of the patient's prostate. Patient information, images and 3D biopsy cartography may be stored or printed for future retrieval and examination.
Technological Characteristics compared with the predicate device:
UROSTATION - 3D PROSTATE SUITE utilizes the same technological characteristics as the predicate device.
- Both are PC-based software applications that provide 2D/3D medical image acquisition and visualization for enhanced observation and analysis of the prostate gland.
- Both devices provide in addition patient and clinical data management features. -
- Both deal with ultrasound images received from commercially available imaging devices.
UROSTATION - 3D PROSTATE SUITE also presents differences with the predicate device. These differences do not raise any safety or effectiveness concern.
- The software architecture follows a linear workflow adapted to the physician's . practice.
- The system implements a dedicated algorithm to visualize biopsies with respect to . a single reference image of the prostate.
- The system has no measurement features.
{2}------------------------------------------------
K100793 - Page 3 of
Substantial Equivalence Comparison Chart
| KOELIS | EIGEN LLC | |
|---|---|---|
| System: | 3D PROSTATE SUITE | 3-D Imaging Workstation |
| 510(k) number: | K081093 | |
| Function | • 2D/3D image- acquisition- viewing/reviewing- processing- storage• Multi-Planar Reformatting (MPR)• Patient and clinical data management | • Volume rendering• Segmentation• Measurements |
| Function: | • Printing | |
| Intended Use: | Process, visualize and record 3D digital ultrasoundimages of the prostate | |
| Data Source: | 3D TRUS scanners | |
| Physical Characterization: | Software package- operates on off the shelf hardware- Windows operating system |
Conclusion:
. .
The results of comparing the intended uses, functions and technological characteristics of the UROSTATION - 3D PROSTATE SUITE with its predicate devices shows that the system is as safe and as effective as its predicate device.
Therefore the UROSTATION - 3D PROSTATE SUITE is substantially equivalent to existing products currently on the market.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Cecile Desmulie Quality Manager KOELIS 5, avenue du Grand Sablon La Tronche 38700 FRANCE
SEP 2 2 2010
Re: K100793
Trade Name: Urostation-3D Prostate Suite Regulation Number: 21 CFR $892.2050 . Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 17, 2010 Received: September 20, 2010
Dear Mr. Desmulie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Donald J. Hett
Donald J. St Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device · Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
K10793
SEP 22 2010
510(k) Number (if known):
Unknown
Device Name:
UROSTATION 3D PROSTATE SUITE
Indications For Use:
UROSTATION - 3D PROSTATE SUITE is intended to be used by physicians in the clinic or hospital for 2D and 3D visualization of ultrasound images of the prostate gland. Additional software features include patient data management, multi-planar reconstruction and 3D image registration.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Page 1 of 1
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
K100793
519K
N/A