K Number
K100446

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-03-05

(16 days)

Product Code
Regulation Number
870.5800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CircuFlow 5100 Sequential Compression Device is a manual, sequential, pneumatic compression device, intended for the primary or adjunctive treatment of primary or secondary lymphedema. The device is also intended for the additional or alternate treatment of venous insufficiency, and chronic venous stasis ulcers associated with venous insufficiency, as well as general treatment of swelling of the extremities. The device is intended for home or hospital use.

Device Description

The CircuFlow 5100 is a sequential pneumatic compression device designed to apply compression to a limb. The device is composed of two components:

  • Pneumatic manual pump
  • Limb sleeve or garment composes of 4 chambers
    The CircuFlow 5100 enables different treatment pressures (0-120mmHg) and treatment times (30 minutes to continue) that should be used according to physician prescription. When activated, air flows into chamber, the pump provides gradient pressurization to the chambers (sequential inflation of distal to proximal, with distal chambers inflated to a greater pressure than the proximal ones). After each chamber is inflated, the pressure is held constant until all chambers are inflated, in order to prevent reverse gradient flow. Once all chambers are inflated, they are then all released simultaneously, and the cycle repeats. Pressure within chambers are adjustable -pressure to chamber 1 is controlled by user-adjusted regulator on the pump. Pressure in chambers 2, 3 & 4 are individually lowered according to the default factory settings. A calibrated dial gauge displays pressure in the range of 0-200 mmHg. Garments are available in 8 sizes, and custom garments are available in variations from standard sizes.
AI/ML Overview

The provided text describes a 510(k) premarket notification for the CircuFlow 5100 Sequential Compression Device. This submission primarily focuses on demonstrating substantial equivalence to predicate devices through technological characteristics, intended use, and performance testing, rather than detailed clinical study outcomes to prove specific acceptance criteria met by the device.

Based on the provided document, here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative "acceptance criteria" in the format of a separate table with specific metrics (e.g., pressure output within X mmHg tolerance, cycle time within Y% of target) for various performance aspects. Instead, it lists various "Performance Testing" categories that were performed to assure the product meets its specifications. The specific results of these tests and the acceptance criteria they were measured against are not provided in this summary.

Acceptance Criteria (Inferred from Test Categories)Reported Device Performance
Electrical Safety (e.g., Dielectric Strength, Leakage Current)"assured that the product meets its specifications"
Gradient Performance"assured that the product meets its specifications"
Garment Performance"assured that the product meets its specifications"
Pressure Performance"assured that the product meets its specifications"
Tensile Strength"assured that the product meets its specifications"
Material Biocompatibility (e.g., Cytotoxicity, Sensitization, Irritation)"assured that the product meets its specifications"

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "Bench and laboratory testing was performed," but it does not specify any sample sizes for these tests. It also does not mention data provenance (e.g., country of origin, retrospective or prospective) as these were likely internal lab tests on device prototypes or components. There is no indication of human subject data being used for this submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Given that the performance testing described is "Bench and laboratory testing," there is no mention of "experts" or "ground truth" being established in the context of clinical scenarios, image interpretation, or disease diagnosis. The ground truth for these tests would be the established engineering standards or internal specifications against which the device's physical and electrical performance was measured.

4. Adjudication Method for the Test Set

As the performance testing consists of bench and laboratory tests, the concept of an adjudication method (like 2+1 or 3+1 for expert review) is not applicable. The results of these tests would likely be objective measurements taken by testing personnel.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This document describes a 510(k) submission focused on demonstrating substantial equivalence through technical characteristics and bench testing, not a clinical study involving human readers or comparative effectiveness with or without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

The device described, the CircuFlow 5100, is a sequential pneumatic compression device, a hardware medical device. There is no mention of any algorithms or AI components that would require standalone performance testing (without human-in-the-loop).

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

For the bench and laboratory performance tests, the "ground truth" would be the engineering specifications, established physical limits, and relevant standards (e.g., IEC 60601-1-1, ISO 10993) against which the device's performance was measured. There is no mention of ground truth based on expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set

Since there is no mention of any AI components or algorithms in the CircuFlow 5100, there is no training set and therefore no sample size for a training set.

9. How the Ground Truth for the Training Set Was Established

As there is no training set, this question is not applicable.

{0}------------------------------------------------

SECTION 5

K100446

MAR - 5 2010

510(K) SUMMARY

Submitter:

Devon Medical, Inc.

Contact Person:

Ruth Wu, COO 1100 First Avenue, Suit 100

King of Prussia, PA 19406

Phone: 800.571.3135

Fax: 484.636.3380

Common Classification & Proprietary Names:

Common Names: Sequential Compression Device Proprietary Name: CircuFlow 5100

Date Prepared:

December 31st 2009

Classification

The classification name, 21 CFR Part and Paragraph number, product code and classification of the CircuFlow 5100.

Classification Name21 CFR SectionProduct CodeClass
Compressible Limb Sleeve870.5800JOWII

Predicate Devices:

The CircuFlow 5100 Sequential Compression Device is substantially equivalent to the following.

Predicate DeviceManufacturer510(k)#
SC-3008 Sequential CirculatorBio Compression Systems Inc.K043423
GS-128 Sequential CompressionSystemMedMark TechnologiesK050584

Device Description

The CircuFlow 5100 is a sequential pneumatic compression device designed to apply compression to a limb. The device is composed of two components:

  • Pneumatic manual pump
  • · Limb sleeve or garment composes of 4 chambers

The CircuFlow 5100 enables different treatment pressures (0-120mmHg) and treatment times (30 minutes to continue) that should be used according to physician prescription. When activated, air flows into chamber, the pump provides gradient pressurization to the chambers (sequential inflation of distal to proximal, with distal chambers inflated to a greater pressure than the proximal ones).

After each chamber is inflated, the pressure is held constant until all chambers are inflated, in

{1}------------------------------------------------

SECTION 5

order to prevent reverse gradient flow. Once all chambers are inflated, they are then all released simultaneously, and the cycle repeats. Pressure within chambers are adjustable -pressure to chamber 1 is controlled by user-adjusted regulator on the pump. Pressure in chambers 2, 3 & 4 are individually lowered according to the default factory settings.

A calibrated dial gauge displays pressure in the range of 0-200 mmHg.

Garments are available in 8 sizes, and custom garments are available in variations from standard sizes.

Intended Use:

The CircuFlow 5100 is a sequential gradient compression pneumatic device, intended for the primary or adjunctive treatment of primary or secondary lymphedema. The device is also intended for the additional or alternate treatment of venous insufficiency, and chronic venous static ulcers associated with venous insufficiency, as well as general treatment of swelling of the extremities. The device is intended for home or hospital use.

Technological Characteristics:

The manufacturer believes that the technological characteristic of the CircuFlow 5100 are substantially similar to those of the predicate devices.

The CircuFlow 5100 has very similar components to its predicate devices and very similar principles of operation. The device consists of an electrically generated source of compressed air, tubing to convey the pressurized air to the sleeve, like the predicates, pressure is applied cyclically for a specified period of time, according to the physician's prescription.

Performance Testing

Bench and laboratory testing was performed and assures that the product meets its specifications. The manufacturer believes that the technological characteristics of the CircuFlow 5100 are substantially similar to those of the predicate devices. The performance testing includes the following tests:

Electrical Safety Tests
Dielectric Strength Test
Leakage Current Test
Gradient Performance Test
Garment Tests
Pressure Performance Test
Tensile Strength Test
Material Biocompatibility Tests
Cytotoxicity Test
Sensitization Test
Irritation and Intracutaneous Reactivity Test

{2}------------------------------------------------

SECTION 5

Standards

The CircuFlow 5100 conforms to the following standards: IEC 60601-1-1 IEC 60601-1-2 UL 60601-1 ISO 10993 ISO 14971

Statement of Substantial Equivalence

The CircuFlow 5100 is substantially equivalent in technology, function, operating parameters, and intended use to predicate devices that are currently commercially available and in distribution, and does not raise any new questions of safety or effectiveness.

Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, Devon Medical Inc, believes that the CircuFlow 5100, is safe and effective and substantially equivalent to the predicate devices as described herein.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird with its wings spread, rendered in a simple, abstract design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

MAR - 5 2010

Devon Medical Inc. c/o Mr. Mark Job Regulatory Reviewer Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313

Re: K100446 Circuflow 5100 Sequential Compression Device Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: February 16, 2010 Received: February 17, 2010

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Mark Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Duna R. Ladner

Image /page/4/Picture/7 description: The image shows a signature next to the letters 'Br' and 'Di'. The signature is illegible, but it appears to be a cursive signature. The letters 'Br' and 'Di' are in a simple font and are aligned vertically. The image is black and white and has a simple composition.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

K100446 510(k) Number (if known):

Device Name: the Devon CircuFlow 5100 Sequential Compression Device

Indications for Use:

The CircuFlow 5100 Sequential Compression Device is a manual, sequential, pneumatic compression device, intended for the primary or adjunctive treatment of primary or secondary lymphedema. The device is also intended for the additional or alternate treatment of venous insufficiency, and chronic venous stasis ulcers associated with venous insufficiency, as well as general treatment of swelling of the extremities.

The device is intended for home or hospital use.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

sumer D. Vcliner

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_ki 00446

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).