(269 days)
The ARCHITEX™ Space Maintenance System is indicated for use as temporary implants to stabilize and support autograft, autograft extenders, allograft, and bone void fillers and/or fractured bone segments with or without bone plates or ritanium mesh in bony defects of oral maxillofacial anatomy.
ARCHITEX™ Space Maintenance System is intended for use in oral-maxillofacial surgical reconstruction and dental regeneration procedures; the implants contained in the System are for maintaining space during bone grating procedures and/or to support soft tissue until bone formation.
The ARCHITEX™ Space Maintenance System consists of a variety of shapes and sizes of screws (Socket Preservation, Tenting, and Mesh Fixation Screws) and titanium mesh implants components as well as placement instruments components. These instruments supplied are general class I instruments added as a convenience for doctors to use to place the provided implants.
The ARCHITEX™ Space Maintenance System includes a selection of Socket Preservation screws which are designed to aid in extraction socket grafting while simultaneously supporting the original gingival margins and papilla.
The ARCHITEX™ Space Maintenance System components are fabricated from medical grade stainless steel and medical grade titanium or titanium alloy. Medical grade titanium and titanium alloy may be used together. The subject implants components will be manufactured from medical grade titanium allov described by such standards as ASTM F 136-08el or ISO 5832-3, and/or from medical grade unalloyed titanium described by such standard as ASTM F67-06. The placement instruments components will be manufactured from medical grade stainless steel as described in ASTM F8PP-09. The ARCHITEX™ Space Maintenance System is sold non-sterile.
The purpose of this 510(k) application is to seek marketing clearance for the ARCHITEX™ Space Maintenance System to be used as a root-form endosseous dental implants system.
The ARCHITEX™ Space Maintenance System (ARCHITEX™) a device used for dental implants used for oral-maxillofacial surgical reconstruction and dental regeneration procedures was evaluated for substantial equivalence to its predicate devices in accordance with the FDA 510(k) premarket notification process.
1. Table of Acceptance Criteria & Reported Device Performance
| Device Component | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Titanium Mesh | Geometrically, materially, and mechanically equivalent to TiMesh® (K062348) | Equivalent to TiMesh® (K062348) in static cantilever bending strength. |
| Titanium Mesh Screws | Geometrically, materially, and mechanically equivalent to TiMesh® (K062348) | Equivalent to TiMesh® (K062348) geometrically, materially, and mechanically. |
| Tenting Screws | Geometrically, materially, and mechanically equivalent to ACE Tru-FIX™ (K080074). Bone pull-out strength and screw subsidence equivalent to predicate. | Equivalent to ACE Tru-FIX™ (K080074) in static axial pullout and subsidence strength. |
| Socket Preservation Screws | Bone interfacing portion geometrically, materially, and mechanically equivalent to ACE Tru-FIX™ (K080074). Bone pull-out strength and screw subsidence equivalent to predicate. | Equivalent to ACE Tru-FIX™ (K080074) in static axial pullout and subsidence strength (bone interfacing portion). |
| Head portion of Socket Preservation Screws | Geometrically, materially, and mechanically equivalent to Straumann Dental Implant System (K071585) abutments. | Equivalent to Straumann Dental Implant System (K071585) abutments (head portion). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size for the test set. However, the non-clinical tests involved "Subject Devices and Predicate Devices" to compare their performance. The data provenance is not specified, but these are primarily bench performance tests conducted on the physical devices rather than patient data.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not applicable. The study involved non-clinical bench testing and material/dimensional analysis, not clinical ground truth established by experts.
4. Adjudication Method
This information is not applicable, as the study did not involve expert adjudication of clinical data.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study focused on the physical and mechanical properties of the device and its predicates.
6. Standalone (Algorithm Only) Performance
No, a standalone (algorithm only) performance study was not done. This device is a physical implant system, not a software algorithm.
7. Type of Ground Truth Used
The "ground truth" for this submission was established through:
- Dimensional and Material Analysis: Comparing the physical specifications and materials of the ARCHITEX™ components against the predicate devices.
- Bench Performance Testing: Conducting standardized and non-standardized mechanical tests (Static Axial Pullout, Static Subsidence, Static Cantilever Bending, Static "Removal Torque" and "Torque to Failure") on both the subject and predicate devices. The results of these tests served as the empirical "ground truth" for equivalence claims.
8. Sample Size for the Training Set
This is not applicable, as there was no "training set" in the context of an algorithm. The study compared the device components directly to predicate device components through bench testing.
9. How the Ground Truth for the Training Set Was Established
This is not applicable, as there was no "training set" for an algorithm. The "ground truth" for the non-clinical evaluation was established through direct comparison of material properties, dimensions, and mechanical performance between the ARCHITEX™ System components and the legally marketed predicate devices, using established ASTM standards where available, and conducting comparative testing where industry standards were absent.
{0}------------------------------------------------
ARCHITEX™ Space Maintenance System 510(k) Summarv
December 2010
DEC 1 3 2010
- Company: Medtronic Sofamor Danek USA 1800 Pyramid Place Memphis, Tennessee 38132 Telephone: (901) 396-3133 (901) 346-9738 Fax:
Contact: Ryan Massey, Sr. Regulatory Affairs Specialist
II. Proposed Proprietary Trade Name: ARCHITEX™ Space Maintenance System
-
- Classification Name(s): Implants, Endosseous, Root-Form per 21 CFR 872.4760; Product Code(s): DZE, JEY
IV. Description:
ARCHITEX™ Space Maintenance System is intended for use in oral-maxillofacial surgical reconstruction and dental regeneration procedures; the implants contained in the System are for maintaining space during bone grating procedures and/or to support soft tissue until bone formation.
The ARCHITEX™ Space Maintenance System consists of a variety of shapes and sizes of screws (Socket Preservation, Tenting, and Mesh Fixation Screws) and titanium mesh implants components as well as placement instruments components. These instruments supplied are general class I instruments added as a convenience for doctors to use to place the provided implants.
The ARCHITEX™ Space Maintenance System includes a selection of Socket Preservation screws which are designed to aid in extraction socket grafting while simultaneously supporting the original gingival margins and papilla.
The ARCHITEX™ Space Maintenance System components are fabricated from medical grade stainless steel and medical grade titanium or titanium alloy. Medical grade titanium and titanium alloy may be used together. The subject implants components will be manufactured from medical grade titanium allov described by such standards as ASTM F 136-08el or ISO 5832-3, and/or from medical grade unalloyed titanium described by such standard as ASTM F67-06. The placement instruments components will be manufactured from medical grade stainless steel as described in ASTM F8PP-09. The ARCHITEX™ Space Maintenance System is sold non-sterile.
The purpose of this 510(k) application is to seek marketing clearance for the ARCHITEX™ Space Maintenance System to be used as a root-form endosseous dental implants system.
{1}------------------------------------------------
IV. Indications for Use:
The ARCHITEX™ Space Maintenance System is indicated for use as temporary implants to stabilize and support autograft, autograft extenders, allograft, bone void fillers and/or fractured bone segments with or without bone plates or titanium mesh in bony defects of oral maxillofacial anatomy.
The Indications for Use Statement referenced above does not differ from the listed predicate devices.
V. Identification of the Legally Marketed Predicate Devices Use to Claim Substantial Equivalence:
The design features, materials, and indications for use of the ARCHITEX™ Space Maintenance System incorporates a variety of functional aspects from the following predicate devices: TiMesh® System (K062348, SE 09/08/06), ACE Tru-FIX™ Implant System (K080074, SE 02/14/08), and Straumann Dental Implant System (K071585, SE 07/11/07),
| ARCHITEX™ Implant | Predicate | Substantial equivalence |
|---|---|---|
| Titanium Mesh | TiMesh® (K062348) | Geometrically, materially,and mechanicallyequivalent |
| Titanium Mesh Screws | TiMesh® (K062348) | Geometrically, materially,and mechanicallyequivalent. |
| Tenting Screws | ACE Tru-FIX™ (K080074) | Geometrically, materially,and mechanicallyequivalent. Bone pull-outstrength and screwsubsidence equivalent topredicate. |
| Socket PreservationScrews | ACE Tru-FIX™ (K080074) | Bone interfacing portion ofscrew geometrically,materially, andmechanically equivalent.Bone pull-out strength andscrew subsidenceequivalent to predicate |
| Straumann Dental ImplantSystem (K071585) | Head portion of screwgeometrically, materially,and mechanicallyequivalent to predicateabutments. |
VI. Brief Discussion of the Non-Clinical Tests Submitted
For a determination of substantial equivalence, the following analysis and bench performance tests were performed on Subject Devices and Predicate Devices:
-
Analysis of Dimensional and Material Features
- ·> Statis Axial Pullout Testing in accordance with ASTM F543-02
{2}------------------------------------------------
-
Statis Subsidence Testing (To date there is no industry standards)
-
Statis Cantilever Bending Testing (To date there is no industry standards)
-
Static "Removal Torque" and "Torque to Failure" testing in accordance with ASTM F543-02
VII. Conclusions Drawn from the Non-Clinical Tests
Results of material and dimensional analysis have demonstrated that ARCHITEX™ Space Maintenance System implants are equivalent to the predicate devices in terms of bone/screw interface features.
When compared with predicate devices, results of bench performance testing indicated all acceptance criteria were met, and demonstrated the subject Tenting Screws and Socket Preservation Screws have equivalent static axial pullout and subsidence strength and that the subject Titanium Mesh has equivalent static cantilever bending strength.
The conclusions drawn from the analysis and performance testing, along with the intended use of the subject devices demonstrate that the ARCHITEX™ Space Maintenance System is substantially equivalent to the listed predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DFC 1 3 2010
Ms. Michelle Obenauer Regulatory Affairs Manager Medtronic Sofamor Danek USA, Incorporated 1800 Pyramid Place Memphis. Tennessee 38132
Re: K100779
Trade/Device Name: ARCHITEX™ Space Maintenance System Regulation Number: 21CFR.872.4760 Regulation Name: Bone Plate Regulatorý Class: II Product Code: JEY, DZE Dated: November 30, 2010 Received: December 1, 2010
Dear Ms. Obenauer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Mr. Obenauer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), a olease go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K100779
Device Name: ARCHITEX™ Space Maintenance System
Indications for Use:
The ARCHITEX™ Space Maintenance System is indicated for use as temporary implants to stabilize and support autograft, autograft extenders, allograft, and bone void fillers and/or fractured bone segments with or without bone plates or ritanium mesh in bony defects of oral maxillofacial anatomy.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
_Yes_X__OR Over-The-Counter Use No Prescription Use Per 21 CFR 801.109
Susan Reamer
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K100776
§ 872.4760 Bone plate.
(a)
Identification. A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.(b)
Classification. Class II.