POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)
K100733 · Rayen Safety Products Limited · LYZ · Apr 30, 2010 · General Hospital
Device Facts
| Record ID | K100733 |
| Device Name | POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED) |
| Applicant | Rayen Safety Products Limited |
| Product Code | LYZ · General Hospital |
| Decision Date | Apr 30, 2010 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.6250 |
| Device Class | Class 1 |
Intended Use
powder free vinyl patient examination glove, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile vinyl patient examination glove; worn on examiner's hand or finger. Acts as physical barrier to prevent cross-contamination between patient and examiner during medical procedures. Used in clinical settings by healthcare professionals. Device is a passive barrier; no electronic or mechanical components. Benefits patient and examiner by reducing risk of pathogen transmission.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including physical property testing per ASTM D5250-06, pinhole leak testing per 21 CFR 800.20, and biocompatibility testing (primary skin irritation and dermal sensitization) per ISO 10993-10.
Technological Characteristics
Material: Vinyl. Form factor: Disposable, non-sterile, clear (non-colored) glove. Standards: ASTM D5250-06 (dimensions, physical properties), ASTM D6124-06 (powder residual <2mg/glove), 21 CFR 800.20 (pinholes). Biocompatibility: ISO 10993-10 (non-irritating, non-sensitizing).
Indications for Use
Indicated for use as a disposable, non-sterile patient examination glove worn by examiners to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
- FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves (K032908)
Related Devices
- K100855 — POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED) · Suzhou Shengyuda Plastic Products Co., Ltd. · Apr 22, 2010
- K082597 — POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED) · Tangshan Linyuan Plastic Products Co., Ltd. · Oct 24, 2008
- K102152 — POWDER-FREE VINYL PATIENT EXAMINATION GLOVES · Hebei Tengda Plastic Co., Ltd. · Apr 28, 2011
- K180381 — Powder Free Vinyl Patient Examination Gloves · Shanxi Hongjin Plastic Technology Co., Ltd. · Nov 7, 2018
- K980093 — DISPOSABLE SYNTHETIC PATIENT EXMINATION GLOVES, POWDER FREE · Universal Gloves Corp. · Mar 2, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
# Summary
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: K100 733 . " (applicant leave blank)
## APR 3 0 2010
#### Premarket Notification [510(k)] Summary
[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :
| Submitter's name : | RAYEN SAFETY PRODUCTS LIMITED |
|--------------------------------|-------------------------------------------------------------|
| Submitter's address : | WEST ROUND ROAD, LUANNAN, TANGSHAN, HEBEI,<br>063500, CHINA |
| Phone number : | (86) 315-4169377 |
| Fax number : | (86) 315-4169376 |
| Name of contact person: | Mr. Jin Zongxiang |
| Date the summary was prepared: | Feb. 25nd, 2010 |
#### [(a)(2)). The name of the device, including the trade or proprietary name if applicable the common or usual name, and the classification name, if known
| Device Name: | Powder Free Vinyl Patient Examination Gloves, Clear(non-colored) |
|-------------------------|--------------------------------------------------------------------------------|
| Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves<br>Other clients private labeling |
| Common Name: | Patient examination glove |
| Classification Name: | Patient examination glove |
| Device Classification: | I |
| Regulation Number: | 21 CFR 880.6250 |
| Panel: | General Hospital (80) |
| Product Code: | LYZ |
#### [(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .
Class I* powder free vinyl patient examination gloves , Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .
#### [(a)(4)] A description of the device
Device Description : powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
... ...
{1}------------------------------------------------
#### [(a)(5)] The summary describes the intended use of the device
Device Intended Use: powder free vinyl patient examination glove, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
#### [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.
| Characteristics | Standard | Device performance |
|-----------------------|-------------------------------------------|-----------------------------------------|
| Dimension | ASTM standard D 5250-06 | Meets |
| Physical Properties | ASTM standard D 5250-06 | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Residual | ASTM standard D 5250-06<br>and D6124-06 | <2mg/glove |
| Biocompatability | Primary Skin Irritation in<br>rabbits | Passes<br>Not a Primary Skin Irritation |
| | Dermal sensitization in the<br>guinea pig | Passes<br>Not a Dermal sensitization |
[(b)(1)] A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its pre-defined acceptance criteria and intended uses.
Powder free vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-06, per ASTM D6124-06, per 21 CFR 800.20 and ISO10993-10. , it is safe and effective, and it's performance meets the requirements of its pre-defined acceptance criteria and intended uses.
[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
[(b)(3)] The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).
It can be concluded that the Powder Free Vinyl Patient Examination Gloves . Clear(non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims . It is as safe,as effective, and performed as well the legally marketed device identified in (a)(3).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Rayen Saftey Products Limited C/O Mr. Chu Xiaoan Room 1606 Building 1, Jianxiang Yuan No. 209, Bei Si Huan Zho Beijing CHINA 100083
APR 3 0 2010
Re: K100733
Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYZ Dated: April 23, 2010 Received: April 26, 2010
Dear Mr. Xiaoan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Mr. Xiaoan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reportung (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Susan Runner
Anthony Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE
Applicant: RAYEN SAFETY PRODUCTS LIMITED
510(k) Number (if known): * K100 733
Device Name: Powder Free Vinyl Patient Examination Gloves, Clear(Non-colored)
Indications For Use:
Powder free vinyl patient examination glove is a disposable non-sterile device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
b.
Division Sign-Off
(Division Sign-Off) (Division Sign-Off) Division of Anton, Dental Devices
K100733 510(k) Number: J