K Number
K100287

Validate with FDA (Live)

Date Cleared
2010-07-06

(158 days)

Product Code
Regulation Number
866.3332
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The JBAIDS Influenza A/H5 (Asian lineage) Detection Kit is intended for use in real-time reverse-transcriptase polymerase chain reaction (rRT-PCR) assays on the Joint Biological Agent Identification and Diagnostic System (JBAIDS) instruments for the in vitro qualitative detection of Influenza A/H5 (Asian lineage) viral RNA in patient nasopharyngeal swab (NPS) or throat swab (TS) specimens for the presumptive laboratory identification of Influenza A/H5 (Asian lineage) virus.

Testing with the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit should be in conjunction with other laboratory testing and clinical observations for the following indications:

  1. Providing epidemiological information for the surveillance of human infection with Influenza A/H5 (Asian lineage) virus.
  2. Identifying patients who may be infected with Influenza A/H5 (Asian lineage) virus based on clinical and epidemiological risk factors.

Testing with the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit should not be performed unless the patient meets the most current U.S. Department of Health and Human Services (DHHS) clinical and epidemiologic criteria for testing suspected A/H5 specimens.

The definitive identification of influenza A/H5 (Asian lineage) either directly from patient specimens or from virus cultures requires additional laboratory testing, along with clinical and epidemiological assessment in consultation with national influenza surveillance experts.

Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.

Use is limited to laboratories with appropriate biosafety equipment and containment procedures. It is intended for use by experienced laboratory personnel who have training in standardized molecular testing procedures and expertise in viral diagnosis, and have received training on the JBAIDS Instrument.

If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a biosafety laboratory (BSL) 3+ facility is available to receive and culture specimens.

Device Description

The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Influenza A/H5 (Asian lineage) Detection Kit is a real-time reverse transcriptase polymerase chain reaction (rRT-PCR) test kit, which, when used with the JBAIDS instrument and software, allows the qualitative in vitro detection of Influenza A/H5 (Asian lineage) viral RNA. The kit contains two freeze-dried assays with primer and fluorescent-probe sets for the detection of Influenza A/H5 (Asian lineage) viral RNA. In particular, the two assays specifically target distinct regions of the influenza A hemagglutinin gene of the highly pathogenic H5N1 viruses from the Asian lineage, without detection of other influenza A virus subtypes, including the North American lineage influenza A/H5 viruses. The tests are performed using the previously FDA-cleared JBAIDS instrument and software. A human gene target assay will be used as an inhibition and extraction control.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria / GoalReported Device Performance
Analytical Sensitivity (Limit of Detection - LoD)To detect Influenza A/H5 (Asian lineage) viral RNA at a specified concentration.50 EID50/mL demonstrated for both A/Vietnam/1203/2004 x A/Puerto Rico/8/34 reassortant and A/Anhui/01/2005/ x A/Puerto Rico/8/34 reassortant strains in NPS and TS specimens, confirmed with 20 individual samples at this level.
Analytical Specificity (Inclusivity & Reactivity)To detect various Influenza A/H5 (Asian lineage) strains.Detected all eight tested Influenza A/H5 (Asian lineage) strains. Lowest detection levels ranged from 10 TCID50/mL to 100 TCID50/mL depending on the strain. Spike levels of 500-1000 TCID50/mL showed similar performance to 50 EID50/mL egg-grown viruses.
Analytical Specificity (Exclusivity / Cross-reactivity)To not cross-react with non-target organisms.100% negative results for all 45 tested non-target organisms (including various influenza viruses, other non-influenza viruses, bacteria, and fungi) when spiked at high concentrations.
Clinical SpecificityTo demonstrate high specificity in clinical samples.At least 99% specificity (with 95% confidence) for both NPS and TS specimens purified with either IT 1-2-3™ VIBE or Platinum Path Purification Kits.
Clinical Sensitivity (Surrogate)To detect Influenza A/H5 (Asian lineage) in surrogate clinical samples.98% success (126 out of 128 positive results) for Influenza A/H5 virus-containing surrogate specimens. All seasonal influenza (48 total results) and influenza-negative specimens (24 total results) produced negative results, as expected. This also implicitly supports analytical specificity against seasonal influenza in a clinical context.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Analytical Sensitivity (LoD): 20 individual specimens (from independent donors) per specimen type (NPS, TS) per purification kit (VIBE, Platinum Path) for confirmation (total 80 confirmed LoD tests). The provenance is not explicitly stated as country of origin, but it implies laboratory-prepared specimens. This was a prospective study design for LoD determination.
  • Analytical Specificity (Inclusivity & Reactivity): 8 Influenza A/H5 (Asian lineage) strains, each spiked into TS specimens at multiple concentrations. The exact number of replicates per concentration is not detailed, but it was "multiple concentrations."
  • Analytical Specificity (Exclusivity): 45 non-target organisms, each spiked into TS specimens at high concentrations. The exact number of replicates is not detailed.
  • Clinical Performance (Surrogate Clinical Sensitivity):
    • Influenza A/H5 Strains: 8 strains, each at 4 concentrations (LoD, 5x LoD, 10x LoD, 100x LoD).
    • Seasonal Influenza Strains: 6 strains, each at 2 concentrations (LoD, 100x LoD).
    • Unspiked samples: 6 specimens.
    • All panels were spiked into TS or NPS specimens. Each specimen was purified with two different kits (IT 1-2-3 M VIBE and IT 1-2-3 ™ Platinum Path).
    • This results in a total of:
      • 8 (H5 strains) * 4 (concentrations) * 2 (specimen types) * 2 (purification kits) = 128 H5-containing samples.
      • 6 (seasonal strains) * 2 (concentrations) * 2 (specimen types) * 2 (purification kits) = 48 seasonal influenza samples.
      • 6 (unspiked) * 2 (specimen types) * 2 (purification kits) = 24 unspiked samples.
    • Total test set for surrogate clinical sensitivity: 128 + 48 + 24 = 200 specimens. These were laboratory-prepared "surrogate" clinical samples. This indicates a prospective design for testing these surrogate samples.
  • Clinical Specificity:
    • NPS specimens: 314 (VIBE purified) + 299 (Platinum Path purified) = 613 specimens.
    • TS specimens: 298 (VIBE purified) + 283 (Platinum Path purified) = 581 specimens.
    • The total number of unique specimens used for clinical specificity is not explicitly stated (some could be both NPS and TS from the same patient, but the numbers suggest separate sample pools).
    • Total specimens tested for clinical specificity = 613 + 581 = 1194.
    • Data Provenance: Frozen, banked NPS and TS specimens previously tested for respiratory pathogens. They were obtained and tested at 3 different test sites. No specific country of origin is mentioned, but "US Department of Health and Human Services" context suggests US-based data. This is a retrospective study design using banked specimens.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly mention the number or qualifications of experts used to establish ground truth for the test set.

  • For Analytical Performance (LoD, Inclusivity, Exclusivity) and Surrogate Clinical Sensitivity, the "ground truth" was established by the precise spiking of known quantities of viral strains into samples by the study designers. This is a laboratory-controlled ground truth, not based on expert interpretation.
  • For Clinical Specificity, the ground truth was based on the results of a "CDC rRT Flu Panel" (the predicate device) which had previously tested the banked clinical specimens. The document states "Specimens were removed from the study if they did not have a valid result for the CDC comparator assay," implying the CDC panel's results served as the reference. The expertise involved in the original testing of these banked samples by the CDC panel is not detailed here.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set.

  • For analytical studies and surrogate clinical sensitivity, the ground truth was defined by the experimental setup (spiking concentrations, known positive/negative samples).
  • For clinical specificity, the predicate device's results served as the comparison, implying a direct comparison rather than an adjudication process of ambiguous results. The JBAIDS software provides automated analysis (positive, negative, uncertain), and invalid/uncertain results required retesting, but this is a retest protocol, not an adjudication process involving experts to resolve discrepancies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.

This device is a molecular diagnostic assay (rRT-PCR kit) that provides automated results (positive, negative, uncertain). The output is interpreted by the JBAIDS software, not by multiple human readers. Therefore, there is no "human-in-the-loop" performance that would be assessed in an MRMC study or an effect size demonstrating human reader improvement with AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies presented are effectively standalone performance studies of the algorithm/device. The device and its accompanying software provide automated interpretation (positive, negative, uncertain) of the rRT-PCR results. The performance characteristics described (LoD, inclusivity, exclusivity, clinical specificity, surrogate clinical sensitivity) represent the device's ability to achieve these results directly from the sample, without human interpretative input into the primary result. Human intervention is limited to operating the instrument, preparing samples, and initiating retesting for uncertain/invalid results, not interpreting the raw amplification curves as "positive" or "negative."

7. The Type of Ground Truth Used

  • Analytical Studies (LoD, Inclusivity, Exclusivity): Laboratory-controlled ground truth (known viral strains spiked at specific concentrations).
  • Surrogate Clinical Sensitivity: Laboratory-controlled ground truth (known viral strains spiked into clinical matrices at specific concentrations).
  • Clinical Specificity: Comparator Assay ground truth. The "CDC rRT Flu Panel" was used as the reference standard for the banked clinical specimens.

8. The Sample Size for the Training Set

The document does not provide information about a separate "training set" or its sample size. This is common for molecular diagnostic kits like this. The device is likely developed and validated using a structured process that involves internal data for assay optimization and parameter setting (which could broadly be considered "training" in a development sense), but there isn't a formally defined "training set" in the context of machine learning model development as typically described in AI/ML device studies. The performance characteristics are derived from the test sets described above.

9. How the Ground Truth for the Training Set Was Established

Since no specific "training set" is identified in the document, there's no information on how its ground truth would have been established. Any internal development and optimization would likely have relied on laboratory-controlled samples with known viral presence/absence and concentration.

{0}------------------------------------------------

K 100287

510(k) Summary

JUL - 6 2010

Idaho Technology Inc. JBAIDS Influenza A/H5 (Asian lineage) Detection Kit

Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitted by:

U.S. Army Medical Material Development Activity Division of Regulated Activities and Compliance 1430 Veterans Drive Fort Detrick, MD 21702-9232

Telephone: 301-619-0317 Facsimile: 301-619-0197

Contact Person: Robert E. Miller, Ph.D. Date Prepared: June 15, 2010

Device name:

Trade name:

JBAIDS Influenza A/H5 (Asian lineage) Detection Kit

Common name:

Reverse-transcriptase real-time PCR assay for targeted Influenza A/H5 (Asian lineage) RNA sequences

Classification name: CFR 21.866.3332 Reagent Kit: Reagents for Detection of Specific Novel Influenza A Viruses (Class II)

{1}------------------------------------------------

Device Description: The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Influenza A/H5 (Asian lineage) Detection Kit is a real-time reverse transcriptase polymerase chain reaction (rRT-PCR) test kit, which, when used with the JBAIDS instrument and software, allows the qualitative in vitro detection of Influenza A/H5 (Asian lineage) viral RNA. The kit contains two freeze-dried assays with primer and fluorescent-probe sets for the detection of Influenza A/H5 (Asian lineage) viral RNA. In particular, the two assays specifically target distinct regions of the influenza A hemagglutinin gene of the highly pathogenic H5N1 viruses from the Asian lineage, without detection of other influenza A virus subtypes, including the North American lineage influenza A/H5 viruses. The tests are performed using the previously FDA-cleared JBAIDS instrument and software. A human gene target assay will be used as an inhibition and extraction control.

Assay Principle: Before testing, specimens are purified using Idaho Technology's 1-2-3TM Sample Purification Kits. The resulting purified sample is added to Unknown reagent vials and a Sample Control reagent vial, along with reconstitution buffer. When the viral RNA is present, a fragment of Influenza A/H5 RNA is transcribed and amplified. The amplicon is detected by fluorescence using a specific hydrolysis probe. Each probe is labeled on one end with a fluorescent reporter moiety (6-carboxyfluorescein (6-FAM)) and elsewhere with a quencher moiety (carboxy tetramethylrhodamine (TAMRA)). When the probe is intact, the quencher absorbs the light emitted by the reporter moiety. During PCR. the probe hybridizes to the target sequence before the exonuclease activity of Taq polymerase hydrolyzes the probe, separating the fluorophore from the quencher and permitting detection of the fluorescent signal generated by the reporter. The fluorescent signal increases as additional templates are amplified and more probes are hydrolyzed.

JBAIDS Software analyzes the fluorescence amplification curves and reports results as positive, negative, or uncertain. A failure of the Positive or Negative Control will result in the entire run being called invalid. Failure of the Sample Control vields a result of "sample control failure" when the associated sample has a negative result for the target assay. Retesting is required to resolve uncertain, invalid or sample control failure results.

{2}------------------------------------------------

Intended Use:

The JBAIDS Influenza A/H5 (Asian lineage) Detection Kit is intended for use in real-time reverse-transcriptase polymerase chain reaction (rRT-PCR) assays on the Joint Biological Agent Identification and Diagnostic System (JBAIDS) instruments for the in vitro qualitative detection of Influenza A/H5 (Asian lineage) viral RNA in patient nasopharyngeal swab (NPS) or throat swab (TS) specimens for the presumptive laboratory identification of Influenza A/H5 (Asian lineage) virus.

Testing with the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit should be in conjunction with other laboratory testing and clinical observations for the following indications:

    1. Providing epidemiological information for the surveillance of human infection with Influenza A/H5 (Asian lineage) virus.
    1. Identifying patients who may be infected with Influenza A/H5 (Asian lineage) virus based on clinical and epidemiological risk factors.

Testing with the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit should not be performed unless the patient meets the most current U.S. Department of Health and Human Services (DHHS) clinical and epidemiologic criteria for testing suspected A/H5 specimens.

The definitive identification of influenza A/H5 (Asian lineage) either directly from patient specimens or from virus cultures requires additional laboratory testing, along with clinical and epidemiological assessment in consultation with national influenza surveillance experts.

Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.

Use is limited to laboratories with appropriate biosafety equipment and containment procedures. It is intended for use by experienced laboratory personnel who have training in standardized molecular testing procedures and expertise in viral diagnosis, and have received training on the JBAIDS Instrument.

If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a biosafety laboratory (BSL) 3+ facility is available to receive and culture specimens.

Indications for Use Same as Intended Use

{3}------------------------------------------------

Same as Intended Use

Special conditions for use statement(s)

  • . For prescription use only
  • Although this test has been shown to detect cultured human-derived Influenza A subtype . H5N1 virus (Asian lineage), the performance characteristics of this test with specimens from humans infected with influenza H5N1 (Asian lineage) or other avian influenza viruses are unknown.
  • Clinical specificity was established when influenza A/H3 and A/H1 were the . predominant influenza A viruses in circulation. When other influenza A viruses are emerging, clinical specificity may vary.
  • Optimum specimen types and timing for peak viral levels during infections caused by a . novel influenza A virus have not been determined. Collection of multiple specimens from the same patient may be necessary to detect the virus.
  • . This product can be used only with the JBAIDS instrument.
  • . Results from this test are intended for use with other laboratory tests and in consultation with national influenza experts.
  • A negative result does not exclude the possibility of Influenza A/H5 (Asian lineage) . infection. Negative test results may occur from improper specimen collection, presence of inhibitors, technical error, or sample mix-up. Test results may be affected by concurrent antiviral therapy or levels of organism in the specimen that are below the limit of detection for the test.
  • The performance of the assay has not been established in individuals who received an . H5N1 vaccine.
  • . This test has not been validated for testing specimens other than those specified or for samples that have been purified using other than the recommended IT 1-2-3TM Sample Purification Kits.
  • As for any PCR assay, improper technique can lead to cross-contamination and false . positive results. Good laboratory technique is essential to the proper performance of this assay. The JBAIDS Influenza A/H5 (Asian lineage) Detection Kit is intended for use by laboratory personnel who have received training on the JBAIDS instrument to perform and interpret the results from this procedure.
  • The JBAIDS Influenza A/H5 (Asian lineage) Detection Kit may not generate . reproducibly positive results when testing samples that have influenza A/H5 (Asian

{4}------------------------------------------------

lineage) viral RNA concentrations lower than the LoD concentration, but higher than the assay cutoff concentration.

Substantial Equivalence: The JBAIDS Influenza A/H5 (Asian lineage) Detection Kit is substantially equivalent to the Centers for Disease Control and Prevention (CDC) Human Influenza Virus Real-time RT-PCR Detection and Characterization Panel (rRT-PCR Flu Panel) (K080570).

The CDC rRT-PCR Flu Panel is a panel of oligonucleotide primers and dual-labeled hydrolysis forobes for the qualitative detection and differentiation of influenza viruses. The panel can be used to test nasopharyngeal specimens, respiratory specimens or virus culture.

Prior to testing, the samples are purified with one of four commercial available sample purification methods (see chart below). A master mix is prepared by combining the appropriate quantities of primers and probes from the rRT-PCR Flu Panel with a commercially available reverse transcription master mix. The master mix is aliquoted into a 96 well plate followed by addition of the purified samples and controls. The prepared place on an ABI 7500 Fast Dx Real-Time PCR instrument and thermocycled according to the cycling conditions described in the rRT-PCR Flu Panel product insert.

Each run on the ABI includes a no-template control (nuclease-free water), a H5 virus control (noninfectious reassortant influenza A/H5 virus containing cultured human cells) and a human specimen control (cultured human cells). The human specimen control is extracted with the test samples and is intended to ensure that the extraction process was properly performed. The notemplate and H5 virus control templates are added to the PCR plate prior to thermocycling.

At the conclusion of the run, the operator is required to set a baseline for all assays included in the run. The operator is then required to interpret the test results for each sample based on the results of the controls and the Ct values for each sample.

ItemDevicePredicate
NameJBAIDS Influenza A/H5 (Asianlineage) Detection KitCDC rRT-PCR Flu Panel (K080570)
TechnologyReal-time PCR using hydrolysisprobesReal-time PCR using hydrolysis probes
Intended UseQualitative in vitro detection ofinfluenza A/H5 (Asian lineage)RNAQualitative in vitro detection ofinfluenza A/H1, A/H3, A/H5 (Asianlineage) RNA
SpecimenTypesNasopharyngeal swabs and throatswabsNasopharyngeal swabs, respiratoryspecimens or virus culture

Table 1 Similarities of the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit and CDC Human Influenza Virus Real-time RT-PCR Detection and Characterization Panel

{5}------------------------------------------------

Table 2 Differences of the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit and CDC Human Influenza Virus Real-time RT-PCR Detection and Characterization Panel

ItemDevicePredicate
NameJBAIDS Influenza A/H5(Asian lineage) Detection KitCDC rRT-PCR Flu Panel (K080570)
RequiredInstrumentationJBAIDS instrumentApplied Biosystems 7500 Fast Dx Real-time PCR instrument with SDS software v 1.4
Interpretation ofTest ResultsAutomated analysis of testresults and controlsUser required to interpret test and controlresults
Enzyme MasterMixAssays come in freeze-driedsingle use vials that include allcomponents of master mixInvitrogen SuperScript™ III Platinum ® One-Step Quantitative RT-PCR Kits
ExtractionMethodsIT 1-2-3TM Platinum PathIT 1-2-3TM VIBEQIAamp® Viral RNA Mini Kit,QIAGEN RNeasy® Mini Kit,Roche MagNA Pure Total Nucleic Acid KitRoche MagNA Pure LC RNA Isolation Kit II

Performance Characteristics:

    1. Analytical Specificity Limit of Detection: The limit of detection (LoD) for the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit was determined using TS and NPS specimens spiked with two different live virus strains (see Table 3). The LoD was first estimated by testing NPS and TS specimens soiked with 10-fold dilutions of virus. The specimens were purified with the IT 1-2-3 TM VIBE Sample Purification Kit and the IT 1-2-3 110 Platinum Path Purification Kit and tested. Once estimated. the LoD was confirmed by testing 20 individual specimens (from independent donors) spiked at the estimated LoD of 50 EID50/mL. This level of virus was successfully detected for all 20 NPS and TS specimens purified with either the IT 1-2-31M VIBE or the IT 1-2-3TM Platinum Path Kit Purification Kits.
      Table 3 Limit of Detection by Influenza A/H5 Virus Strain
Influenza A/H5 (Asian lineage) Strain TestedLimit of Detection
A/Vietnam/1203/2004 x A/Puerto Rico/8/34 reassortant50 EID50/mL
A/Anhui/01/2005/ x A/Puerto Rico/8/34 reassortant50 EID50/mL
    1. Analytical Specificity Inclusivity & Reactivity: Eight Influenza A/H5 (Asian lineage) strains were spiked into TS specimens at multiple concentrations and tested using the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit. The lowest level of virus that was detected is shown in Table 4. For all eight strains, a spike level of 500-1000 TCID50/mL was determined to have similar performance (with regard to detection and Cp values) as specimens spiked with 50 EID50/mL of the egg grown viruses used in the LoD study.

{6}------------------------------------------------

Panel MembersCladeLowest Level of DetectionTCID50/mL
Avian precursor1.0100
Yunnan(A/Chicken/Yunnan/1251/03)
A/Vietnam/1203/2004(H5N1)-PR8/CDC-RG1.0100
Avian precursor Hunan (A/Duck/Hunan/795/02)2.1100
A/Chicken/Korea/IS/062.2100
A/Scaly Breasted Munia/Hong Kong/45/20062.3100
A/Japanese white-eye/Hong Kong/1038/20062.310
A/Common Magpie/Hong Kong/645/20062.3100
A/Anhui/01/2005(H5N1)-PR8-IBCDC-RG2.310

Table 4 Influenza A/H5 (Asian lineage) Viruses Used to Evaluate Inclusivity

    1. Analytical Specificity Exclusivity: To ensure that the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit does not cross react with other organisms, TS specimens spiked with high concentrations of the non-target organisms listed in Table 5 were prepared and tested. Specimens spiked with all 45 (100%) organisms were negative with both the Target 1 and Target 2 assays.
      Table 5 Exclusivity Panel
Influenza VirusesNon-Influenza VirusesBacteria/Fungi
Influenza A/H1N1 A/Brisbane/59/07Enterovirus 71Bacteria
Influenza A/H1N1 A/Denver/1/57Adenovirus type 1 Ad. 71Bordetella pertussis A639
Influenza A/H1N1 A/PR/8/34Adenovirus type 7a S-1058Mycoplasma pneumoniae M129
Influenza A/H1N1 A/FM/1/47Coronavirus OC43Moraxella catarrhalis Ne 11
Influenza A/H1N1 A/NWS/33Coronavirus 299EPseudomonas aeruginosa
Influenza A/H3N2 A/Wisconsin/67/2005Rhinovirus 1AStaphylococcus aureus COL
Influenza A/H3N2 A/Victoria/3/75Parainfluenza virus, type 2Streptococcus pneumoniae Type 59
Influenza A/H3N2 A/Port Chalmers/1/73Parainfluenza virus, type 3Legionella micdadei Tatlock
Influenza A/H3N8 A/MAL/ALB/16/87RSV AMycobacterium tuberculosis H37Ra-
Influenza A/H5N1 A/TY/MA/40550/87-BEL/42hMPV-8 Peru6-2003Escherichia coli O157:H7
Influenza A/H5N1 A/DK/PA/4560069-9/06Neisseria elongata
Influenza A/H5N1 A/MUTESWAN/MI/451072-2/06Staphylococcus epidermidis
Influenza A/H5N2 A/DK/SC/318328-3/04Streptococcus pyogenes
Influenza A/H6N2 A/Chicken/CA/32213-1/2000Fungi
Influenza A/H7N3 A/TY/UT/24721-10/95Candida albicans
Influenza A/H7N7 A/Mallard/Netherlands/12/2000 IB-CDC-1
Influenza A/H9N2 A/Turkey/Wisconsin/1966
Influenza B B/FL/04/06
Influenza B B/Lee/40
Influenza B B/Taiwan/2/62
Influenza B B/GL/1739/54

{7}------------------------------------------------

    1. Clinical Performance: Due to the rarity of infection with the influenza A/H5 (Asian lineage) virus, it was not possible to evaluate the clinical sensitivity of the test system using true clinical samples. As a result, a surrogate clinical sensitivity assessment was performed and a clinical specificity assessment was conducted using influenza-negative clinical specimens spiked with influenza A/H5 (Asian lineage) virus.
    1. Surrogate Clinical Sensitivity: A panel of surrogate samples consisting of eight strains of Influenza A/H5 (Asian lineage) and six strains of seasonal influenza (Table 6) were spiked into TS or NPS specimens collected in viral transport medium (VTM). The complete panel, for both TS and NPS specimens, consisted of the following: Influenza A/H5 viruses at limit of detection (LoD), 5× LoD, 10× LoD, and 100× LoD (LoD was estimated to be 500 TCIDso/mL); seasonal influenza viruses at LoD and 100× LoD; and six specimens that were not spiked with any virus. Each specimen was purified with the IT 1-2-3 M VIBE and IT 1-2-3 ™ Platinum Path Sample Purification Kits. As expected, all seasonal influenza (48 total results) and influenza-negative specimens (24 total results) produced negative results. Of the 128 Influenza A/H5 virus-containing specimens. 126 produced positive results (98% success).

Table 6 Clinical Sensitivity Panel

Panel Members
Influenza A/H5 Strains
Avian precursor Yunnan(A/Chicken/Yunnan/1251/03)
A/Vietnam/1203/2004(H5N1)-PR8/CDC-RG
Avian precursor Hunan (A/Duck/Hunan/795/02)
A/Chicken/Korea/IS/06
A/Scaly Breasted Munia/Hong Kong/45/2006
A/Japanese white-eye/Hong Kong/1038/2006
Seasonal Influenza Strains
A/New Caledonia/20/1999 (H1N1)
A/Hawaii/15/2001 (H1N1)
A/New York/55/2004 (H3N2)
A/Wisconsin/67/2005 (H3N2)
B/Ohio/01/2005 (Victoria/2/87-like)
B/Florida/07/2004 (Yamagata/16/88-Like)
A/Common Magpie/Hong Kong/645/2006
A/Anhui/01/2005(H5N1)-PR8-IBCDC-RG
Unspiked samples
unspiked #1
unspiked #2
unspiked #3
unspiked #4
unspiked #5
unspiked #6

{8}------------------------------------------------

    1. Clinical Specificity: A multicenter clinical study was conducted over a 4-month period using frozen, banked NPS and TS specimens that had been previously tested for respiratory pathogens. The specimens were obtained and tested at 3 different test sites.
      Three hundred fourteen (314) NPS specimens purified using the IT 1-2-3"M VIBE Sample Purification Kit and 299 NPS specimens purified using the IT 1-2-3TM Platinum Path Purification kit had valid negative results for both the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit and the CDC rRT Flu Panel. Similarly, 298 VIBE purified and 283 Platinum Path purified TS specimens had valid negative results with both kits. These results demonstrate that the clinical specificity for NPS and TS specimens purified with the IT 1-2-3 100 VIBE or Platinum Path Purification Kits and tested with the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit is at least 99% with 95% confidence. Results are summarized in Table 7.
SampleTypePurificationKitSiteSpecimensJBAIDS ResultsClinicalSpecificityClaim
TotalTestedRemovedFromStudyaPositiveNotDetectedInconclusiveSCFailureInvalid
Nasopharyngeal SwabsIT 1-2-3TMVIBESite 116032012701099-100%Specificity,95% CI
Site 3201001870212b
Total36132 (8.9%)03140312
Nasopharyngeal SwabsIT 1-2-3TMPlatinumPathSite 115321013200099-100%Specificity,95% CI
Site 317400167016b
Total32721 (6.4%)0299016
Throat SwabsIT 1-2-3TMVIBESite 217130166011b,c99-100%Specificity,95% CI
Site 313200132000
Total3033 (1.0%)0298011
Throat SwabsIT 1-2-3TMPlatinumPathSite 2193140170009b99-100%Specificity,95% CI
Site 3140001130522b
Total33314 (4.2%)02830531
Table 7 Influenza A/H5 Comparison Results
-----------------------------------------------

4 Specimens were removed from the study if they did not have a valid result for the CDC comparator assay.

6 Failure of external extraction control used only for clinical study.

& The positive control for the Target 2 assay failed and was not retested.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Robert E. Miller, Ph.D., RAC Director Division of Regulated Activities and Compliance U.S. Army Medical Materiel Development Activity Office of the Surgeon General, US Army 1430 Veterans Drive, Fort Detrick, Maryland, 21702-9232

JUL 06 2010

Re: K100287

Trade/Device Name: JBAIDS Influenza A/H5 (Asian lineage) Detection Kit Regulation Number: 21 CFR §866.3332 Reagents for detection of specific novel influenza A viruses Regulation Name: Regulatory Class: Class II NXD Product Code: June 21, 2010 Dated: Received: June 28, 2010

Dear Dr. Miller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{10}------------------------------------------------

Page 2 - Dr. Robert Miller

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Vally, a Agns

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Indications for Use

510(k) Number: K100287

Device Name: JBAiDS Influenza A/H5 (Asian lineage) Detection Kit

Indications for Use:

The JBAIDS Influenza A/H5 (Asian lineage) Detection Kit is intended for use in real-time reverse-transcriptase polymerase chain reaction (rRT-PCR) assays on the Joint Biological Agent Identification and Diagnostic System (JBAIDS) instruments for the in vitro qualitative detection of Influenza A/H5 (Asian Iineage) viral RNA in patient nasopharyngeal swab (NPS) or throat swab (TS) specimens for the presumptive laboratory identification of Influenza A/H5 (Asian lineage) virus.

Testing with the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit should be in conjunction with other laboratory testing and clinical observations for the following indications:

    1. Providing epidemiological information for the surveillance of human infection with Influenza A/H5 (Asian lineage) virus.
    1. Identifying patients who may be infected with Influenza A/H5 (Asian lineage) virus based on clinical and epidemiological risk factors.

Testing with the JBAIDS Influenza A/H5 (Asian lineage) Detection Kit should not be performed unless the patient meets the most current U.S. Department of Health and Human Services (DHHS) clinical and epidemiologic criteria for testing suspected A/H5 specimens.

The definitive identification of influenza A/H5 (Asian lineage) either directly from patient specimens or from virus cultures requires additional laboratory testing, along with clinical and epidemiological assessment in consultation with national influenza surveillance experts.

Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.

Use is limited to laboratories with appropriate biosafety equipment and containment procedures. It is intended for use by experienced laboratory personnel who have training in standardized molecular testing procedures and expertise in viral diagnosis, and have received training on the JBAIDS Instrument.

If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses

{12}------------------------------------------------

and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a biosafety laboratory (BSL) 3+ facility is available to receive and culture specimens.

X Prescription Use_ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety

(OIVD)

Uve Schif

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

2 100287 510(k).

§ 866.3332 Reagents for detection of specific novel influenza A viruses.

(a)
Identification. Reagents for detection of specific novel influenza A viruses are devices that are intended for use in a nucleic acid amplification test to directly detect specific virus RNA in human respiratory specimens or viral cultures. Detection of specific virus RNA aids in the diagnosis of influenza caused by specific novel influenza A viruses in patients with clinical risk of infection with these viruses, and also aids in the presumptive laboratory identification of specific novel influenza A viruses to provide epidemiological information on influenza. These reagents include primers, probes, and specific influenza A virus controls.(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Reagents for Detection of Specific Novel Influenza A Viruses.” See § 866.1(e) for information on obtaining this document.
(2) The distribution of these devices is limited to laboratories with experienced personnel who have training in standardized molecular testing procedures and expertise in viral diagnosis, and appropriate biosafety equipment and containment.