K Number
K100622

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-08-27

(176 days)

Product Code
Regulation Number
880.6880
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Belimed Steam Sterilizer MST-V 3-3-6 is designed for sterilization of non porous and porous heat and moisture-stable materials used in healthcare facilities ... The Belimed Steam Sterilizer MST-V 3-3-6 is equipped with the following factory-programmed Sterilization cycles and cycle values (Table 1).

Device Description

The Belimed Steam Sterlilzers MST-V 3-3-6 is intended for use in hospital and health care facilities and is intended to be used in an identical manner as the Belimed Steam Sterilizer Model 5-5-9. The small chamber size incorporates flexibility, and allows operating the sterilizers in the limited space conditions. The Belimed Steam Sterilizer MST-V 3-3-6 is designed to be used for the thermal sterilization of porous and non-porous, heat and moisture stabile materials in the healthcare facilities. Depending of the chosen cycle materials as different as textiles, unwrapped or wrapped instrument trays can be sterilized. The Belimed Steam Sterilizer MST-V 3-3-6 is factory equipped with cycles which have been tested in accordance with AAMI/ANS ST8:2008 under defined load conditions. The predicate device with a chamber volume 73 I has been validated previously.

AI/ML Overview

The provided 510(k) summary describes a steam sterilizer, a device that does not typically involve AI/ML technology, human readers, or image interpretation. Therefore, many of the requested categories in your prompt are not applicable to this type of device submission.

However, I can extract the acceptance criteria and the summary of the studies performed to demonstrate the device meets these criteria.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from AAMI/ANSI ST8:2008 & ST79:2006 Standards)Reported Device Performance
Sterility Assurance Level (SAL): At least 10⁻⁶ reduction of biological indicators. This is generally demonstrated by achieving a time-at-temperature (Fo) sufficient to produce an Fo of at least 12 minutes by half cycle for relevant cycles.Achieved for all validated cycles: - Empty Chamber Testing (PreVac, PreVac Flash, Gravity cycles): Demonstrated the sterilizer's capability to provide steady-state thermal conditions and sterility assurance level within the chamber corresponding to predicted tooling (Section 5.4.2.5 of ANSI/AAMI ST8). - PreVac Cycles with Fabric Test Pack: Achieved a sterility assurance level of at least 10⁻⁶ through time-at-temperature sufficient to produce an Fo of at least 12 minutes by half cycle, moisture retention of less than 3% increase in pre-sterilization test pack weight, and exhibited no wet spots (Section 5.5.2.5 ANSI/AAMI ST8). - PreVac Cycles with Full Load Instruments Trays: Achieved a sterility assurance level of at least 10⁻⁶ through time-at-temperature sufficient to produce an Fo of at least 12 minutes by half cycle, moisture retention of less than 20% increase in pre-sterilization weight, and exhibited no wet spots on the outer wrapper (Section 5.5.4 of ANSI/AAMI ST8). - Flash Cycles (Unwrapped or Single Wrapped Instrument Trays): Achieved a sterility assurance level of at least 10⁻⁶ through time-at-temperature sufficient to produce an Fo of at least 12 minutes by half cycle (Section 5.5.5.2 AAMI/ANSI ST8:2008 and ANSI/AAMI ST79:2006 section 10.7). - Gravity Cycles (Unwrapped Instrument Tray): Achieved a sterility assurance level of at least 10⁻⁶ through time-at-temperature sufficient to produce an Fo of at least 12 minutes by half cycle (Section 5.5.5.1 AAMI/ANSI ST8:2008).
Bowie-Dick Test Pack Performance: Attainment of a uniform color change throughout the test sheet and the normal minimum pressure phase (Section 5.6 of AAMI/ANSI ST8).Achieved: The Bowie-Dick Test cycle was verified, attaining the required color change throughout the test sheet and normal minimum pressure phase, and exposure temperature within 70 seconds (Section 5.6 of AAMI/ANSI ST8).
Software Validation: Performed according to FDA's moderate level of concern recommendations.Achieved: Software validation for cycle operation was performed according to FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005)" recommendations.
Safety and Performance Requirements: Compliance with relevant safety standards (UL 61010-1:2005, IEC EN 61010-2-040:2005, IEC EN 60601-1-2:2001 +A1:2006, IEC EN 62304:2006, ASME Section VIII, Division 1, EN ISO 14971:2007).Achieved: The device has been designed, constructed, and tested to meet the mentioned safety and performance requirements.
Quality Management System: Certified according to ISO 9001 and ISO 13485.Achieved: Belimed has a certified quality management system according to ISO 9001 and ISO 13485.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify a numerical sample size for the "test set" in terms of number of sterilizer cycles or biological indicator runs. Instead, it refers to qualification studies conducted "as described in" various sections of the AAMI/ANSI ST8:2008 and ST79:2006 standards. These standards typically outline the protocols for validation, which involve performing a sufficient number of cycles to demonstrate reproducibility and efficacy.
  • Data Provenance: The studies were conducted by Belimed Sauter AG, likely at their facilities in Switzerland (given the submitter's address). The data is presumably prospective, as it describes verification studies performed specifically for this device model. The regulatory standards (AAMI/ANSI) are international, though the submission is for the US market.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • This information is not applicable. The "ground truth" for a steam sterilizer's performance is objective - the complete kill of biological indicators and the achievement of a specific Sterility Assurance Level (SAL) confirmed through laboratory testing and physical measurements (e.g., temperature, pressure, time). It does not involve expert interpretation or consensus in the context of medical imaging or diagnostic devices.

4. Adjudication Method for the Test Set

  • This information is not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements among human readers or expert reviewers, typically in diagnostic studies. The evaluation of a sterilizer's performance is based on quantifiable physical and biological outcomes, not subjective interpretations requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC or comparative effectiveness study involving human readers with or without AI assistance was done. This type of study is irrelevant for a steam sterilizer device.

6. Standalone (Algorithm Only) Performance

  • No standalone algorithm performance study was done. This device is a physical sterilizer, not an AI algorithm. Its "performance" is its ability to effectively sterilize materials, not to produce diagnostic outputs. While it does have "software validation," this refers to the control system's proper functioning, not a diagnostic algorithm.

7. Type of Ground Truth Used

  • The ground truth used for validating the sterilizer's performance is:
    • Biological Indicators: Complete kill of robust test organisms (e.g., bacterial spores like Geobacillus stearothermophilus) used in biological indicator preparations.
    • Physical Parameters: Achievement and maintenance of specific temperature, pressure, and time conditions within the sterilizer chamber.
    • Chemical Indicators: Proper color changes or responses of chemical indicators and Bowie-Dick test packs.
    • Moisture Retention: Measurement of moisture content in test packs post-sterilization.
    • These are objective, measurable outcomes in accordance with established sterilization standards (AAMI/ANSI).

8. Sample Size for the Training Set

  • This device does not involve a "training set" in the context of AI/ML. The sterilizer's design and factory-programmed cycles are developed based on engineering principles and established sterilization science, not by training a model on a dataset. The "validation" performed is more akin to verification and performance testing against predetermined standards.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no "training set" for this device. The physical and biological ground truths described in point 7 are used for validation and verification, not for training a machine learning model.

{0}------------------------------------------------

510(k) Summary

K10 0622

Date:August, 10, 2010
Submitter's Name / Address:Belimed Sauter AGZelgstrasse 8CH-8583 Sulgen / SwitzerlandAUG 2 7 2010
Contact Person:Hans Stadler, Head of R&DEmail: hans.stadler@belimed-sauter.chPhone: +41 71 644 85 97Fax: +41 71 644 85 11
Trade Name:Belimed Steam Sterilizer MST-V
Model:3-3-6
Classification:Steam Sterilizer – Class II, as listed per 21 CFR880.6880Product Code 80 FLE
Predicate Device:Belimed Steam Sterilizer TOP 5000Model 5-5-9 (K060337)

DEVICE DESCRIPTION:

The Belimed Steam Sterlilzers MST-V 3-3-6 is intended for use in hospital and health
ed Steam Sterist be interested to be used in an identical manner as the Belimed Steam The Belimed Steam Sterlizers MST-V 3-3-6 is mended for accurrent in the procession
care facilities and is intended to be used in an identical manner as the Belimed Steam
Care

Sterilizer Model 5-5-9.
The small chamber size incorporates flexibility, and allows operating the sterilizers in the The small omited space conditions.

... .. ..

nonclinical comparison to the predicate device:

The Belimed Steam Sterilizer MST-V 3-3-6, is very similar to the predicate device.
The Belimed Steam Sterity on discts device include: The Belimed Steam Otenilizer MOV.
Modifications made from the predicate device include:

  • Smaller chamber size .
  • Optional Steam Generator Integrated .
  • Optional Otcam Scham Scharating panel) .

CLINICAL DATA:

No clinical data is required for this device classification submission.

INTENDED USE:

INTERESED and Sterilizer MST-V 3-3-6 is designed for sterilization of non-porous and
The Belimed Steam Sterials materials used in health care facilities.. The Belimed Steam Sterlizer MST-V 5-3-0 is dolgined for the believed in health care facilities...
porous heat and moisture-stable materials used in health care facilities...

The Belimed Steam Sterilizer MST-V 3-3-6 is equipped with the following factorythe Steam Sterling System and cycle values (Table 1). The Belimed Steam Stemizer MST 7 - 8 - 8 - 8 - 8 - 8 - 8 - 4 - 4 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - -

. ".

:年!

Belimed Steam Sterilizer MST-V 3-3-6 FDA 510(K) Author: HST/ KHA

Release 1.2 / August 10, 2010 Page 1 of 5

{1}------------------------------------------------

Table 1: Factory programmed sterilization cycles
No.CYCLEPRE-TREATMENTSTERILIZETEMP. &PRESSURESTERILIZETIMEDRYTIMERECOMMENDED LOAD
1PreVac 270long Dry3 vacuum pulses:90/100/100 mbar2 Steam pulses:1500/1500 mbar270 °F2880mbar4 min30 minTwo double wrapped instrumenttrays, maximum total weight of50 lbs or 25lbs / tray
2PreVac 270short Dry3 vacuum pulses:90/100/100 mbar2 Steam pulses:1500/1500mbar270 °F2880mbar4 min5 minTwo Fabric Packs total weight ofMaximum 9 lbs orOne single wrapped instrument traywith non- porous single instrument
3Prevac Flash2701 vacuum pulse:90mbar270 °F2880mbar4 min1 minOne unwrapped instrument tray,maximum weight of 25 lbs orOne unwrapped instrument tray withsingle instrument
4Gravity 250Purge time 4 min& Temp. > 98°C250 °F2050mbar25 min0 minOne unwrapped non-porousinstrument tray,maximum weight of 25 lbs
5Bowie-DickTest3 vacuum pulses:90/100/100 mbar2 Steam pulses:1500/1500mbar273 °F3030mbar3.5 min1 minBowie-Dick- or other FDA clearedair removal test pack
6Leak TestVacuum:65mbarTest time: 15min-Empty chamber
7Warm up &leak testVacuum:65mbarTest time: 15min270 °F2880mbar3 min3 minEmpty chamber
8PreVac270 long Dry3 vacuum pulses:90/100/100 mbar2 Steam pulses:1500/1500mbar270 °F2880mbar10 min30 minOne double wrapped instrument tray,maximum total weight of 25 lbs
9PreVacFlash 2701 vacuum pulse:90mbar270 °F2880mbar10 min1 minOne unwrapped instrument tray,maximum weight of 25 lbs orOne single wrapped instrument traywith non- porous single instrument

mmed sterilization cycles

Note on table 1: Factory set dry time is recommended minimum. Extended dry time may be required.
The saling and seel conditions

Note on table 1: Paciory Ser dry time is recomment
Fabric load should be preconditions:
Fabric load should be preconditioned between 68°F and at a relative humidity of at 60% for at least 2 hours.

.

{2}------------------------------------------------

The Belimed Steam Sterilizer MST-V 3-3-6 is offered in the following configurations:

Table 2: Dimensions
-----------------------
Modelsingle doordouble doorConfigurationChamber size(Volume)(L)Chamber size(H x W x D)(inch)Chamber size(H x W x D)(mm)Overall Dimensions(H x W x D)(inch)Overall Dimensions(H x W x D)(mm)
3-3-6 VS13-3-6 VS21 door2 door7313"x 13"x 25"335 x335 x 64060"x 25"x 31"1530 x640 x 800

The Belimed Steam Sterilizer MST-V 3-3-6 is designed to be used for the temmal The Belimed Steam Sterlizer MST-V 3-5-6 1s designed to 10-2001
sterilization of porous and non-porous, heat and moisture stabile materials in the healthcare facilities.

healthcare facilities.
Depending of the chosen cycle materials as different as textiles, unwrapped or wrapped Depending of the chosen oyolo materials assisted as a be sterilized.

The Belimed Steam Sterlizer MST-V 3-3-6 is factory equipped with cycles which have
s The Belimed States of the AMANSLCTR:2008 under defined load conditions. The The Belimed Steam Stellizer WS L-V S-3-0 1s accory oquipped on the Unditions. The
been tested in accordance with AAMI/ANS ST8:2008 under defined headly been tested in accordance with A thin INST Oven validated previously.
predicate device with a chamber volume 73 I has been validated previously.

EFFECTIVENESS:

Efficacy of sterilizer function and exposure time recommendations are ultimately shown Efficacy of sterilizer function and exposure undersens and version of
by complete kill of biological indicators and verifying an appropriate safety factor of by complete kill of biological indications and welling and sparter AG validates its
sterility assurance level (SAL) of at least 10° reduction. Belimed Sauter AG validates its sterility assurance level (SAL) of at least 10 - 1east 10 guidelines developed by
sterilization cycles by recommended practices, standards and guidelines developed of Medical sterilization cycles by recommended practices, standation for Advancement of Medical
various independent organizations such as the Association Sterilizer MSTvarious independent organizations such as the Association of Arantes of Steam Steam Steam Steam Steam Steam
Instrumentation (AAMI). Prior release, Belimed Steam Steam Steam S instrumentation (7V ann).

The results of the Belimed Steam Sterilizer MST-V 3-3-6 verfication studies demonstrate
s for the Belimed Steam Sterial and and are oummarized as follows: The results of the Bellined Steam Otenized MS - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

  • Empty chamber testing performed as described in Section 5.4.2.5 of ANSI/AMMI
    Charles of the Market Marker Mas Flash, ond Gravity, cycles, These cycles, These cycles Empty chamber testing periormed as described in Ocouity cycles. These oycles
    ST8:2008, for the PreVac, PreVac, PreVac, PreVac, Themal Conditions . ST8:2008, for the PreVac, FreVac Flash and Shawly State the Bronditions
    demonstrated the sterlizer's capability to provide steady state the assurance level demonstrated the sterlizers capability to provide sterility assurance level within the chamber that are corresponding with the predicted tooling as and one of the steel of Section 4.4.2.5 of ANSI/AAMI ST8.
  • of ANSI/AAMI ST8.
    All PreVac cycles verified using the fabric test pack, as described in Section 5.5.2.5 ANSI/AAMI ST8. All PreVac cycles verified using the fabric test pack, as asociated in ST8.
    AAMI/ANSI ST8:2008 were qualified according to section 5.5.2.5 ANSI/AAMI ST8. ● AAMI/ANSI ST8:2008 were qualified assurance level of at least 10° through
    These cycles demonstrated a sterility assurance level of at least 10° through These cycles demonstrated a stellity assurance level of an Far of at least 12
    achievement of a time-at-temperature sufficient to produce an Foresterilization achievement of a time-at-temperature retention of less than 3 % increase in pre-sterilization
    minutes by half cycle, moisture retention of less than 3 % increase in pre-steri test pack weight, and exhibited no wet spots.
  • test pack weight, and exhibited no were spots.
    All PreVac cycles verified using full load instruments trays were qualified a sterility All PreVac cycles verified using tull load instruments trays world in the mail.
    section 5.5.4 of ANSI/AAM ST8:2008 These cycles demonstrature . section 5.5.4 of ANSIAAMI S 18:2008 These cycles donited assurance level of at least 10° through achievement of a time and the produce and

{3}------------------------------------------------

less than 20 % increase in pre-sterilization weight of the towel, and exhibited no wet spots on the outer wrapper.

  • spots on the outer wrapped or single wrapped or single wrapped instrument trays . All Flash cycles vernied using the unwrapped of eingle star.
    were qualified according to section 5.5.5.2 AAMI/ANSL ST8:2008 and ANSI/AAM were qualified according to Section 2.0.12 19:00:21 assurance level of at
    ST79:2006 section 10.7. These cycles demonstrated a sterlity assurance level of For S / 79.2006 section 10.7. These cycles demonstrated a commy
    least 10° through achievement of a time-at-temperature sufficient to produce an F。of at least 12 minutes by half cycle.
  • All Gravity cycles verified using the unwrapped instrument tray, were qualified . according to section 5.5.5.1 AAMI/ANSI ST8:2008. These cycles demonstrated a according to section 5.3.3.1 Anni/ANOT Of Schievement of a time at temperature
    sterility assurance level of at least 10 fthrough actually half a time at temperature sufficient to produce an Fo of at least 12 minutes by half cycle.
  • The Bowie Dick Test cycle was verified using the Bowie-Dick Test Pack was qualified . according to section 5.6 of AAMI/ANSI STB, and democrative devices, attained the according to section 5.6 of AnimANO. One temperature devices attained the change infoughout the test sheet and the normal similar made prosure phase.
  • exposure temperature within To seconds or progress performed according to FDA's
    The software validation for the cycle operation was performed according to FDA's . moderate level of concern recommendations provided in the document "Guidance for moderate level of concern recommendations provided in the are and in Medical Devices (May 2005)".

SAFETY:

Belimed Sauter AG's sterilizers including the Belimed Steam Steam Stealizer MST-V 3-6 have Belimed Sauter AG's stemizes including the Belined Cetuand performance requirements
been designed, constructed and tested to meet the safety and performance requirements been designed, constructed and tested to hieet the Saley and porchile of the most of the Steam Sterilizer MST-V complies with the following safety standards:

  • UL 61010-1:2005 1.
  • IEC EN 61010-2-040:2005 2.
  • IEC EN 60601-1-2:2001 +A1:2006 3.
  • IEC EN 62304:2006 4.
  • IEC EN 62304:2006
    American Society of Mechanical Engineers (ASME), Section VIII, Division 1 for
    American Society of Mechanical Engineers (Add, 2008 5. unfired pressure vessels Ed. 2007 +Add. 2008
  • EN ISO 14971:2007 6.

Belimed has a certified quality management system according to ISO 9001 and ISO 13485.

HAZARDS-FAILURE OF PERFORMANCES

Failure of the sterilization process can lead to incidence of cross contamination, the Failure of the sterilization process can lead to includes or order common and the world.
transmission of potentially infectious organisms from one infected person to another was not otherwise infected prior to the incident.

was not otherwise intected pror to the includent.
To avoid failure, the user music ensure the materials, instructions for use are To avoid failure, the user must ensure the manufacturer's instructions for use are
sterilized are thoroughly cleaned, that the manufacturer's instructions for use are sterilized are thoroughly cleaned, that the manufacturer of has been validated, that
followed, that the cycle to be used for each type of sterliizer's manufacturer's followed, that the cycle to be used to each type of sterliizers manufacturers manufacturers manufacturers and is the sterilizer has been mantained in accordance with the time startiizer load is
maintenance schedule and is operating properly, and themical sterilization process maintenance schedule and is operating propeny, and that oach end of the are and interest in indicators.

Today, there are many steam sterlizers in daily use in hospitals throughout the United
The Sun States of Schooline assefyration or sterilization process failure is relativel Today, there are many steam sterlizers in daily use in nospitals and organisations is relatively

{4}------------------------------------------------

rare considering the wide spread use of steam sterilizers. Further, there are no known reports in the literature of patient infection that have resulted from steam sterilizer failure. The technology designed in Belimed Steam Sterilizer MST-V provides microprocessor controller safeguards that abort the cycle and give appropriate signals, alerts and warnings when required conditions have not been met or when a malfunction occurs.

USER INFORMATION

Belimed Sauter AG provides information to the user that is intended to insure safe and effective use of steam sterilization in its detailed User's Manual and other labeling. Belimed Sauter AG also recommends the use and periodic review of the AAMI steam sterilization standards to ensure further assurance of the safe and effective use of steam sterilization equipment in health care facilities.

CONCLUSION

The Belimed Steam Sterilizer MST-V 3-3-6 is a substantially equivalent device to that of the predicate device. There have been no substantial changes in technology and intended use to the predicate device model 5-5-9 (K060337). This steam sterilizer MST-V 3-3-6 meets the applicable requirements of the applicable standards.

Based on the provided information in this premarket notification, it can be concluded that the subject device is substantial equivalent to the predicate device and is safe and effective when used as intended.

Belimed Steam Sterilizer MST-V 3-3-6 FDA 510(K) Author: HST/ KHA

Release 1.2 / August 10, 2010 Page 5 of 5

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Belimed Sauter AG C/O Mr. Stefan Preiss Responsible Third Party Official TÜV SÜD America 1775 Old Highway 8 NW New Brighton, Minnesota 55112-1891

AUG 27 2010

Re: K100622

Trade/Device Name: Belimed Steam Sterilizer MST-V 3-3-6 Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: August 16, 2010 Received: August 23, 2010

Dear Mr. Preiss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include, requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2- Mr. Preiss

. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing . practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Susan Kanrer

Anthony D. Watson, B.S. M.S. M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

K100622
M10-9

AUG 2 7 2010

510(k) Number: K100622

Device Name: Belimed Steam Sterilizer MST-V 3-3-6

Indications for Use:

The Belimed Steam Sterilizer MST-V 3-3-6 is designed for sterilization of non porous and porous heat and moisture-stable materials used in healthcare facilities ...

The Belimed Steam Sterilizer MST-V 3-3-6 is equipped with the following factory-programmed Sterilization cycles and cycle values (Table 1).

Table 1: Factory programmed sterilization cycles

No.CYCLEPRE-TREATMENTSTERILIZETEMP. &PRESSURESTERILIZETIMEDRYTIMERECOMMENDED LOAD
1PreVac 270long Dry3 vacuum pulses:90/100/100 mbar2 Steam pulses:1500/1500 mbar270 °F2880mbar4 min30 minTwo double wrapped instrumenttrays, maximum total weight of50 lbs (max. 25lbs / tray)
2PreVac 270short Dry3 vacuum pulses:90/100/100 mbar2 Steam pulses:1500/1500mbar270 °F2880mbar4 min5 minTwo Fabric Packs total weight ofMaximum 9 lbs orOne single wrapped instrument traywith non- porous single instrument
3Prevac Flash2701 vacuum pulse:90mbar270 °F2880mbar4 min1 minOne unwrapped instrument tray,maximum weight of 25 lbs orOne single wrapped instrument traywith non- porous single instrument
4Gravity 250Purge time 4 min& Temp. > 98°C250 °F2050mbar25 min0 minOne unwrapped non-porousinstrument tray,maximum weight of 25 lbs
5Bowie-DickTest3 vacuum pulses:90/100/100 mbar2 Steam pulses:1500/1500mbar273 °F3030mbar3.5 min1 minBowie-Dick- or other FDA clearedair removal test pack
6Leak TestVacuum:65mbarTest time: 15minEmpty chamber

Page 1 of 2

{8}------------------------------------------------

No.CYCLEPRE-TREATMENTSTERILIZETEMP. &PRESSURESTERILIZETIMEDRYTIMERECOMMENDED LOAD
7Warm up &leak testVacuum:65mbarTest time: 15min270 °F2880mbar3 min3 minEmpty chamber
8PreVac270 long Dry3 vacuum pulses:90/100/100 mbar2 Steam pulses:1500/1500mbar270 °F2880mbar10 min30 minOne double wrapped instrumenttray,maximum total weight of 25 lbs
9PreVacFlash 2701 vacuum pulse:90mbar270 °F2880mbar10 min1 minOne unwrapped instrument tray,maximum weight of 25 lbs orOne single wrapped instrument traywith non- porous single instrument

Note on table 1: Factory set dry time is recommended minimum. Extended dry time may be required depending on local conditions.

local conditions.
Fabric load should be preconditioned between 68°F and at a relative humidity of at least 35% to 60% for at least 2 hours.

The Belimed Steam Sterilizer MST-V 3-3-6 is offered in the following configurations:

Table 2: Dimensions

Modelsingle doordouble doorConfigurationChamber size(Volume)(L)Chamber size(H x W x D)Overall Dimensions(H x W x D)
3-3-6 VS13-3-6 VS21 door2 door73(inch)13"x 13"x 25"(mm)335x 335x 640(inch)60"x 25"x 31"(mm)1530x640x800

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use x_OTC (21 CFR 807 Subpart C)

Concurrence of CDRH, Office of Device Evaluation (ODE)

f Device Evaluation (ODE)

Susa Rusa

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K006623

Page 2 of 2

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).